1. Home
  2. Medical Devices
  3. Medical Device Safety
  4. Medical Device Recalls and Early Alerts
  5. Incubator/Warmer Correction: GE Healthcare Updates Use Instructions for Giraffe OmniBed and Giraffe Omnibed CareStation due to Risk of Heater Door Falling and Causing Infant Injury
  1. Medical Device Recalls and Early Alerts

Incubator/Warmer Correction: GE Healthcare Updates Use Instructions for Giraffe OmniBed and Giraffe Omnibed CareStation due to Risk of Heater Door Falling and Causing Infant Injury

This recall involves updating instructions for using certain devices, and does not involve removing them from where they are used or sold.  The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it without following the updated instructions.

Affected Product

Correct Assembly: Thread shank not visible, only shoulder can be seen.

Figure 1: Correct Assembly: Thread shank not visible, only shoulder can be seen.

Incorrect Assembly: Shoulder and thread shank both are visible.

Figure 2: Incorrect Assembly: Shoulder and thread shank both are visible.

  • Product Names: Giraffe OmniBed and Giraffe OmniBed CareStation 
    • Giraffe OmniBed: All models
  • Giraffe OmniBed CareStation: 00840682116862/All models
  • Lot/Serial Numbers: See full list of affected devices

What to Do

  • Do not move the canopy if a high priority alarm goes off and canopy movement stops.
  • Follow all user manual instructions for pre-use steps, which include checking the function of the heater doors before clinical use. 

Reason for Recall

GE HealthCare and subsidiary Datex-Ohmeda Inc. are updating use instructions for the Giraffe OmniBed and Giraffe OmniBed CareStation after reports that the screw securing the warmer heater doors may not be properly tightened, causing the heater doors to become loose. If doors become loose, a High Priority alarm goes off and canopy movement stops. If users attempt to force the canopy to keep moving, the doors may become damaged and could potentially fall on the patient.

The use of affected product may cause serious adverse health consequences from even minor trauma in the neonatal patient, including deep soft tissue injuries, traumatic brain injuries, and death.

There have been no reported injuries and no reports of death.

Device Use

Giraffe OmniBed and Giraffe OmniBed CareStations are combination infant incubator and infant warmer. The device can be operated as either an incubator or as a warmer and can transition from one mode to the other on demand. It cannot be operated in both modes at the same time. Incubators and warmers provide controlled warmth for newborns (neonates) who are unable to regulate their own body temperatures. Incubators provide an enclosed, temperature-controlled environment and warmers provide controlled heat in an open environment. These devices may also be used for short periods of time to help a neonate transition from the uterus to the external environment. 

Contact Information

Customers in the U.S. with questions about this recall should contact GE Healthcare Service at 1-800-437-1171 or a local service representative.

