COMPANY ANNOUNCEMENT
BD Announces Voluntary Recall on Intraosseous Products
When a company announces a recall, market withdrawal, or safety alert, the FDA posts the company's announcement as a public service. FDA does not endorse either the product or the company.
Read AnnouncementSummary
- Company Announcement Date:
- FDA Publish Date:
- Product Type:
- Medical Devices
- Reason for Announcement:
-
Recall Reason DescriptionProduct issues may result in delays in care due to limited or non-functioning intraosseous access or could also lead to needlestick injuries
- Company Name:
- BD
- Brand Name:
-
Brand Name(s)BD
- Product Description:
-
Product DescriptionIntraosseous Needle Set Kits, Intraosseous Manual Driver Kits, Intraosseous Powered Drivers
Company Announcement
FRANKLIN LAKES, N.J. (June 22, 2022) – BD (Becton, Dickinson and Company), a leading global medical technology company, announced a voluntary recall on the BD™
Intraosseous Needle Set Kits, BD™ Intraosseous Manual Driver Kits and BD™ Intraosseous Powered Drivers.
Certain lots within the expiration date of these intraosseous products may result in the following:
These issues may result in delays in care due to limited or non-functioning intraosseous access or could also lead to needlestick injuries.
Impacted products include:
Product Name |
Catalog No. |
UDI |
Lot No. |
Expiration Date |
Product Package Size |
---|---|---|---|---|---|
Needle Kit for Powered Driver 15mm x 15Ga | D015151NK | See appendix 1 | See appendix 1 | See appendix 1 | 1 device per package |
Manual Driver Needle Kit 15mm x 15Ga | D015151MK | See appendix 1 | See appendix 1 | See appendix 1 | 1 device per package |
Needle Kit for Powered Driver 25mm x 15Ga | D015251NK | See appendix 1 | See appendix 1 | See appendix 1 | 1 device per package |
Manual Driver Needle Kit 25mm x 15Ga | D015251MK | See appendix 1 | See appendix 1 | See appendix 1 | 1 device per package |
Needle Kit for Powered Driver 35mm x 15Ga | D015351NK | See appendix 1 | See appendix 1 | See appendix 1 | 1 device per package |
Manual Driver Needle Kit 35mm x 15Ga | D015351MK | See appendix 1 | See appendix 1 | See appendix 1 | 1 device per package |
Needle Kit for Powered Driver 45mm x 15Ga | D015451NK | See appendix 1 | See appendix 1 | See appendix 1 | 1 device per package |
Manual Driver Needle Kit 45mm x 15Ga | D015451MK | See appendix 1 | See appendix 1 | See appendix 1 | 1 device per package |
Needle Kit for Powered Driver 55mm x 15Ga | D015551NK | See appendix 1 | See appendix 1 | See appendix 1 | 1 device per package |
Manual Driver Needle Kit 55mm x 15Ga | D015551MK | See appendix 1 | See appendix 1 | See appendix 1 | 1 device per package |
Intraosseous Powered Driver (drill) | D001001 | 00801741163579 | All | All | 1 device per package |
Customers should immediately review their inventory for the catalog and lot numbers listed above. Affected needle kits should be destroyed in compliance with the health care institution’s process for disposal. The use of affected intraosseous powered drivers should be paused until a BD representative provides instruction that it is safe for use. BD representatives will be reaching out to customers for inspection and will repair the devices if required.
There are no replacement products currently. BD recommends that customers evaluate their clinical needs and consider obtaining and using an alternative intraosseous product. BD will notify customers when replacement products become available.
Full customer and distributor letters can be found on the recall notification section of BD.com.
BD remains fully committed to ensuring patient and user safety. Customers requiring additional assistance or guidance may contact a service representative at 1-844-823-5433 Monday through Friday between 8 a.m. and 5 p.m. Central Time by saying “Recall” when prompted or via email at productcomplaints@bd.com.
Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
- Complete and submit the report Online
- Regular Mail or Fax: Download form or call 1- 800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178
Appendix 1
Catalog Number |
Lot / Serial Number |
UDI (GTIN, DI + PI), if applicable |
Date of Manufacture |
Expiration Date |
---|---|---|---|---|
D015151NK | 121460 | 801741163586 | 09/10/2019 | 09/10/2022 |
D015251NK | 121496 | 801741163678 | 09/13/2019 | 09/13/2022 |
D015451NK | 121593 | 801741163609 | 09/26/2019 | 09/26/2022 |
D015251NK | 122096 | 801741163678 | 12/06/2019 | 01/01/2023 |
D015151NK | 122145 | 801741163586 | 12/11/2019 | 01/01/2023 |
D015451NK | 122169 | 801741163609 | 12/17/2019 | 01/01/2023 |
D015251MK | 122193 | 801741163630 | 12/30/2019 | 01/02/2023 |
D015151MK | 122359 | 801741163623 | 01/21/2020 | 01/21/2023 |
D015451MK | 122375 | 801741163654 | 01/22/2020 | 01/22/2023 |
D015451NK | 123305 | 801741163609 | 05/18/2020 | 05/18/2023 |
D015251NK | 123306 | 801741163678 | 05/18/2020 | 05/18/2023 |
D015151NK | 123307 | 801741163586 | 05/18/2020 | 05/18/2023 |
D015451MK | 123433 | 801741163654 | 06/02/2020 | 06/02/2023 |
D015451NK | 125016 | 801741163609 | 12/07/2020 | 12/07/2023 |
D015151NK | 125298 | 801741163586 | 01/22/2021 | 01/22/2024 |
D015251NK | 125299 | 801741163678 | 01/22/2021 | 01/22/2024 |
D015451NK | 125517 | 801741163609 | 02/09/2021 | 02/09/2024 |
D015251NK | 125533 | 801741163678 | 02/10/2021 | 02/10/2024 |
D015151NK | 125534 | 801741163586 | 02/10/2021 | 02/10/2024 |
D015251NK | 125535 | 801741163678 | 02/10/2021 | 02/10/2024 |
D015151NK | 125662 | 801741163586 | 02/25/2021 | 02/25/2024 |
D015351NK | 125721 | 801741163593 | 03/02/2021 | 03/30/2024 |
D015451NK | 125770 | 801741163609 | 03/05/2021 | 03/05/2024 |
D015551NK | 125841 | 801741163616 | 03/10/2021 | 03/10/2024 |
D015351MK | 125842 | 801741163647 | 03/10/2021 | 03/10/2024 |
D015551MK | 125843 | 801741163661 | 03/10/2021 | 03/10/2024 |
D015451NK | 125977 | 801741163609 | 03/22/2021 | 03/22/2024 |
D015251NK | 126075 | 801741163678 | 03/30/2021 | 03/30/2024 |
D015151NK | 126095 | 801741163586 | 04/05/2021 | 04/08/2024 |
D015251NK | 126149 | 801741163678 | 04/06/2021 | 04/06/2024 |
D015251NK | 126178 | 801741163678 | 04/12/2021 | 04/12/2024 |
D015451NK | 126299 | 801741163609 | 04/20/2021 | 04/20/2024 |
D015451NK | 126300 | 801741163609 | 04/20/2021 | 04/20/2024 |
D015451NK | 126340 | 801741163609 | 04/27/2021 | 04/27/2024 |
D015151NK | 126467 | 801741163586 | 05/10/2021 | 05/10/2024 |
D015251NK | 126569 | 801741163678 | 05/17/2021 | 05/17/2024 |
D015451NK | 126575 | 801741163609 | 05/17/2021 | 05/17/2024 |
D015151NK | 126677 | 801741163586 | 05/25/2021 | 05/25/2024 |
D015251NK | 126686 | 801741163678 | 05/26/2021 | 05/26/2024 |
D015551MK | 126834 | 801741163661 | 06/07/2021 | 06/07/2024 |
D015351NK | 126901 | 801741163593 | 06/11/2021 | 06/11/2024 |
D015351NK | 126939 | 801741163593 | 06/16/2021 | 06/16/2024 |
D015551NK | 126951 | 801741163616 | 06/17/2021 | 06/17/2024 |
