Predetermined Change Control Plans

On this page


Information About Predetermined Change Control Plans (PCCPs)

What is a PCCP?

A Predetermined Change Control Plan (PCCP) describes certain planned changes or modifications that may be made to the device, how the manufacturer will develop, test, and implement those modifications, and how the manufacturer will assess the impact of those modifications. Manufacturers may submit, and the FDA may authorize, a PCCP as part of a marketing submission for a device.

This webpage provides general information about PCCPs, a list of FDA-authorized devices with PCCPs, and other available resources. The FDA intends to update and add more resources for stakeholders about PCCPs as they are available.

What is the value of a PCCP?

By including a PCCP in a device’s marketing submission, a manufacturer can seek FDA authorization for certain future device modifications. If the FDA authorizes the PCCP, the manufacturer may make those planned changes to the device without submitting a new marketing submission for each change, as long as the modifications are included in and implemented in accordance with the authorized PCCP.

PCCPs can help facilitate innovation by allowing manufacturers to make planned device improvements more quickly and allowing device advancements to reach patients faster, while continuing to ensure the safety and effectiveness of the device. It may also reduce the regulatory burden on manufacturers, by reducing the need for additional marketing submissions for each planned change, and on the FDA, by reducing the number of additional marketing submissions for review.

What modifications can be included in a PCCP?

A PCCP can include planned modifications to the subject device of the marketing submission that:

  • Would otherwise require a new marketing submission,
  • Maintain the device within the device’s intended use, and
  • For devices subject to 510(k) requirements, allow the device to remain substantially equivalent to the predicate device.

See section 515C of the FD&C Act, 21 CFR 807.81(a)(3), and 21 CFR 814.39(a).

What information should be included in a PCCP?

As described in FDA guidance, a PCCP generally includes:

  1. Description of Modifications: A description of the planned modifications to the device, including changes to the device characteristics, performance, or other specifications.
  2. Modification Protocol: A description of:
    • the performance evaluation activities, including pre-defined acceptance criteria, for each modification to ensure continued safety and effectiveness of the device
    • the update procedures, including how a manufacturer will update their device to implement the modification and, when applicable, how users will be notified about device updates
  3. Impact Assessment: A description of:
    • the individual and cumulative impact of each modification
    • how implementation of one modification impacts the implementation of another modification
    • the benefits and risks of each modification
    • how the Modification Protocol will continue to ensure the safety and effectiveness of the device

The level of detail included in a PCCP is similar to the level of detail included in other sections of the marketing submission for the device.

What is the difference between a PCCP and a traditional device change?

A traditional device change is a modification that is made after a device has been authorized for marketing. Depending on the nature of the change, the manufacturer may need to submit a new or supplemental marketing submission to the FDA before implementing the change. See 21 CFR 807.81(a)(3) or 21 CFR 814.39(a).

In contrast, a PCCP allows a manufacturer to seek authorization for certain future device modifications as part of a device marketing submission. After FDA authorization of a PCCP as part of a device marketing submission, a manufacturer can make planned changes to a device without submitting a new marketing submission for each change when the changes are included in and implemented in accordance with the authorized PCCP. A PCCP therefore provides a mechanism for making specified future modifications to the device while maintaining the device’s safety and effectiveness.

Which marketing submission types can include a PCCP?

PCCPs are reviewed and authorized as part of the device marketing submission. PCCPs may be included in marketing submissions for devices reviewed through the 510(k), De Novo, or Premarket Approval (PMA) pathways, including device-led combination products. See section 513(f)(2) and 515C of the FD&C Act.

Where is a PCCP included in a marketing submission?

A PCCP is included in a standalone section and discussed in the cover letter of your device marketing submission. For eSTAR submissions, a PCCP can be identified in the General Device Characteristics section of the Device Description by using the appropriate checkbox and attachment option.


