Recall Alert

An update on the Cyclospora outbreak available here.

Insulin Pump Recall: Insulet Removes Certain Omnipod 5 Pods

The affected product lots and information about the issue below have changed. This recall involves removing certain devices from where they are used or sold. The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it.

Affected Product

Image shows box of Omnipod 5 Pods.

What to Do

Do not use affected Pods. If you do not have enough Pods due to this issue, talk to your health care provider about other methods of insulin delivery.

On March 12, Insulet sent all affected customers a letter recommending the following actions:

  • Do not use pods from affected lots.  
  • Check the lot number on the Pod tray lid, the Pod box or the Pod itself to determine if your Pods are affected. For more information on how to find the lot number, go to Appendix A on this page: https://www.omnipod.com/mdc-3-26
  • Discontinue use of any impacted Pod immediately.  If the Pod you are currently using is from an affected lot, it is important that you immediately change your Pod to resume insulin delivery.  When changing your Pod, confirm that the new Pod is not from an affected lot.  
  • If you do not have enough Pods due to this issue, talk with your health care provider to find other methods of insulin delivery while waiting for Pods to be replaced. Always follow your health care provider's guidance on appropriate glucose monitoring. 
  • Always check the expiration date before using a Pod. Do not use any Pod that is past its expiration date, as expired Pods may not function as intended.  
  • Visit omnipod.com/check-pods to acknowledge receipt and check to see if your lot is affected. 
  • Contact Insulet to request replacement Pods and return unused Pods from affected lots or to receive additional support by:  
    • Utilizing Insulet’s live agent chat by clicking the Podder Support button in the lower right of this page: https://www.omnipod.com/current-podders  
    • Calling Insulet’s Product Support phone line at 1-800-641-2049, available 24/7

On April 10, 2026, Insulet sent all affected customers an updated letter expanding the list of affected Pod lots to include Pods that have expired and providing additional details on this issue to clarify that the Pod defect may not be detectable by users. Insulet is also sending targeted communications to users who activate a Pod from an impacted lot with information on Pod discontinuation.

Reason for Recall

Insulet stated that Pods from certain lots may have a small tear in the internal tubing that delivers insulin. If this happens, insulin may leak inside the Pod instead of being fully infused in the body as intended, potentially leading to under-delivery of insulin.  

If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), a serious medical condition that requires prompt medical treatment and can be life-threatening if not treated. The risk of under-delivery increases if users apply more than one affected Pod in a row.

If there is a fluid leak inside the Pod, users may receive a hazard alarm notifying them to remove the Pod.  In some cases, there may be no alarm or alert, and the issue may go unnoticed, potentially leading to under-delivery of insulin with the potential for prolonged high blood glucose levels. If this occurs, the user’s blood glucose levels may increase and may not respond as expected to additional insulin delivery. This could result in an Automated Delivery Restriction alert advising users to troubleshoot.  

Because this problem may happen without triggering an alarm or alert, users could receive less insulin than needed without realizing it. For this reason, affected users should not rely only on alarms to know if something is wrong. If users have Pods from an affected lot, they should switch to a Pod from a nonaffected lot as soon as possible or talk with their healthcare provider about other options for insulin delivery.

This issue does not affect continuous glucose monitoring systems (CGM) or CGM readings.

As of April 17, Insulet has reported 29 serious injuries and no deaths associated with this issue.

Device Use

Omnipod is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.

Contact Information

Customers in the U.S. with adverse reactions, quality problems, or questions about this issue should contact Insulet at 1-800-641-2049 or visit https://www.omnipod.com/current-podders to reach a live agent chat.

Additional FDA Resources 

Additional Company Resources 

Company-provided information is posted here by the FDA as a public service.  

  1. Urgent Medical Device Correction | Omnipod  [03/12/2026] 
  2. Insulet Corporation - Insulet Initiates Voluntary Medical Device Correction for Certain Omnipod® 5 Pods in the U.S. [03/12/2026]

Unique Device Identifier (UDI)  

The unique device identifier (UDI) helps identify individual medical devices sold in the United States from manufacturing through distribution to patient use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified, and problems potentially corrected more quickly.

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program.

Timeline of Communication Updates

DateActions
04/30/2026The FDA updated this communication to correct the number of serious injuries reported by Insulet based on new information from the firm, and to clarify the inclusion of expired affected lots.  No additional lots were added with this update.
04/29/2026The FDA updated this communication to inform the public to include expired lots communicated by Insulet to customers on April 10th, provide updated information from the firm, and confirm that this issue has been classified as a Class I Recall. No additional lots were added with this update beyond what Insulet communicated to customers about expired lots on April 10th.
03/18/2026The FDA issued an Early Alert communication to notify the public of a potentially high-risk device issue.

Subscribe to Medical Device Safety and Recalls

Sign up to receive email updates on medical device recalls, safety communications, and other safety information.
Back to Top