  5. FDA-Cleared Sterilants and High Level Disinfectants with General Claims for Processing Reusable Medical and Dental Devices
  1. Reprocessing of Reusable Medical Devices: Information for Manufacturers

Recall Alert

An update on the Cyclospora outbreak available here.

FDA-Cleared Sterilants and High Level Disinfectants with General Claims for Processing Reusable Medical and Dental Devices

Section VI. of FDA’s Final Guidance for Industry and FDA Staff: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling outlines six criterion that should be addressed in reprocessing instructions. Criterion 4 recommends that reprocessing instructions should include devices and accessories that are legally marketed. On this page is a table of FDA-cleared liquid chemical sterilants and high level disinfectants, last updated December 2023.

Product / Approval Number Manufacturer Active Ingredient(s) Sterilant Contact Conditions High Level
Disinfectant Contact Conditions
Tristel Duo ULT High Level Disinfectant Foam for Ultrasound Probes (DEN220041) Tristel Solutions Limited 320 ppm Chlorine dioxide No indication for device sterilization.
Passes AOAC Sporicidal Activity Test in 12 hrs. at 20°C. 
For Manual application using DUO Wipes:
2 min at 20°C
Single use – foam generated on site
For processing of endocavity transvaginal, transrectal and skin surface transducers only.
Contact conditions established by simulated use testing with transducer.
Acecide-C High Level Disinfectant and Sterilant (K091210) Best Sanitizers, Inc. 3100-3400 ppm peracetic acid Indication for device sterilization
2 hrs at 20ºC
5 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test and by simulated use testing with endoscopes.
7 min at 20ºC
5 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
Extended Use Aldahol High Level Disinfectant (K09189) Healthpoint, Ltd. 3.4% glutaraldehyde
20.1% isopropanol
Indication for device sterilization
For processing in a legally marketed Automated Endoscope Reprocessor capable of maintaining solution temperature at 25ºC:

6 hrs at 25ºC
14 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test and by simulated use testing with endoscopes.
For processing in a legally marketed Automated Endoscope Reprocessor capable of maintaining solution temperature at 25ºC:
5 min at 25ºC
14 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
Resert XL HLD High Level Disinfectant (K091022) Steris Corporation 2.0% hydrogen peroxide No indication for device sterilization.
Passes the Modified AOAC Sporicidal Activity Test in 6 hrs at 20ºC.
8 min at 20ºC
21 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.
 

Steris S40 Sterilant Concentrate (K090036 and K173256)
Note: Cleared for use with the Steris System 1E Liquid Chemical Sterilant Processing System

Steris Corporation ≥1820 mg/L peracetic acid Indication for device sterilization
6 min at 46-55°C Single use – diluted by system
For use in the Steris System 1E and endo Liquid Chemical Sterilant Processing Systems only
Contact conditions based on AOAC Sporicidal Activity Test and simulated use testing with endoscopes.
No indication for high level disinfection
Aldahol V High Level Disinfectant (K113015) Microchem Laboratories, Inc. 3.4% glutaraldehyde
20.1% isopropanol
Indication for device sterilization.
8 hrs at 20ºC
14 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test and by simulated use testing with endoscopes.
 
10 min at 20ºC
14 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.
Rapicide OPA-28 High Level Disinfectant (K120306) Minntech Corporation 0.575% ortho-phthalaldehyde No indication for device sterilization.
Passes the Modified AOAC Sporicidal Activity Test in 32 hrs at 20ºC and 25°C.
For Manual Processing:
10 min at 20ºC
28 days Maximum Reuse
For processing in a legally marketed Automated Endoscope Reprocessor capable of maintaining solution temperature at 25ºC:
5 min at 25ºC
28 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
 
OPA 30 Solution (K140703) Ciden Technologies, LLC 0.6% ortho-phthalaldehyde No indication for device sterilization.
Passes the Modified AOAC Sporicidal Activity Test in 32 hrs at 20°C and 25°C.
For Manual Processing:
12 min at 20ºC
30 days Maximum Reuse
For processing in a legally marketed Automated Endoscope Reprocessor capable of maintaining solution temperature at 25ºC:
5 min at 25ºC
14 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
Acecide High Level Disnfectant and Sterilant (K041984) Best Sanitizers, Inc. 3300-3800 ppm peracetic acid No indication for device sterilization.
5 hrs at 25°C
5 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test and by simulated use testing with endoscopes.
5 min at 25ºC
5 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
TD-5 High-level Disinfectant (K051305) CS Medical, L.L.C. 2.65% glutaraldehyde

No indication for device sterilization.