List of Authorized Devices with PCCPs

Date Final DecisionSubmission NumberDeviceCompanyPanel (Lead)Primary Product Code
7/21/2023K231173IRREGULAR RHYTHM NOTIFICATION FEATURE (IRNF)APPLE INC.CARDIOVASCULARQDB
8/31/2023K22359723ANDME PERSONAL GENOME SERVICE (PGS) GENETIC HEALTH RISK REPORT FOR BRCA1/BRCA2 (SELECTED VARIANTS)23ANDME, INC.PATHOLOGYQAZ
9/28/2023K232699LOW EJECTION FRACTION AI-ECG ALGORITHMANUMANA, INC.CARDIOVASCULARQYE
12/22/2023K233027NANOZOOMER S360MD SLIDE SCANNER SYSTEMHAMAMATSU PHOTONICS K.K.PATHOLOGYPSY
1/3/2024K231779REMI AI DISCRETE DETECTION MODULEEPITEL, INC.NEUROLOGYOMB
1/13/2024K233216CLEWICU SYSTEMCLEW MEDICAL LTD.CARDIOVASCULARQNL
2/1/2024P190016/S007TULA SYSTEMTUSKER MEDICAL, INC.EAR, NOSE, AND THROATQJA
3/1/2024K233030BONEMRIMRIGUIDANCE B.VRADIOLOGYQIH
3/8/2024K233438SLEEPSTAGEMLBEACON BIOSIGNALS, INC.NEUROLOGYOLZ
3/15/2024K233986BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - GN CIPROFLOXACIN (0.0156-4 µG/ML)BECTON, DICKINSON AND COMPANYMICROBIOLOGYLON
3/22/2024K231806THERMO SCIENTIFIC OXOID REZAFUNGIN DISC (5 µG) RZF5OXOID LIMITED (PART OF THERMO FISHER SCIENTIFIC)MICROBIOLOGYJTN
4/2/2024K211815LIFESCALE GRAM NEGATIVE KIT (LSGN) WITH THE LIFESCALE AST SYSTEMAFFINITY BIOSENSORSMICROBIOLOGYSAN
4/12/2024P160026/S040LIFEPAK 1000 DEFIBRILLATOR, LIFEPAK 20 DEFIBRILLATOR/ MONITOR, LIFEPAK 20E DEFIBRILLATOR/ MONITOR, LIFEPAK 15 MONITOR/DEFIBRILLATOR, LIFEPAK 12 DEFIBRILLATOR/ MONITORPHYSIO-CONTROL. INC.CARDIOVASCULARMKJ
5/23/2024K232603CAMAPS FXCAMDIAB LTD.CLINICAL CHEMISTRYQJI
6/7/2024K231010CORVAIRALIVECOR, INC.CARDIOVASCULARMHX
6/14/2024K233955CLARIUS OB AICLARIUS MOBILE HEALTH CORP.RADIOLOGYIYN
6/26/2024P150004/S071AXIUM NEUROSTIMULATOR SYSTEMABBOTT MEDICALNEUROLOGYPMP
6/26/2024P140009/S095BRIO NEUROSTIMULATION SYSTEMABBOTT MEDICALNEUROLOGYMHY
6/26/2024P010032/S213GENESIS AND EON FAMILY NEUROSTIMULATION (IPG) SYSTEMSABBOTT MEDICALNEUROLOGYLGW
7/2/2024K240555TYTO INSIGHTS FOR CRACKLES DETECTIONTYTO CARE LTD.ANESTHESIOLOGYPHZ
7/11/2024K241006NATURAL CYCLESNATURALCYCLES NORDIC ABOBSTETRICS AND GYNECOLOGYPYT
7/12/2024K234009ACORN 3D SOFTWARE (AC-SEG-4009); ACORN 3DP MODEL (AC-101-XX)MIGHTY OAK MEDICALRADIOLOGYQIH
7/18/2024K241060HARDYDISK AST CEFIDEROCOL 30µG (FDC30)HARDY DIAGNOSTICSMICROBIOLOGYJTN
