| 07/13/2026 | Pulmonary Valve Catheter Delivery System Recall: Medtronic Removes Harmony Catheter Delivery Catheter System (DCS) | Pulmonary Valve Catheter Delivery System | Recall Confirmed |
| 07/09/2026 | Early Alert: Convenience Kit Issue from Medical Action Industries | Convenience Kits | Early Alert – FDA collecting information |
| 07/09/2026 | Early Alert: Convenience Kits Issue from Medline | Convenience Kits | Early Alert – FDA collecting information |
| 07/07/2026 | Infusion Pump Correction: Fresenius Kabi Issues Correction for Ivenix Large Volume Infusion Pump | Infusion Pump | Recall Confirmed |
| 07/07/2026 | Infusion Pump Correction: Fresenius Kabi Issues Correction for dropped or jarred Ivenix Large Volume Infusion Pumps | Infusion Pump | Recall Confirmed |
| 06/26/2026 | First Aid Kit Recall: North American Rescue Issues Correction for First Aid Kits Containing TRUE METRIX Devices | First Aid Kit | Recall Confirmed |
| 06/18/2026 | Early Alert: Catheter Introducer Issue from Abiomed and Oscor | Catheter Introducer | Early Alert – FDA collecting information |
| 07/06/2026 | Convenience Kit Correction: Windstone Medical Packaging, Inc. Issues Correction for Convenience Kit | Convenience Kit | Recall Confirmed |
| 06/13/2026 | Convenience Kit Correction: Medline Issues Correction for Kits Containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP | Convenience Kit | Recall Confirmed |
| 06/11/2026 | Heart Pump Correction: Abiomed Issues Correction for Automated Impella Controllers | Heart Pump | Recall Confirmed |
| 06/11/2026 | Early Alert: Positive Pressure Breathing Device Issue from Baxter | Positive Pressure Breathing Device | Early Alert – FDA collecting information |
| 07/02/2026 | Breathing Circuit Set Correction: Hamilton Medical Updates Use Instructions for Breathing Circuit Set | Breathing Circuit | Recall Confirmed |
| 06/22/2026 | Infant Resuscitation System Correction: GE HealthCare Issues Correction for Giraffe and Panda Infant Resuscitation Systems and Warmers with a M1091607-R Blender | Infant Resuscitation System | Recall Confirmed |
| 07/02/2026 | Insulin Pump Recall: Insulet Removes Omnipod Pods | Insulin Pump | Recall Confirmed |
| 06/03/2026 | Anesthesia Kit Correction: Becton Dickinson Issues Correction for BD Spinal Trays Containing Bupivacaine Ampules | Convenience Kit | Recall Confirmed |
| 06/02/2026 | Early Alert: Catheter Introducer Kit Issue from Abiomed | Catheter Introducer Kit | Early Alert – FDA collecting information |
| 05/29/2026 | Pediatric Care Bed Correction: KayserBetten Issues Correction for KayserBett IDA Beds | Pediatric Care Bed | Recall Confirmed |
| 05/27/2026 | Early Alert: IV Tubing Set Issue from ICU Medical | IV Tubing Set | Early Alert – FDA collecting information |
| 07/02/2026 | Heart Pump Recall: Abiomed Removes Impella CP Sets with SmartAssist | Heart Pump | Recall Confirmed |
| 05/26/2026 | Manifold Recall: Medline Removes Namic Star Off Handle Manifolds | Manifold | Recall Confirmed |
| 05/26/2026 | Dialysis Catheter Kit Recall: Arrow International Removes Dialysis Catheter Kits Containing Merit Medical Splittable Sheath Introducers | Dialysis Catheter Kit | Recall Confirmed |
| 06/22/2026 | Heart Pump Controller Correction: Abiomed Updates Use Instructions for Automated Impella Controllers | Heart Pump Controller | Recall Confirmed |
| 05/21/2026 | Ventilator Recall: React Health Removes VOCSN V+Pro Ventilators | Ventilator | Recall Confirmed |
| 05/21/2026 | Automated Compounding System Recall: Omnicell Removes Syringe Labels Used with the i.v.STATION Automated Compounding System | Automated Compounding System | Recall Confirmed |
| 06/15/2026 | Anesthesia Machine Correction: Draeger, Inc. Issues Correction for Atlan A350 and A350XL Anesthesia Workstations | Anesthesia Machine | Recall Confirmed |
| 05/07/2026 | Pacemaker Correction: Boston Scientific Issues Correction for ACCOLADE Pacemakers and CRT-Ps | Pacemaker, CRT-P | Recall Confirmed |