Passes the AOAC Sporicidal Activity Test in 10 hrs at 22 °C.

5 min at 37.8ºC
Single use to be used exclusively with the TD-100 Transesophageal Probe Disinfector.
Contact conditions established by simulated use testing with endoscopes.
Aldahol III High Level Disinfectant (K041360) Healthpoint LTD 3.4% glutaraldehyde
26% isopropanol

Indication for device sterilization.

10 hrs at 20° C
Contact conditions based on AOAC Sporicidal Activity Test only.

10 min at 20ºC
14 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
Banicide Advanced for Sterilization and High Level Disinfection (K012889) Pascal Company, Inc. 3.5% glutaraldehyde

Indication for device sterilization.

10 hrs at 25°C
30 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.

45 min at 25°C
30 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
Cidex® OPA Concentrate (K032959) Advanced Sterilization Products 5.75% ortho-phthaldehyde (OPA)

No indication for device sterilization.

In-Use solution (0.05% OPA) passes the AOAC Sporicidal Activity Test in 32 hrs at 50°C.

5 min at 50ºC
Single use – diluted by system
For use in the EvoTech Integrated Endoscope Disinfection System only.
Contact conditions established by simulated use testing with endoscopes.
Cidex OPA Solution (K030004) Advanced Sterilization Products 0.55% ortho-phthaldehyde No indication for device sterilization. Passes the AOAC Sporicidal Activity Test in 32 hrs at 20°C and 25°C.

Manual Processing
12 min at 20°C
14 days Maximum Reuse

Automated Endoscope Reprocessor (AER)
5 min at 25°C
14 days Maximum Reuse
(For processing in an AER only with FDA-cleared capability to maintain solution temperature at 25°C.) Contact conditions established by simulated use testing with endoscopes.

