U.S. flag An official website of the United States government
  1. Home
  2. Medical Devices
  3. Medical Device Safety
  4. Letters to Health Care Providers
  5. Non-sterile, Single-use Pneumatic Tourniquet Cuffs Conservation Strategies - Letter to Health Care Providers
  1. Letters to Health Care Providers

Non-sterile, Single-use Pneumatic Tourniquet Cuffs Conservation Strategies - Letter to Health Care Providers

June 5, 2023

The U.S. Food and Drug Administration (FDA) is aware of health care facilities and providers experiencing localized supply constraints of non-sterile, single-use pneumatic tourniquet cuffs in the United States. These devices are mainly used in elective limb surgeries and in emergency and trauma settings. During interruption in the supply of these devices, the FDA is recommending health care facilities and health care providers implement conservation strategies such as using alternative devices and reusing existing devices after appropriate sterilization/disinfection procedures have been appropriately followed.

Recommendations

Health care facilities and providers should use non-sterile, single use pneumatic tourniquet cuffs as supplies allow. When non-sterile, single use pneumatic tourniquet cuffs are not available in sufficient supply, consider the following recommendations, including conservation strategies, to maintain the quality and safety of patient care:

  • Implement conservation strategies such as limiting use of pneumatic tourniquets, based on surgical flow needs. 
  • Reuse non-sterile, single-use pneumatic tourniquet cuffs by following either your institution’s procedures, if available, or the steps outlined below as you deem appropriate: 
    • Ensure tourniquet cuffs can be reused. Discard if broken, damaged, heavily contaminated (with blood or other fluids), or if a cleaned cuff is unable to hold the pressure at 220-300mmHg range. 
    • Clean tourniquet cuffs even though they may not have visible contamination or debris, including particulate removal (using a lint roller), stain removal (brush and formulated enzymatic cleaner), and hose cleaning (with adhesive remover). 
    • Sterilize or disinfect tourniquet cuffs: 
      • To sterilize tourniquet cuffs, contact the manufacturer for appropriate sterilization methods.
      • To disinfect the tourniquet cuff, leave the device soaked in 70%-90% ethyl or isopropyl alcohol for 1 minute, or wet spray the cuff with an exposure time of at least one minute. Allow the alcohol to evaporate and then air dry completely. There is no need to rinse with clean water. Do not use disinfectant wipes or high-level disinfectants. 
    • Keep sterilized or disinfected tourniquet cuffs clean until next use. 
    • Separate the sterilized tourniquet cuffs from disinfected (may not be sterile) ones to avoid cross contamination during surgeries. You may use a non-permeable drape placed over the tourniquet cuff to reduce a potential risk of contamination with body fluids. 
    • Collect used cuffs and repeat steps above as needed. You may number the tourniquet cuffs to keep track of re-sterilized and re-disinfected tourniquet cuffs. Consult with the manufacturer for maximum number of reuses.
  • Use non-pneumatic tourniquet cuffs in elective limb surgical procedures.
    • Although non-pneumatic tourniquets are most frequently used in the prehospital, point-of-care setting, they can also be used in the other care settings, including the operating room.
    • Read device instructions and ensure operating room staff are properly trained in their use. 
    • Never place the non-pneumatic tourniquet cuff over joints or around the neck.
    • In elective surgeries, never place the tourniquet directly on the skin.
    • Dispose of the tourniquet after each use. Non-pneumatic tourniquets should not be re-used or reprocessed (that is, disinfected or sterilized).
  • Ensure the maximum number of times each cuff brand can be safely reused, after cleaning and sterilization or cleaning and disinfection, as stated above.
  • Encourage the use of appropriate third-party facilities to reprocess single-use devices for reuse, especially among health care providers or health care settings that may lack the capabilities to reprocess medical devices labeled for single use.
  • Contact the FDA at deviceshortages@fda.hhs.gov as well as your group purchasing organization, local product representative, distributor, or account manager if the conservation strategies are not able to maintain sufficient supply. 
  • Consider recommendations from the FDA as well as relevant professional organizations for other strategies that might be appropriate for your organization.

Background 

A recent recall of Non-Sterile Disposable Tourniquet Cuffs in an already constrained market has led to localized supply constraints of these devices. Tourniquet cuff manufacturers are aware of this situation and working towards increasing device availability. 

FDA Actions

The FDA will continue to monitor the situation to evaluate possible manufacturing disruptions and help ensure availability of non-sterile, single use pneumatic tourniquet cuffs for health care facilities and providers.

The FDA will keep health care facilities, providers, and the public informed if significant new information becomes available.

Reporting Problems to the FDA

The FDA encourages health care providers to report any adverse events or suspected adverse events experienced with any non-sterile, single-use pneumatic tourniquet cuffs. 

Prompt reporting of adverse events can help the FDA identify and better understand the risks associated with medical devices.

Contact Information

If you have questions about this letter, contact the Division of Industry and Consumer Education (DICE).
 

 

Subscribe to Medical Device Safety and Recalls

Sign up to receive email updates on medical device recalls, safety communications, and other safety information.

Back to Top