2024 Scorecard Statistics
- December 2024
- November 2024
- October 2024
- September 2024
- August 2024
- July 2024
- June 2024
- May 2024
- April 2024
- March 2024
- February 2024
- January 2024
December 2024
| Certified facilities, as of October 1, 2024 | 8,931 | ||
| Certification statistics, as of December 1, 2024 | |||
| Total certified facilities / Total accredited units | 8,959 / 26,392 | ||
| Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,958 / 13,699 | ||
| Certified facilities with DBT units3,4/ Accredited DBT units | 8,210 / 12,693 | ||
| FY 2024 inspection statistics, as of December 1, 2024 | |||
| Facilities inspected | 1,413 | ||
| Total units at inspected facilities | 4,019 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 noncompliance | 1.1% | ||
| Level 2 noncompliance | 21.2% | ||
| Percent of inspections with no noncompliance | 77.7% | ||
| Total annual mammography procedures reported, as of December 1, 20241 | 42,682,819 | ||
November 2024
| Certified facilities, as of October 1, 2024 | 8,931 | ||
| Certification statistics, as of November 1, 2024 | |||
| Total certified facilities / Total accredited units | 8,939 / 26,315 | ||
| Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,938 / 13,672 | ||
| Certified facilities with DBT units3,4/ Accredited DBT units | 8,174 / 12,643 | ||
| FY 2024 inspection statistics, as of November 1, 2024 | |||
| Facilities inspected | 676 | ||
| Total units at inspected facilities | 19,000 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 noncompliance | 1% | ||
| Level 2 noncompliance | 20% | ||
| Percent of inspections with no noncompliance | 79% | ||
| Total annual mammography procedures reported, as of November 1, 20241 | 42,646,438 | ||
October 2024
| Certified facilities, as of October 1, 2023 | 8,834 | ||
| Certification statistics, as of October 1, 2024 | |||
| Total certified facilities / Total accredited units | 8,931 / 26,193 | ||
| Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,931 / 13,625 | ||
| Certified facilities with DBT units3,4/ Accredited DBT units | 8,142 / 12,568 | ||
| FY 2024 inspection statistics, as of October 1, 2024 | |||
| Facilities inspected | 8,531 | ||
| Total units at inspected facilities | 24,512 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 noncompliance | 1.3% | ||
| Level 2 noncompliance | 11.7% | ||
| Percent of inspections with no noncompliance | 87% | ||
| Total annual mammography procedures reported, as of October 1, 20241 | 42,440,214 | ||
September 2024
| Certified facilities, as of October 1, 2023 | 8,834 | ||
| Certification statistics, as of September 1, 2024 | |||
| Total certified facilities / Total accredited units | 8,919 / 26,117 | ||
| Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,919 / 13,615 | ||
| Certified facilities with DBT units3,4/ Accredited DBT units | 8,107 / 12,502 | ||
| FY 2024 inspection statistics, as of September 1, 2024 | |||
| Facilities inspected | 7,912 | ||
| Total units at inspected facilities | 22,688 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 noncompliance | 1.4% | ||
| Level 2 noncompliance | 11.3% | ||
| Percent of inspections with no noncompliance | 87.3% | ||
| Total annual mammography procedures reported, as of September 1, 20241 | 42,232,409 | ||
August 2024
| Certified facilities, as of October 1, 2023 | 8,834 | ||
| Certification statistics, as of August 1, 2024 | |||
| Total certified facilities / Total accredited units | 8,915 / 26,045 | ||
| Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,915 / 13,603 | ||
| Certified facilities with DBT units3,4/ Accredited DBT units | 8,070 / 12,442 | ||
| FY 2024 inspection statistics, as of August 1, 2024 | |||
| Facilities inspected | 6,882 | ||
| Total units at inspected facilities | 19,750 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 1.5% | ||
| Level 2 violation | 11.4% | ||
| Percent of inspections with no violation | 87.2% | ||
| Total annual mammography procedures reported, as of August 1, 20241 | 42,081,655 | ||
July 2024
| Certified facilities, as of October 1, 2023 | 8,834 | ||
| Certification statistics, as of July 1, 2024 | |||
| Total certified facilities / Total accredited units | 8,894 / 25,971 | ||
| Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,894 / 13,585 | ||
| Certified facilities with DBT units3,4/ Accredited DBT units | 8,039 / 12,386 | ||
| FY 2024 inspection statistics, as of July 1, 2024 | |||
| Facilities inspected | 6,198 | ||
| Total units at inspected facilities | 17,873 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 1.5% | ||
| Level 2 violation | 11.3% | ||
| Percent of inspections with no violation | 87.2% | ||
| Total annual mammography procedures reported, as of July 1, 20241 | 41,836,128 | ||
June 2024
| Certified facilities, as of October 1, 2023 | 8,834 | ||
| Certification statistics, as of June 1, 2024 | |||
| Total certified facilities / Total accredited units | 8,872 / 25,866 | ||
| Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,872 / 13,552 | ||
| Certified facilities with DBT units3,4/ Accredited DBT units | 7,981 / 12,314 | ||
| FY 2024 inspection statistics, as of June 1, 2024 | |||
| Facilities inspected | 5,549 | ||
| Total units at inspected facilities | 15,947 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 1.6% | ||
| Level 2 violation | 11.1% | ||
| Percent of inspections with no violation | 87.3% | ||
| Total annual mammography procedures reported, as of June 1, 20241 | 41,725,229 | ||
May 2024
| Certified facilities, as of October 1, 2023 | 8,834 | ||
| Certification statistics, as of May 1, 2024 | |||
| Total certified facilities / Total accredited units | 8,868 / 25,757 | ||
| Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,867 / 13,515 | ||
| Certified facilities with DBT units3,4/ Accredited DBT units | 7,955 / 12,242 | ||
| FY 2024 inspection statistics, as of May 1, 2024 | |||
| Facilities inspected | 4,845 | ||
| Total units at inspected facilities | 13,920 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 1.5% | ||
| Level 2 violation | 11.0% | ||
| Percent of inspections with no violation | 87.5% | ||
| Total annual mammography procedures reported, as of May 1, 20241 | 41,497,527 | ||
April 2024
| Certified facilities, as of October 1, 2023 | 8,834 | ||
| Certification statistics, as of April 1, 2024 | |||
| Total certified facilities / Total accredited units | 8,854 / 25,639 | ||
| Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,852 / 13,474 | ||
| Certified facilities with DBT units3,4/ Accredited DBT units | 7,926 / 12,165 | ||
| FY 2024 inspection statistics, as of March 1, 2024 | |||
| Facilities inspected | 4,215 | ||
| Total units at inspected facilities | 12,080 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 1.6% | ||
| Level 2 violation | 10.7% | ||
| Percent of inspections with no violation | 87.6% | ||
| Total annual mammography procedures reported, as of March 1, 20241 | 41,350,402 | ||
March 2024
| Certified facilities, as of October 1, 2023 | 8,834 | ||
| Certification statistics, as of March 1, 2024 | |||
| Total certified facilities / Total accredited units | 8,852 / 25,585 | ||
| Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,850 / 13,474 | ||
| Certified facilities with DBT units3,4/ Accredited DBT units | 7,901 / 12,111 | ||
| FY 2024 inspection statistics, as of March 1, 2024 | |||
| Facilities inspected | 3,446 | ||
| Total units at inspected facilities | 9,846 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 1.8% | ||
| Level 2 violation | 10.9% | ||
| Percent of inspections with no violation | 87.3% | ||
| Total annual mammography procedures reported, as of March 1, 20241 | 41,040,355 | ||
February 2024
| Certified facilities, as of October 1, 2023 | 8,834 | ||
| Certification statistics, as of February 1, 2024 | |||
| Total certified facilities / Total accredited units | 8,862 / 25,547 | ||
| Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,861 / 13,471 | ||
| Certified facilities with DBT units3,4/ Accredited DBT units | 7,900 / 12,076 | ||
| FY 2024 inspection statistics, as of February 1, 2024 | |||
| Facilities inspected | 2,740 | ||
| Total units at inspected facilities | 7,748 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 2% | ||
| Level 2 violation | 10.3% | ||
| Percent of inspections with no violation | 87.8% | ||
| Total annual mammography procedures reported, as of February 1, 20241 | 40,869,997 | ||
January 2024
| Certified facilities, as of October 1, 2023 | 8,834 | ||
| Certification statistics, as of January 1, 2024 | |||
| Total certified facilities / Total accredited units | 8,858 / 25,439 | ||
| Certified facilities with 2D digital units2 / Accredited 2D digital units | 8,857 / 13,443 | ||
| Certified facilities with DBT units3,4/ Accredited DBT units | 7,865 / 11,996 | ||
| FY 2024 inspection statistics, as of January 1, 2024 | |||
| Facilities inspected | 2,115 | ||
| Total units at inspected facilities | 5,910 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 2.2% | ||
| Level 2 violation | 10.1% | ||
| Percent of inspections with no violation | 87.7% | ||
| Total annual mammography procedures reported, as of January 1, 20241 | 40,533,714 | ||
1 This number is an aggregate of the total number of procedures performed annually as reported by facilities to their accreditation bodies at the time of their re-accreditation, which takes place once every three years. We have aggregated only the numbers reported by MQSA-certified, non-Veterans Hospital Administration facilities. The aggregate number may not reflect the current number of procedures performed at these facilities.
2Includes Full Field Digital Mammography and Computed Radiography Units.
3DBT - Digital Breast Tomosynthesis.
4Facilities with DBT also have FFDM so the DBT facilities count is included within the digital facilities count.