Sterilox Liquid High Level Disinfectant System (K013280) Sterilox, Technologies, Inc. Hypochlorite and
Hypochlorous acid 650-675 ppm Active free chlorine
No indication for device sterilization. Passes the Modified AOAC Sporicidal Activity Test in 24 hrs at 25°C. 10 min at 25°C
Single use - generated on site
Contact conditions established by simulated use testing with endoscopes.
Sporicidin Sterilizing and Disinfecting Solution (K003087) Sporicidin International 1.12% glutaraldehyde
1.93% phenol/phenate
Indication for device sterilization.
12 hrs at 25°C
14 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.
20 min at 25°C
14 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
Rapicide High Level Disinfectant and Sterilant (K993042) MediVators, Inc. 2.5% glutaraldehyde Indication for device sterilization.
7 hrs 40 min at 35°C
28 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes and additional supporting information.
Automated Endoscope Reprocessor
5.0 min at 35°C
28 days Maximum Reuse
(For processing in an AER only with FDA-cleared capability to maintain the solution temperature at 35°C.)
Contact conditions established by simulated use testing with endoscopes.
°Cidex®OPA Solution High Level Disinfectant (K991487) Advanced Sterilization Products 0.55% ortho-phthaldehyde No indication for device sterilization. Passes the AOAC Sporicidal Activity Test in 32 hrs at 20°C. 12 min at 20°C
14 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.
°Cetylcide-G Concentrate and Diluent Concentrate (K974188) Cetylite Industries, Inc. 3.2% glutaraldehyde Indication for device sterilization.
10 hrs at 20°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.
40 min at 20°C
28 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.
MedSci 3% Glutaraldehyde (K974062) MedSci, Inc. 3% glutaraldehyde Indication for device sterilization. 10 hrs at 25°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.
25 min at 25°C
28 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.
EndoSpor Plus Sterilizing and Disinfecting Solution (K972708)
Note: Due to the lack of test strips for monitoring the concentrations of the active ingredients, the reuse period is limited to 14 days.
Cottrell Limited 7.35% hydrogen peroxide
0.23% peracetic acid
Indication for device sterilization.
180 min at 20°C
14 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
15 min at 20°C
14 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.
SporoxTM Sterilizing & Disinfection Solution (K970230) Reckitt & Colman Inc. 7.5% hydrogen peroxide Indication for device sterilization.
6 hrs at 20°C
21 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.
30 min at 20°C
21 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.
PeractTM 20 Liquid Sterilant/Disinfectant (K960513) Minntech Corporation 1.0% hydrogen peroxide
0.08% peracetic acid
Indication for device sterilization.
8 hrs at 20°C
14 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
25 min at 20°C
14 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
Procide 14 N.S. (K932922) Cottrell Limited 2.4% glutaraldehyde Indication for device sterilization.
10 hrs at 20°C
14 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.
45 min at 20°C
14 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.
OmnicideTM Long Life Activated Dialdehyde Solution (K932922) Cottrell Limited 2.4% glutaraldehyde Indication for device sterilization.
10 hrs at 20°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.
45 min at 20°C
28 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.
OmnicideTM Plus (K932922) Cottrell Limited 3.4% glutaraldehyde Indication for device sterilization.
10 hrs at 20°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.
45 min at 20°C
28 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.
Metricide Plus 30 Long-Life Activated Dialdehyde Solution (K931592) Metrex Research, Inc. 3.4% glutaraldehyde Indication for device sterilization.
10 hrs at 25°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only. FDA accepted Metricide Plus as identical to Cidex Plus.
90 min at 25°C
28 days Maximum Reuse
FDA accepted Metricide Plus as identical to Cidex Plus.
Metricide 28 Long-Life Activated Dialdehyde Solution (K931052) Metrex Research, Inc. 2.5% glutaraldehyde Indication for device sterilization.
10 hrs at 25°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only. FDA accepted Metricide 28 as identical to Cidex Formula 7.
90 min at 25°C
28 days Maximum Reuse
FDA accepted Metricide 28 as identical to Cidex Formula 7.
Metricide Activated Dialdehyde Solution (K930284) Metrex Research, Inc. 2.6% glutaraldehyde Indication for device sterilization.
10 hrs at 25°C
14 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only. FDA accepted Metricide as identical to Cidex.
45 min at 25°C
14 days Maximum Reuse
FDA accepted Metricide as identical to Cidex.
Cidex™ Activated Dialdehyde Solution (K924434) Johnson & Johnson Medical Products 2.4% glutaraldehyde Indication for device sterilization.
10 hrs at 25°C
14 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.
45 min at 25°C
14 days Maximum Reuse Contact conditions based on literature references.
Cidex Formula 7™ Long-Life Activated Dialdehyde Solution (K924334) Johnson & Johnson Medical Products 2.5% glutaraldehyde Indication for device sterilization.
10 hrs at 20-25°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.
90 min at 25°C
28 days Maximum Reuse
Contact conditions based on literature references.
Cidex PlusTM 28 Day Solution (K923744) Johnson & Johnson Medical Products 3.4% glutaraldehyde Indication for device sterilization.
10 hrs at 20-25°C
28 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.
20 min at 25°C
28 days Maximum Reuse Contact conditions based on literature references.
Wavicide - 01 (K914749) Wave Energy Systems 2.5% glutaraldehyde Indication for device sterilization.
10 hrs at 22°C
30 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test only.
45 min at 22°C
30 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.
Resert XL HLD High Level Disnfectant (K080420) Steris Corporation 2.0% Hydrogen peroxide No indication for device sterilization.
In-Use solution passes the AOAC Sporicidal Activity Test in 6 hrs at 20 °C
8 min at 20ºC
21 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
Opaciden Solution (K070627) Ciden Technologies, LLC

0.60% ortho-phthalaldehyde (OPA)

No indication for device sterilization.
In-Use solution passes the AOAC Sporicidal Activity Test in 32 hrs at 21 ° C and 32 hrs at 25-27°C.
For Manual Processing:
12 min at 20ºC
14 days Maximum Reuse
For processing in a legally marketed Automated Endoscope Reprocessor capable of maintaining solution temperature at 25ºC:
5 min at 25ºC
14 days Maximum Reuse
Contact conditions established by simulated use testing with endoscopes.
Sterilox High Level Disinfectant System (K063159) Sterilox, Technologies, Inc. Hypochlorous acid/hypochlorite
400-450 ppm Active free chlorine
No indication for device sterilization.
Passes the Modified AOAC Sporicidal Activity Test in 32 hrs at 30 ° C.
10 min at 30ºC
Single use - generated on site
Contact conditions established by simulated use testing with endoscopes.`
Cidex Activated Dialdehyde Solution (K060618) Advanced Sterilization Products 2.4% glutaraldehyde 10 hrs at 25 °C
14 days Maximum Reuse
Contact conditions based on AOAC Sporicidal Activity Test and by simulated use testing with endoscopes.
45 min at 25 °C
14 days Maximum Reuse Contact conditions established by simulated use testing with endoscopes.

 