8/1/2024DEN230046PGDX ELIO PLASMA FOCUS DXPERSONAL GENOME DIAGNOSTICS, INC.PATHOLOGYSBY
8/5/2024K240953AI PLATFORM 2.0 (AIP002)EXO IMAGINGRADIOLOGYQIH
8/19/2024K241504IFUSE TORQ TNT IMPLANT SYSTEMSI-BONE INC.ORTHOPEDICOUR
8/21/2024K240290AIMIFY (1.X)SUBTLE MEDICAL, INC.RADIOLOGYLLZ
8/27/2024K241684OVERJET CHARTING ASSISTOVERJET, INCRADIOLOGYQIH
9/11/2024K241380FETOLY-HEARTDIAGNOLYRADIOLOGYIYN
9/12/2024DEN230081HEARING AID FEATURE (HAF)APPLE INC.EAR, NOSE, AND THROATSCR
9/13/2024K240929SLEEP APNEA NOTIFICATION FEATURE (SANF)APPLE INC.ANESTHESIOLOGYQZW
10/23/2024K241324LIFESCALE GRAM NEGATIVE KIT (LSGN) WITH THE LIFESCALE AST SYSTEMAFFINITY BIOSENSORS, LLCMICROBIOLOGYSAN
11/1/2024K241828AFFERA INTEGRATED MAPPING SYSTEM; IMPEDANCE LOCALIZATION PATCH KITMEDTRONIC, INC.CARDIOVASCULARDQK
2/6/2025P240010/S001MI CANCER SEEKCARIS LIFE SCIENCESPATHOLOGYPQP
2/10/2025K243201THE SENSITITRE 20-24 HOUR HAEMOPHILUS INFLUENZAE/STREPTOCOCCUS PNEUMONIAE (HP) MIC OR BREAKPOINT SUSCEPTIBILITY SYSTEM WITH PENICILLIN IN THE DILUTION RANGE OF 0.015-32 UG/MLTHERMOFISHER SCIENTIFICMICROBIOLOGYJWY
2/10/2025K242905THE SENSITITRE 20-24 HOUR HAEMOPHILUS INFLUENZAE/STREPTOCOCCUS PNEUMONIAE (HP) MIC OR BREAKPOINT SUSCEPTIBILITY SYSTEM WITH LINEZOLID IN THE DILUTION RANGE OF 0.12-32 UG/MLTHERMOFISHER SCIENTIFICMICROBIOLOGYJWY
3/19/2025K243987TRIMED ELBOW AND FOREARM SYSTEMTRIMED, INC.ORTHOPEDICHRS
3/19/2025K243999TRIMED ANKLE FIXATION SYSTEMTRIMED, INC.ORTHOPEDICHRS
3/20/2025P150033/S201MEDTRONIC MICRA TRANSCATHETER PACEMAKER SYSTEMMEDTRONIC INC.CARDIOVASCULARPNJ
3/20/2025P120017/S038MODEL 5071 LEADMEDTRONIC INC.CARDIOVASCULARDTB
3/20/2025P010031/S884CONCERTO/INSYNC SENTRY/INSYNC MAXIMO IMPLANTABLE CARDIOVASCULAR DEFIBRILLATORS WITH CARDIAC RESYNCHICNIZATIONMEDTRONIC CARDIAC RHYTHM DISEASE MANAGEMENTCARDIOVASCULARNIK
3/20/2025P080006/S190MEDTRONIC ATTAIN ABILITY MODEL 4196 LEADMEDTRONIC INC.CARDIOVASCULAROJX
3/20/2025P060039/S118ATTAIN STARFIX MODEL 4195 LEADMEDTRONIC INC.CARDIOVASCULAROJX
3/20/2025P850051/S085ACTIVITRAX MODELS 8400, 8402, 8403MEDTRONIC INC.CARDIOVASCULARLWO
3/20/2025P090013/S336REVO MRI SURESCAN IPG AND PACING SYSTEMMEDTRONIC, INCCARDIOVASCULARLWP
3/20/2025P010015/S558MEDTRONIC INSYNC(TM) BIVENTRICAL PACING SYSTEMMEDTRONIC INC.CARDIOVASCULARLWP
3/20/2025P820003/S144VERSATRAX MODEL 7000 UNIVERSAL A-V PULSE GENERATORMEDTRONIC CARDIAC RHYTHM DISEASE MANAGEMENTCARDIOVASCULARLWP
