Medical Devices
Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions; Final Guidance for Industry and FDA Staff
Document issued on: November 13, 2000
This document supersedes Guidance for Cardiopulmonary Bypass Oxygenators 510(k)
Submissions; Final January 17, 2000
![]() |
U.S. Department of Health and Human Services Circulatory Support and Prosthetic Devices Branch |
Preface
Public Comment:
Comments and suggestions may be submitted at any time for Agency consideration to Dockets Management Branch, Division of Management Systems and Policy, Office of Human Resources and Management Services, Food and Drug Administration, 5630 Fishers Lane, Room 1061, (HFA-305), Rockville, MD, 20852. When submitting comments, please refer to the exact title of this guidance document. Comments may not be acted upon by the Agency until the document is next revised or updated.
For questions regarding the use or interpretation of this guidance contact Catherine Wentz at (240) 276-4141 or by e-mail at catherine.wentz@fda.hhs.gov.
Additional Copies:
Additional copies are available from the Internet or CDRH Facts-on-Demand. In order to receive this document via your fax machine, call the CDRH Facts-On-Demand system at 800-899-0381 or 301-827-0111 from a touch-tone telephone. Press 1 to enter the system. At the second voice prompt, press 1 to order a document. Enter the document number 1361 followed by the pound sign (#). Follow the remaining voice prompts to complete your request.
Table of Contents
4.2 Preparation of the Test Oxygenator
4.3 Biological Compatibility
4.4 Physical Characterization/Integrity
4.4.2 Heat Exchanger Fluid Pathway Integrity
4.4.3 Gas Pathway Integrity
4.4.4 Blood Volume Capacity of Oxygenator
6.1.2 Blood Damage Testing Protocol
6.1.3 Blood Damage Data Reporting
6.3 Biological/ Material Compatibility
Guidance for Cardiopulmonary Bypass Oxygenator 510(k) Submissions
| This document is intended to provide guidance. It represents the Agency’s current thinking on this topic. It does not create or confer any rights for or on any person and does not operate to bind Food and Drug Administration (FDA) or the public. An alternative approach may be used if such approach satisfies the requirements of the applicable statute, regulations, or both |
This guidance document describes a means by which cardiopulmonary bypass oxygenator devices may comply with the requirement of special controls for class II devices. Designation of this guidance document as a special control means that manufacturers attempting to establish that their device is substantially equivalent to a predicate cardiopulmonary bypass oxygenator device should demonstrate that the proposed device complies with either the specific recommendations of this guidance or some alternate control that provides equivalent assurances of safety and effectiveness.
The Least Burdensome Approach
The issues identified in this guidance document represent those that we believe need to be addressed before your device can be approved/cleared for marketing. In developing the guidance, we carefully considered the relevant statutory criteria for Agency decision-making. We also considered the burden that may be incurred in your attempt to comply with the guidance and address the issues we have identified. We believe that we have considered the least burdensome approach to resolving the issues presented in the guidance document. If, however, you believe that information is being requested that is not relevant to the regulatory decision for your pending application or that there is a less burdensome way to address the issues, you should follow the procedures outlined in the "A Suggested Approach to Resolving Least Burdensome Issues" document.
| 1. |
Device Identification"Cardiopulmonary bypass oxygenator (21 CFR §870.4350) - A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery." The device is intended for use up to six hours in duration. |
||||||||||||||
2. |
|
||||||||||||||
3. |
|
Test oxygenator |
the oxygenator submitted for 510(k) regulatory clearance. |
Aged test oxygenator |
the test oxygenator after it has undergone appropriate accelerated or real time aging. |
Predicate oxygenator |
a similar oxygenator to the test oxygenator that has previously been cleared by FDA and is currently in commercial distribution in the United States. |
Control blank circuit |
a recirculation circuit that does not include an oxygenator but is otherwise identical to those circuits used to test oxygenators for blood damage. |
Static control blood |
a small volume of blood that is not circulated in a circuit and used to monitor autohemolysis. |
Index of Hemolysis |
the mass of hemoglobin released from the red blood cells per 100L of blood pumped through the device. |
Modified Index of Hemolysis |
the ratio of the amount of hemoglobin released into the plasma normalized by the total amount of hemoglobin pumped through the device. |
4.
Oxygenator Testing
To evaluate substantial equivalence of a cardiopulmonary bypass oxygenator, provide data addressing the biological, material, physical, and performance characteristics over the expected storage (shelf life) and use lifetime of the oxygenator. The potential failure modes for oxygenators, which should be investigated thoroughly, include leaks, toxicity, loss of gas transfer efficiency, gas embolism, thromboembolism, and blood damage. The oxygenator should be tested over its entire performance specification range under expected use conditions for six hours in duration. Integrity testing should be conducted for 6 hours. In vitro testing should be conducted according to an established protocol using at least five devices for each test. Submit the testing protocols along with the results of the tests reported in a statistically meaningful manner. Include the range of values, mean, standard deviation and standard error of the mean for each data set. For any comparative test, provide the p-value or similar measure indicating statistical significance of the comparison. Calibrate all instruments and equipment used in conducting these tests, to minimize the limit of testing error. Accuracy of the test apparatus should conform to that in section 11.1.2 of standard CAN/CSA-Z 364.3M90 (1990) or equivalent.