AnchorFull List of Affected Devices

HDGR52546 

HDGR52736 

HDGR52802 

HDGR52861 

HDGR52910 

HDGR52958 

HDGR53007 

HDGR53058 

HDGR53110 

HDGR53185 

HDGR52547 

HDGR52737 

HDGR52803 

HDGR52862 

HDGR52911 

HDGR52959 

HDGR53008 

HDGR53059 

HDGR53112 

HDGR53186 

HDGR52558 

HDGR52739 

HDGR52804 

HDGR52863 

HDGR52912 

HDGR52960 

HDGR53009 

HDGR53060 

HDGR53113 

HDGR53187 

HDGR52609 

HDGR52740 

HDGR52805 

HDGR52864 

HDGR52913 

HDGR52961 

HDGR53010 

HDGR53061 

HDGR53114 

HDGR53188 

HDGR52610 

HDGR52741 

HDGR52806 

HDGR52865 

HDGR52914 

HDGR52962 

HDGR53011 

HDGR53062 

HDGR53115 

HDGR53189 

HDGR52613 

HDGR52742 

HDGR52813 

HDGR52866 

HDGR52915 

HDGR52963 

HDGR53012 

HDGR53063 

HDGR53116 

HDGR53190 

HDGR52620 

HDGR52743 

HDGR52814 

HDGR52867 

HDGR52916 

HDGR52964 

HDGR53013 

HDGR53064 

HDGR53117 

HDGR53191 

HDGR52621 

HDGR52744 

HDGR52815 

HDGR52868 

HDGR52917 

HDGR52965 

HDGR53014 

HDGR53065 

HDGR53119 

HDGR53192 

HDGR52622 

HDGR52745 

HDGR52816 

HDGR52869 

HDGR52918 

HDGR52966 

HDGR53015 

HDGR53066 

HDGR53120 

HDGR53193 

HDGR52623 

HDGR52746 

HDGR52817 

HDGR52870 

HDGR52919 

HDGR52967 

HDGR53016 

HDGR53067 

HDGR53125 

HDGR53194 

HDGR52624 

HDGR52747 

HDGR52818 

HDGR52871 

HDGR52920 

HDGR52968 

HDGR53018 

HDGR53068 

HDGR53126 

HDGR53195 

HDGR52626 

HDGR52748 

HDGR52820 

HDGR52872 

HDGR52921 

HDGR52969 

HDGR53019 

HDGR53069 

HDGR53127 

HDGR53196 

HDGR52657 

HDGR52749 

HDGR52821 

HDGR52873 

HDGR52922 

HDGR52970 

HDGR53021 

HDGR53070 

HDGR53128 

HDGR53197 

HDGR52658 

HDGR52750 

HDGR52822 

HDGR52874 

HDGR52923 

HDGR52971 

HDGR53022 

HDGR53071 

HDGR53129 

HDGR53213 

HDGR52664 

HDGR52751 

HDGR52823 

HDGR52875 

HDGR52924 

HDGR52972 

HDGR53023 

HDGR53072 

HDGR53134 

HDGR53214 

HDGR52665 

HDGR52752 

HDGR52824 

HDGR52876 

HDGR52925 

HDGR52973 

HDGR53024 

HDGR53073 

HDGR53135 

HDGR53215 

HDGR52668 

HDGR52753 

HDGR52825 

HDGR52877 

HDGR52926 

HDGR52974 

HDGR53025 

HDGR53074 

HDGR53136 

HDHR53216 

HDGR52669 

HDGR52754 

HDGR52827 

HDGR52878 

HDGR52927 

HDGR52975 

HDGR53026 

HDGR53075 

HDGR53137 

HDGR53217 

HDGR52670 

HDGR52755 

HDGR52828 

HDGR52879 

HDGR52928 

HDGR52976 

HDGR53027 

HDGR53076 

HDGR53138 

HDGR53219 

HDGR52671 

HDGR52756 

HDGR52829 

HDGR52880 

HDGR52929 

HDGR52977 

HDGR53028 

HDGR53077 

HDGR53139 

HDGR53220 

HDGR52672 

HDGR52757 

HDGR52830 

HDGR52881 

HDGR52930 

HDGR52978 

HDGR53029 

HDGR53078 

HDGR53140 

HDGR53221 

HDGR52673 

HDGR52758 

HDGR52831 

HDGR52882 

HDGR52931 

HDGR52979 

HDGR53030 

HDGR53079 

HDGR53141 

HDGR53222 

HDGR52674 

HDGR52759 

HDGR52832 

HDGR52883 

HDGR52932 

HDGR52980 

HDGR53031

HDGR53080 

HDGR53142 

HDGR53223 

HDGR52675 

HDGR52760 

HDGR52833 

HDGR52884 

HDGR52933 

HDGR52981 

HDGR53032 

HDGR53081 

HDGR53145 

HDGR53224 

