U.S. flag An official website of the United States government
  1. Home
  2. Medical Devices
  3. Device Advice: Comprehensive Regulatory Assistance
  4. Guidance Documents (Medical Devices and Radiation-Emitting Products)
  5. Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submissions - Final Guidance for Industry and FDA
  1. Guidance Documents (Medical Devices and Radiation-Emitting Products)

Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submissions - Final Guidance for Industry and FDA

PDF Printer Version
(118 KB)

Document issued on: November 29, 2000

This document supersedes documents "Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submissions;
February 21, 2000

Image of CDRH Logo

U.S. Department of Health and Human Services
Food and Drug Administration
Center for Devices and Radiological Health

Circulatory Support and Prosthetic Devices Branch
Division of Cardiovascular and Respiratory Devices
Office of Device Evaluation


Preface

Public Comment

Comments and suggestions may be submitted at any time for Agency consideration to Dockets Management Branch, Division of Management Systems and Policy, Office of Human Resources and Management Services, Food and Drug Administration, 5630 Fishers Lane, Room 1061, (HFA-305), Rockville, MD, 20852. When submitting comments, please refer to the exact title of this guidance document. Comments may not be acted upon by the Agency until the document is next revised or updated.

For questions regarding the use or interpretation of this guidance contact Catherine Wentz at (240) 276-4141 or by e-mail at catherine.wentz@fda.hhs.gov.

Additional Copies

Additional copies are available from the Internet. You may also send an e-mail request to CDRH-Guidance@fda.hhs.gov to receive a copy of the guidance. Please use the document number 1622 to identify the guidance you are requesting.


Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submissions;
Final Guidance for Industry and FDA

This document is intended to provide guidance. It represents the Agency’s current thinking on the above. It does not create or confer any rights for or on any person and does not operate to bind FDA or the public. An alternative approach may be used if such approach satisfies the requirements of the applicable statute, regulations, or both.

Introduction:

This guidance document describes a means by which cardiopulmonary bypass arterial line blood filter devices may comply with the requirement of special controls for class II devices. Designation of this guidance document as a special control means that manufacturers attempting to establish that their device is substantially equivalent to a predicate cardiopulmonary bypass arterial line blood filter device should demonstrate that the proposed device complies with either the specific recommendations of this guidance or some alternate control that provides equivalent assurances of safety and effectiveness.

This guidance document has been developed as a special control to support a change in classification from class III to class II. It identifies relevant material on preclinical studies and labeling to include in a 510(k) premarket notification application. We intend for it be used in conjunction with the FDA guidance documents listed below. These are also special controls. All FDA requirements regarding premarket notification submissions are not repeated in this document.

  • Use of International Organization for Standardization (ISO) 10993 "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing"
  • 510k Sterility Review Guidance

The Least Burdensome Approach:

The issues identified in this guidance document represent those that we believe need to be addressed before your device can be approved/cleared for marketing. In developing the guidance, we carefully considered the relevant statutory criteria for Agency decision-making. We also considered the burden that may be incurred in your attempt to comply with the guidance and address the issues we have identified. We believe that we have considered the least burdensome approach to resolving the issues presented in the guidance document. If, however, you believe that information is being requested that is not relevant to the regulatory decision for your pending application or that there is a less burdensome way to address the issues, you should follow the procedures outlined in "The Least Burdensome Provisions: Concept and Principles; Guidance for Industry and Food and Drug Administration Staff" document.

Scope:

This guidance is limited to the preclinical and labeling aspects of cardiopulmonary bypass arterial line blood filter devices. A cardiopulmonary bypass arterial line blood filter is a device used as part of a gas exchange (oxygenator) system to filter non-biologic particles and emboli out of the blood. It is used in the arterial return line, (21 CFR 870.4260, DTM).

The following table sets forth the risks to health associated with this device that were identified in the proposed classification ruling (dated February 26, 1979), as well as additional adverse event reporting since the classification ruling. Next to each risk is a description of the special control suggested to address the risk.

RISK TO HEALTH SPECIAL CONTROLS
  1. Damage to formed blood elements, e.g., clotting, hemolysis

Blood Studies: Evaluate hemolysis and cell depletion (i.e., blood cell counts) over a 6 hour circulation period for the subject, predicate and dynamic control circuits. In addition, blood component (platelet) functionality should be performed if new technology, materials, or surface characteristics are introduced. Blood studies should be performed at the maximum labeled flow rate. In some cases where flow dynamics at low flow rates may play a role in blood trauma, the data should also be presented for the minimum labeled flow rate

Visual Inspection: Gross inspection for thrombus.

  1. Excessive pressure drop resulting in inadequate blood flow, damage to device, structural integrity, damage to the line

Pressure Integrity Testing: Perform burst pressure for test device using sustained static pressure at 1.5 times the maximum anticipated pressure for the intended use for 6 hours. Observe for leaks, tears, and structural integrity. Use Water or saline as the test medium.