| 05/07/2026 | Convenience Kit Recall: Medical Action Industries Removes Convenience Kits Containing Namic RA Syringes | Convenience Kit | Recall Confirmed |
| 05/07/2026 | Convenience Kit Recall: American Contract Systems Removes Convenience Kits Containing Namic RA Syringes | Convenience Kit | Recall Confirmed |
| 05/05/2026 | Surgical Stapler Reload Recall: Intuitive Surgical Removes 8mm SureForm 30 Gray Reloads | Surgical Stapler | Recall Confirmed |
| 05/05/2026 | Sizing Catheter Recall: Cook Medical Removes Various Centimeter Sizing Catheters | Sizing Catheter | Recall Confirmed |
| 04/28/2026 | Blood Glucose Monitor Recall: Trividia Health Issues Correction for TRUE METRIX Blood Glucose Monitoring Systems | Blood Glucose Monitor | Recall Confirmed |
| 05/20/2026 | Thoracic Stent Graft Recall: Bolton Medical Removes Relay Pro System | Thoracic Stent Graft | Recall Confirmed |
| 05/11/2026 | Convenience Kit Recall: Aligned Medical Solutions Removes Convenience Kits Containing Namic RA Syringes | Convenience Kit | Recall Confirmed |
| 04/22/2026 | Hemodialysis Bloodline Set Correction: B. Braun Medical Inc. Issues Correction for Streamline Airless System Hemodialysis Bloodlines and B3 Low Volume Bloodlines | Bloodline Set | Recall Confirmed |
| 04/22/2026 | Insulin Pump Correction: Tandem Diabetes Care Issues Correction for Tandem Mobi Insulin Pumps | Insulin Pump | Recall Confirmed |
| 04/15/2026 | Reconstruction Prosthesis Correction: Stryker Updates Use Instructions for Patient-Fitted TMJ Implants | Reconstruction Prosthesis | Recall Confirmed |
| 05/28/2026 | Ventilator Correction: Philips Issues Correction for Trilogy Evo Platform Ventilators | Ventilator | Recall Confirmed |
| 04/15/2026 | Ventilator Tubing Correction: Percussionaire Corporation Updates Use Instructions for Phasitron In-Line Valve | Ventilator Tubing | Recall Confirmed |
| 04/14/2026 | Anesthesia Workstation Correction: Draeger Issues Correction for Atlan A350 and A350 XL | Anesthesia Workstation | Recall Confirmed |
| 04/14/2026 | Dialysis Catheter Recall: Merit Medical Removes 16F Dual-Valved Splittable Sheath Introducer | Dialysis Catheter | Recall Confirmed |
| 05/12/2026 | Convenience Kit Recall: AVID Medical Removes Convenience Kits Containing Namic RA Syringes | Convenience Kit | Recall Confirmed |
| 04/09/2026 | Ventilator Correction: Baxter Updates Use Instructions for Volara Single-Patient Use Circuits | Ventilator | Recall Confirmed |
| 04/09/2026 | Angiographic Syringe Recall: Medline Industries Removes Namic Angiographic Rotating Adaptor Control Syringes | Angiographic Syringe | Recall Confirmed |
| 03/25/2026 | Heating Pad Recall: Navajo Manufacturing Company Removes Handy Solutions Neck & Shoulders Heating Pad | Heating Pad | Recall Confirmed |
| 04/29/2026 | Insulin Pump Recall: Insulet Removes Certain Omnipod 5 Pods | Insulin Pump | Recall Confirmed |
| 05/28/2026 | Cryoprobe Recall: Erbe USA Removes Flexible Cryoprobes | Cryoprobe | Recall Confirmed |
| 03/06/2026 | Wound and Burn Dressing Recall: Integra LifeSciences Removes Certain MediHoney and CVS Wound and Burn Products | Wound and Burn Dressing | Recall Confirmed |
| 03/05/2026 | Catheter Recall: Medline Industries Removes Reprocessed Electrophysiology and Ultrasound Catheters | Electrophysiology Catheter and Ultrasound Catheter | Recall Confirmed |
| 06/18/2026 | Catheter Recall Expansion: Medline Industries Removes Reprocessed Electrophysiology and Ultrasound Catheters | Electrophysiology Catheter and Ultrasound Catheter | Recall Confirmed |
| 02/25/2026 | Infusion Pump Software Correction: Fresenius Kabi Issues Correction for Ivenix Large Volume Pump Software | Infusion Pump Software | Recall Confirmed |
| 04/03/2026 | Purge Cassette Recall: Abiomed Removes Impella Heart Pump Purge Cassettes | Heart Pump | Recall Confirmed |
| 02/13/2026 | Electronic Homecare Bed Correction: Medline Updates Use Instructions for Homecare Beds | Electronic Homecare Bed | Recall Confirmed |