HDGR52678 

HDGR52761 

HDGR52835 

HDGR52885 

HDGR52934 

HDGR52982 

HDGR53033 

HDGR53085 

HDGR53146 

HDGR53225 

HDGR52682 

HDGR52762 

HDGR52836 

HDGR52886 

HDGR52935 

HDGR52983 

HDGR53034 

HDGR53086 

HDGR53147 

HDGR53226 

HDGR52683 

HDGR52763 

HDGR52839 

HDGR52887 

HDGR52936 

HDGR52984 

HDGR53035 

HDGR53087 

HDGR53152 

HDGR53231 

HDGR52684 

HDGR52764 

HDGR52840 

HDGR52888 

HDGR52937 

HDGR52985 

HDGR53036 

HDGR53088 

HDGR53153 

HDGR53232 

HDGR52685 

HDGR52765 

HDGR52841 

HDGR52890 

HDGR52938 

HDGR52986 

HDGR53037 

HDGR53089 

HDGR53154 

HDGR53235 

HDGR52686 

HDGR52768 

HDGR52842 

HDGR52891 

HDGR52939 

HDGR52987 

HDGR53038 

HDGR53090 

HDGR53155 

HDGR53241 

HDGR52687 

HDGR52769 

HDGR52843 

HDGR52892 

HDGR52940 

HDGR52988 

HDGR53039 

HDGR53091 

HDGR53156 

HDGR53242 

HDGR52700 

HDGR52770 

HDGR52844 

HDGR52893 

HDGR52941 

HDGR52990 

HDGR53040 

HDGR53092 

HDGR53157 

HDGR53251 

HDGR52709 

HDGR52771 

HDGR52845 

HDGR52894 

HDGR52942 

HDGR52991 

HDGR53041 

HDGR53093 

HDGR53158 

HDGR52714 

HDGR52775 

HDGR52846 

HDGR52895 

HDGR52943 

HDGR52992 

HDGR53042 

HDGR53094 

HDGR53159 

HDGR52718 

HDGR52776 

HDGR52847 

HDGR52896 

HDGR52944 

HDGR52993 

HDGR53043 

HDGR53095 

HDGR53160 

HDGR52720 

HDGR52777 

HDGR52848 

HDGR52897 

HDGR52945 

HDGR52994 

HDGR53044 

HDGR53096 

HDGR53161 

HDGR52721 

HDGR52778 

HDGR52849 

HDGR52898 

HDGR52946 

HDGR52995 

HDGR53045 

HDGR53097 

HDGR53162 

HDGR52723 

HDGR52779 

HDGR52850 

HDGR52899 

HDGR52947 

HDGR52996 

HDGR53047 

HDGR53098 

HDGR53163 

HDGR52726 

HDGR52780 

HDGR52851 

HDGR52900 

HDGR52948 

HDGR52997 

HDGR53048 

HDGR53099 

HDGR53164 

HDGR52727 

HDGR52781 

HDGR52852 

HDGR52901 

HDGR52949 

HDGR52998 

HDHR53049 

HDGR53100 

HDGR53165 

HDGR52728 

HDGR52782 

HDGR52853 

HDGR52902 

HDHR52950 

HDGR52999 

HDGR53050 

HDGR53102 

HDGR53166 

HDGR52729 

HDGR52789 

HDGR52854 

HDGR52903 

HDGR52951 

HDGR53000 

HDGR53051 

HDGR53103 

HDGR53167 

HDGR52730 

HDGR52790 

HDGR52855 

HDGR52904 

HDGR52952 

HDGR53001 

HDGR53052 

HDGR53104 

HDGR53168 

HDGR52731 

HDGR52791 

HDGR52856 

HDGR52905 

HDGR52953 

HDGR53002 

HDGR53053 

HDGR53105 

HDGR53169 

HDGR52732 

HDGR52792 

HDGR52857 

HDGR52906 

HDGR52954 

HDGR53003 

HDGR53054 

HDGR53106 

HDGR53171 

HDGR52733 

HDGR52793 

HDGR52858 

HDGR52907 

HDGR52955 

HDGR53004 

HDGR53055 

HDGR53107 

HDGR53175 

HDGR52734 

HDGR52794 

HDGR52859 

HDGR52908 

HDGR52956 

HDGR53005 

HDGR53056 

HDGR53108 

HDGR53176 

HDGR52735 

HDGR52799 

HDGR52860 

HDGR52909 

HDGR52957 

HDGR53006 

HDGR53057 

HDGR53109 

HDGR53181

HDHR52549 

HDHR52708 

HDHR52774 

HDHR52881 

HDHR53029 

HDHR52551 

HDHR52709 

HDHR52830 

HDHR52882 

HDHR53030 

HDHR52552 

HDHR52710 

HDHR52831 

HDHR52883 

HDHR53031 

HDHR52553 

HDHR52711 

HDHR52832 

HDHR52884 

HDHR53032 