3/20/2025P220012/S043AURORA EV-ICD SYSTEMMEDTRONIC, INC.CARDIOVASCULARLWS
3/20/2025P980050/S145MEDTRONIC(R) JEWEL(R) AF 7250 DUAL CHAMBER IMPLANTABLE CARDIOVERTER DEFIBRILLATOR, MODEL 9961 PROGRAMMER APPLICATION SOFMEDTRONIC INC.CARDIOVASCULARLWS
3/20/2025P980016/S918VIRTUSO/ENTRUST/MAXIMO/INTRINSIC/MARQUIS/IMPLANTABLE CARDIVERTER DEFIBRILLATORSMEDTRONIC CARDIAC RHYTHM DISEASE MANAGEMENTCARDIOVASCULARLWS
3/20/2025P900061/S175MEDTRONIC PCD TACHYARRHYTHMIA CONTROL SYSTEMMEDTRONIC CARDIAC RHYTHM DISEASE MANAGEMENTCARDIOVASCULARLWS
3/20/2025P830061/S230STEROID TIP(TM) MODEL 4503&4003 TRANSVENOUS PACINGMEDTRONIC CARDIAC RHYTHM DISEASE MANAGEMENTCARDIOVASCULARNVN
3/20/2025P930039/S270MEDTRONIC(R) CAPSUREFIX LEAD MODEL 4068,4067,4568MEDTRONIC, INC.CARDIOVASCULARNVN
3/20/2025P850089/S171CAPSURE SP, CAPSURE, CAPSURE 2 LEADS, EXCELLENCE S, IMPULSE, IMPLUSE II EXCELLENCE SS, LEADSMEDTRONIC CARDIAC RHYTHM DISEASE MANAGEMENTCARDIOVASCULARNVN
3/20/2025P030036/S151MEDTRONIC SELECTSECURE LEAD MODEL 3830MEDTRONIC CARDIAC RHYTHM DISEASE MANAGEMENTCARDIOVASCULARNVN
3/20/2025P950024/S114MEDTRONIC(R) CAPSURE(R) EPI PACING LEAD MODEL 4695MEDTRONIC INC.CARDIOVASCULARNVN
3/20/2025P920015/S297MEDTRONIC(R) TRANSVENE LEAD SYSTEMMEDTRONIC INC.CARDIOVASCULARNVY
3/20/2025P930022/S020MEDTRONIC(R) LEGEND PLUS(TM) PULSE GENERATORMEDTRONIC INC.CARDIOVASCULARNVZ
3/20/2025P970012/S101MEDTRONIC.KAPPA 400 SERIES PACEMAKERSMEDTRONIC INC.CARDIOVASCULARNVZ
3/20/2025P890003/S475SYNERGYST II PULSE GENERATOR MODELS 7070 & 7071MEDTRONIC, INC.CARDIOVASCULARNVZ
3/20/2025P980035/S807MEDTRONIC KAPPA 700/600 SERIES PULSE GENERATORS AND MODEL 9953 SOFTWAREMEDTRONIC INC.CARDIOVASCULARNVZ
3/20/2025P990001/S123DIVA PLATFORM IMPLANTABLE PULSE GENERATORS AND PRO VIT APPLICATION SOFTWARE VERSION 3.3.2MEDTRONIC INC.CARDIOVASCULARNVZ
4/3/2025K242551SYNGO DYNAMICS (VERSION VA41D)SIEMENS HEALTHCARE GMBHRADIOLOGYQIH
4/4/2025K242807HEARTFOCUS (V.1.1.1)DESKIRADIOLOGYQJU
4/7/2025K243567TYTO INSIGHTS FOR RHONCHI DETECTIONTYTO CARE LTD.ANESTHESIOLOGYPHZ
4/9/2025K241589CERIBELL SEIZURE DETECTION SOFTWARECERIBELL, INC.NEUROLOGYOMB
4/10/2025K243943TRIMED COMPRESSION SCREWSTRIMED, INC.ORTHOPEDICHWC
4/11/2025K243558CANVAS DXCOGNOA, INC.NEUROLOGYQPF
