Convenience Kit Recall: Medline Removes Convenience Kits Affected by BD ChloraPrep Applicator Recall
The affected products and recommendations for what to do with the devices below have not changed. This recall involves removing certain devices from where they are used or sold. The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it.
Affected Product
The FDA is aware that Medline has issued a letter to affected customers recommending certain convenience kits affected by the BD ChloraPrep Applicator Recall be removed from where they are used or sold.
Affected devices:
- Medline Convenience Kits that contain certain BD ChloraPrep products:
- ChloraPrep Clear 1 mL Single Sterile applicators
- Item Number: 930480
- Lot: 4032183
- FREPP Clear 1.5 mL applicators with paper lidding
- Item Number: 930299
- Lot: 4073005
- ChloraPrep Clear 1 mL Single Sterile applicators
- Full List of Affected Product
What to Do
Review inventory for convenience kits affected by this issue, quarantine, and destroy the entire kit after completion of response forms. Be aware that these kits were previously identified as part of a recall and may have over-labels instructing users to remove the discard the affected component at point of use.
On August 11, Medline sent all affected customers a letter recommending the following actions:
- Be aware that the actions required for certain products identified in a prior Medline notification have changed:
- Customers are instructed to destroy the entire convenience kit for products identified in this communication due to potential contamination.
- The prior Medline notification issued on June 12, 2026, instructed customers to over-label affected convenience kits with a sticker indicating that the affected component should be removed and discarded prior to use.
- This updated action applies only to the items listed in this notification and does not affect any other convenience kits included in the June 12, 2026, notification.
- Immediately check your stock for convenience kits affected by this issue. Quarantine all affected product.
- Destroy affected product upon completion of your response form.
- If you are a distributor or have resold or transferred this product to another company or individual, notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers’ quantities on your response form.
Reason for Recall
Medline stated that certain convenience kits containing recalled BD ChloraPrep applicators should be removed from use due to fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides.
This action applies to products previously identified in a Medline notification sent on June 12. Be advised that the actions required for certain lots in the original notification have changed based on additional recall details received from BD (Becton, Dickinson and Company) regarding their ChloraPrep applicators. Following the initial notification, BD provided additional information regarding its recall of item number 930480, lot 4032183 of ChloraPrep Clear 1 mL Single Sterile applicators and item number 930299, lot 4073005 of FREPP Clear 1.5 mL applicators with paper lidding.
Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death. If the fungus is introduced into a patient's bloodstream during placement of an intravascular catheter, the catheter would likely need to be removed, requiring an additional procedure. Infection at a surgical site may also necessitate medical and surgical intervention, as well as long term treatment with antifungal medications.
As of August 17, Medline has not reported any serious injuries or deaths associated with this issue.
Device Use
Convenience kits contain various surgical instruments, dressings and/or other materials, and pharmaceutical components that are intended to be used during various surgical procedures.
Contact Information
Customers in the U.S. with adverse reactions, quality problems, or questions about this issue should contact Medline at 866-359-1704 or recalls@medline.com.
Additional FDA Resources
- FDA Enforcement Report
- CDRH Medical Device Recall Database
- BD ChloraPrep Applicator Recall: FDA Enforcement Report
- Convenience Kit Correction: Medline Issues Correction for Convenience Kits Containing BD ChloraPrep Applicators
Unique Device Identifier (UDI)
The unique device identifier (UDI) helps identify individual medical devices sold in the United States from distribution to use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified more quickly, and as a result, problems potentially resolved more quickly.
- How do I recognize a UDI on a label?
- AccessGUDID database - Identify Your Medical Device
- Benefits of a UDI System
How do I report a problem?
Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program.
Timeline of Communication Updates
| Date | Actions |
|---|---|
| 10/13/2026 | The FDA updated this communication to inform the public that this issue has been classified as a Class I Recall. |
| 08/24/2026 | The FDA issued an Early Alert communication to notify the public of a potentially high-risk device issue. |