  5. Hintermann Series H3 Total Ankle Replacement Has a Higher-Than-Expected Risk of Device Failure: FDA Safety Communication
  1. Safety Communications

Hintermann Series H3 Total Ankle Replacement Has a Higher-Than-Expected Risk of Device Failure: FDA Safety Communication

Post-approval Study Data Update: October 31, 2025

The FDA is providing an update on post-approval study data for the Hintermann Series H3 Total Ankle Replacement (TAR) system.

As provided by the manufacturer in the final post-approval study (PAS) report for the long-term follow-up of patients who were enrolled in the premarket clinical studies, the estimated rate of additional surgery associated with the implanted device to remove or revise metal components is 14.9% at 10 years, compared to the rate at 5 years which was 6.5%. When additional surgery to remove or revise all components (polyethylene [plastic] components as well as metal components) is included, the estimated rate is 31.8% at 10 years, compared to the rate at 5 years which was 19.2%. These rates are based on a statistical calculation of the probability of device component removal or revision using the available time to additional surgery data of patients enrolled in the study.  The final data summary for the post-approval study results can be found on the FDA post-approval study webpage.

At this time, the FDA recommendations provided below have not changed. The FDA continues to work with the manufacturer to evaluate all available information about the performance of Hintermann Series H3 TAR systems and the risk of device failure. In addition, the FDA will continue to work with the manufacturer to ensure that patients and providers are aware of the PAS data on the long-term risk of device failure. We will continue to keep the public informed if significant new information becomes available.

Español 简体中文 (Simplified Chinese) Tagalog

Date Issued: February 29, 2024

The U.S. Food and Drug Administration (FDA) is alerting patients, caregivers, and health care providers about a higher-than-expected risk of device failure with the Hintermann Series H3 Total Ankle Replacement (TAR) system, manufactured by DT MedTech LLC.

The FDA is evaluating interim post-approval study (PAS) results for the Hintermann Series H3 TAR system and other real-world data. For patients with the Hintermann Series H3 TAR system, the results suggest a higher rate of failure, specifically, additional surgery (removals or revisions of metal components, at least 16.1%) associated with the implanted device compared with the rate in the premarket clinical studies (9.9%).  When all types of revisions are included in the interim PAS results (such as revisions of the plastic component as well as the metal component), the rate of additional surgery is at least 28.5%.

The FDA is working with the manufacturer to evaluate data from all available sources to better understand potential causes of the higher failure rate.

Recommendations for Patients and Caregivers

  • Patients who are considering a Hintermann Series H3 TAR system:
    • Discuss all available treatment options for painful arthritic ankle joints with your health care provider.
    • Know there are benefits and risks associated with all joint replacement medical devices and procedures.
  • Patients who have a Hintermann Series H3 TAR system:
    • If the system is functioning well, and you have no new or worsening pain or symptoms, the FDA does not recommend surgery to remove it.
    • Contact your health care provider if you are experiencing any of the following:
      • any new or worsening pain or swelling,
      • inability to use your ankle or bear weight,
      • grinding or other noise, or
      • weakness around your implanted device.
    • Be aware, your health care provider may perform a physical examination of your operated ankle and obtain X-rays to evaluate it. In some instances, a CT scan may be necessary to assess if the plastic component in your Hintermann Series H3 TAR system is broken.
    • Report any problems or complications experienced with your TAR system to the FDA. Your report, along with information from other sources, can provide information that helps improve patient safety.

Recommendations for Health Care Providers

  • Review and discuss the Recommendations for Patients and Caregivers above with your patients.
  • As part of shared decision-making, discuss the benefits and risks of all relevant treatment options for painful arthritic ankle joints with your patients.
  • When making treatment recommendations, consider that there is a higher risk of device failure with the Hintermann Series H3 TAR system compared with the rate in the premarket clinical studies.
  • Read and carefully follow the Instructions for Use for the Hintermann Series H3 TAR system.
  • Monitor patients with the Hintermann Series H3 TAR system for device problems such as loosening and fractures of the implant components of the device.
  • For suspected device problems, such as a fractured plastic (polyethylene) component, consider performing X-rays to further evaluate the device integrity.
    • Be aware that changes on X-rays can be subtle. If X-rays are negative and polyethylene fracture is still suspected, a CT scan may be needed to determine whether a plastic component fracture has occurred.
    • Be aware that the clinical presentation and the signs or symptoms of fracture in plastic materials such as polyethylene can be subtle even in a CT scan.
  • Report any problems or complications experienced by patients with Hintermann Series H3 TAR systems to the FDA.

Device Description

Mobile-bearing TARs consist of a metal plate in the lower leg (tibial), a mobile plastic (polyethylene) component (inlay), and a metal ankle (talar) component. They are indicated for use as a non-cemented total ankle prosthesis ("artificial joint") and used to replace a painful arthritic ankle joint due to osteoarthritis, post-traumatic osteoarthritis, or rheumatoid arthritis.

The Hintermann Series H3 TAR system was approved in the United States in 2019 with the following two post-approval studies (PAS) required by the FDA:

  • a long-term study with 10-year follow-up of the cohort of 298 patients who were enrolled in the premarket clinical studies and
  • a new enrollment study with five-year follow-up.

The FDA has approved a total of two mobile-bearing TARs – the Scandinavian Total Ankle Replacement devices (STAR Ankle) in 2009, and the Hintermann Series H3 TAR system.

