Inspection and Field Testing of Radiation-Emitting Electronic Products: Attachment B: Specific Instructions for Laser Product Inspections and Tests
Background
The Laser Products Performance Standard (the standard), promulgated in August 1976, was designed to protect the public from unnecessary radiation hazards associated with the use of these products. The radiation emitted from these laser products can pose varying degrees of hazards depending upon the type, magnitude, and accessibility of the radiation and upon the particular functions or operations they perform. The standard was last amended in 1985. Since then, the CDRH has intended to harmonize the requirements of the standard with those of the international standard IEC 60825-1: 2001. As an interim step the CDRH published its Laser Notice 50 in 2001 stating that it would not object to compliance with specified requirements of the international standards in lieu of comparable requirements of the CDRH standard.
Specific Instructions
High-risk laser products and their manufacturers should be inspected or tested as a priority. Examples of high-risk laser products and manufacturers include:
- Class IIIb and IV medical lasers (e.g. surgical)
- Class IIIb and IV industrial lasers used in material processing
- Class IIIb and IV lasers used in law enforcement or military applications
- Manufacturers with known or suspected problems based on previous inspection, field tests or complaints
- New manufacturers not yet inspected
- Manufacturers introducing new technology to the US market
- Manufacturers with a large potion of the US market share for any laser product. Class I low risk laser products, such as optical disk drives or laser printers, should not be inspected or tested.
Electro-optics specialists have been specifically trained in general EPRC requirements and also have specialized training in the laser product performance standards. EOS’s should perform these inspections and field tests, and may train additional field staff or accompany a medical device investigator to conduct joint EPRC/medical device inspections. If an EOS has training in both EPRC and QSIT inspections, a single EOS may conduct both portions of the inspection.
CDRH is responsible for review of laser manufacturer inspection and product field test observations and initiating administrative or regulatory follow-up.
References
Frequently Asked Questions about Lasers.
Performance Standard-Lasers and Products Incorporating Lasers
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/CFRSearch.cfm?FR=1040.10
Performance Standard-Specific Laser Products (Includes Display, Survey, and Medical Laser Products)
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/CFRSearch.cfm?FR=1040.11
Reporting Guide-Radiation Safety Product Report for Laser Products
Reporting Guide-Radiation Safety Product Report for Laser Light Shows/Displays
Laser Quality Control Guide
Refer to the laser products main page for guidance documents and additional information.
Laser Product Codes
| Translation of 2-Digit Code | Product Name | Product Code | CFR | Definition | |
|---|---|---|---|---|---|
| Other Laser Products | Automotive Accessory, Automobile or Transport Vehicle, Laser | 95 | RDV | 1040.10 | A laser product or product containing a laser that is an automotive or other transport vehicle accessory. |
| Other Laser Products | General Purpose Laser Products, Non-Medical | 95 | RDW | 1040.10 | Product, laser, general, emit beam; A laser or product containing a laser that is intended for general purpose use with no medical claims. |
| Other Laser Products | Other | 95 | RZZ | Unk | A laser or product containing a laser for which its intended use is not previously defined. |
| Laser Light Show/Display Products | Low-Power Laser Light Show Projector | 95 | RDZ | 1040.10; 1040.11 | Product, laser, demo, projector, laser light show/display, Class IIIa/3R and lower; Laser projection system that incorporates a laser having a maximum radiation output of less than 5mw. |
| Laser Light Show/Display Products | High-Power Laser Light Show Projector (Output > 5mW) | 95 | REA | 1040.10; 1040.11 | Product, laser, demo, projector, laser light show/display, Class IIIb/IV/3B/4; Laser projection system that incorporates a laser having a maximum radiation output power greater than 5mw. |
| Laser Light Show/Display Products | High-Power Laser Light Show | 95 | REB | 1040.10; 1040.11 | Product, laser, demo, laser light show/display, Class IIIb/IV/3B/4; Laser light show or demonstration using laser projection equipment having an output that exceeds 5mW |
| Laser Light Show/Display Products | Laser Video Projector | 95 | REC | 1040.10; 1040.11 | Product, laser, demo, projector, display, video; A laser used in conjunction or incorporated in a video display system or projector. |
| Laser Light Show/Display Products | Laser Advertising Display System | 95 | RED | 1040.10; 1040.11 | Product, laser, demo, system, display, advertising |
| Laser Light Show/Display Products | Laser Visual Display - Display Retinal Image, Non-Medical Display Product | 95 | REE | 1040.10; 1040.11 | Product, laser, display, system, images, direct to retina |
| Laser Light Show/Display Products | Other | 95 | RZZ | Unk | Other laser products used in light shows or demonstrations that are not otherwise defined. |
| Medical Laser Products | Laser, Ophthalmic | 86 | HQF | 1040.10; 1040.11 | |
| Medical Laser Products | Laser Instrument, Surgical, Powered | 79 | GEX | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, Surgical, Gynecologic | 85 | HHR | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, ENT Microsurgical Carbon-Dioxide | 77 | EWG | 1040.10; 1040.11 | |
| Medical Laser Products | Photocoagulator and Accessories | 86 | HQB | 1040.10; 1040.11 | |