4.5
|
|
| 4.5.1 |
Use fresh whole animal blood collected and refrigerated for less than 24 hours for performance testing. Bovine blood is most commonly used in this type of testing. To minimize the effects of interanimal variation, it is advised that the common blood pool be composed of blood from more than one animal. To simulate clinical usage, FDA and ISO 7199 recommend anticoagulation of the blood with heparin (e.g. 4500 units of heparin per liter of blood). |
||||||
4.5.2 |
The oxygenator's gas transfer characteristics, and blood side pressure drop are evaluated using operational variable settings that span the manufacturer's recommended operating range of six hours.
|
|
TABLE 1. Oxygen Gas Transfer Data Set* *Provide a table of data for each of the three different blood flow rates (minimum, nominal, and maximum) per the manufacturer’s recommendation, and according to the labeling. FDA suggests the table format given below.
+ Oxygen Transfer Rate [ml O2 (STPD)/min] = Q x (CaO2 outlet - CvO2 inlet CvO2 (or CaO2) [ml O2/L blood] = 12 g Hb/ 100 ml blood x 1.34 ml O2/ 1 Gm Hb x 1000 ml blood/ 1 L blood x (% O2 saturation) + (PO2 mmHg x 0.00314 ml O2/ (100 ml blood) (mmHg) x 1000 ml blood/ 1 L blood) TABLE 2. Carbon Dioxide Gas Transfer Data Set* * Provide a table with data for each of the three different blood flow rates (min, max, and nominal). FDA suggests the table format given below.
The gas flow entering and leaving the oxygenator should be analyzed for CO2. The CO2 transfer rate is calculated as follows: + CO2 Transfer Rate [ml CO2 (STPD)/min] = F x % CO2 F = gas flow rate, exiting the blood-gas exchange device, L/min. TABLE 3. Blood Pressure Drop Data Set* *Provide a table of data for each of the three different blood flow rates (minimum, nominal, and maximum). Hydrostatic differences between the locations of the inlet and outlet pressure transducers should be accounted for in the table. FDA suggests the table format given below.
Provide graphs along with the above tables showing gas transfer rate (O2 and CO2) as a function of blood flow rate and gas flow rate. Determine the blood flow rate at which the gas exchange is sufficient to cause the outlet O2 saturation to be 95% O2 saturation. Provide a graph along with the above data tables showing the blood side pressure drop as a function of blood flow rate. To better simulate the clinical use of the device, submit data that shows the effects on gas transfer and pressure drop when the blood side outlet of the oxygenator experiences a mean minimum backpressure of 150 mmHg. |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5. |
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 5.1 |
General TestingIdentify the components of the testing circuit (described in the text and in figures) and include the following components as a minimum:
For determining the performance factor of the oxygenator's heat exchanger, a heat-exchanging device is used in the loop to maintain the same venous inlet blood temperature conditions to each of the oxygenators under evaluation. Blood sampling ports directly before the inlets and directly after the outlets of each of the heat exchangers are used to measure the blood temperature as a function of the operating variables. Monitor the heat exchange and waterside pressure drop over a six-hour period, in addition to recording blood and water parameters as stated above. During the testing, blood should be circulated at the maximum recommended flow rate. The specified time intervals for tabular and graphical data are at 10 min, 1 hr, 2 hr, 4 hr, and 6 hr.
|
TABLE 4. Heat Exchanger Data Set*
*Provide a table of data for each of the three different blood flow rates (minimum, nominal, and maximum). FDA suggests the table format given below.
| Water Flow Rates (L/min) |
| Blood Flow Rate = (L/min) |
Minimum |
Nominal |
Maximum |
| Inlet Blood Temperature (°C) | |||
| Outlet Blood Temperature (°C) | |||
| Inlet Water Temperature (°C) | |||
| Outlet Water Temperature (°C) | |||
| Calculate Performance Factor+ |
| R = | (Bo - Bi)/ (Wi - Bi) | |
Where |
Bo = |
blood temperature at the oxygenator outlet |
| Bi = | blood temperature at the oxygenator inlet | |
| Wi = | water temperature at the oxygenator inlet |
TABLE 5. Water Pressure Drop Data Set*
*Provide a table of data for each of the three different blood flow rates (minimum, nominal, and maximum). Hydrostatic differences between the locations of the inlet and outlet pressure transducers should be accounted for in the table. FDA suggests the table format given below.
| Water Flow Rates (L/min) |
| Blood Flow Rate = (L/min) |
Minimum |
Nominal |
Maximum |
| Inlet Water Pressure (mmHg) | |||
| Outlet Water Pressure (mmHg) | |||
| Water Pressure Drop (mmHg) |
Provide a graph along with the above data tables showing the heat exchanger performance factor as a function of both blood flow rate and water flow rate. Provide a graph along with the data tables showing the waterside pressure drop as a function of blood flow rate and water flow rate.
