Update on Alert: Endoscope Accessories Forceps/Irrigation Plug Issue from Olympus

This communication is part of the Communications Pilot to Enhance the Medical Device Recall Program. This recall involves removing certain devices from where they are used or sold. The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it. The affected products and recommendations for what to do with the devices below have not changed. 

Affected Product

MAJ-891 Forceps/Irrigation Plug

The FDA is aware that Olympus has issued a letter to affected health care providers recommending the forceps/irrigation plug accessory to certain endoscopes be removed from use:

  • MAJ-891 Forceps/Irrigation Plug (Isolated Type)
  • All lots
  • UDI: 04953170063114

Full Listing of Olympus Endoscopes compatible with the MAJ-891 and Compatible Alternative Devices to MAJ-891
Scope SeriesModel NumberModel NameCompatible with MAJ-2092Compatible with BPS-Y
Cystoscope 
(CYF Series)
CYF-240CYF-240 Flexible Video CystoNephroscopeNoYes
Cystoscope 
(CYF Series)
CYF-5CYF-5 CYSTOSCOPE EVOLUTIONTIYesYes
Cystoscope 
(CYF Series)
CYF-5AN3627870 CYF-5A CYSTOSCOPE EYesYes
Cystoscope 
(CYF Series)
CYF-5RCYF-5R Flex CystoNephro Fiberscope ReverYesYes
Cystoscope 
(CYF Series)
CYF-V2CYF-V2 VISERA Cysto-Nephro vYesYes
Cystoscope 
(CYF Series)
CYF-V2RCYF-V2R Flex CystoNephro Videoscope RevYesYes
Cystoscope 
(CYF Series)
CYF-VA2CYF-VA2 VISERA Cysto-Nephro VideoscopeYesYes
Cystoscope 
(CYF Series)
CYF-VHCYF-VH HD Flex CystoNephro VideoscopeYesYes
Cystoscope 
(CYF Series)
CYF-VHACYF-VHA HD FlexCystoNephroVideoscopeYesYes
Cystoscope 
(CYF Series)
CYF-VHRCYF-VHR HD FlexCystoNephroVideoscopeYesYes
Ureterscope 
(URF Series)
URF-P5N3627930 OLA URF-P5 URF-P5 FLEXIBLE URETYesYes
Ureterscope 
(URF Series)
URF-P6URF-P6 SUPER-SLIM FLEXIBLE FIBEROPTIC URYesYes
Ureterscope 
(URF Series)
URF-P6RURF-P6R SUPER-SLIM FLEXIBLE FIBEROPTIC UYesYes
Ureterscope 
(URF Series)
URF-P7URF-P7 Flex Uretero FiberscopeYesYes
Ureterscope 
(URF Series)
URF-P7RURF-P7R Flex Uretero FiberscopeYesYes
Ureterscope 
(URF Series)
URF-VURF-V EndoEYE Flexible UreteoscopeYesYes
Ureterscope 
(URF Series)
URF-V2URF-V2 SLIM FLEXIBLE VIDEO URETEROSCOPEYesYes
Ureterscope 
(URF Series)
URF-V2RURF-V2R SLIM FLEX VID URETERO SCOPE, REVYesYes
Ureterscope 
(URF Series)
URF-V3URF-V3 URETEROSCOPE, FLEX, VIDEO, STDYesYes
Ureterscope 
(URF Series)
URF-V3RURF-V3R URETEROSCOPE, FLEX, VIDEO, REVYesYes
Choledochoscope 
(CHF Series)
CHF-P20QCHF-P20Q CHOLEDOCHOINEPHRO CYS TOFIBERSCOPENoNo
Choledochoscope 
(CHF Series)
CHF-CB30LCHF-CB30L TRANSLAPAROSCOPIC CHOLEDOCHOSCOPENoNo
Choledochoscope 
(CHF Series)
CHF-BP30CHF-BP30 OES CHOLEDOCHOFIBERNoNo
Choledochoscope 
(CHF Series)
CHF-P60CHF-P60 OES CHOLEDOCHOSCOPENoNo
Hysteroscope 
(HYF Series)
HYF-1TOES HYSTEROFIBERSCOPE OLYMPUS HYF TYPE 1TNoNo

What to Do

  • On December 18, 2024, Olympus sent all affected customers an Urgent: Medical Device Advisory Notice recommending the following actions:
    • Due to the risk of infection that may result from improper MAJ-891 reprocessing, alternatives to the MAJ-891 should be used instead.
    • Available alternative devices to the MAJ-891 for Olympus cystoscopes (CYF series) and ureteroscopes (URF series) include:
      • Luer-Split model MAJ-2092 + Seal-Port
      • Adjustable Biopsy Port Seal Y-Adapter model BPS-Y

        A full listing of Olympus Endoscopes compatible with the MAJ-891 and Compatible Alternative Devices to MAJ-891 is shown above.

    • There are currently no alternative Olympus irrigation plugs for use with CHF and HYF endoscopes. Use an alternative scope in those situations or a non-Olympus irrigation plug provided it has been validated for use with the Olympus scope by the plug manufacturer.
    • If no alternative for the MAJ-891 is available, it is important to closely follow the reprocessing instructions for both the endoscope(s) and the MAJ-891 Forceps/Irrigation Plug, especially detaching the MAJ-891 from the instrument channel port of the endoscope and disassembling before it is cleaned, disinfected or sterilized (see image provided under the “Affected Product” section). The reprocessing instructions for the MAJ-891 can be found in Chapters 5 through 7 of the MAJ-891 Instructions for Use (IFU).
    • Users should also inspect the biopsy valve (MAJ-579) (see image provided under the “Affected Product” section) for damage or deformation and replace the biopsy valve if damage is identified, as instructed in the IFU. A damaged or deformed biopsy valve may impact the reprocessing efficacy.

Reason for Alert

Patient infection as a result of exposure to a contaminated device can occur when reprocessing of the MAJ-891 Forceps/Irrigation Plug is improper and/or incomplete, such as not disconnecting the MAJ-891 from the endoscope and disassembling it before reprocessing. This exposure could result in patient injury including infection, urinary tract infection, or sepsis, and, in some cases, could result in death. These harms may require inpatient hospitalization/monitoring, and treatment with oral or intravenous antibiotics.

Olympus has reported 120 injuries and 1 report of death due to infection following procedures in which the MAJ-891 was used with a cystoscope (CYF scope).

Device Use

The MAJ-891 is endoscope accessory attached to the instrument channel port of certain Olympus endoscopes, including cystoscopes (CYF series), ureteroscopes (URF series), choledochoscopes (CHF series), and hysteroscopes (HYF series) to allow both irrigation and the use of endo-therapy accessories. The MAJ-891 was discontinued in 2022 from the US market.

Contact Information

Customers in the U.S. with adverse reactions, quality problems, or questions about this recall should contact Olympus’s Technical Assistance Center at 1-800-848-9024, option 1.

Additional FDA Resources

Unique Device Identifier (UDI)

The unique device identifier (UDI) helps identify individual medical devices sold in the United States from distribution to use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified more quickly, and as a result, problems potentially resolved more quickly.

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program.

Subscribe to Medical Device Safety and Recalls

Sign up to receive email updates on medical device recalls, safety communications, and other safety information.
Back to Top