4/15/2025K242658THE SENSITITRE 18-24 HOUR MIC OR BREAKPOINT SUSCEPTIBILITY SYSTEM WITH MOXIFLOXACIN IN THE DILUTION RANGE OF 0.008-16 UG/MLTHERMOFISHER SCIENTIFICMICROBIOLOGYJWY
4/16/2025K243853CLARIUS PROSTATE AICLARIUS MOBILE HEALTH CORP.RADIOLOGYQIH
4/30/2025K250274ETEST IMIPENEM/RELEBACTAM P. AERUGINOSA (IRPA) (0.008/4-128/4 µG/ML)BIOMERIEUX, SAMICROBIOLOGYJWY
5/7/2025K243684BRIGHTHEART VIEW CLASSIFIERBRIGHTHEARTRADIOLOGYQIH
5/8/2025K250226CLARIUS MEDIAN NERVE AICLARIUS MOBILE HEALTH CORP.RADIOLOGYQIH
5/15/2025K250454HARDYDISK AZTREONAM/AVIBACTAM 30/20µG (AZA50)HARDY DIAGNOSTICSMICROBIOLOGYJTN
5/19/2025K250447BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - GN IMIPENEM-RELEBACTAM (0.0625/4 - 16/4 µG/ML)BECTON, DICKINSON AND COMPANYMICROBIOLOGYLON
5/21/2025K250798CONTROL-IQ+ TECHNOLOGYTANDEM DIABETES CARE, INC.CLINICAL CHEMISTRYQJI
5/22/2025K242659THE SENSITITRE 18-24 HOUR MIC OR BREAKPOINT SUSCEPTIBILITY SYSTEM WITH VANCOMYCIN IN THE DILUTION RANGE OF 0.25-128 UG/MLTHERMOFISHER SCIENTIFICMICROBIOLOGYJWY
6/5/2025K251456BRIGHTHEART VIEW CLASSIFIERBRIGHTHEARTRADIOLOGYQIH
6/6/2025K242737EMPATICA HEALTH MONITORING PLATFORM; EMBRACEPLUS; EMPATICA CARE; CARE PORTALEMPATICA S.R.L.NEUROLOGYMWI
6/12/2025K250856ETEST AZTREONAM/AVIBACTAM (AZA) (0.016/4-256/4 µG/ML)BIOMðRIEUXMICROBIOLOGYJWY
6/12/2025K250789THERMO SCIENTIFIC OXOID AZTREONAM/AVIBACTAM DISC (30/20 µG) AZA50THERMO FISHER SCIENTIFIC (OXOID LTD.)MICROBIOLOGYJTN
6/20/2025K242843THE SENSITITRE 20-24 HOUR HAEMOPHILUS INFLUENZAE/STREPTOCOCCUS PNEUMONIAE (HP) MIC OR BREAKPOINT SUSCEPTIBILITY SYSTEM WITH CEFTOBIPROLE IN THE DILUTION RANGE OF 0.008-16 UG/MLTHERMOFISHER SCIENTIFICMICROBIOLOGYJWY
6/23/2025K250885CEFTOBIPROLE BPR 5 µG DISCLIOFILCHEM INC.MICROBIOLOGYJTN
6/26/2025K243391AISIGHT DXPATHAI, INC.PATHOLOGYQKQ
6/26/2025K250956HARDYDISK AST GEPOTIDACIN 10µG (GEP10)HARDY DIAGNOSTICSMICROBIOLOGYJTN
6/27/2025K250984MAESTRO SYSTEM (REF100)MOON SURGICALGENERAL AND PLASTIC SURGERYQZB
6/30/2025K250990THE SENSITITRE 18-24 HOUR MIC OR BREAKPOINT SUSCEPTIBILITY SYSTEM WITH CIPROFLOXACIN IN THE DILUTION RANGE OF 0.002-64 UG/MLTHERMOFISHER SCIENTIFICMICROBIOLOGYJWY
7/1/2025K251032MINIMED 780G INSULIN PUMPMEDTRONIC MINIMED INC.CLINICAL CHEMISTRYQFG
7/2/2025P240040ONCOMINE DX EXPRESS TESTLIFE TECHNOLOGIES CORPORATIONPATHOLOGYPQP
7/2/2025K250524MENDAERA GUIDANCE SYSTEMMENDAERA, INC.RADIOLOGYITX
