1. Home
  2. Medical Devices
  3. Medical Device Safety
  4. Medical Device Recalls and Early Alerts
  5. Alert: Siemens Healthineers Issues Correction for 3 Tesla MRI Systems
  1. Medical Device Recalls and Early Alerts

Alert: Siemens Healthineers Issues Correction for 3 Tesla MRI Systems

This recall involves correcting certain devices and does not involve removing them from where they are used or sold. The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it without correction.  

Affected Products

Product Name 

Material Number

UDI-DI

MAGNETOM Verio10276755 N/A
MAGNETOM Skyra (DE)1043291504056869006703
Biograph mMR 10433372 04056869006741
MAGNETOM Verio Dot 1068433304056869006673
MAGNETOM Prisma 10849582 04056869006727
MAGNETOM Skyra fit  1084958004056869006710
MAGNETOM Verio Dot Upgr.1068433404056869006680
MAGNETOM Vida (DE) 11060815 04056869039176
MAGNETOM Skyra (CN) 11410327  04056869227139
MAGNETOM Lumina (DE)1134491604056869230740
MAGNETOM Vida Fit 1141048104056869245188
MAGNETOM Vida (CN)1151615204056869260143
MAGNETOM Lumina (CN)11516153 04056869260136
MAGNETOM Skyra Fit BioMatrix1151621704056869299273
MAGNETOM Cima.X Upgr. (DE)  1168930404056869975245
MAGNETOM Cima.X (DE) 1164715804056869975221
MAGNETOM Connectom.X11371480N/A
BIOGRAPH One (DE) 11689172  04068151020472
MAGNETOM Cima.X Fit  1164715904056869975238
MAGNETOM Cima.X (CN) 11689616 04056869996509

What to Do

If the error message “Magnet Supervision: The average heater power is above the alarm threshold. Please contact Siemens Healthineers Service” appears on your host computer, no further scans can be performed. Follow additional safety actions immediately. 

On August 27, 2025, Siemens Healthineers sent all affected customers an Urgent Medical Device Correction letter recommending the following actions:

  • If the error message “Magnet Supervision: The average heater power is above the alarm threshold. Please contact Siemens Healthineers Service” appears on the host computer, take the following safety actions:
    • Do not reboot the system. 
    • Remove the patient from the system. 
    • Lock/secure all access to the scanning room with the system remaining energized at field. 
    • Post a “DO NOT USE” notice on or near the device.
    • Call the facility’s local Siemens Healthineers service organization.
    • Make sure all relevant personnel are aware of this issue.
  • Be aware that the Magnet Stop is only intended for emergencies. DO NOT initiate a manual quench of the magnet unless there is an emergency.
  • Ensure all designated emergency exit routes remain unobstructed at all times and are properly marked. In particular, the exit from the examination room must be maintained at all times.
  • Grant access to the system so an inspection can be carried out as directed by Siemens Healthineers service organization. 
  • Make sure all users of affected products and others who may need to be informed receive the relevant safety information provided within the notice and comply with its recommendations.
  • If the equipment is no longer at the facility, forward the notice to the equipment’s new owner. 
  • You may continue to use the system until an inspection has been performed.

Reason for Correction

Siemens Healthineers is correcting 3 Tesla Siemens Healthineers MRI systems due to the potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a sudden loss of superconductivity, or magnet quench, helium gas may be unable to escape through designated venting paths, leading to pressure build-up within the helium containment system. This pressure build up could rupture the helium containment system, potentially resulting in a massive helium leak into the scanning room. Rupture of the helium containment system may also manifest as an explosion, with overpressure in the MRI room and debris expelled forcefully.

The use of affected product may cause serious adverse health consequences, including cold burns, asphyxiation, trauma, and death.  

As of August 29, Siemens Healthineers has not reported any serious injuries or deaths related to this issue.

Device Use

Affected 3 Tesla Siemens Healthineers MRI systems include magnetic resonance diagnostic devices (MRDD) and mMR systems, which combine MRDD and Positron Emission Tomography (PET). These devices display the internal structure and/or function of the head, body, or extremities, and may be utilized by appropriately trained health care professionals to aid in the detection, localization, and diagnosis of diseases and disorders.

Contact Information

Customers in the U.S. with questions about this recall should contact the Siemens Healthineers service organization at 1-800-888-7436.

Additional FDA Resources:  

Unique Device Identifier (UDI)  

The unique device identifier (UDI) helps identify individual medical devices sold in the United States from manufacturing through distribution to patient use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified, and problems potentially corrected more quickly.  

How do I report a problem?  

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program. 

Subscribe to Medical Device Safety and Recalls 

Sign up to receive email updates on medical device recalls, safety communications, and other safety information.  
Back to Top