Early Alert: Convenience Kit Issue from AVID Medical

CDRH is issuing this Early Alert to notify the public of a potentially high-risk device issue. The FDA will keep the public informed and update this web page as significant new information becomes available.  

Affected Product

The FDA is aware that AVID Medical has issued an Urgent Medical Device Recall Notification to affected customers recommending certain angiographic syringes included in convenience kits be removed from where they are used or sold. Affected devices: 

Description (Tray Name)Model/Item NumberUDI-DI
CARDIAC CATH PACKEAMC1000-0510809160377314
SAMMC ANGIOGRAPHYSAMM066-1510809160388495
HEART CATH, SELF REGIONALSELF131-0510809160408650
VA DETROIT CATH LAB PACKVMCD02910809160388969
CATH LABSACL75AM20809160481377
CARDIAC CATH TRAY SHANDS JAXSHND401-0410809160395615
CARDIAC CATH PACKWAFB208-0210809160424759
PERC TRAYSLPC34-0110809160511213
CARDIAC CATH PACK SAN23CARDQDRCC3610809160469088
NEURO IR PERIPHERAL (PS 64405)UINR44-0120809160454439
CATH LABSACL75-0120809160481377

Full List of Affected Lots

What to Do

Identify kits with affected syringes. Apply over-labels to all affected kits on hand stating that the affected syringes must be removed and discarded from further use.  

On March 13, AVID Medical sent all affected customers a letter recommending the following actions:

  • Identify, segregate, and quarantine all affected product.
  • Add warning labels to all affected kits stating that the affected syringes must be removed and discarded from further use.
    • Use the labeling template provided by AVID to print labels. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling.
  • If you are a distributor or have resold or transferred this product, notify them of this recall communication.
  • Per Medline, due to the risk of serious injury or death, customers are directed to remove and destroy all Namic RA Syringes. If use is unavoidable because failure to proceed would result in patient harm, the Namic RA syringe must be used with extreme caution and vigilance, including manual stabilization of syringe-to-manifold connections and continuous, 100% monitoring during setup and throughout the entire procedure.
  • All usage guidelines and the Instructions for Use are provided in Medline’s recall letter. Strict adherence to all usage guidelines and the Instructions for Use are required.

Check this web page for updates. The FDA is currently reviewing information about this potentially high-risk device issue and will keep the public informed as significant new information becomes available.  

Reason for Alert

AVID Medical initiated an urgent Medical Device Recall notification for convenience kits that contain syringes affected by Medline’s Namic Angiographic Control Syringes with Rotating Adaptor (“Namic RA Syringes”) recall.  

Medline Industries stated that the syringe rotating adaptor on affected devices may unwind during use, resulting in a loose connection or full disconnection between the syringe and manifold. If unwinding occurs, there is potential for biohazard exposure, blood loss, and infection. Loose connections or disconnections may introduce air into the line, which could result in air embolism. These conditions may lead to serious injury or death. All units have the potential to exhibit this failure mode.

As of March 13, Medline Industries has reported four serious injuries and no deaths associated with this issue.

Device Use

Convenience kits contain various surgical instruments, dressings and/or other materials, and pharmaceutical components that are intended to be used during various surgical procedures.  

Contact Information

Customers in the U.S. with adverse reactions, quality problems, or questions about this issue should contact AVID Medical at complaints@owens-minor.com.  

Additional FDA Resources  

Unique Device Identifier (UDI)

The unique device identifier (UDI) helps identify individual medical devices sold in the United States from distribution to use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified more quickly, and as a result, problems potentially resolved more quickly.

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program. 