7/3/2025K243679MAMMOSCREEN (4)THERAPIXELRADIOLOGYQDQ
7/11/2025K243274CDC HUMAN INFLUENZA VIRUS REAL-TIME RT-PCR DIAGNOSTIC PANEL: INFLUENZA A/B TYPING KIT (VER 2); INFLUENZA A SUBTYPING KIT (VER 4); INFLUENZA B GENOTYPING KIT (VER 1.1 AND 2); AND INFLUENZA A/H5 SUBTYPING KIT (VER 4)CENTERS FOR DISEASE CONTROL AND PREVENTIONMICROBIOLOGYOZE
7/18/2025K250902HEARTFLOW ANALYSISHEARTFLOW, INC.CARDIOVASCULARPJA
7/18/2025K250084MICROSCAN DRIED GRAM-NEGATIVE MIC/COMBO PANELS WITH AZTREONAM (AZT) (0.5-64 µG/ML)BECKMAN COULTER, INC.MICROBIOLOGYLTT
7/24/2025P160035/S052EXCOR PEDIATRIC VENTRICULAR ASSIST DEVICEBERLIN HEART INC.CARDIOVASCULARDSQ
7/24/2025P250002QUEST DIAGNOSTICS AAVRH74 ANTIBODY ELISA CDXQUEST DIAGNOSTICS NICHOLS INSTITUTEIMMUNOLOGYQWQ
7/28/2025K250652ECG-AI LOW EJECTION FRACTION (LEF) 12-LEAD ALGORITHM (1010)ANUMANA, INC.CARDIOVASCULARQYE
7/31/2025DEN240068ARTERAAI PROSTATEARTERA INC.PATHOLOGYSFH
8/8/2025K251713BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - GN ERAVACYCLINE (0.125-2 µG/ML)BECTON, DICKINSON AND COMPANYMICROBIOLOGYLON
8/12/2025K251879SULOPENEM SPM 2 µGLIOFILCHEMMICROBIOLOGYJTN
8/15/2025K250036MICROSCAN DRIED GRAM-POSITIVE MIC/COMBO PANELS WITH DAPTOMYCIN (DAP) (0.06-32 µG/ML)BECKMAN COULTER, INC.MICROBIOLOGYLTT
8/20/2025K243804MICROSCAN DRIED GRAM-NEGATIVE MIC/COMBO PANELS WITH CEFEPIME (CPE) (0.12-64 µG/ML) (MICROSCAN)BECKMAN COULTER, INC.MICROBIOLOGYLTT
8/21/2025K251579VITEK 2 AST-GRAM NEGATIVE CEFAZOLIN (=<1-=>32 µG/ML)BIOMERIEUX, INC.MICROBIOLOGYLON
8/22/2025K243685MAMMOSCREEN BDTHERAPIXELRADIOLOGYQIH
8/28/2025K252014THE SENSITITRE 18-24 HOUR MIC OR BREAKPOINT SUSCEPTIBILITY SYSTEM WITH AMIKACIN IN THE DILUTION RANGE OF 0.25-256 UG/MLTHERMO FISHER SCIENTIFICMICROBIOLOGYJWY
8/29/2025K251217SMARTGUARD TECHNOLOGY; PREDICTIVE LOW GLUCOSE TECHNOLOGYMEDTRONIC MINIMED INCCLINICAL CHEMISTRYQJI
9/2/2025K250753VELMENI FOR DENTISTS (V4D)VELMENI INC.RADIOLOGYMYN
9/4/2025P240013/S010SPHERE-9 CATHETER AND AFFERA ABLATION SYSTEMMEDTRONIC INC.CARDIOVASCULARQZI
9/5/2025K250575THE SENSITITRE 18 - 24 HOUR MIC OR BREAKPOINT SUSCEPTIBILITY SYSTEM WITH AZTREONAM/AVIBACTAM IN THE DILUTION RANGE OF 0.008/4 - 64/4 UG/MLTHERMOFISHER SCIENTIFICMICROBIOLOGYJWY
9/7/2025P140009/S121BRIO NEUROSTIMULATION SYSTEMABBOTT MEDICALNEUROLOGYMHY