| Medical Laser Products | Lens, Surgical, Laser, Accessory, Ophthalmic Laser | 86 | LQJ | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, Neurosurgical | 84 | LKW | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, Neurosurgical, Argon | 84 | LLF | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, Neodymium: YAG, Pulmonary Surgery | 73 | LLO | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, Neodymium: YAG, for Gynecologic Use | 85 | LLW | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, Neodymium: YAG, Ophthalmic for Uses Other than Posterior Capsulotomy and Cutting Pupil | 86 | LXS | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, Neodymium: YAG, Ophthalmic for Uses Other than Posterior Capsulotomy | 86 | LOI | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, Neodymium: YAG, Optical, Pumped Parametric Oscillator | 86 | MVQ | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, Microsurgical Argon, for Use in Otology | 77 | LXR | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, Microsurgical Argon, for Uses Other Than Otology | 77 | LMS | 1040.10; 1040.11 | |
| Medical Laser Products | Laser for Gastro-Urology Use | 78 | LNK | 1040.10; 1040.11 | |
| Medical Laser Products | Device, Angioplasty, Laser, Coronary | 74 | LPC | 1040.10; 1040.11 | |
| Medical Laser Products | Device, Laser Peripheral Angioplasty | 74 | LWX | 1040.10; 1040.11 | |
| Medical Laser Products | Catheter, Coronary Laser Myoplasty | 74 | MGC | 1040.10; 1040.11 | |
| Medical Laser Products | System, Laser, Transmyocardial Revascularization | 74 | MNO | 1040.10; 1040.11 | |
| Medical Laser Products | Instrument, Visual Field, Laser | 86 | HPJ | 1040.10; 1040.11 | |
| Medical Laser Products | Laser for Pain Therapy | 84 | LLP | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, System, Excimer | 86 | LZS | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, Dental | 76 | LYB | 1040.10; 1040.11 | |
| Medical Laser Products | Photodynamic Therapy (PDT) | 79 | MVF | 1040.10; 1040.11 | |
| Medical Laser Products | Photodynamic Therapy (PDT), Fiber Optic | 79 | MVG | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, Fluorescence Caries Detection | 76 | NBL | 1040.10; 1040.11 | |
| Medical Laser Products | Laser for Wound Healing | 79 | LXU | 1040.10; 1040.11 | |
| Medical Laser Products | Ophthalmoscope, Laser Scanner | 86 | MYC | 1040.10; 1040.11 | |
| Medical Laser Products | Laser, Phacolysis | 86 | MXO | 1040.10; 1040.11 | |
| Medical Laser Products | Caries Detector, Laser, Light, Transmission | 76 | NTK | 1040.10; 1040.11 | |
| Medical Laser Products | Other | 95 | RZZ | Unk | A laser or laser product intended for medical treatment or other uses on humans, not previously defined. |
| Other Demonstration Laser Products | Laser Science Education Products | 95 | REI | 1040.10; 1040.11 | Product, laser, demo, education, illustrate science principles |
| Other Demonstration Laser Products | Other | 95 | RZZ | Unk | Laser products used for demonstrations that are not otherwise defined. |
| Toy, Novelty, Play Laser Products | Toy, Novelty, Play Laser Product | 95 | REJ | 1040.10; 1040.11 | Product, laser, toy/novelty |
| Research, Scientific, Laboratory Laser Products | Research Laser, Scientific, Laboratory Laser Products | 95 | REK | 1040.10 | Product, laser, research/laboratory; A laser under development in and of itself. A laser used for conducting research during development of new data or to improve a process would not be considered a research laser although it is being used in research. |
| Research, Scientific, Laboratory Laser Products | Guide-Star Laser System, Research, Scientific, Laboratory Laser Products | 95 | REL | 1040.10 | Product, laser, adaptive-optics telescope focusing accessory, generate artificial star; A laser used for alignment of optical telescopes. |
| Research, Scientific, Laboratory Laser Products | Spectroscopy Instrument, Laser, Research, Scientific, Laboratory Laser Products | 95 | REM | 1040.10 | Product, laser, instrument, spectroscopy; An instrument incorporating a laser for spectroscopic testing or examination with no medical claims. |
| Research, Scientific, Laboratory Laser Products | Particle-Size Measuring Instrument, Laser, Scientific, Laboratory Laser Products | 95 | REN | 1040.10 | Product, laser, instrument, particle size measurement; An instrument or system incorporating a laser for determining the size or number of particles of particles a test sample. |
| Research, Scientific, Laboratory Laser Products | Analytical Measuring and Detection, Research, Scientific, Laboratory Laser Products | 95 | REO | 1040.10 | Product, laser, instrument, analyze/detect chemical species |
| Research, Scientific, Laboratory Laser Products | Other | 95 | RZZ | Unk | Laser products used in scientific and laboratory applications that are not otherwise defined. |
| Surveying, Leveling, Alignment Laser Products | Surveying Laser Product, Leveling, Alignment Laser Products | 95 | REP | 1040.10 | Product, laser, surveying, instrument, determine position by measurement of angles |
| Surveying, Leveling, Alignment Laser Products | Ranging (Geodimeter) Laser Products | 95 | REQ | 1040.10 | Product, laser, ranging, instrument, measure distance by time-of-flight |
| Surveying, Leveling, Alignment Laser Products | Alignment Laser Product, Surveying, Leveling, Alignment Laser Products | 95 | RER | 1040.10; 1040.11 | Product, laser, alignment, aid positioning or adjusting parts in relation to each other |
| Surveying, Leveling, Alignment Laser Products | Laser Pointer, Surveying, Leveling, Alignment Laser Products | 95 | RES | 1040.10; 1040.11 | Product, laser, pointer, indicate point of interest; A laser product intended specifically to define a spot or surface for drawing attention to a viewer. |
| Surveying, Leveling, Alignment Laser Products | Laser Target Designator, Surveying, Leveling, Alignment Laser Products | 95 | RET | 1040.10 | Product, laser, target designator; An optical devices, using a visible beam of laser light that permits the alignment of a gun, cannon or rocket system with its target. |
