U.S. flag An official website of the United States government
  1. Home
  2. Safety
  3. Recalls, Market Withdrawals, & Safety Alerts
  4. ReFlow Medical Issues Recall of Specific Lots of Wingman35 Crossing Catheters
  1. Recalls, Market Withdrawals, & Safety Alerts

COMPANY ANNOUNCEMENT

ReFlow Medical Issues Recall of Specific Lots of Wingman35 Crossing Catheters

This recall has been completed and FDA has terminated this recall.

When a company announces a recall, market withdrawal, or safety alert, the FDA posts the company's announcement as a public service. FDA does not endorse either the product or the company.

Read Announcement

Summary

Company Announcement Date:
FDA Publish Date:
Product Type:
Medical Devices
Reason for Announcement:
Recall Reason Description
Due to tip splitting or separation
Company Name:
ReFlow Medical
Brand Name:
Brand Name(s)
Wingman35
Product Description:
Product Description
Catheters

Company Announcement

On April 22, 2017, ReFlow Medical initiated a lot-specific voluntary recall of the Wingman35 Crossing Catheters.

The FDA is classifying this as a Class 1 recall, the most serious recall situation where the patient is exposed to a reasonable likelihood of death or a serious injury. The Wingman35 Crossing Catheters have been found to exhibit tip splitting or separation, which has resulted in two adverse events. ReFlow has received 2 complaints of catheter-tip splitting and/or separation. A total of 2327 Wingman catheters are in distribution.

Tip splitting has the potential to lead to loss of device function. Tip separation may require medical intervention to retrieve a separated segment or may occlude blood flow to end organs.

The Wingman Crossing Catheters in this recall were distributed between March 2015 and March 2017.

ReFlow Medical has notified its customers and distributors by recall notification letters. The letters requested that all customers and distributors quarantine and discontinue use of all potentially affected units and return the affected product to the company as soon as possible for credit.

FDA and other regulatory agencies around the world have been notified of this action.

Consumers with questions may contact ReFlow Medical Customer Relations at 1-949-481-0399, Monday through Friday, between 8:00 a.m. and 4:30 p.m. Pacific time or by email at info@reflowmedical.com.

Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

  • Complete and submit the report Online
  • Regular Mail or Fax: Download form or call 1- 800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178

This recall is being conducted with the knowledge of the U.S. Food and Drug Administration.

 
 

Company Contact Information

Consumers:
Isa Rizk
949-481-0399
 
Back to Top