| 02/05/2026 | Detachable Coil System Recall: J&J MedTech/CERENOVUS Inc. Remove CEREPAK Detachable Coil System | Detachable Coil System | Recall Confirmed |
| 02/04/2026 | Glucose Monitor Sensor Recall: Abbott Diabetes Care Removes Certain FreeStyle Libre 3 and FreeStyle Libre 3 Plus | Continuous Glucose Monitor Sensor | Recall Confirmed |
| 04/03/2026 | Heart Pump Correction: Abiomed Updates Use Instructions for Impella RP with SmartAssist Devices | Heart Pump | Recall Confirmed |
| 02/23/2026 | Insufflation Unit Recall: Olympus Removes High Flow Insufflation Units | Insufflation Unit | Recall Confirmed |
| 02/25/2026 | Stent Recall: Boston Scientific Removes Certain AXIOS Stents and Electrocautery-Enhanced Delivery Systems | Stent | Recall Confirmed |
| 01/16/2026 | Anesthetic Vaporizer Recall: Draeger Removes Vapor 2000 and Vapor 3000 Vaporizers | Anesthetic Vaporizer | Recall Confirmed |
| 01/16/2026 | Anesthesia Kit Recall: Medline Removes Anesthesia Circuits and Anesthesia Circuit Kits | Anesthesia Kit | Recall Confirmed |
| 02/11/2026 | Emergency Response Safety Kit Recall: Airlife Removes Certain Broselow Pediatric Emergency Rainbow Tapes and Broselow ALS Organizers | Broselow Rainbow Tape | Recall Confirmed |
| 03/16/2026 | Dialysis Tubing Set Correction: Vantive Updates Use Instructions for Prismaflex and Oxiris Sets | Dialyzer Accessory | Recall Confirmed |
| 01/09/2026 | Organ Recovery Pack Correction: AVID Medical Issues Correction for Medical Convenience and Organ Recovery Kits | Convenience Kit | Recall Confirmed |
| 01/05/2026 | Infusion Pump Software Recall: Baxter Removes Sigma Spectrum Infusion System Platforms | Infusion Pump Software | Recall Confirmed |
| 01/05/2026 | Esophageal pH Monitoring Capsule Recall: Medtronic and Given Imaging Remove Bravo CF Capsule Delivery Devices | Esophageal pH Monitoring Capsule | Recall Confirmed |
| 01/05/2026 | Diagnostic Intravascular Catheter Recall: Conavi Removes Novasight Hybrid Catheters | Diagnostic Intravascular Catheter | Recall Confirmed |
| 01/05/2026 | Intravascular PICC Catheter Recall: Bard Removes PowerPICC Intravascular Catheters | Intravascular PICC Catheter | Recall Confirmed |
| 01/05/2026 | Aspiration System Correction: Calyxo Updates Use Instructions for CVAC Aspiration Systems | Aspiration System | Recall Confirmed |
| 01/05/2026 | Infusion Pump Recall: Fresenius Kabi Removes Ivenix Large Volume Pumps | Infusion Pump | Recall Confirmed |
| 01/02/2026 | Custom Surgical Pack Recall: Alcon Removes Custom Pak Ophthalmic Procedure Packs | Custom Surgical Pack | Recall Confirmed |
| 12/30/2025 | Intravascular Administration Set Recall: ICU Medical Removes IV Gravity Burette Set | Intravascular Administration Set | Recall Confirmed |
| 12/29/2025 | Continuous Glucose Monitoring Software Correction: Dexcom Issues Correction for Dexcom G6 and G6 Pro Software | Continuous Glucose Monitoring Software | Recall Confirmed |
| 12/23/2025 | Catheter Mount Recall: Draeger Removes ErgoStar Catheter Mounts | Catheter Mount | Recall Confirmed |
| 12/23/2025 | Endovascular Graft Recall: Cook Medical Removes Zenith Alpha 2 Thoracic Endovascular Graft | Endovascular Grafts | Recall Confirmed |
| 12/23/2025 | System Correction: IMPRIS Imaging Inc Issues Correction for IMRIS Neuro III-SV | MRI Systems | Recall Confirmed |
| 03/16/2026 | Convenience Kit Recall: Integra LifeSciences Removes Tuohy Needle | Convenience Kit | Recall Confirmed |
| 12/29/2025 | Anesthesia System Correction: GE HealthCare Updates Use Instructions for CareStation Anesthesia Systems | Anesthesia Systems | Recall Confirmed |
| 12/03/2025 | Correction Alert: Olympus Updates Use Instructions for Ligating Device | Esophageal Ligator | Recall Confirmed |
| 12/01/2025 | Recall Alert: Glycar SA Removes Pericardial Patch | Intracardiac Patch | Recall Confirmed |
| 11/28/2025 | Correction Alert: Becton Dickinson Updates Use Instructions for BD Alaris Pump Issue | Infusion Pumps | Recall Confirmed |