| Surveying, Leveling, Alignment Laser Products | Laser Aiming Product, Visible, Surveying, Leveling, Alignment Laser Products | 95 | REU | 1040.10; 1040.11 | Product, laser, aiming, visible, attached to weapon; An optical devices, using a visible beam of laser light that permits the alignment of a gun, cannon or rocket system with its target |
| Surveying, Leveling, Alignment Laser Products | Laser Aiming Product, Non-Visible, Surveying, Leveling, Alignment Laser Products | 95 | REV | 1040.10; 1040.11 | Product, laser, aiming, infrared, attached to weapon, viewed with night-vision equipment; An optical devices, using an invisible beam of laser light that permits the alignment of a gun, cannon or rocket system with its target. |
| Surveying, Leveling, Alignment Laser Products | Other | 95 | RZZ | Unk | Other laser products used for surveying, leveling and alignment that are not otherwise defined. |
| Safety, Security, Surveillance Laser Products | IR Laser Illuminator with Alignment Aid/Night Vision System, Safety, Security, Surveillance Laser Products | 95 | REW | 1040.10; 1040.11 | Product, laser, infrared, illuminator with alignment aid, viewed through night-vision equipment |
| Safety, Security, Surveillance Laser Products | IR Laser Illuminator Only/Night Vision System, Safety, Security, Surveillance Laser Products | 95 | REX | 1040.10 | Product, laser, infrared, illuminator only, viewed through night-vision equipment |
| Safety, Security, Surveillance Laser Products | Collision-Avoidance Laser System, Safety, Security, Surveillance Laser Products | 95 | REY | 1040.10 | Product, laser, infrared, collision-avoidance system |
| Safety, Security, Surveillance Laser Products | Laser Traffic Signal, Safety, Security, Surveillance Laser Products | 95 | REZ | 1040.10 | Product, laser, traffic signal/control |
| Safety, Security, Surveillance Laser Products | Laser Automotive Lighting & Signals, Safety, Security, Surveillance Laser Products | 95 | RFA | 1040.10 | Product, laser, automotive, lighting/signals |
| Safety, Security, Surveillance Laser Products | IR Laser Intrusion Detection/Security System, Safety, Security, Surveillance Laser Products | 95 | RFB | 1040.10 | Product, laser, infrared, intrusion detecting, security system |
| Safety, Security, Surveillance Laser Products | Laser Radar (Lidar) or Speed Measurement, Safety, Security, Surveillance Laser Products | 95 | RFC | 1040.10 | Product, laser, infrared, Doppler or time-of-flight speed measurement |
| Safety, Security, Surveillance Laser Products | Other | 95 | RZZ | Unk | Laser products used in safety, security, surveillance applications not otherwise defined |
| Safety, Security, Surveillance Laser Products | Laser Weapon (Military or Police), Safety, Security, Surveillance Laser Products | 95 | RFD | 1040.10 | Product, laser, weapon (military/police) |
| Material Processing Laser Products | Laser Cutter, Material Processing Laser Products | 95 | RFE | 1040.10 | A high power laser intended to cut or drill a variety of materials in an industrial or commercial environment. |
| Material Processing Laser Products | Laser Welder, Material Processing Laser Products | 95 | RFF | 1040.10 | A high power laser intended to weld (join) materials in an industrial or commercial environment. |
| Material Processing Laser Products | Microelectronic Mask or Chip Checking/Repair, Material Processing Laser Products | 95 | RFG | 1040.10 | A laser intended to inspect and/or repair microelectronic components in an industrial or commercial environment. |
| Material Processing Laser Products | UV Curing, Material Processing Laser Products | 95 | RFH | 1040.10 | An ultraviolet wavelength laser used to illuminate a material of a certain composition such that the laser "cures" or causes a chemical reaction to change the material in a desired fashion with no medical claims. Typical materials are adhesives, plastics, potting compounds, etc. |
| Material Processing Laser Products | Print Industry Plate Maker, Material Processing Laser Products | 95 | RFI | 1040.10 | A laser intended to etch, engrave or otherwise create printer's plates used in an industrial or commercial environment. |
| Material Processing Laser Products | Process Control, Material Processing Laser Products | 95 | RFJ | 1040.10 | A laser used for inspection, counting, or other application intended to monitor a part of the manufacturing process in an industrial or commercial environment. Often incorporated in an automated process system. |
| Material Processing Laser Products | Laser Vision, Material Processing | 95 | RFK | 1040.10 | A laser used for positioning, focusing, inspection, counting, or other application in an industrial or commercial environment. Often incorporated in an automated assembly line system. |
| Material Processing Laser Products | Laser Micrometer, Material Processing | 95 | RFL | 1040.10 | A laser used in high precision dimensional measurements in materials processing. |
| Material Processing Laser Products | Laser-Based Material Positioning System | 95 | RFM | 1040.10 | A laser used in precision positioning of materials in manufacturing in an industrial or commercial environment. |
| Material Processing Laser Products | Other | 95 | RZZ | Unk | A laser used in materials processing not otherwise defined. |
| Material Processing Laser Products | General Industrial Use Material Processing Laser Products | 95 | RZN | 1040.10 | A laser used in industrial manufacturing or materials processing not otherwise defined. |
| Data Measurement, Transmit, Control Laser Products | Fiber Optic Communication and Data Transfer, Laser | 95 | RFN | 1040.10 | A laser used in fiber optic communications to transmit data and information. |
| Data Measurement, Transmit, Control Laser Products | IR Free-Space Data Transmit/Control, Laser | 95 | RFO | 1040.10 | A laser used in free space (open air) communications to transmit data and information. |