HDHR52554 

HDHR52720 

HDHR52833 

HDHR52885 

HDHR53039 

HDHR52555 

HDHR52721 

HDHR52834 

HDHR52894 

HDHR53040 

HDHR52556 

HDHR52722 

HDHR52835 

HDHR52900 

HDHR53041 

HDHR52557 

HDHR52723 

HDHR52836 

HDHR52901 

HDHR53042 

HDHR52559 

HDHR52724 

HDHR52837 

HDHR52902 

HDHR53043 

HDHR52561

HDHR52725 

HDHR52838 

HDHR52903 

HDHR53044 

HDHR52562 

HDHR52726 

HDHR52839 

HDHR52904 

HDHR53045 

HDHR52563 

HDHR52727 

HDHR52840 

HDHR52905 

HDHR53046 

HDHR52564 

HDHR52728 

HDHR52841 

HDHR52906 

HDHR53047 

HDHR52565 

HDHR52729 

HDHR52842 

HDHR52907 

HDHR53048 

HDHR52567 

HDHR52730 

HDHR52843 

HDHR52908 

HDHR53049 

HDHR52568 

HDHR52731 

HDHR52844 

HDHR52909 

HDHR53030 

HDHR52569 

HDHR52732 

HDHR52845 

HDHR52951 

HDHR52910 

HDHR52572 

HDHR52733 

HDHR52847 

HDHR52911 

HDHR53052 

HDHR52573 

HDHR52734 

HDHR52848 

HDHR52912 

HDHR53053 

HDHR52575 

HDHR52735 

HDHR52849 

HDHR52913 

HDHR53054 

HDHR52576 

HDHR52736 

HDHR52850 

HDHR52914 

HDHR53055 

HDHR52577 

HDHR52737 

HDHR52851 

HDHR52915 

HDHR53059 

HDHR52578 

HDHR52738 

HDHR52852 

HDHR52926 

HDHR53062 

HDHR52579 

HDHR52739 

HDHR52853 

HDHR52927 

HDHR52580 

HDHR52740 

HDHR52854 

HDHR52928 

HDHR52581 

HDHR52741 

HDHR52855 

HDHR52929 

HDHR52588 

HDHR52742 

HDHR52857 

HDHR52971 

HDHR52589 

HDHR52743 

HDHR52858 

HDHR52972 

HDHR52650 

HDHR52744 

HDHR52859 

HDHR52974 

HDHR52653 

HDHR52745 

HDHR52860 

HDHR52975 

HDHR52654 

HDHR52746 

HDHR52861 

HDHR52976 

HDHR52689 

HDHR52747 

HDHR52862 

HDHR52977 

HDHR52690 

HDHR52748 

HDHR52863 

HDHR52978 

HDHR52691 

HDHR52749 

HDHR52864 

HDHR52979 

HDHR52692 

HDHR52750 

HDHR52865 

HDHR52981 

HDHR52693 

HDHR52751 

HDHR52866 

HDHR52982 

HDHR52694 

HDHR52752 

HDHR52867 

HDHR52983 

HDHR52695 

HDHR52753 

HDHR52868 

HDHR52985 

HDHR52696 

HDHR52754 

HDHR52869 

HDHR52986 

HDHR52697 

HDHR52759 

HDHR52870 

HDHR52987 

HDHR52698 

HDHR52760 

HDHR52871 

HDHR52988 

HDHR52699 

HDHR52767 

HDHR52872 

HDHR52989 

HDHR52700 

HDHR52768 

HDHR52873 

HDHR52990 

HDHR52701 

HDHR52769 

HDHR52874 

HDHR52991 

HDHR52702 

HDHR52770 

HDHR52875 

HDHR52992 

HDHR52703 

HDHR52771 

HDHR52876 

HDHR52993 

HDHR52704 

HDHR52772 

HDHR52877 

HDHR52994 

HDHR52707 

HDHR52773 

HDHR52880 

HDHR52995

2082844-001-526493, 1TABX6226911180201 (TABX62269)

2082844-001-526493, 1TABX6215911180122 (TABX62159) 

2082844-001-526493, 1TABX6219811180129 (TABX62198) 

2082844-001-526493, 1TABX6227011180201 (TABX62270) 

2082844-001-526493, 1TABX6229111180202 (TABX62291) 

2082844-001-526493, 1TABX6215611180122 (TABX62156) 

2082844-001-526493, 1TABX6226011180201 (TABX62260) 

2082844-001-526493, 1TABX6213511180119 (TABX62135) 

2082844-001-526493, 1TABX6218911180124 (TABX62189) 

2082844-001-526493, 1TABX6219611180124 (TABX62196) 

2082844-001-526493, 1TABX6218811180124 (TABX62188) 