| 11/26/2025 | Recall Alert: BALT USA Removes MEGA Ballast Distal Access Platform | Ventilator Systems | Recall Confirmed |
| 11/26/2025 | Alert: Baxter Permanently Removes Life2000 Ventilation System | Ventilator Systems | Recall Confirmed |
| 11/25/2025 | Max Mobility/Permobil Removes All SpeedControl Dials Used with SmartDrive MX2+ Power Assist Devices | Wheelchair Power Assist | Recall Confirmed |
| 11/21/2025 | Recall Alert: Intersurgical Inc Removes i-View Video Laryngoscope | Laryngoscope | Recall Confirmed |
| 11/21/2025 | Correction Alert: Medline Industries, LP Updates Use Instructions for Medline Kits Containing Flexicare BritePro Solo Fiber Optic Laryngoscope Handles | Medical Convenience Kit | Recall Confirmed |
| 11/20/2025 | Alert: NOxBOX Ltd. Issues Correction for NOxBOXi Nitric Oxide Delivery System Due to Risk for Nitric Oxide Dosing Fluctuations | Nitric Oxide Delivery System | Recall Confirmed |
| 11/20/2025 | Alert: NOxBOX Ltd. Issues Correction for NOxBOXi Nitric Oxide Delivery System Due to Rapid Command Input Issue | Nitric Oxide Delivery System | Recall Confirmed |
| 11/20/2025 | Alert: Medline Industries, LP Removes Certain Sterile Medline Convenience Kits | Convenience Kits | Recall Confirmed |
| 11/19/2025 | Alert: Siemens Healthineers Issues Correction for 3 Tesla MRI Systems | 3T MRI Systems | Recall Confirmed |
| 11/19/2025 | Alert: Update to Use Instructions for B Braun Hemodialysis Blood Tubing Set | Blood Tubing | Recall Confirmed |
| 10/10/2025 | Alert: Automated Impella Controller Correction due to Cybersecurity Issue from Abiomed | Automated Impella Controller | Recall Confirmed |
| 09/26/2025 | Update on Early Alert: Infusion Set Performance Issue from BD | Infusion Set | Recall Confirmed |
| 09/24/2025 | Blood and Plasma Warming Device Correction: 3M Company Issues Correction for Ranger Blood/Fluid Warming System | Blood and Plasma Warming Devices | Recall Confirmed |
| 09/24/2025 | Endoscopic Aspiration Needle Recall: Olympus Removes Certain ViziShot 2 FLEX (19G) Needles Due to Potentially Deformed A-traumatic Tips | Bronchoscope Accessory | Recall Confirmed |
| 09/23/2025 | Catheter Controller Recall: Abiomed Removes Automated Impella Controllers due to Purge Retainer Issues | Catheter Controllers | Recall Confirmed |
| 09/18/2025 | Ablation Catheter Correction: Abbott Updates Use Instructions for TactiFlex Ablation Catheter | Ablation Catheters | Recall Confirmed |
| 09/16/2025 | Electrical Wheelchair Component Correction: mo-Vis BVBA Issues Correction for R-net Joysticks due to a Firmware Error | Electrical Wheelchair Components | Recall Confirmed |
| 09/16/2025 | Convenience Kits Recall: Medline Removes Kits that May Contain Recalled Medtronic Cardiac Cannulas | Cardiac Cannula | Recall Confirmed |
| 09/15/2025 | Insulin Pump Correction: Tandem Diabetes Care Issues Correction for Certain t:slim X2 Insulin Pumps Due to Risk for Faulty Speaker Wiring That May Cause Malfunction and Stop Insulin Delivery | Insulin Pump | Recall Confirmed |
| 09/12/2025 | Continuous Glucose Monitor Apps Correction: Dexcom, Inc. Issues Correction for G7 Apps and ONE+ Apps due to a Software Design Error That Does Not Alert Users of Unexpected Sensor Failure | Glucose Monitors | Recall Confirmed |
| 09/12/2025 | Update on Alert: Defibrillation Lead Issue from Boston Scientific | Defibrillation Systems | Recall Confirmed |
| 09/12/2025 | Ventilator Recall: Philips Respironics Removes Certain DreamStation Devices Due to Programming Errors That Can Result in Failed Therapy Modes | Ventilators | Recall Confirmed |
| 08/29/2025 | Mobile Power Unit AC Power Cord Recall: Abbott Medical Removes AC Power Cord Associated with HeartMate Mobile Power Unit (MPU) Due to Potential Risk of the Cord Not Locking into Place | Heart Pump Accessory | Recall Confirmed |
| 08/29/2025 | Breathing Circuit Set Recall: Hamilton Medical, Inc. Removes Coaxial Breathing Circuit Set Due to Possible Inner Blue Tube Cracks That May Impact Ventilation | Breathing Circuit Sets | Recall Confirmed |