| Data Measurement, Transmit, Control Laser Products | Remote Controller, Laser, Data Measurement, Transmit | 95 | RFP | 1040.10 | A laser used to transmit signals and/or information in order to operate equipment or machinery remotely. |
| Data Measurement, Transmit, Control Laser Products | Interferometric Position Measuring Product, Laser | 95 | RFQ | 1040.10 | A laser used as an interferometer for high precision positioning and/or measurements. |
| Data Measurement, Transmit, Control Laser Products | Product Incorporating Certified Class 1 Laser Data Measurement, Transmit, Control | 95 | RFR | 1040.10 | A data measurement, data transmission, or remote control product that incorporates a certified Class 1 laser. |
| Data Measurement, Transmit, Control Laser Products | Other | 95 | RZZ | Unk | A data measurement, data transmission, or remote control product that incorporates a laser other than a certified Class 1 laser. |
| Utility/Peripheral Laser Products | Reprographics, Laser, Utility/Peripheral Laser Products | 95 | RFS | 1040.10 | A reprographics machine that incorporates a laser utilized to expose internal sensitive components or materials for photocopying text and graphics. |
| Utility/Peripheral Laser Products | Laser Printer, Utility/Peripheral Laser Products | 95 | RFT | 1040.10 | A printing machine that incorporates a laser utilized in printing images on paper with no medical claims. |
| Utility/Peripheral Laser Products | Laser FAX Machine, Utility/Peripheral Laser Products | 95 | RFU | 1040.10 | A printing machine that incorporates a laser utilized in printing facsimiles of images on paper. |
| Utility/Peripheral Laser Products | CD, CD-ROM Player, Laser Utility/Peripheral Laser Products | 95 | RFV | 1040.10 | A CD or CD-ROM player that utilizes a laser to read data on the compact disc. |
| Utility/Peripheral Laser Products | DVD, DVD-ROM Player, Laser Utility/Peripheral Laser Products | 95 | RFW | 1040.10 | A DVD or DVD-ROM player that utilizes a laser to read data on the digitally recorded video disc. |
| Utility/Peripheral Laser Products | CD-R, CD-RW Recorder, Utility/Peripheral Laser Products | 95 | RFX | 1040.10 | A CD-R or CD-RW recorder machine that utilizes a laser to read and/or write data on the compact disc. |
| Utility/Peripheral Laser Products | DVD-R, DVD+R, DVD-RAM, DVD+RW, DVD-RW Recorder, Utility/Peripheral Laser Products | 95 | RFY | 1040.10 | A DVD recorder machine that utilizes a laser to read and write or read, write, and erase data on a digitally recorded video disc in any of the data formats: DVD-R, DVD+R, DVD-RAM, DVD-RW, or DVD+RW. |
| Utility/Peripheral Laser Products | UPC Reader (Bar Code Reader), Utility/Peripheral Laser Products | 95 | RFZ | 1040.10 | A laser used to scan across a bar code to identify the product. Bar code readers can be hand-held accessories, under-counter components incorporated in store check-out systems, or laser scanner systems incorporated in assembly lines used for identification and inventory purposes in manufacturing facilities, warehouses and storage facilities, or other consumer, industrial, health care, or commercial locations. |
| Utility/Peripheral Laser Products | Home/Office Machine Incorporating Utility/Peripheral Laser | 95 | RZP | 1040.10 | A laser utilized in the home or office environment not otherwise defined. |
| Utility/Peripheral Laser Products | Product Incorporating Certified Class 1 Data Utility/Peripheral Laser Products | 95 | RGA | 1040.10 | A utility/peripheral laser product that incorporates a certified Class 1 laser. |
| Utility/Peripheral Laser Products | Other | 95 | RZZ | Unk | A utility/peripheral laser product that incorporates a laser other than a certified Class 1 laser. |
| In Vitro and Other Medical Laser Products | Veterinary Laser, In Vitro and Other Medical Laser Products | 95 | RGB | 1040.10; 1040.11 | A laser used for treatment of animals other than human |
| In Vitro and Other Medical Laser Products | Separator, Automated, Blood Cell, Diagnostic | 81 | GKT | 1040.10 | |
| In Vitro and Other Medical Laser Products | Automated Differential Cell Counter | 81 | GKZ | 1040.10 | |
| In Vitro and Other Medical Laser Products | Cell Particle Counter (Automated) | 81 | GKL | 1040.10 | |
| In Vitro and Other Medical Laser Products | Urine Particle Counter | 88 | LKM | 1040.10 | |
| In Vitro and Other Medical Laser Products | System, Separation, Hematopoietic Stem Cell | 81 | MZK | 1040.10 | |
| In Vitro and Other Medical Laser Products | Test, Urea (Breath or Blood) for H. Pylori Test | 83 | MSQ | 1040.10 | |
| In Vitro and Other Medical Laser Products | Multipurpose System for In-vitro Coagulation | 81 | JPA | 1040.10 | |
| In Vitro and Other Medical Laser Products | System, Laser Assisted Hatching | 85 | MRX | 1040.10; 1040.11 | |
| In Vitro and Other Medical Laser Products | Sorter, Cell | 81 | KEX | 1040.10 | |
| In Vitro and Other Medical Laser Products | Separator, Semi-Automated, Blood Component | 81 | MYY | 1040.10 | |
| In Vitro and Other Medical Laser Products | Other | 95 | RZZ | Unk | A laser used for in vitro applications or other medical applications that do not expose patients to the laser radiation. |
| Positioning Medical Laser Products | X-Ray Field Indicator Light (Laser), Positioning Medical Laser Products | 95 | RGC | 1020.30; 1040.10; 1040.11 | A laser incorporated in a diagnostic x-ray system that is irradiated onto the film screen area indicating the x-radiation area. The beam is usually scanned to show a rectangular region for patient placement. |
| Positioning Medical Laser Products | Monitor, Patient Position, Light Beam | 90 | IWE | 1040.10; 1040.11 | |
| Positioning Medical Laser Products | Positioning Medical Laser Product | 95 | RZS | 1040.10; 1040.11 | A laser used for positioning in medical applications not otherwise defined. |
Classification of Non-compliant Items
| Performance requirements | |||
|---|---|---|---|
| 1040.10(d) | Classified in higher class | Minor, Concern | Class B, C |
| 1040.10(d) | Classified in lower class | Major | Class A |