2082844-001-526493, 1TABX6216011180122 (TABX62160) 

2082844-001-487119, To be provided (TABV60550) 

2082844-001-526493, 1TABW6398711171213 (TABW63987) 

2082844-001-526493, 1TABX6214111180119 (TABX62141) 

2082844-001-526493, 1TABW6409911171219 (TABW64099) 

2082844-001-526493, 1TABW6400911171213 (TABW64009) 

2082844-001-526493, 1TABW6396611171211 (TABW63966) 

2082844-001-526493, 1TABW6400811171213 (TABW64008) 

2082844-001-526493, 1TABX6215411180122 (TABX62154) 

2082844-001-526493, 1TABW6408611171219 (TABW64086) 

2082844-001-526493, 1TABX6210811180117 (TABX62108) 

2082844-001-526493, 1TABX6213011180119 (TABX62130) 

2082844-001-526493, 1TABW6406911171219 (TABW64069) 

2082844-001-526493, 1TABW6408411171219 (TABW64084) 

2082844-001-526493, 1TABW6404311171218 (TABW64043) 

2082844-001-526493, 1TABW6409811171219 (TABW64098) 

2082844-001-526493, 1TABW6396411171211 (TABW63964) 

2082844-001-526493, 1TABW6404911171218 (TABW64049) 

2082844-001-526493, 1TABW6395411171211 (TABW63954) 

2082844-001-526493, 1TABW6408711171219 (TABW64087) 

2082844-001-487119, Not Applicable (TABV60078) 

2082844-001-526493, 1TABX6232011180207 (TABX62320) 

2082844-001-526493, 1TABX6227211180201 (TABX62272) 

2082844-001-526493, 1TABX6211811180119 (TABX62118) 

2082844-001-526493, 1TABX6214611180119 (TABX62146) 

2082844-001-526493, 1TABW6398311171213 (TABW63983) 

2082844-001-526493, 1TABW6399411171213 (TABW63994) 

2082844-001-526493, 1TABX6220011180129 (TABX62200) 

2082844-001-526493, 1TABW6396211171211 (TABW63962) 

2082844-001-526493, 1TABX6227611180201 (TABX62276) 

2082844-001-487119, Not Applicable (TABV60092) 

2082844-001-526493, 1TABW6393811171211 (TABW63938) 

2082844-001-526493, 1TABW6399511171213 (TABW63995) 

2082844-001-526493, 1TABX6213111180119 (TABX62131) 

2082844-001-526493, 1TABX6225711180201 (TABX62257) 

2082844-001-526493, 1TABX6214011180119 (TABX62140) 

2082844-001-526493, 1TABX6226411180201 (TABX62264) 

2082844-001-526493, 1TABX6231911180207 (TABX62319) 

2082844-001-526493, 1TABX6230111180202 (TABX62301) 

2082844-001-526493, 1TABX6228711180202 (TABX62287) 

2082844-001-526493, 1TABX6229611180202 (TABX62296) 

2082844-001-526493, 1TABX6230011180202 (TABX62300) 

2082844-001-526493, 1TABX6233311180207 (TABX62333) 

2082844-001-526493, 1TABX6233111180207 (TABX62331) 

2082844-001-526493, 1TABX6229011180202 (TABX62290) 

2082844-001-526493, 1TABX6229311180202 (TABX62293) 

2082844-001-526493, 1TABX6216311180123 (TABX62163) 

2082844-001-526493, 1TABX6223511180130 (TABX62235) 

2082844-001-526493, 1TABX6222711180130 (TABX62227) 

2082844-001-526493, 1TABX6216811180123 (TABX62168) 

2082844-001-526493, 1TABX6216411180123 (TABX62164) 

2082844-001-526493, 1TABX6214911180122 (TABX62149) 

2082844-001-526493, 1TABX6217811180123 (TABX62178) 

2082844-001-526493, 1TABX6220611180129 (TABX62206) 

2082844-001-526493, 1TABX6216511180123 (TABX62165) 

2082844-001-526493, 1TABX6225011180201 (TABX62250) 

2082844-001-526493, 1TABX6217311180123 (TABX62173) 

2082844-001-526493, 1TABX6216211180123 (TABX62162) 

2082844-001-526493, 1TABX6216611180123 (TABX62166) 

2082844-001-526493, 1TABX6223411180130 (TABX62234) 

