Correction Alert: Medline Industries, LP Updates Use Instructions for Medline Kits Containing Flexicare BritePro Solo Fiber Optic Laryngoscope Handles

This recall involves updating instructions for using certain devices and does not involve removing them from where they are used or sold. The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it without following the updated instructions.

Affected Products

This recall contains Medline medical convenience kits containing the Flexicare BritePro Solo, including convenience kits labeled as:

Device NameKit NumberUDI-DILots
NEWBORN RESUSCITATION KITACC010381C10193489922127 (EA), 40193489922128 (CS)21LDA450, 21KDA979, 21JDB271, 21LDA608
DRAWER 5 - POSITION 3ACC01053210193489902358 (EA), 40193489902359 (CS)21FDA234
DRAWER #4C ADULTACC01067110195327351090 (EA), 40195327351091 (CS)23EDC536
DRAWER #2ACC010728UDI/DI 10195327506544 (EA), 40195327506545 (CS)24EDC276, 23KDC368, 23KDB032
ADULT INTUBATION TRAYACC010369B10193489777512 (EA) 40193489777513(CA)21DDA738, 21CDA614, 21BDA717
ADULT ICU INTUBATION TRAYACC010379A10193489777529 (EA) 40193489777520(CA)21BDB738, 21BDB493
INTUBATION TRAYACC01039210193489218053 (EA) 40193489218054(CA)24JDA206, 21HDB930, 21HDA284, 21EDB901, 21EDB869, 21EDB805
INTUBATION KITACC010467A10193489834550 (EA) 40193489834551(CA)21EDA887
ADULT AIRWAY TRAYACC01046910193489414677 (EA) 40193489414678(CA)22BDA430, 21KDA741, 21CDA522, 21BDA753
TRAY 2 INTUBATION DRAWERACC01048010193489499933 (EA) 40193489499934(CA)21HDA876, 21FDB951, 21FDB019, 21EDA065, 21CDB872, 21BDB094
TRAY 2 INTUBATION DRAWERACC010480A10195327009366 (EA) 40195327009367(CA)21JDB163
ADULT INTUBATIONACC01054010193489977547 (EA) 40193489977548(CA)21IDC276
DIFFICULT AIRWAY TRAY #1ACC010593A10195327168995 (EA) 40195327168996(CA)22EDB290
DIFFICULT AIRWAY TRAY #1ACC010593B10195327195472 (EA) 40195327195473(CA)22FDB947
INTUBATION TRAYDYNDA1847A10193489481259 (EA) 40193489481250(CA)21FBG208, 21CBL507
INTUBATION TRAY (ADULT)DYNJAA245A10193489804584 (EA) 40193489804585(CA)21LBV006, 21IBV349, 21HBT154, 21DBI703, 21CBN261

What to Do

Place the over-labels provided by Medline Industries on affected kits to indicate that the affected component should be removed and discarded prior to using the kit.

On July 25, 2025, Medline Industries, LP sent all affected customers an immediate action required medical device recall recommending the following actions:  

  • Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 
  • Please complete the recall response form. Please list the quantity of affected product in inventory on the form. Even if there is no affected product in inventory, please complete and submit the form. 
  • Upon receipt of the submitted response form, over-labels will be provided to be placed on affected inventory, with instructions for staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 
  • Alert any person or company to whom this kit may have been resold or transferred of this recall communication. Please include customers’ quantities on the response form.  Check this web page for updates. The FDA is currently reviewing information about this potentially high-risk device issue and will keep the public informed as significant new information becomes available.

Reason for Updated Use Instructions

Medline Industries, LP is updating the use instructions for their medical convenience kits containing the previously recalled component Flexicare BritePro Solo due to the light not illuminating as intended. If the light malfunctions, it may prevent effective illumination of the patient airway. 

The use of affected product may cause serious adverse health consequences, including delay in treatment or requiring intubation using a different device and possibly death.   

As of August 8, 2005, Medline Industries has not reported any serious injuries or deaths associated with this issue. 

Device Use

This recall is for Medline medical convenience kits containing the Flexicare BritePro Solo. The BritePro Solo devices are intended for direct vision laryngoscopy to obtain an illuminated view of the vocal cords and the glottis by inserting the laryngoscope blade into the oral cavity to manipulate the tongue or epiglottis.

Contact Information

Customers in the U.S. with adverse reactions, quality problems, or questions about this recall should contact the Medline Recall Department at 866-359-1704 or recalls@medline.com.

Additional FDA Resources:  

Additional Company Resources: 

Company provided information on a recall, is posted here by the FDA as a public service.  

Unique Device Identifier (UDI)  

The unique device identifier (UDI) helps identify individual medical devices sold in the United States from manufacturing through distribution to patient use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified, and problems potentially corrected more quickly.

How do I report a problem?  

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program.