| 08/29/2025 | Manual Resuscitator Recall: Ambu Inc. Removes SPUR II Resuscitators Due to Blocked Manometer Port | Resuscitators | Recall Confirmed |
| 08/28/2025 | Ventilator Correction: Hamilton Medical AG Corrects HAMILTON-C6 Due to Risk of Ventilation Interruption from Defective Circuit Board | Ventilators | Recall Confirmed |
| 08/28/2025 | Applicator Recall: Integra LifeSciences Removes Extended Tip Applicator Due to Potential Sterility and Endotoxin Concerns | Applicators | Recall Confirmed |
| 08/28/2025 | Infusion Pump Correction: ICU Medical, Inc. Issues Correction for Plum Duo Infusion System Due to Software Possibly Resulting in Pump Becoming Unresponsive | Infusion Pump | Recall Confirmed |
| 08/27/2025 | Catheter Controller Recall: Abiomed Removes Automated Impella Controllers due to Pump Driver Circuit Assembly Issues | Catheters | Recall Confirmed |
| 08/22/2025 | Vascular Stent Recall: Boston Scientific Removes Carotid WALLSTENT Monorail Endoprosthesis Due to Risk of Resistance During Device Withdrawal | Vascular Stent | Recall Confirmed |
| 12/22/2025 | Cardiac Cannula Recall: Medtronic Removes DLP Left Heart Vent Catheter | Cardiac Cannula | Recall Confirmed |
| 08/15/2025 | Infusion Pumps Correction: Baxter Issues Correction for Novum IQ Infusion Pumps | Infusion Pump | Recall Confirmed |
| 08/12/2025 | Recall: Draeger Removes SafeStar and TwinStar Filters due to Risk of Misleading Carbon Dioxide (CO₂) Readings | Breathing System Filters | Recall Confirmed |
| 08/06/2025 | Vascular Access Systems Correction: Boston Scientific Updates Use Instructions for WATCHMAN Access Systems | Vascular and Transseptal Access Systems | Recall Confirmed |
| 08/04/2025 | Continuous Ventilator Correction: Philips Respironics Updates Use Instructions for BiPAP A30, A40, and V30 Devices Due to Interruptions and/or Loss of Therapy | Continuous Ventilator | Recall Confirmed |
| 07/25/2025 | Disposable Surgical Stapler Correction: Ethicon Endo-Surgery, LLC Issues Correction for Endopath Echelon to Address Inadvertent Lockout During Surgical Procedures | Disposable Surgical Staplers | Recall Confirmed |
| 07/24/2025 | Arterial Cannula Recall: Edwards Lifesciences Removes Arterial Cannula due to Risk of Wire Exposure | Arterial Cannula | Recall Confirmed |
| 07/23/2025 | Mobile Lift Component Recall: Baxter Healthcare Corporation Removes Mobile Lift Component due to Risk of Improper Attachment | Mobile Lift Component | Recall Confirmed |
| 12/22/2025 | Infusion Pump Correction: Baxter Issues Correction for Novum IQ Large Volume Pumps | Infusion Pump | Recall Confirmed |
| 07/22/2025 | Medical Procedure Kits Correction: Medline Industries, LP Issues Correction for Medline Craniotomy Kits Containing Codman Disposable Perforators Due to Risk for Device Disassembly | Medical Procedure Kits | Recall Confirmed |
| 07/21/2025 | Continuous Ventilator (Respirator) Correction: Maquet Critical Care AB Updates Use Instructions for Servo Ventilator Systems Due to Risk of Inaccurate Compliance Measurement Leading to Improper Tidal Volume Delivery in Neonatal Patients | Continuous Ventilator | Recall Confirmed |
| 07/21/2025 | Applicator Recall: Integra LifeSciences Removes MicroMyst Applicators Due to Potential Sterility Concerns | Piston Syringe | Recall Confirmed |
| 07/18/2025 | Manual Resuscitator Recall: SunMed Holdings, LLC Removes Adult Manual Resuscitator Devices Due to Incorrect Assembly of B/V Filter | Manual Resuscitator | Recall Confirmed |
| 07/17/2025 | Continuous Glucose Monitor Receiver Recall: Dexcom, Inc. Removes Certain Dexcom G6, G7, ONE, and ONE+ Receivers Due to Speaker Malfunction That May Cause Missed Alerts for Dangerous Blood Sugar Levels | Continuous Glucose Monitor | Recall Confirmed |
| 07/17/2025 | Portable Oxygen Concentrator Car Adapter Recall: Drive DeVilbiss Healthcare Removes iGo DV6X-619 DC Car Adapter Due to Cord Possibly Becoming Hot to the Touch and/or Melting While Being Used | Oxygen Concentrator | Recall Confirmed |
