  5. Early Alert: Catheter Introducer Issue from Abiomed and Oscor
  1. Medical Device Recalls and Early Alerts

Early Alert: Catheter Introducer Issue from Abiomed and Oscor

CDRH is issuing this Early Alert to notify the public of a potentially high-risk device issue. The FDA will keep the public informed and update this web page as significant new information becomes available.

Affected Product

The FDA is aware that Abiomed and its manufacturing partner Oscor have issued a letter to affected customers recommending all Introducer Kits used with Impella heart pumps be corrected prior to continued use.

Affected product contained in Impella Pump sets:

Introducer Product CodeIntroducer Product DescriptionPump Set Product CodeGTINPump Set Description
2000342Kit, 23Fr Introducer, 11cm,   Sterile100032300813502012811Impella RP Flex with SmartAssist   Set
0052-3021Introducer Kit, 23Fr0046-003500813502011869Impella RP Set with SmartAssist
0052-3006Axillary Insertion Introducer100010000813502012828Impella 5.5 Set with SmartAssist
0052-3046Kit, 14Fr Introducer, 13cm&25cm, Sterile0048-000300813502012279
00813502011876
Impella CP Set

Affected product sold in individual packaging:

Introducer   Product CodeIntroducer Product DescriptionSingle Sale Product CodeGTINSingle Sale Description
2000342Kit, 23Fr Introducer, 11cm, Sterile100044100813502012910Introducer Kit
23 Fr, 11 cm
0052-3021Introducer Kit, 23Fr0052-000200813502011319Introducer Kit 23 Fr for Impella RP
0052-3006Axillary Insertion Introducer0052-001100813502010015Axillary Insertion Kit
0052-3046Kit, 14Fr Introducer, 13cm&25cm, Sterile100055000813502013474Impella CP Introducer
Kit 14 Fr, 13 and 25 cm
0052-3052Kit, 14Fr Introducer, 13cm, Long Taper0052-003800813502011708Impella CP Introducer
Kit 14 Fr, 13 cm
0052-3053Kit, 14Fr Introducer, 25cm, Long Taper0052-003900813502011715Impella CP Introducer
Kit 14 Fr, 25 cm

What to Do

Hospitals may continue to use existing inventory. Consider a device exchange or using the repositioning sheath to minimize blood loss if potential leakage is observed.

On June 10, 2026, Abiomed and Oscor sent all affected customers a letter recommending the following actions:

  • Hospitals may continue to use existing inventory. Physicians should continue to use appropriate clinical awareness, monitoring, and patient support as needed. 
  • Consider a device exchange or using the repositioning sheath to minimize blood loss if potential leakage from the sidearm and under the sheath cap and along the hub score lines in the hub region is observed. 
  • Forward the firm’s notice to anyone in your facility who manages, transports, stores, stocks, or uses the subject products
  • If any of the subject products have been forwarded to another facility, contact that facility and provide them with the firm’s notice.
  • Post a copy of the firm’s notice in a visible area for awareness.

Check this web page for updates. The FDA is currently reviewing information about this potentially high-risk device issue and will keep the public informed as significant new information becomes available.

Reason for Alert

Abiomed has identified a potential for introducer sheath leakage in 14Fr and 23Fr Introducers from the sidearm, under the sheath cap, and along the hub score lines in the hub region. The unexpected leakage in these locations is due to certain manufacturing issues that increase the risk of access-site bleeding. 

If this issue were to occur, the user would see blood loss at the sidearm and hub of the introducer sheath as shown below on the 14Fr introducers. Leak locations are the same for both 14Fr and 23Fr introducers.

Catheter Introducer

Introducer leakage may result in access site bleeding at the time of sheath insertion or early device manipulation. This may cause blood loss requiring medical intervention, such as device exchange or removal, manual compression, and, in some cases, blood transfusion. 

In certain circumstances, particularly patients undergoing large bore vascular access with systemic anticoagulation or limited physiologic reserve (with a bleed that went undetected), exposure to this hazard may result in life-threatening hemorrhage and hemodynamic instability which may lead to death.

As of April 22, Abiomed has reported eight instances of major bleeding and three complaints that have corresponding patient deaths. Abiomed determined that the deaths are not attributable to introducer leakage.

Device Use

The Abiomed Introducer is intended to introduce catheters to the body.

Contact Information

Customers in the U.S. with adverse reactions, quality problems, or questions about this issue should contact Abiomed at OneMD-Field-Actions@its.jnj.com.

Unique Device Identifier (UDI)  

The unique device identifier (UDI) helps identify individual medical devices sold in the United States from manufacturing through distribution to patient use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified, and problems potentially corrected more quickly.

How do I report a problem?  

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program.