| 1040.10(f)(1) | Protective housing allows unnecessary body access to Class IV or high IIIb radiation | Major | Class A |
| 1040.10(f)(1) | Protective housing allows unnecessary straight line access to interior Class IV or high IIIb radiation With high risk of exposure (IV or IIIb product) With low risk of exposure (IV or IIIb product) With any risk of exposure (I, IIa, II, or IIIa product) | Major Minor Major | Class A Class B Class A |
| 1040.10(f)(1) | Protective housing allows unnecessary body access to low Class IIIb or IIIa radiation In a Class IV or IIIb product In a Class I, IIa, II, or IIIa product | Minor Major | Class B Class A |
| 1040.10(f)(1) | Protective housing allows necessary body access to Class IIIa or IIIa radiation In a Class IV or IIIb product In a Class I, IIa, II, or IIIa product | Concern Minor | Class C Class B |
| 1040.10(f)(1) | Protective housing allows unnecessary body access to Class II radiation In a Class II product In a Class I product | Concern Minor | Class C Class B |
| 1040.10(f)(2) | Safety interlocks absent when required | Major | Class A |
| 1040.10(f)(2) | Single safety interlock when redundant required | Major | Class A |
| 1040.10(f)(2) | Single component with multiple contacts when redundant required | Minor | Class B |
| 1040.10(f)(2) | Defeatable safety interlocks lacks indication | Minor | Class B |
| 1040.10(f)(2) | Defeatable safety interlocks fails to prevent replacement of protective housing during defeat | Minor | Class B |
| 1040.10(f)(3) | No remote interlock connector | Major | Class A |
| 1040.10(f)(4) | No key control | Major | Class A |
| 1040.10(f)(4) | Key control removable when on | Major | Class A |
| 1040.10(f)(5) | No emission indicator | Major | Class A |
| 1040.10(f)(5) | No delay preceding radiation emission | Minor | Class B |
| 1040.10(f)(5) | Shorter delay than required | Minor | Class B |
| 1040.10(f)(5) | Remote control lacks emission indicator | Major | Class A |
| 1040.10(f)(6) | Beam attenuator without approvable alternate | Major | Class A |
| 1040.10(f)(6) | Beam attenuator with approvable alternate | Concern | Class C |
| 1040.10(f)(8) | Viewing optics Hazardous Non-hazardous for viewing period | Major Concern | Class A Class C |
| 1040.10(f)(9) | No scanning guards | Major | Class A |
| 1040.10(f)(10 | No manual reset | Major | Class A |
| 1040.10(g)(1), (2), and (3) | Warning logotype None Classification too low Classification too high | Major Major Minor, Concern | Class B Class B Class B |
| 1040.10(g)(4) | Warning logotype output information | Minor | Class B |
| 1040.10(g)(5) | No aperture label | Minor | Class B |
| 1040.10(g)(5) | Aperture label not in close proximity to aperture | Minor | Class B |
| 1040.10(g)(5) | Aperture label wording incorrect | Concern | Class C |
| 1040.10(g)(6), (7) | No protective housing labels | Minor | Class B |
| 1040.10(g)(6), (7) | Protective housing placement inappropriate | Minor | Class B |
| 1040.10(g)(6), (7) | Protective housing wording wrong | Concern | Class C |
| 1040.10(g)(8) | Invisible radiation warning on labels | Minor | Class B |
| 1040.10(g)(9), (10) | Label positioning and legibility | Minor | Class B |
| 1040.10(h)(1) | User instructions (i) Promoting unsafe practices Inadequate instructions to avoid exposure (ii) Inadequate radiometric specifications (iii) Inadequate reproductions and locations (iv) Inadequate listing of controls Inadequate caution statement | Major Minor Minor Minor Minor Concern | Class A Class B Class B Class B Class B Class C |
| 1040.10(h)(2)(i) | Reproduction of warning logotype not in catalogs | Minor | Class B |
| 1040.10(h)(2)(ii) | Service information inadequate | Minor | Class B |
| Specific product requirements | |||
| 1040.11(a)(1) | Means to measure medical laser output None Inaccurate | Major Major | Class A Class A |
| 1040.11(a)(2) | Inadequate calibration procedure/schedule | Major | Class A |
| 1040.11(a)(3) | Aperture label | Minor | Class B |
| 1040.11(b) | Excessive output on surveying lasers | Major | Class A |
| 1040.11(c) | No variance for demonstration Class IIIb or Class IV lasers | Major | Class A |
Sample Laser Product Inspection and Field Test Checklist
LASER PRODUCT TEST RECORD
MANUFACTURER____________________ CLASS____________
MODEL_______________________ SERIAL NUMBER_______
Status of Unit Examined (Circle one): Prototype/Production unit
Status of Assembly (circle one): Complete/Incomplete
Manufactured Date:_________________________
A. Product Description: (Include basic configuration and size of product, reference to photos and/or
diagrams, basic functions to be performed during operation and during maintenance.)
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
Product Report: Has the product been reported to CDRH?
Yes_________ No_________
If yes, what is the Accession Number? _____________________________
Summary of Product Evaluation:
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
C. Certification/Identification Requirements. If possible, obtain a sample of each required label and attach it to this report. Otherwise, quote pertinent information, especially any noncompliant items.
1. Certification label (1010.2)
a. Is the label permanently affixed? Yes___ No___ ND___ NA___
b. Is the Label readily viewable? Yes___ No___ ND___ NA___
Location:______________________
c. Is the label properly stated? Yes___ No___ ND___ NA___
(Note: Products under variance require modified certification labels 1010.4(d))
d. Remarks: ________________________________________________
__________________________________________________________
2. Identification label (1010.3)
a. Is the label permanently affixed? Yes___ No___ ND___ NA___
b. Is the label readily viewable? Yes___ No___ ND___ NA___
Location: _______________________
c. Does the label contain the full name and address?
Yes___ No___ ND___ NA___
d. Does the label contain the place of manufacture (in full or in code)?
Yes___ No___ ND___ NA___
e. If coded, has CDRH been provided the code?