2082844-001-526493, 1TABX6220811180129 (TABX62208) 

2082844-001-526493, 1TABW6395211171211 (TABW63952) 

2082844-001-526493, 1TABW6395111171211 (TABW63951) 

2082844-001-526493, 1TABW6407811171219 (TABW64078) 

2082844-001-526493, 1TABX6211011180117 (TABX62110) 

2082844-001-526493, 1TABX6211311180117 (TABX62113) 

2082844-001-526493, 1TABX6233211180207 (TABX62332) 

2082844-001-526493, 1TABX6228211180201 (TABX62282) 

2082844-001-526493, 1TABX6228311180201 (TABX62283) 

2082844-001-526493, 1TABW6404611171218 (TABW64046) 

2082844-001-526493, 1TABW6400011171213 (TABW64000) 

2082844-001-526493, 1TABW6409511171219 (TABW64095) 

2082844-001-526493, 1TABW6409611171219 (TABW64096) 

2082844-001-526493, 1TABW6393311171211 (TABW63933) 

2082844-001-526493, 1TABX6232411180207 (TABX62324) 

2082844-001-526493, 1TABX6229211180202 (TABX62292) 

2082844-001-526493, 1TABX6225411180201 (TABX62254) 

2082844-001-526493, 1TABX6225511180201 (TABX62255) 

2082844-001-526493, 1TABX6225811180201 (TABX62258) 

2082844-001-487119, Not Applicable (TABV60433) 

2082844-001-526493, 1TABX6219711180124 (TABX62197) 

2082844-001-526493, 1TABX6212811180119 (TABX62128) 

2082844-001-526493, 1TABW6394411171211 (TABW63944) 

2082844-001-526493, 1TABW6398211171213 (TABW63982) 

2082844-001-526493, 1TABW6394811171211 (TABW63948) 

2082844-001-526493, 1TABW6394011171211 (TABW63940) 

2082844-001-526493, 1TABX6222611180130 (TABX62226) 

2082844-001-526493, 1TABW6393711171211 (TABW63937) 

2082844-001-526493, 1TABX6224011180130 (TABX62240) 

2082844-001-526493, 1TABX6223911180130 (TABX62239) 

2082844-001-526493, 1TABW6399611171213 (TABW63996) 

2082844-001-526493, 1TABX6224311180130 (TABX62243) 

2082844-001-526493, 1TABX6224211180130 (TABX62242) 

2082844-001-526493, 1TABX6224411180201 (TABX62244) 

2082844-001-526493, 1TABX6222911180130 (TABX62229) 

2082844-001-526769, 1TABW6397711171212 (TABW63977) 

2082844-001-526493, 1TABW6401311171213 (TABW64013) 

2082844-001-526493, 1TABW6395911171211 (TABW63959) 

2082844-001-526493, 1TABW6408211171219 (TABW64082) 

2082844-001-526493, 1TABW6402311171213 (TABW64023) 

2082844-001-526493, 1TABW6396511171211 (TABW63965) 

2082844-001-526493, 1TABX6233511180207 (TABX62335) 

2082844-001-526493, 1TABX6229511180202 (TABX62295) 

2082844-001-526493, 1TABX6224611180201 (TABX62246) 

2082844-001-526493, 1TABX6225311180201 (TABX62253) 

2082844-001-526493, 1TABX6222311180130 (TABX62223) 

2082844-001-526493, 1TABX6225111180201 (TABX62251) 

2082844-001-526493, 1TABX6225211180201 (TABX62252) 

2082844-001-526493, 1TABX6223311180130 (TABX62233) 

2082844-001-526493, 1TABX6223211180130 (TABX62232) 

2082844-001-526493, 1TABX6223611180130 (TABX62236) 

2082844-001-526493, 1TABX6224811180201 (TABX62248) 

2082844-001-526493, 1TABX6223811180130 (TABX62238) 

2082844-001-526493, 1TABX6224911180201 (TABX62249) 

2082844-001-526493, 1TABX6224511180201 (TABX62245) 

2082844-001-526493, 1TABX6210911180117 (TABX62109) 

2082844-001-526493, 1TABX6219411180124 (TABX62194) 

2082844-001-526493, 1TABX6219311180124 (TABX62193) 

2082844-001-526493, 1TABX6213911180119 (TABX62139) 