| 12/22/2025 | Microbore Extension Set Recall: B. Braun Medical Inc. Removes Microbore Extension Sets | Microbore Extension Set | Recall Confirmed |
| 04/23/2026 | Cranial Drill Recall: Integra LifeSciences Recalls Codman Disposable Perforators Due to Risk of Device Disassembly | Cranial Drill | Recall Confirmed |
| 07/10/2025 | Closed Suction Catheter Recall: Avanos Medical, Inc. Removes Ballard Closed Suction Systems Due to Risk of Non-Sterility | Closed Suction Catheter | Recall Confirmed |
| 07/09/2025 | Infant Breathing System Recall: AirLife/Vyaire Removes Infant Heated Wire Circuits Due to Risk for Inadvertent Adapter Disconnection During Ventilation | Infant Breathing System | Recall Confirmed |
| 07/09/2025 | Broselow Pediatric Emergency Rainbow Tape Recall: AirLife Removes Certain Broselow Pediatric Emergency Rainbow Tapes due to Misprinted Information | Pediatric Emergency Rainbow Tape | Recall Confirmed |
| 12/22/2025 | Blood Pump Controller Correction: Abiomed Updates Use Instructions for Automated Impella Controllers | Blood Pump Controller | Recall Confirmed |
| 06/25/2025 | Angiographic Catheter Recall: Cook Removes Beacon Tip Angiographic Catheters due to Tip Separation | Angiographic Catheter | Recall Confirmed |
| 06/23/2025 | Resuscitation System Recall: ZOLL Circulation, Inc. Recalls AutoPulse NXT Resuscitation System Due to a Failure Code | Resuscitation System | Recall Confirmed |
| 06/23/2025 | Medical Procedure Kits Correction: Medline Industries, LP Issues Correction for Medline Procedure Kits Containing Medtronic Aortic Root Cannula due to Potential Excess Material in Male Luers | Medical Procedure Kits | Recall Confirmed |
| 06/23/2025 | Anesthesia Delivery Systems Recall: GE HealthCare Issues Correction for Certain Carestations due to Risk of Ineffective Ventilation | Anesthesia Delivery System | Recall Confirmed |
| 06/17/2025 | Blood Products Administration Set Recall: Fresenius Kabi Removes Large Volume Pump Blood Products Administration Set due to Incorrect Assembly | Blood Products Administration Set | Recall Confirmed |
| 06/17/2025 | Convenience Kit Recall: Medline Neonatal and Pediatric Kits Containing Smiths Medical ORAL/NASAL Endotracheal Tubes due to Unexpectedly Smaller Tube Diameter | Neonatal and Pediatric Convenience Kits | Recall Confirmed |
| 06/17/2025 | Aspiration Catheter Recall: Q’Apel Medical, Inc. Removes Hippo 072 Aspiration System, Cheetah Delivery Tool | Aspiration Catheter | Recall Confirmed |
| 06/16/2025 | Infusion Pump Recall: Zyno Medical Removes Certain Z-800 Series Infusion Pumps due to Software Issue | Infusion Pump | Recall Confirmed |
| 06/13/2025 | Intra-Operative Positioning System Guidewire Recall: Centerline Biomedical Removes Certain IOPS Guidewires due to Delamination | IOPS Guidewire | Recall Confirmed |
| 06/13/2025 | Liquid Bicarbonate Concentrate Recall: Nipro Removes MedicaLyte Liquid Bicarbonate Concentrate due to Contamination | Liquid Bicarbonate Concentrate | Recall Confirmed |
| 06/05/2025 | Medtronic Removes Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula Due to Risk for Disconnection of the Flange from the Device Cannula | Tube tracheostomy and tube cuff | Recall Confirmed |
| 06/05/2025 | Anesthesia Breathing Circuit Correction: Draeger, Inc. Updates Use Instructions for VentStar Flex and Anesthesia Circuit Kits Due to Risk of Inadequate Ventilation from Cracked Hoses | Anesthesia Breathing Circuit Kits | Recall Confirmed |
| 06/05/2025 | Infusion Pump Correction: Baxter Issues Correction for Novum IQ Large Volume Pump due to Potential for Underinfusion | Infusion Pump | Recall Confirmed |
| 06/03/2025 | False Alarm Issue with Infusion Pump from Smiths Medical | Infusion Pump | Recall Confirmed |
| 06/03/2025 | Infusion Pump Thermal Damage Issue from Smiths Medical | Infusion Pump | Recall Confirmed |