D015551NK | 127003 | 801741163616 | 06/22/2021 | 06/22/2024 |
D015151NK | 127040 | 801741163586 | 06/24/2021 | 06/24/2024 |
D015451NK | 127092 | 801741163609 | 06/28/2021 | 06/28/2024 |
D015251NK | 127118 | 801741163678 | 07/01/2021 | 07/01/2024 |
D015451NK | 127119 | 801741163609 | 07/01/2021 | 07/01/2024 |
D015251NK | 127170 | 801741163678 | 07/07/2021 | 07/07/2024 |
D015451NK | 127300 | 801741163609 | 07/22/2021 | 07/22/2024 |
D015251NK | 127417 | 801741163678 | 08/04/2021 | 08/04/2024 |
D015551NK | 127496 | 801741163616 | 08/11/2021 | 08/11/2024 |
D015451NK | 127516 | 801741163609 | 08/12/2021 | 08/12/2024 |
D015151NK | 127711 | 801741163586 | 08/27/2021 | 08/27/2024 |
D015251NK | 127756 | 801741163678 | 09/01/2021 | 09/01/2024 |
D015451NK | 127808 | 801741163609 | 09/07/2021 | 09/07/2024 |
D015251NK | 128001 | 801741163678 | 09/21/2021 | 09/21/2024 |
D015151NK | 128002 | 801741163586 | 09/21/2021 | 09/21/2024 |
D015451NK | 128120 | 801741163609 | 10/01/2021 | 10/01/2024 |
D015151NK | 128208 | 801741163586 | 10/12/2021 | 10/12/2024 |
D015351NK | 128385 | 801741163593 | 10/27/2021 | 10/27/2024 |
D015251NK | 128482 | 801741163678 | 11/03/2021 | 11/03/2024 |
D015451NK | 128487 | 801741163609 | 11/03/2021 | 11/03/2024 |
D015551NK | 128724 | 801741163616 | 12/01/2021 | 12/01/2024 |
D015451NK | 128725 | 801741163609 | 12/01/2021 | 12/01/2024 |
D015151NK | 128778 | 801741163586 | 12/06/2021 | 12/06/2024 |
D015251NK | 128829 | 801741163678 | 12/07/2021 | 12/07/2024 |
About BD
BD is one of the largest global medical technology companies in the world and is advancing the world of health by improving medical discovery, diagnostics and the delivery of care. The company supports the heroes on the frontlines of health care by developing innovative technology, services and solutions that help advance both clinical therapy for patients and clinical process for health care providers. BD and its 75,000 employees have a passion and commitment to help enhance the safety and efficiency of clinicians' care delivery process, enable laboratory scientists to accurately detect disease and advance researchers' capabilities to develop the next generation of diagnostics and therapeutics. BD has a presence in virtually every country and partners with organizations around the world to address some of the most challenging global health issues. By working in close collaboration with customers, BD can help enhance outcomes, lower costs, increase efficiencies, improve safety and expand access to health care. For more information on BD, please visit bd.com or connect with us on LinkedIn at www.linkedin.com/company/bd1/ and Twitter @BDandCo.
Contacts:
Media:
Troy Kirkpatrick
VP, Public Relations
858.617.2361
troy.kirkpatrick@bd.com
Investors
Francesca DeMartino
SVP, Head of Investor Relations
201.847.5743
francesca.demartino@bd.com
Company Contact Information
- Consumers:
- 1-844-823-5433
- productcomplaints@bd.com.
- Media:
- Troy Kirkpatrick, VP, Public Relations
- 858-617-2361
- troy.kirkpatrick@bd.com