Recalls, Market Withdrawals, & Safety Alerts
The list below provides information gathered from press releases and other public notices about certain recalls of FDA-regulated products. Not all recalls have press releases or are posted on this page. See Additional information about recalls for a more complete listing.
The Recalls, Market Withdrawals & Safety Alerts are available on FDA’s website for three years before being archived. To search archived content, visit Search FDA Archive and input the name of the product and/or company name in the Search terms box as well as the year to get the most inclusive search results. To scroll through archived Recalls, Market Withdrawals & Safety Alerts content by year, see the Recall and Safety Alerts Archive.
* FDA Employees: Use Chrome to Subscribe until further notice.
| Date | Brand Name(s) | Product Description | Product Type | Recall Reason Description | Company Name | Terminated Recall | Excerpt |
|---|---|---|---|---|---|---|---|
| Aligned Medical Angio Pack | Angio Packs (AMS6908E and AMS6908F) | Medical Devices | Risk of the syringe rotating adapter unwinding during use, which may result in a loose connection and/or full disconnection between the syringes and manifold, potentially causing biohazard exposure, blood loss, infection, or air embolism. | Windstone Medical Packaging dba Aligned Medical Solutions | … Aligned Medical Angio Pack … Windstone Medical Packaging dba Aligned Medical Solutions … On April 02, 2026, … |