GUIDANCE DOCUMENT

Recommendations for Reducing the Risk of Transfusion-Transmitted Hepatitis B Virus Guidance for Industry November 2026

Final
Docket Number:
FDA-2024-D-5942
Issued by:
Guidance Issuing Office
Center for Biologics Evaluation and Research

We, FDA, are providing you, blood establishments that collect blood and blood components, including Source Plasma, with recommendations for testing blood and blood components for hepatitis B virus (HBV) to reduce the risk of transfusion-transmitted HBV. The recommendations contained in this guidance apply to the collection of Whole Blood and blood components, including Source Plasma.

HBV is a relevant transfusion-transmitted infection (RTTI) (Title 21 of the Code of Federal Regulations (CFR) 630.3(h)(1)(ii)) and, therefore, subject to the testing requirements in 21 CFR 610.40, the donor deferral requirements in 21 CFR 610.41 and the donor notification requirements in 21 CFR 630.40.

Under 21 CFR 610.40(a)(1), an establishment that collects blood and blood components must test each donation of human blood or blood components for transfusion or for use in manufacturing a product, including donations intended as a component of, or used to manufacture, a medical device, for evidence of infection due to HBV. In addition, under 21 CFR 610.40(b), establishments must use, in accordance with the manufacturer’s instructions, one or more screening tests that FDA has licensed, approved or cleared as necessary to meet the testing requirement and to reduce adequately and appropriately the risk of transmission of RTTIs. This guidance provides recommendations for testing blood donations to reduce adequately and appropriately the risk of HBV transmission under these regulations.


Submit Comments

You can submit online or written comments on any guidance at any time (see 21 CFR 10.115(g)(5))

If unable to submit comments online, please mail written comments to:

Dockets Management
Food and Drug Administration
5630 Fishers Lane, Rm 1061
Rockville, MD 20852

All written comments should be identified with this document's docket number: FDA-2024-D-5942.

Questions?

Contact Point
Center for Biologics Evaluation and Research (CBER)
Food and Drug Administration
10903 New Hampshire Ave Bldg. 71, Rm. 3103
Silver Spring, MD 20993-0002
industry.biologics@fda.hhs.gov
(800) 835-4709
(240) 402-8010
Back to Top