Yes___ No___ ND___ NA___
f. Are the month and year of manufacture stated in full?*
Yes___ No___ ND___ NA___
Month and year: __________________________
g. Remarks:____________________________________________________
____________________________________________________________
*Note: Serialization is acceptable in lieu of month and year for consumer electronic products.
D. Special Purpose Products (1040.11)
1. Is the product a medical laser product?
Yes___ No___ ND___ NA___
Note: In inspecting manufacturers of not only medical laser products but also laser products that are medical devices, verify compliance with other applicable requirements including but not limited to current registration and listing, 510k market clearances, device master record or quality system, current complaint and service records, etc.
a. Does the product include a means of measurement of levels of radiation intended for irradiation of the human body?
Yes___ No___ ND___ NA___
b. How is this accomplished? Measure beam prior to delivery system and determine output levels via calibration constant _________;
Measure output of delivery system ___________;
Other ___________.
c. Indication: power ________; energy________; time _______.
d. Type of indicator: energy/power select switch ___; "Test shot" display (remains constant until next best shot) ___; Real time display (displays level at all times) ___; Other ___.
e. If test shot is available only at initiation of procedure or if a select switch is used, does the product have an internal monitoring system capable of maintaining output levels to within ± 20% of displayed value?
Yes ___ No___
f. Is display analog ____; or digital ____? If digital, are there sufficient significant digits to allow ± 20% accuracy? Yes ___ No ___
g. Is the total measurement error within ± 20% (see Attachment G) Yes___ No___ ND___ NA___
h. Is there a laser aiming beam? Yes ___ No ___. Is there a means to measure the level of the aiming beam if the product is ophthalmic and the aiming beam may exceed 1 mW or if the product is not ophthalmic but the aiming beam may exceed 5 mW?
Yes ___ No ___
i. Remarks: _____________________________________________________
_____________________________________________________________
2. Is the product a surveying, leveling, and alignment product?
Yes___ No___ ND___ NA___
a. Is access prevented for wavelengths of 400 nm to 710 nm to radiation power in excess of 5.0 mW for any duration greater than 3.8 x 10-4 seconds?
b. Is access prevented to radiation levels in excess of Class I limits for any other combination of emission Duration and wave length range?
Yes___ No___ ND___ NA___
c. Remarks: ____________________________________________________
____________________________________________________________
3. Is the product a demonstration laser product?
Yes___ No___ ND___ NA___
a. Does the product prevent human access to radiation in excess of the Class IIIa (3R) limit?
Yes___ No___ ND___ NA___
b. Remarks: ____________________________________________________
____________________________________________________________
E. Label Requirements. See instruction in paragraph B.
1. Warning logotypes* (1040.10(g)(1),(2),(3),(4),(5),(9), and (10)
a. Is the logotype the correct logotype?
Yes___ No___ ND___ NA___
b. Is the label properly worded for its class designation?
Yes___ No___ ND___ NA___
c. Does the label have the proper color?
Yes___ No___ ND___ NA___
d. Is the output information present and correct?
(Maximum output stated ________.) Yes___ No___ ND___ NA___
e. Is the media or wavelength information present and correct?
Yes___ No___ ND___ NA___
f. Is the label permanently affixed and clearly visible during operation, maintenance, and
service?
Yes___ No___ ND___ NA___
g. Is the label positioned so as to make exposure unnecessary during reading?
Yes___ No___ ND___ NA___
Location: ____________________
h. Does the label include a warning for "invisible" or "invisible and/or visible" radiation?
Yes___ No___ ND___ NA___
i. Remarks: __________________________________________________________
__________________________________________________________________
__________________________________________________________________
Note: Warning labels in accordance with IEC 60825-1 including product classification are acceptable.
2. Aperture label (for Classes II, IIIa, IIIb, IV, 3R, 3B and 4) 1040.10(g)(5),(8),(9), and (10)
a. Is a label present and in proximity to each aperture?
Yes___ No___ ND___ NA___
b. Is the label properly worded? Yes___ No___ ND___ NA___
c. Is the label permanently affixed and clearly visible?
Yes___ No___ ND___ NA___
d. Is the label positioned so as to make exposure unnecessary during reading?
Yes___ No___ ND___ NA___
(Location: ______________)
e. Does the label include a warning for "invisible" or "invisible and/or visible" radiation?
Yes___ No___ ND___ NA___
f. Remarks: __________________________________________________________
__________________________________________________________________
__________________________________________________________________
3. Noninterlocked protective housing label (1040.10(g)(6),(8),(9), and (10)
a. Are the labels on or near all appropriate panels or covers which are removed for operation, maintenance, or service?
Yes___ No___ ND___ NA___
b. Are all labels visible prior to removal of such portions of the protective housing?
Yes___ No___ ND___ NA___
c. Are all labels visible after opening?
Yes___ No___ ND___ NA___
d. Are all labels correctly worded? Yes___ No___ ND___ NA___
e. Are all labels permanently affixed and clearly visible?
Yes___ No___ ND___ NA___
f. Do all labels contain a warning for "invisible" or "invisible and/or visible" radiation? Yes___ No___ ND___ NA___
g. Remarks: __________________________________________________________
__________________________________________________________________
__________________________________________________________________
4. Defeatably interlocked housing labels 1040.l0(g)(7),(8),(9), and (10)
a. Are labels provided for each defeatably interlocked panel or cover which is removed for operation, maintenance, or service?
Yes___ No___ ND___ NA___
b. Are all labels visible prior to interlock defeat?
Yes___ No___ ND___ NA___
c. Are all labels visible during interlock defeat?
Yes___ No___ ND___ NA___
d. Are all labels correctly worded?
Yes___ No___ ND___ NA___
e. Are all labels permanently affixed and clearly visible?
Yes___ No___ ND___ NA___
f. Do all labels contain a warning for "invisible" or invisible and/or visible" radiation?
