Send Medical Device eSTAR and eCopy Premarket Submissions Online

October 3, 2022 - The FDA is announcing that you may now send electronic copy (eCopy) or electronic Submission Template And Resource (eSTAR) premarket submissions online through the CDRH Customer Collaboration Portal ("CDRH Portal").

Building on the progress tracker for 510(k) submissions launched in 2021 and the trial process of electronic uploads launched in July 2022, the CDRH Portal now allows anyone to register for a CDRH Portal account to send CDRH eCopy or eSTAR premarket submissions online.

Starting October 1, 2023, all 510(k) submissions, unless exempted, must be submitted as electronic submissions using eSTAR.

Introduction

Each person who wants to market in the U.S. a Class I, II, or III device intended for human use, for which a Premarket Approval application (PMA) is not required, must, if not submitting a De Novo request, submit a 510(k) to FDA unless the device is exempt from the 510(k) requirements of the Federal Food, Drug, and Cosmetic Act (the FD&C Act) and does not exceed the limitations of exemptions in .9 of the device classification regulations (e.g., 21 CFR 862.9, 21 CFR 864.9). There is no 510(k) form; however, 21 CFR 807 Subpart E describes the requirements for a 510(k) submission. Before marketing a device, each submitter must receive an order, in the form of a letter, from FDA which finds the device to be substantially equivalent to a legally marketed device. This order “clears” the device for commercial distribution (see The 510(k) Program Guidance).

A 510(k) is a premarket submission made to FDA to demonstrate that the device to be marketed is substantially equivalent to a legally marketed device (section 513(i)(1)(A) FD&C Act) which is not subject to PMA. Submitters must compare their device to one or more similar legally marketed devices and make and support their substantial equivalence claims. A legally marketed device to which a new device may be compared for a determination regarding substantial equivalence is a device: 1) that was legally marketed prior to May 28, 1976 (preamendment device); 2) which has been reclassified from Class III to Class II or I (such as a device that was granted marketing authorization via the De Novo classification process under section 513(f)(2) of the FD&C Act); or 3) which has been found to be substantially equivalent through the 510(k) process. The legally marketed device(s) to which substantial equivalence is drawn is commonly known as the “predicate.” Although devices recently cleared under 510(k) are often selected as the predicate to which substantial equivalence is claimed, any legally marketed device that is not subject to PMA may be used as a predicate. The predicate device must not have been removed from the market at the initiative of FDA nor have been determined to be misbranded or adulterated by a judicial order.

Until the submitter receives an order declaring a device substantially equivalent to a legally marketed device, the submitter may not proceed to market the device. Once the device is determined to be substantially equivalent to a legally marketed device, it can then be marketed in the U.S. The substantial equivalence determination is usually made within 90 days and is made based on the information submitted by the submitter.

Please note that FDA does not typically perform 510(k) pre-clearance facility inspections. The submitter may market the device immediately after 510(k) clearance is granted. The manufacturer should be prepared for an FDA quality management system (21 CFR 820) inspection at any time after 510(k) clearance.

What is Substantial Equivalence

A 510(k) requires demonstration of substantial equivalence to a legally marketed device not subject to PMA.

A device is substantially equivalent if, in comparison to a predicate it:

  • has the same intended use as the predicate, and
  • has the same technological characteristics as the predicate;
    or
  • has the same intended use as the predicate, and
  • has different technological characteristics and does not raise different questions of safety and effectiveness than the predicate, and
  • the information submitted to FDA demonstrates that the device is as safe and effective as the predicate.

A claim of substantial equivalence does not mean the new and predicate devices need to be identical. FDA first establishes that the new and predicate devices have the same intended use and any differences in technological characteristics do not raise different questions of safety and effectiveness. If there are differences in technological characteristics but they do not raise different questions of safety and effectiveness, FDA then reviews the proposed scientific methods for evaluating the new or different characteristics’ effects on safety and effectiveness. If FDA determines that the methods are acceptable, it evaluates performance data. This performance data can include clinical data and non-clinical bench performance data, including engineering performance testing, sterility, electromagnetic compatibility, software validation, and biocompatibility evaluation, among other data.

If FDA determines that a device is not substantially equivalent, the submitter may:

Who is Required to Submit a 510(k)

You generally are required to submit a 510(k) if your device is subject to the requirements arising under section 510(k) of the FD&C Act and you are a manufacturer, including a:

  1. Finished device manufacturer that manufactures a device according to its own specifications and markets it in the U.S.
  2. Manufacturer of components or accessories to finished devices that are ready to be used for any intended health-related purpose and packaged or labeled for commercial distribution for such health-related purpose, or devices which of necessity must be further processed by a licensed practitioner or other qualified person to meet the needs of a particular patient. Manufacturers of device components are not required to submit a 510(k) unless such components are promoted for sale to an end user, such as replacement parts. A manufacturer can be either domestic or foreign.
  3. Remanufacturer.
  4. Reprocessor of single-use devices.
  5. Specifications developer. 

