U.S. flag An official website of the United States government
  1. Home
  2. Medical Devices
  3. Science and Research | Medical Devices
  4. Endoscope Geometric Distortion- A Method for Quantitative Evaluation
  1. Science and Research | Medical Devices

Endoscope Geometric Distortion- A Method for Quantitative Evaluation

Catalog of Regulatory Science Tools to Help Assess New Medical Devices

Technical Description

This tool is a method to evaluate endoscope geometric distortion. Current methods to evaluate endoscope geometric distortion are either too difficult to understand or too simple to accurately describe complex geometric distortion. The method overcomes these limitations, has clear physical meaning, and can facilitate lesion size estimation during diagnosis.

As one of the several important optical performance characteristics, endoscope geometric distortion can negatively affect size estimation and feature identification related diagnosis. Therefore, a quantitative and simple method to evaluate endoscope geometric is imperative for both the endoscope industry and medical device regulatory agents.

Fig. Distortion measurement setup: (a) test target and (b) test setup

Fig. Distortion measurement setup: (a) test target and (b) test setup

Intended Purpose

The tool can be used to measure and compare geometric distortion of endoscopes. While the tool was original designed for measuring geometric distortion of endoscopes, it can be extended to any optical imaging cameras.

Testing

The method has been evaluated with an endoscopic system (Olympus EVIS EXERA II). Details are described in peer-reviewed publication: Development of the local magnification method for quantitative evaluation of endoscope geometric distortion.

Limitations

Most endoscopes have geometric distortion that are circularly symmetric (that is, rotationally symmetric with respect to any angle). If an endoscope has defects and thus doesn’t have a symmetric distortion, the evaluation should be performed in different directions (for example, horizontal, vertical, diagonal) and the results should be compared.

Supporting Documentation

Contact

Tool Reference

In addition to citing relevant publications please reference the use of this tool using DOI: 10.5281/zenodo.7883653

For more information:

Note: This tool was previously listed in the catalog as “Quantitative evaluation of endoscope geometric distortion.”

 
Back to Top