Yes___ No___ ND___ NA___
g. Remarks: _____________________________________________________
_____________________________________________________________
F. Performance Requirements (1040.l0(f)):
1. Protective Housing (1040.l0(f)(l))
a. Does the housing prevent access at all times to laser radiation above Class I not necessary for operation of the product?
Yes___ No___ ND___ NA___
b. Does the housing prevent access at all times to collateral optical radiation above Class I not necessary for operation of the product?
Yes___ No___ ND___ NA___
c. Has x-radiation been evaluated?
Yes___ No___ ND___ NA___
d. Does the housing prevent access to x-radiation levels in excess of 0.5 mR/hr at all times during operation of the product?
Yes___ No___ ND___ NA___
e. Remarks: _____________________________________________________
_____________________________________________________________
2. Safety Interlocks (1040.10(f)(2)) (Complete for each interlock. Identify the portion of
removable or displaceable housing and interlock described.)
a. Do operation or maintenance functions require moving portions of the housing which could allow access to radiation?
Yes___ No___ ND___ NA___
Describe: _____________________________________________________
_____________________________________________________________
b. Class of radiation to which access could be gained?
Class ______.
c. Is a fail safe or multiple interlock required (including 1040.l(f)(2)(iii)?
Yes___ No___ ND___ NA___
Where? ______________________________________________________
_____________________________________________________________
d. Are safety interlock(s) present? where?_______________
1. TYPE: Microswitch____; Mercury switch____;
male-female plug____; mechanical shutter____; other____.
Describe: _____________________________________________
e. Method of limiting access: directly interrupts primary laser power____; interrupts primary laser power through relay, contactor, switching tube or transistor____; spoils the cavity____; shutter beam via solenoid____; other____.
f. Is there a multiple or fail safe interlock on each housing for which an interlock is required?
Yes___ No___ ND___ NA___
g. Is the interlock defeatable? Yes___ No___ ND___ NA___
h. Is there an indication of defeat? Yes___ No___ ND___ NA___
Describe: _______________________________________________________
i. Does the interlock preclude replacement of the housing while the interlock is defeated?
Yes___ No___ ND___ NA___
j. Are non-safety interlocks present? Where?_____________
Yes___ No___ ND___ NA___
k. Remarks: _________________________________________________
_________________________________________________________
3. Remote Interlock Connector (1040.l0(f)(3), Class IIIb or IV systems only)
a. Is a remote control connector present?
Yes___ No___ ND___ NA___
b. Type? Describe: ________________________________________
c. Is the voltage across the connector less than 130 volts RMS?
Yes___ No___ ND___ NA___
d. Is the access to laser and collateral radiation prevented when the terminals are not joined?
Yes___ No___ ND___ NA___
e. Method of operation: Directly interrupts laser power____; interrupts laser power through relay, etc.____; shutters beam or interrupts cavity____.
f. Does the emission delay reactivate when the remote control circuit is interrupted?
Yes___ No___ ND___ NA___
g. Must the emission be manually restarted following interruption via the remote interlock connector?
Yes___ No___ ND___ NA___
h. Remarks: ____________________________________________________
____________________________________________________________
4. Key Control (1040.10(f)(4), Class IIIb, IV, 3B, or 4 systems only)
a. Is a key control present? Yes___ No___ ND___ NA___
Describe: ___________________________________________________
_____________________________________________________________
b. Is a key removable in the "on" position?
Yes___ No___ ND___ NA___
c. Is operation prevented when the key is removed?
Yes___ No___ ND___ NA___
d. How? ____________________________________________________
____________________________________________________________
e. Remarks: _________________________________________________
____________________________________________________________
5. Beam Attenuator (1040.l0(f)(6), Class IIIb, IV, 3B or 4 systems only)
a. Is a beam attenuator present? Yes___ No___ ND___ NA___
b. Type: mechanically operated shutter____; electrically operated____; aperture cap or cover____; other____.
Describe: ___________________________________________________
____________________________________________________________
c. Is the attenuator permanently attached?
Yes___ No___ ND___ NA___
d. Does the attenuator prevent access by any part of the body to radiation in excess of Class I limits?
Yes___ No___ ND___ NA___
e. If there is no beam attenuator, has the manufacturer requested and obtained approval of an alternate means of safety?
Yes___ No___ ND___ NA___
f. Remarks: _________________________________________________
_____________________________________________________________
6. Emission Indicator (1040.10(f)(5), Class, IIIb, IV, 3B or 4 Systems only)
a. Is an emission indicator present on the laser product?
Yes___ No___ ND___ NA___
Where? ________________________________________________
b. Type: tungsten lamp(s)____; neon lamp(s)____; LED(s)____; other____.
Describe: _________________________________________________
_______________________________________________________
c. If the indicator is visible, is it visible through the protective eyewear that is normally supplied or recommended?
Yes___ No___ ND___ NA___
d. Can the indicator be viewed without exposure to radiation in excess of Class I limits?
Yes___ No___ ND___ NA___
e. Is there a delay between an indication of emission and the beginning of emission?
Yes___ No___ ND___ NA___
f. How is emission delay achieved? Thermal relay___; inherent in the lasing process___; delay circuit___; other___.
Describe: ________________________________________________
_________________________________________________________
g. Length of delay? ____________________
h. Is the power source or operation control separable from the laser by greater than 2 meters when assembled for use?
Yes___ No___ ND___ NA___
i. If separated greater than 2 meters, is an emission indicator present on the energy source or controller?
Yes___ No___ ND___ NA___
Where? _____________________________________________________
j. Type: Tungsten lamp(s)___; neon lamp(s)___; LED(s)___; bell or buzzer___; meter or display___; mechanical flag___; other___.
Describe: ______________________________________________
______________________________________________________
k. Is there a delay between an indication of emission and the beginning of emission?
Yes___ No___ ND___ NA___
l. How is emission delay achieved? Thermal relay___; inherent in the lasing process___; delay circuit___; other___.