The FDA previously issued a Safety Communication in 2021 about the higher risk of device failure related to breakage of the polyethylene (plastic) component of the STAR Ankle system.

FDA Actions

The FDA is working with the manufacturer to evaluate all available information about the performance of Hintermann Series H3 TAR systems and the risk of device failure.

The FDA will continue to work with mobile-bearing TAR manufacturers to better understand the factors that contribute to device failures.

The FDA is also collaborating with international regulatory agencies to review data from registries for mobile-bearing TARs and further evaluate device failure rates.

The FDA will keep the public informed if significant new information becomes available.

Reporting Problems with Your Device

If you think you had a problem with your device, the FDA encourages you to report the problem through the MedWatch Voluntary Reporting Form.

Health care personnel employed by facilities that are subject to the FDA's user facility reporting requirements should follow the reporting procedures established by their facilities.

Unique Device Identifier

The unique device identifier (UDI) helps identify individual medical devices sold in the United States from manufacturing through distribution to patient use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified, and problems potentially corrected more quickly.

You can find the UDI provided by DT MedTech LLC for Hintermann Series H3 TAR systems by checking the table below.

Version or Model NumberDevice DescriptionDevice Identifier Number
300105H3 PE INLAY SIZE 1 - 5MMB095300105
300106H3 PE INLAY SIZE 1 - 6MMB095300106
300107H3 PE INLAY SIZE 1 - 7MMB095300107
300109H3 PE INLAY SIZE 1 - 9MMB095300109
300205H3 PE INLAY SIZE 2 - 5MMB095300205
300206H3 PE INLAY SIZE 2 - 6MMB095300206
300207H3 PE INLAY SIZE 2 - 7MMB095300207
300209H3 PE INLAY SIZE 2 - 9MMB095300209
300305H3 PE INLAY SIZE 3 - 5MMB095300305
300306H3 PE INLAY SIZE 3 - 6MMB095300306
300307H3 PE INLAY SIZE 3 - 7MMB095300307
300309H3 PE INLAY SIZE 3 - 9MMB095300309
300405H3 PE INLAY SIZE 4 - 5MMB095300405
300406H3 PE INLAY SIZE 4 - 6MMB095300406
300407H3 PE INLAY SIZE 4 - 7MMB095300407
300409H3 PE INLAY SIZE 4 - 9MMB095300409
300505H3 PE INLAY SIZE 5 - 5MMB095300505
300506H3 PE INLAY SIZE 5 - 6MMB095300506
300507H3 PE INLAY SIZE 5 - 7MMB095300507
300509H3 PE INLAY SIZE 5 - 9MMB095300509
300605H3 PE INLAY SIZE 6 - 5MMB095300605
300606H3 PE INLAY SIZE 6 - 6MMB095300606
300607H3 PE INLAY SIZE 6 - 7MMB095300607
300609H3 PE INLAY SIZE 6 - 9MMB095300609
301111TALAR COMPONENT RIGHT SIZE 1B095301111
301112TALAR COMPONENT RIGHT SIZE 2B095301112
301113TALAR COMPONENT RIGHT SIZE 3B095301113
301114TALAR COMPONENT RIGHT SIZE 4B095301114
301115TALAR COMPONENT RIGHT SIZE 5B095301115
301116TALAR COMPONENT RIGHT SIZE 6B095301116
301121FC TALAR COMPONENT RIGHT SIZE 1B095301121
301122FC TALAR COMPONENT RIGHT SIZE 2B095301122
301123FC  TALAR COMPONENT RIGHTSIZE 3B095301123
301124FC TALAR COMPONENT RIGHT SIZE 4B095301124
301125FC TALAR COMPONENT RIGHT SIZE 5B095301125
301201H3 TIBIAL COMPONENT RIGHT SIZE 1B095301201
301202H3 TIBIAL COMPONENT RIGHT SIZE 2B095301202
301203H3 TIBIAL COMPONENT RIGHT SIZE 3B095301203
301204H3 TIBIAL COMPONENT RIGHT SIZE 4B095301204
301205H3 TIBIAL COMPONENT RIGHT SIZE 5B095301205
301206H3 TIBIAL COMPONENT RIGHT SIZE 6B095301206
302111TALAR COMPONENT LEFT SIZE 1B095302111
302112TALAR COMPONENT LEFT SIZE 2B095302112
302113TALAR COMPONENT LEFT SIZE 3B095302113
302114TALAR COMPONENT LEFT SIZE 4B095302114
302115TALAR COMPONENT LEFT SIZE 5B095302115
302116TALAR COMPONENT LEFT SIZE 6B095302116
302121FC TALAR COMPONENT LEFT SIZE 1B095302121
302122FC TALAR COMPONENT LEFT SIZE 2B095302122
302123FC TALAR COMPONENT LEFT SIZE 3B095302123
302124FC TALAR COMPONENT LEFT SIZE 4B095302124
302125FC TALAR COMPONENT LEFT SIZE 5B095302125
302201H3 TIBIAL COMPONENT LEFT SIZE 1B095302201
302202H3 TIBIAL COMPONENT LEFT SIZE 2B095302202
302203H3 TIBIAL COMPONENT LEFT SIZE 3B095302203
302204H3 TIBIAL COMPONENT LEFT SIZE 4B095302204
302205H3 TIBIAL COMPONENT LEFT SIZE 5B095302205
302206H3 TIBIAL COMPONENT LEFT SIZE 6B095302206

Questions?

If you have questions, contact CDRH's Division of Industry and Consumer Education (DICE).