2082844-001-526493, 1TABX6216111180122 (TABX62161) 

2082844-001-526493, 1TABX6213611180119 (TABX62136) 

2082844-001-526493, 1TABX6212011180119 (TABX62120) 

2082844-001-526493, 1TABX6215811180122 (TABX62158) 

2082844-001-526493, 1TABX6213211180119 (TABX62132) 

2082844-001-526493, 1TABX6217111180123 (TABX62171) 

2082844-001-526493, 1TABX6221811180130 (TABX62218) 

2082844-001-526493, 1TABW6405911171219 (TABW64059) 

2082844-001-526493, 1TABX6224711180201 (TABX62247) 

2082844-001-526493, 1TABX6223111180130 (TABX62231) 

2082844-001-403566, Not Applicable (TABU67011) 

2082844-001-499499, 1TABW6259811170606 (TABW62598) 

2082844-001-430326, Not Applicable (TABV60027) 

2082844-001-430326, Not Applicable (TABV60030) 

2082844-001-430326, Not Applicable (TABV60028) 

2082844-001-430326, Not Applicable (TABU67004) 

2082844-001-462725, Not Applicable (TABU67040) 

2082844-001-462725, Not Applicable (TABU67039) 

2082844-001-462725, Not Applicable (TABU67056) 

2082844-001-462725, Not Applicable (TABU67041) 

2082844-001-462725, Not Applicable (TABU67053) 

2082844-001-462725, Not Applicable (TABU67045) 

2082844-001-462725, Not Applicable (TABU67047) 

2082844-001-462725, Not Applicable (TABU67050) 

2082844-001-462725, Not Applicable (TABU67043) 

2082844-001-460023, Not Applicable (TABU67014) 

2082844-001-510290, 010084068211685521TABV61236160902 (TABV61236) 

2082844-001-558377, 1TABX6238011180219 (TABX62380) 

2082844-001-534467, 1TABX6330711180628 (TABX63307) 

2082844-001-534467, 1TABX6330611180628 (TABX63306) 

2082844-001-534467, 1TABX6331411180628 (TABX63314) 

2082844-001-534467, 1TABX6236011180213 (TABX62360) 

2082844-001-534467, 1TABX6331211180628 (TABX63312) 

2082844-001-534467, 1TABX6331111180628 (TABX63311) 

2082844-001-534467, 1TABX6330911180628 (TABX63309) 

2082844-001-534467, 1TABX6331511180628 (TABX63315) 

2082844-001-501649, 1TABX6330411180626 (TABX63304) 

2082844-001-501649, 1TABX6330311180626 (TABX63303) 

2082844-001-501649, 1TABX6330511180626 (TABX63305) 

2082844-001-467192, Not Applicable (TABV60033) 

2082844-001-495642, Not Applicable (TABV60053) 

2082844-001-508846, 1TABX6247711180301 (TABX62477) 

2082844-001-508846, 1TABX6247611180301 (TABX62476) 

2082844-001-508846, 1TABX6246211180228 (TABX62462) 

2082844-001-508846, 1TABX6246311180228 (TABX62463) 

2082844-001-508846, 1TABX6248411180305 (TABX62484) 

2082844-001-508846, 1TABX6248511180305 (TABX62485) 

2082844-001-498552, 1TABX6248911180305 (TABX62489) 

2082844-001-498552, 1TABX6248611180305 (TABX62486) 

2082844-001-498552, 1TABX6248711180305 (TABX62487) 

2082844-001-498552, 1TABX6248811180305 (TABX62488) 

2082844-001-508846, 1TABX6332211180629 (TABX63322) 

2082844-001-508846, 1TABX6332111180629 (TABX63321) 

2082844-001-508846, 1TABX6221111180130 (TABX62211) 

2082844-001-508846, 1TABX6221211180130 (TABX62212) 

2082844-001-485099, Not Applicable (TABV60013) 

2082844-001-485099, Not Applicable (TABV60102) 

2082844-001-485099, Not Applicable (TABV60055) 

2082844-001-485099, Not Applicable (TABV60054) 

2082844-001-485099, Not Applicable (TABV60065) 

2082844-001-485099, Not Applicable (TABV60015) 

2082844-001-485099, Not Applicable (TABV60072) 

2082844-001-485099, Not Applicable (TABV60069) 