| 06/03/2025 | Infusion Pump Intermittent Connection Issue from Smiths Medical | Infusion Pump | Recall Confirmed |
| 05/23/2025 | Esophagogastric Tube Recall: BD Issues Correction for Esophagogastric Balloon Tamponade Tubes due to Challenges Removing Plastic Plugs from Rubber Lumen | Esophagogastric Tube | Recall Confirmed |
| 04/24/2025 | Heart Pump Accessory Removal: Abbott Removes HeartMate Mobile Power Unit due to Instances of Sudden Power Loss | Heart Pump Accessory | Recall Confirmed |
| 03/31/2025 | Vascular Cannula Recall: Medtronic Removes Aortic Root Cannula Due to Unexpected Loose Material in the Male Luer | Aortic Cannula | Recall Confirmed |
| 03/25/2025 | Endotracheal Tube Recall: Smiths Medical Removes Intubation ORAL/NASAL Endotracheal Tubes Due to Smaller Than Expected Tube Diameter That May Cause Underventilation | Endotracheal Tube | Recall Confirmed |
| 03/20/2025 | Infusion Pump Software Correction: Becton, Dickinson and Company (BD) Issues Correction for BD Alaris Systems Manager and Care Coordination Engine Infusion Adapter Software Due to Risk for Outdated Automated Programming Requests to Load | Infusion Pump Software | Recall Confirmed |
| 03/19/2025 | Implantable Port Recall: Smiths Medical Removes ProPort Plastic Implantable Ports Due to Manufacturing Error that May Cause Separation | Implantable Port | Recall Confirmed |
| 03/18/2025 | Embolization Device Recall and Correction: Medtronic Removes Unused 027 Compatible Pipeline Vantage Embolization Device with Shield Technology, Updates Use Instructions for 021 Compatible Pipeline Vantage Embolization Devices due to Increased Risk of Incomplete Wall Apposition and Braid Deformation | Embolization Device | Recall Confirmed |
| 03/20/2025 | Update on Alert: Infusion Pump Issue from Baxter Healthcare Corporation | Infusion Pump | Recall Confirmed |
| 03/03/2025 | Endovascular System Recall: Philips Removes and Discontinues Distribution of Tack Endovascular System Due to User Challenges That Caused Additional Procedures to Reposition or Remove Implant | Endovascular System | Recall Confirmed |
| 03/03/2025 | Vaporizer Recall: Getinge Removes Vaporizer Sevoflurane Quick-Fil and Expands Recall of Vaporizer Sevoflurane Maquet Filling due to Risk of Patient and Health Care Professional Exposure to Toxic Chemical Hydrogen Fluoride | Vaporizer | Recall Confirmed |
| 02/28/2025 | Ablation Catheter Correction: Biosense Webster Updates Use Instructions for Varipulse due to High Rate of Stroke or Transient Ischemic Attack | Ablation Catheter | Recall Confirmed |
| 02/28/2025 | Patient Table Correction: Philips Updates Use Instructions for Allura and Azurion Systems due to Patient Fall Risk from Incorrectly Positioned Mattresses | Patient Table | Recall Confirmed |
| 02/28/2025 | Power Assist Device Recall: Max Mobility/Permobil Removes SpeedControl Dial Component Used with SmartDrive MX2+ Power Assist Device Due to Risk for Motor to Be Unresponsive to the User | Control Dial | Recall Confirmed |
| 02/28/2025 | Regard Newborn Kit Recall: ROi CPS, LLC Removes Certain Newborn Kits Due to a Recalled Component, the Neo-Tee T-Piece Resuscitator With An Undersized Controller Spring | Resuscitator | Recall Confirmed |
| 02/28/2025 | Endoscope Instrument Recall: Olympus Removes Single Use Guide Sheath Kits Due to Risk for Radiopaque Guide Sheath Tip Detaching During Procedures | Sheath Kit | Recall Confirmed |
| 02/21/2025 | Pacemaker Recall: Boston Scientific Corporation Recalls Accolade Pacemaker Devices Due to a Manufacturing Issue That May Lead to Early Device Replacement | Pacemaker | Recall Confirmed |
| 02/14/2025 | Heart Pump Recall: Abiomed, Inc. Updates Use Instructions for Impella RP with SmartAssist and Impella RP Flex with SmartAssist Due to A Risk That the Tip of Guidewires or Other Medical Devices May Come into Contact with The Impella Pump During Insertion, Adjustment, Or Removal | Heart Pump | Recall Confirmed |