To better simulate the clinical use of the device, data should also be submitted showing the effects on heat transfer and water pressure drop when the blood side outlet of the oxygenator experiences a mean minimum backpressure of 150 mmHg during use.
6.
Blood Damage Performance Evaluation
The oxygenator will be evaluated for in vitro blood damage by monitoring the plasma hemoglobin concentration, WBC and platelets for six hours in duration.
| 6.1 |
General TestingIdentify the components of the testing circuit (described in the text and in figures) and include the following components as a minimum:
The aged, test oxygenators and the predicate oxygenators should be tested in paired fashion using identical circuits with the same pool of blood. The total volume of blood in the test circuits should be identical and minimized so that the sensitivity of the testing for the blood damage is increased. In general, the total circuit blood volume should be 500 mL - 2500 mL depending on the maximum recommended blood flow rate. Investigators should also consider running a "control blank circuit" concurrently with the test and predicate device circuits to determine the baseline blood damage caused by the components of the recirculation circuit when an oxygenator is not present (using a long tubing compressive clamp to create the back pressure on the pump and diffuse external heating of the blood tubing or reservoir to maintain proper temperature). Besides creating baseline blood damage data, running a "control blank circuit" with each blood pool also creates a means to evaluate variations in blood fragility between tests and to make comparisons to published values.
|
TABLE 6. Parameter Sampling Schedule
|
Sampling Schedule (minutes) |
||||||
| Parameter |
Main Blood Pool |
Baseline* |
90 |
180 |
270 |
360 |
| Plasma Hemoglobin Concentration | X | X | X | X | X | X |
| Activated Clotting Time (ACT) | X | X | X | X | ||
| Hematocrit | X | X | X | X | ||
| Platelets and WBC | X | X | X | X | X | X |
| Total Blood Hemoglobin | X | X | X | |||
| Temperature | X | X | X | X | X | |
| Blood Gas Values (pO2, pCO2, pH,) | X | X | X | X | X | |
| Blood and Gas Flow Rates | X | X | X | X | X |
* "Baseline" refers to the time (at "zero minutes" on the sampling schedule) after all of the blood has been introduced into the circuit, all bubbles have been removed from the circuit, the recirculation pump has been run at the proper blood flow rate for at least 5 minutes to insure complete mixing throughout the circuit, and the blood gas values, gas flow rate, and gas composition conditions have been established.
A traditional normalized "index of hemolysis" (IH), interpreted as the mg of hemoglobin released from the red blood cells per 100L of blood pumped through the device, should be calculated using the following formula (be sure to check the units of each entity in the equation):
IH [mg/ 100L] = [(plHgb/t) x ((100 - Hct)/100) x Vol. x 1000] / Q
plHgb/t (mg/dl min) = the slope of the plasma hemoglobin concentration (mg/dl) versus time (min.) plot obtained from a linear best-fit to this data
Hct (%) = average hematocrit of the blood circuit
Vol (L) = average blood volume of the circuit
Q (L/min) = blood flow rate during testing
Traditionally, only the plasma hemoglobin concentration at time zero and time 360 min were used to calculate the IH. However, using the slope of the plasma hemoglobin concentration versus time plot allows all of the data to be equally weighted in the determination of the IH. Although the IH has been used by several groups to "normalize" their blood damage data, it actually varies directly with the amount of cellular hemoglobin being pumped through the circuit. For this reason, the "modified index of hemolysis" (MIH), which is the ratio of the amount of hemoglobin released into the plasma normalized by the total amount of hemoglobin pumped through the device, should also be calculated in the submission as follows:
MIH [mg/ mg] = [(plHgb/t) x ((100 - Hct)/100) x Vol x 106] / (Q x Hb)
Where Hb (mg/dl) = average total hemoglobin concentration in the circuit
Although a standardized testing protocol has not been established, supplemental testing for up to six hours in duration at the manufacturer’s specified minimum blood flow rate (with blood with an ACT of 300-450 seconds and a normal to high platelet count) is also recommended. In-line monitoring for microparticle formation and post-test oxygenator evaluation for evidence of areas of blood stasis, thrombus deposition, and maldistribution of flow are encouraged.
6.2 |
|
6.3 |
|
6.4 |
|
6.5 |
|
7.
Packaging
7.1 |
|
7.2 |
|