You may also be required to submit a 510(k) if you currently have a device in commercial distribution or are reintroducing it into commercial distribution and the device is about to be significantly changed or modified in design, components, method of manufacture, or intended use. A significant change or modification includes a change or modification that could significantly affect the safety or effectiveness of the device, or a major change or modification in the intended use of the device. For example, while most repackagers and relabelers are not required to submit a 510(k), a repackager or relabeler may be required to submit a 510(k) if they make significant labeling changes such as modifying device manuals by adding a new intended use, or deleting or adding warnings or contraindications, etc. They may also be required to submit a 510(k) if their operations, such as sterilization, alter the condition of a device in a manner that significantly affects its safety or effectiveness.

Please note that all manufacturers (including specifications developers) of Class II and III devices and select Class I devices must comply with the requirements in Design and Development (ISO 13485:2016(E), Clause 7.3 and its subclauses, incorporated by reference into 21 CFR 820.7(b), and 21 CFR 820.10). The holder of a 510(k) must have design and development files available for FDA review during a site inspection. In addition, any changes to the device specifications or manufacturing processes must be made in accordance with the Quality Management System regulation (21 CFR 820) and may be subject to a new 510(k). Additional information is found on the webpage “Is a new 510(k) required for a modification to the device?”  

When a 510(k) is Required

A 510(k) is required when:

  1. Unless exempt or subject to a PMA or De Novo classification, a person required to register is introducing a device into commercial distribution (marketing) for the first time. After May 28, 1976 (effective date of the Medical Device Amendments to the Act), anyone who wants to sell a device in the U.S. is required to make a 510(k) submission at least 90 days prior to offering the device for sale in interstate commerce, even though it may have been under development or clinical investigation before that date. If your device was not marketed by your firm before May 28, 1976, a 510(k) may be required.
  2. There is a significant change or modification to a legally marketed device’s design, components, method of manufacture, or intended use, including a change or modification in the device that could significantly affect its safety or effectiveness, or a major change or modification in the intended use of the device. Any modifications must be made in accordance with the Quality Management System regulation, 21 CFR 820, and records of changes must be maintained. The manufacturer should provide an appropriately robust justification of a decision that submission of a new 510(k) is not required.

    See the webpage “Is a new 510(k) required for a modification to the device?” for additional information.

When a 510(k) is Not Required

  1. If you sell unfinished devices to another firm for further processing or sell components to be used in the assembling of devices by other firms, you might not be required to submit a 510(k). However, if you manufacture components or accessories that are ready to be used for any intended health-related purpose and are packaged or labeled for commercial distribution for such health-related purpose, including as replacement parts, or if your devices of necessity must be further processed by a licensed practitioner or other qualified person to meet the needs of a particular patient, a 510(k) may be required.
  2. Your device is not being marketed or commercially distributed. Please note that if you perform clinical investigations of your device to determine safety and effectiveness, you are subject to the Investigational Device Exemptions (IDE) regulation (21 CFR 812).
  3. You distribute a device from the original place of manufacture to the person who makes the final delivery or sale of the device to the ultimate consumer or user and do not manufacture, repackage, process, or relabel the device.
  4. In most cases, if you are a contract manufacturer, contract sterilizer, repackager or relabeler and the device is not significantly changed or modified in design, components, method of manufacture, or intended use. You must use the FDA-assigned 510(k) number of the cleared 510(k) if listing the device.
  5. Your device was legally in commercial distribution before May 28, 1976 and has not been significantly changed or modified in design, components, method of manufacture, or intended use. These devices are “grandfathered.” You may use Form FDA 3752 to submit documentation to CDRH for the purpose of establishing the pre-amendment status of a device.
  6. You are an importer of a foreign made medical device and the foreign manufacturer submitted a 510(k) and received marketing authorization. 
  7. Your device is exempted from 510(k) by regulation (21 CFR 862-892). That is, certain Class I or II devices can be marketed for the first time without having to submit a 510(k). A list of the Class I and II exempted devices can be found on Medical Device Exemptions 510(k) and GMP Requirements. However, if the device exceeds the limitations of exemptions in .9 of the device classification regulation chapters (e.g., 21 CFR 862.9, 21 CFR 864.9), such as if the device has a different intended use or operates using a different fundamental scientific technology than a legally marketed device in that generic type of device, then a 510(k) must be submitted to market the device.

Preamendment Devices

The term “preamendment device” refers to a device that was in commercial distribution before May 28, 1976. 

If a preamendment device has not been significantly changed or modified in design, components, method of manufacture, or intended use, and is not subject to a regulation requiring a PMA application published by FDA, it is considered a “grandfathered” device and does not require a 510(k). If the device is labeled for a different intended use than the intended use from before May 28, 1976, the device is considered a new device and a 510(k) might be required for marketing clearance.

A firm may use Form FDA 3752 to submit documentation to CDRH for the purpose of establishing the preamendment status of a device. See Preamendment Status for information on documentation requirements.

Third Party Review Program

The Center for Devices and Radiological Health (CDRH) has implemented a Third Party Review Program. This program provides an option to manufacturers of certain devices of submitting their 510(k) to private parties (Recognized Third Parties) identified by FDA for review instead of submitting directly to CDRH. For more information on the program, eligible devices and a list of Recognized Third Parties go to Third Party Review Program Information page.

Additional Information