| 02/10/2025 | Breathing Circuit Kit Recall: Sentec/Percussionaire Removes VDR4 Phasitron Breathing Circuits due to Venturi Component Malfunctions that May Reduce Pressure and Volume Flow | Breathing Circuit | Recall Confirmed |
| 02/06/2025 | Arterial Catheter Recall: Medline Industries, LP, Removes Integrated Arterial Catheters due to Excess Material at Catheter Hub that May Detach | Catheter Kit | Recall Confirmed |
| 03/20/2025 | Update on Alert: Atherectomy Catheter System Issue from Bard Peripheral Vascular | Atherectomy Catheter System | Recall Confirmed |
| 02/05/2025 | Oxygen Concentrator Recall: JIANGSU JUMAO X-CARE MEDICAL EQUIPMENT CO LTD Removes JMC5A Ni/TruAire-5 Oxygen Concentrator due to the devices spontaneously catching fire | Oxygen Concentrator | Recall Confirmed |
| 02/05/2025 | Continuous Ventilator Correction: Baxter Healthcare Corporation Issues Correction for Life2000 Ventilator due to a nonconforming battery charger, which triggers a battery alarm and renders the ventilator inoperable | Ventilator | Recall Confirmed |
| 02/03/2025 | Pressure Monitoring Device Recall: Medtronic Neurosurgery Issues Correction for Becker and Exacta External Drainage and Monitoring Systems Due to Cracks And/Or Leaks in Stopcocks | Monitoring System | Recall Confirmed |
| 01/31/2025 | Gas Powered Emergency Resuscitator Recall: Mercury Medical Removes Neo-Tee T-Piece Resuscitator Due to An Undersized Spring in The Controller That May Prevent the Device from Delivering the Required Pressure Levels Needed for Effective Ventilation | Resuscitator | Recall Confirmed |
| 01/30/2025 | Glucose and Glucose/Ketone Meter Correction: Nova Biomedical Corporation Issues Software Correction for StatStrip Glucose and Glucose/Ketone Hospital Meters Due to Risk for Transmission of Incorrect Patient Results | Glucose Strip/Meter | Recall Confirmed |
| 03/05/2025 | Update on Alert: Infusion Pump Software Issue from Fresenius Kabi USA | Infusion Pump Software | Recall Confirmed |
| 07/15/2025 | Endoscopic Vessel Harvesting (EVH) System Correction: Getinge and Maquet Cardiovascular Update Use Instructions for VasoView HemoPro 2 (VH-4000 and VH-4001) EVH Systems due to Risk for Bent or Detached Heater Wires and Silicone Peeling or Detaching During Use | Endoscopic Vessel Harvesting (EVH) System | Recall Confirmed |
| 01/13/2025 | Outpatient Telemetry Correction: Philips Issues Correction for Monitoring Service Application used with Mobile Cardiac Outpatient Telemetry Due to Potential for Missed Information or Notifications That May Impact Patient Care | Mobile Cardiac Telemetry Monitoring | Recall Confirmed |
| 03/05/2025 | Update on Alert: Solution Set Issue from Baxter Healthcare Corporation | Solution Sets | Recall Confirmed |
| 01/31/2025 | Update to Alert: Fluid Delivery Set Issue from Medline | Fluid Delivery Set | Recall Confirmed |
| 01/31/2025 | Update to Alert: Extracorporeal Blood Circuit Issue from Nuwellis | Single Use Blood Circuit | Recall Confirmed |
| 01/31/2025 | Update to Alert: Endoscope Accessories Forceps/Irrigation Plug Issue from Olympus | Endoscope Accessory | Recall Confirmed |
| 12/20/2024 | Infusion Pump Battery Recall: ICU Medical Removes Some CSB Batteries Intended For Use With Plum 360, A+, and A+3 Infusion Systems Due to Reports of Allegedly Counterfeit, Untested Batteries In Use | Infusion Pumps | Recall Confirmed |
| 01/31/2025 | Update to Alert: Nephroscope Sheath Issue from Trokamed GmbH | Nephroscope Sheath | Recall Confirmed |
| 12/18/2024 | Catheter Correction: Boston Scientific Updates Use Instructions for POLARx and POLARx FIT Cryoablation Balloon Catheters due to Higher than Anticipated Number of Reports of Atrio-Esophageal Fistula | Balloon Catheter | Recall Confirmed |
| 12/18/2024 | Implantable Marker Recall: Hologic Removes BioZorb 3D Bioabsorbable Markers due to Risk for Patient Complications | Implantable Radiographic Marker | Recall Confirmed |
| 01/31/2025 | Update to Alert: Infusion Pump Issue from Fresenius Kabi USA | Infusion Pumps | Recall Confirmed |