Pressure Drop: Perform pressure drop testing to steady state, on the test and predicate devices, at highest rated flow rate. Use blood or a blood analog as the testing medium.

  1. Structural Integrity; i.e., connections pull apart
Test and state the pull strength required to separate the connections; compare to predicate device and anticipated worst case clinical conditions.
  1. Excessive pressure gradients; i.e., blood damage, inadequate blood flow
Flow rate capacity: Determine the flow rate limitation(s) to assure safe and effective performance.
  1. Does not provide efficient removal of solid and gaseous emboli

Filtration Efficiency: Determine filtration efficiency over the labeled range of particle size at maximum flow rate.

Air Handling: Assess the ability of the device to remove free circulating gas bubbles at several specified flow rates. Challenge the filter with gaseous emboli introduced proximal to the filter. State the total volume of gas introduced, flow rate of the gas introduced, flow rate of the circuit, temperature and distance from the test device. Use a bubble detector distal to the device to verify adequate air handling.

Labeling: Recommend the use of the attached Checklist.

  1. User error

Labeling: Include clear, concise instructions for use. Describe the human factors review e.g., inclusion of a troubleshooting guide, easy formatting of instructions for use, etc. Use the attached Checklist as a guide.

Provide flow rate, duration of use (e.g., 6 hours), and other pertinent information obtained through performance testing within the labeling, to facilitate correct use of the device.

The use of a bubble detector downstream from the arterial line filter is strongly recommended.

  1. Blood incompatibility

Biocompatibility testing: Perform testing on aged product, as recommended in the FDA guidance on ISO 10993: Use of International Standard ISO 10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, dated May 1, 1995 to assure that the materials used are non-toxic for the intended use. Include sensitization, pyrogenicity, acute systemic toxicity, mutagenicity, cytotoxity, irritiation, and hemocompatibility/hemolysis testing.

If biocompatibility testing is performed on the unaged product, then submit an FT-IR and results of cytotoxicity testing on the aged product as well, that demonstrates that the physical and chemical properties of the material(s) have not changed.

  1. Incompatibility of the product when exposed to circulating blood; and infection.
Sterilization: Perform sterilization validation to ensure that the sterilization process is capable of providing a Sterility Assurance Limit (SAL) of 10-6. Perform biological indicator, pyrogen, and bioburden testing to ensure acceptable limits of biological contaminants.
  1. Insufficient device performance, material incompatibility, and lack of sterility over a period of time

Shelf-life: Submit the protocol and data from real or accelerated aging studies for the intended labeled shelf-life for the device. If accelerated studies are used, validate (for your own files) the results with real-time data.

Perform package integrity and barrier property assessment: use validated physical and microbial-based methods.

ATTACHMENT: Human Factors/Instruction for Use Checklist

This Checklist is a human factors tool to aid the manufacturer when designing the device and developing the instructions for use manual. Using this checklist will reduce many of the risks to health discussed in this guidance; especially those that can addressed by adequate instructions for use.

  1. EQUIPMENT ASSEMBLY

    __ Mount motor drive correctly, relative to venous reservoir.

    __ Check that all electrical connections are secure.

    __ Test control module power and display.

    __ Check date and integrity of sterile centrifugal pump (and disposable probe) package(s).

    __ Check that flow transducer/sensor (and disposable probe) are sized properly.

    __ Assemble perfusion circuit in a sterile manner.

    __ Allow sufficient tubing length for standby pumping unit.

    __ Connect flow transducer/sensor (and disposable probe) to circuit in correct location and flow direction.

  2. PRIME PUMP AND CIRCUIT

    __ CO2 flush pump and circuit, if indicated; turn off CO2.

    __ Gravity-prime and debubble pump and perfusion circuit.

    __ Check pump for leaks, irregular motion, and noise.

    __ Check circuit for visible air.

    __ Check that all tubing connections are secure.

    __ Clamp pump outlet line completely.

    __ Clamp venous return line completely.

  3. OPERATING PARAMETERS

    __ Calibrate transducers/sensors according to manufacturer’s instructions.

    __ Set low/high flow alarms.

    __ Verify flow alarms.

  1. EMERGENCY BACKUP EQUIPMENT

    __ Backup power available.

    __ Hand crank available.

    __ Pump (and disposable probe) available.

    __ Control module or roller pump available.

  1. PERFUSION

    __ Achieve minimum pump speed prior to unclamping lines.

    __ Monitor control module for messages and alarms.

    __ Monitor perfusion circuit for visible air.

    __ Maintain minimum pump speed prior to clamping lines.

  1. CLEANUP

    __ Turn off power.

    __ Discard disposable components.

    __ Cover magnetic motor drive.

    __ Clean console and transducer/sensor.

  1. CHECK EQUIPMENT

    __ Inspect and verify that equipment is operational.

    __ Maintain indicated preventive maintenance schedule

    __ Recharge batteries to full capacity (if necessary).

 
Back to Top