Describe: ______________________________________________
______________________________________________________
m. Length of delay? ____________________________________________
n. Remarks: __________________________________________________
____________________________________________________________
____________________________________________________________
6. Location of Controls (1040.10(f)(7))
a. Are the controls located so that exposure is unnecessary for operation or adjustments?
Yes___ No___ ND___ NA___
b. Remarks: __________________________________________________
____________________________________________________________
____________________________________________________________
7. Viewing optics (1040.l0(f)(8))
a. Are viewing optics or viewports present?
Yes___ No___ ND___ NA___
b. Type: microscope___; telescope___; viewport___; display screen___; other___.
Describe: ___________________________________________________
____________________________________________________________
____________________________________________________________
c. Where? ___________________________________________________
____________________________________________________________
d. Do the viewing optics attenuate radiation at all times during operation or maintenance to levels less than Class I limits?
Yes___ No___ ND___ NA___
e. Do the viewing optics employ a shutter or variable attenuator?
Yes___ No___ ND___ NA___
f. Upon failure of the shutter of the variable attenuator is access to radiation levels greater than the Class I limits prevented?
Yes___ No___ ND___ NA___
g. Remarks: _________________________________________________
____________________________________________________________
____________________________________________________________
8. Scanning Safeguard (1040.10(f)(9))
a. Is the radiation emitted by the product scanned?
Yes___ No___ ND___ NA___
b. Is the classification of the product based on the level of scanned radiation?
Yes___ No___ ND___ NA___
c. In the event of scan failure, is human access to laser radiation in excess of the product class prevented?
Yes___ No___ ND___ NA___
d. Remarks: _________________________________________________
_____________________________________________________________
9. Manual Reset Mechanism (1040.l0(f)(l0) Class IV laser systems)
Describe the operation of the Manual Reset. ___________________
____________________________________________________________
How is it achieved? (latching relay, etc.) ________________
____________________________________________________________
10. Removable laser system (1040.10(c)(2))
a. Does the product incorporate a laser system?
Yes___ No___ ND___ NA___
b. Is the laser system removable?
Yes___ No___ ND___ NA___
c. If removable, is the laser system independently certified?
Yes___ No___ ND___ NA___
d. If not removable, specify how removability is prevented: hard wiring___; modified connector___; assembled internally from components___; other (specify). __________________________
e. Remarks: __________________________________________________
____________________________________________________________
____________________________________________________________
G. Laser Product Measurements
Model # ________________________ Serial # _______________________
Manufacturer's Claimed Classification: ___________________________________
Brief description of product: ___________________________________________
___________________________________________________________________
___________________________________________________________________
Test Instrument(s) Used: _______________________________________________
___________________________________________________________________
Circle radiometric quantity tested and specify units below (Radiance (W cm-2 sr-1), Radiant Energy (J), Power (W), etc.)
| Measurement No. | Wavelength (nm) | Instrument reading, R (units ________) | Calibration factor, K (units ________) | Corrected value, R*K (units ________) |
|---|---|---|---|---|
Calculations (as needed):
__________________________________________________________________
__________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
H. Results of FDA measurements: ______________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
I. Compliance with other requirements (e.g., conditions of a variance, labeling for medical devices, etc.)
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
J. Information requirements (Directions: Complete this section only if the information and requirements are reviewed during the inspection).
1. User Information (1040.l0(h)(l))
a. Does the manual contain adequate instructions for assembly,
operation, and maintenance? Yes___ No___ ND___ NA___
b. Does it contain clear warnings to avoid exposure?
Yes___ No___ ND___ NA___
c. Does it contain a statement of output parameters?
Yes___ No___ ND___ NA___
d. Does it contain legible reproductions of all labels and hazard warnings?
Yes___ No___ ND___ NA___
e. Does it include the corresponding position of each label on the product?
Yes___ No___ ND___ NA___
f. Does it contain listing of controls, adjustments, and procedures for operation and maintenance?
Yes___ No___ ND___ NA___
g. Does it contain a schedule of maintenance?
Yes___ No___ ND___ NA___
h. Does it contain the "Caution - use of controls..." warning?
Yes___ No___ ND___ NA___
i. Does it contain a compatibility statement (laser source or laser system not supplied with the product?
Yes___ No___ ND___ NA___
j. Does it contain a calibration schedule (medical laser product)?
Yes___ No___ ND___ NA___
k. Does it include a warning not to point the laser radiation at the audience (especially Class IIIa demonstration laser products)?
Yes___ No___ ND___ NA___
l. Does it include information to determine nominal hazard zone(class IV multi-axis workstations)?
Yes___ No___ ND___ NA___
m. Remarks: __________________________________________________________
__________________________________________________________________
__________________________________________________________________
2. Purchasing Information (1040.10(h)(2))
a. Are legible reproductions of the logotype required to be affixed to the product (including information required for positions 1, 2, and 3) contained in catalogues, specification sheets, and descriptive brochures?
Yes___ No___ ND___ NA___
3. Servicing Information (1040.l0(h)(2))
a. Are adequate instructions for service adjustments and service procedures available?
Yes___ No___ ND___ NA___
b. Are clear warnings and precautions to avoid possible exposure included?
Yes___ No___ ND___ NA___
c. Is a schedule of maintenance necessary to keep the product in compliance included?
Yes___ No___ ND___ NA___
d. Are controls and procedures which would he used by reasons other than the manufacturer or his agent to increase accessible emission levels listed?
Yes___ No___ ND___ NA___
e. Is a clear description of the locations of displaceable portions of the protective housing provided?
Yes___ No___ ND___ NA___
f. Do these instructions provide legible reproductions of required labels and hazard warnings?
Yes___ No___ ND___ NA___
g. Do these instructions include protective procedures for service personnel?
Yes___ No___ ND___ NA___
h. Remarks: __________________________________________________________
__________________________________________________________________
__________________________________________________________________