2082844-001-485099, Not Applicable (TABV60016) 

2082844-001-512022, 1TABX6332411180702 (TABX63324) 

2082844-001-531627, 1TABX6254211180309 (TABX62542) 

2082844-001-512022, 1TABX6254111180307 (TABX62541) 

2082844-001-495786, Not Applicable (TABV60056) 

2082844-001-565598, 1TABX63177180605 (TABX63177) 

2082844-001-565598, 1TABX63176180605 (TABX63176) 

2082844-001-565598, 1TABX63175180605 (TABX63175) 

2082844-001-535110, 1TABX6332011180629 (TABX63320) 

2082844-001-534358, 1TABX6237111180215 (TABX62371) 

2082844-001-534358, 1TABX6266511180313 (TABX62665) 

2082844-001-523533, 1TABW62153170215 (TABW62153) 

2082844-001-547388, 1TABX63316180628 (TABX63316) 

2082844-001-498396, 1TABX6331811180629 (TABX63318) 

2082844-001-498396, 1TABX6331911180629 (TABX63319) 

2082844-001-505050, 1TABX62459180227 (TABX62459) 

2082844-001-505050, 1TABX62460180227 (TABX62460) 

2082844-001-505050, 1TABW63210170815 (TABW63210) 

2082844-001-507094, 010084068211685521TABV61052160804 (TABV61052) 

2082844-001-510416, 1TABX62540180307 (TABX62540) 

2082844-001-500909, 1TABX62457180227 (TABX62457) 

2082844-001-500909, 1TABX62458180227 (TABX62458) 

2082844-001-477964, Not Applicable (TABU67058) 

2082844-001-516038, 1TABX6246111180227 (TABX62461) 

2082844-001-501478, 1TABX62431180222 (TABX62431) 

2082844-001-471681, Not Applicable (TABU67028) 

2082844-001-471681, Not Applicable (TABU67034) 

2082844-001-471681, Not Applicable (TABU67032) 

2082844-001-462005, Not Applicable (TABU67021) 

2082844-001-462005, Not Applicable (TABU67026) 

2082844-001-516778, 1TABX6248011180302 (TABX62480) 

2082844-001-516778, 1TABX6248111180302 (TABX62481) 

2082844-001-516778, 1TABX6247811180302 (TABX62478) 

2082844-001-516778, 1TABX6247911180302 (TABX62479) 

2082844-001-516778, 1TABV6231611161114 (TABV62316) 

2082844-001-500008, 1TABX62365180215 (TABX62365) 

2082844-001-500008, 1TABX62363180215 (TABX62363) 

2082844-001-500008, 1TABX62362180215 (TABX62362) 

2082844-001-500008, 1TABX62367180215 (TABX62367) 

2082844-001-500008, 1TABX62364180215 (TABX62364) 

2082844-001-500008, 1TABX62366180215 (TABX62366) 

2082844-001-500008, 1TABX62369180215 (TABX62369) 

2082844-001-500008, 1TABX62368180215 (TABX62368) 

2082844-001-500008, 010084068211685521TABV62289161114 (TABV62289) 

2082844-001-500008, 010084068211685521TABV62294161114 (TABV62294) 

2082844-001-499372, 1TABX62377180215 (TABX62377) 

2082844-001-499372, 1TABX62376180215 (TABX62376) 

2082844-001-499372, 1TABX62373180215 (TABX62373) 

2082844-001-499372, 1TABX62374180215 (TABX62374) 

2082844-001-517864, 1TABX63236180619 (TABX63236) 

2082844-001-517864, 1TABX63252180619 (TABX63252) 

2082844-001-403565, Not Applicable (TABV60039) 

2082844-001-559202, 1TABX6248211180302 (TABX62482) 

2082844-001-403565, Not Applicable (TABV60097) 

2082844-001-403565, Not Applicable (TABV60062) 

2082844-001-403565, Not Applicable (TABV60003) 

2082844-001-403565, Not Applicable (TABV60060)

Additional FDA Resources

Unique Device Identifier (UDI)

The unique device identifier (UDI) helps identify individual medical devices sold in the United States from manufacturing through distribution to patient use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified, and problems potentially corrected more quickly.

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program.

Subscribe to Medical Device Safety and Recalls

Sign up to receive email updates on medical device recalls, safety communications, and other safety information.
Back to Top