9/7/2025P010032/S240GENESIS AND EON FAMILY NEUROSTIMULATION (IPG) SYSTEMSABBOTT MEDICALNEUROLOGYLGW
9/9/2025K251682MUSCLEVIEW 2.0SPRINGBOK, INC. (DBA SPRINGBOK ANALYTICS)RADIOLOGYLNH
9/11/2025K250507HYPERTENSION NOTIFICATION FEATURE (HTNF)APPLE INC.CARDIOVASCULARSFR
9/12/2025K251368FETOLYDIAGNOLYRADIOLOGYIYN
9/15/2025K250237INFEROPERATE SUITEBEIJING INFERVISION HEALTHCARE MEDICAL TECHNOLOGY CO., LTD.RADIOLOGYQIH
9/17/2025K251902VITAFORM PROCEDURAL MASK - BLUE (VITAFORM BLUE); VITAFORM PROCEDURAL MASK WITH SHIELD (VITAFORM-FS)VITACORE INDUSTRIES INC.GENERAL HOSPITALFXX
9/18/2025K250369AXIAL3D INSIGHTAXIAL MEDICAL PRINTING LIMITEDRADIOLOGYQIH
9/25/2025K252062THE SENSITITRE 18-24 HOUR MIC OR BREAKPOINT SUSCEPTIBILITY SYSTEM WITH COLISTIN IN THE DILUTION RANGE OF 0.12-16 UG/MLTHERMOFISHER SCIENTIFICMICROBIOLOGYJWY
9/30/2025K252072FRANCISELLA TULARENSIS REAL-TIME PCR ASSAYCENTERS FOR DISEASE CONTROL AND PREVENTIONMICROBIOLOGYSGA
10/2/2025K252114MTS AZTREONAM-AVIBACTAM 0.016/4 - 256/4 µG/MLLIOFILCHEM S. R. L.MICROBIOLOGYJWY
10/10/2025K251866TEGADERM POST-OP TRANSPARENT ANTIMICROBIAL DRESSINGSOLVENTUM US, LLCGENERAL AND PLASTIC SURGERYFRO
10/24/2025K252357GLUCOSE2ABBOTT IRELANDCLINICAL CHEMISTRYCFR
10/27/2025K251875LIFESCALE GRAM NEGATIVE KIT (LSGN) WITH THE LIFESCALE AST SYSTEMAFFINITY BIOSENSORS, LLCMICROBIOLOGYSAN
11/3/2025K250392ZAP-X RADIOSURGERY SYSTEM (ZAP-X)ZAP SURGICAL SYSTEMS, INC.RADIOLOGYIYE
11/6/2025K253265BRIEFCASE-TRIAGEAIDOC MEDICAL, LTD.RADIOLOGYQAS
11/7/2025K252041FIBRESOLVE (WITH PCCP)IMVARIA, INC.RADIOLOGYQWO
11/20/2025K253105HARDYDISK AST CEFIDEROCOL 30µG (FDC30)HARDY DIAGNOSTICSMICROBIOLOGYJTN
11/21/2025K252070CERIBELL INFANT SEIZURE DETECTION SOFTWARECERIBELL, INC.NEUROLOGYOMB
11/24/2025K252366A2Z-UNIFIED-TRIAGEA2Z RADIOLOGY AI INC.RADIOLOGYQAS
12/5/2025K251337THERMO SCIENTIFIC OXOID GEPOTIDACIN DISC (10 µG) GEP10THERMO FISHER SCIENTIFIC (OXOID LTD.)MICROBIOLOGYJTN
12/8/2025K251936CERIBELL DELIRIUM MONITOR SYSTEMCERIBELL, INC.NEUROLOGYNCG
12/17/2025K251416ULTRASIGHT GUIDANCEULTRASIGHT LTD.RADIOLOGYQJU
12/18/2025K251562OTSCNEO SYSTEM SET (100.03N-14N, 100.27N-31N)OVESCO ENDOSCOPY AGGASTROENTEROLOGY-UROLOGYPKL
12/18/2025K252235PVAD IQ SOFTWAREULTRASIGHT LTD.RADIOLOGYQIH
12/19/2025K252284EMURMUR HEART AICSD LABSCARDIOVASCULARDQD
12/19/2025K251152DBLG2DIABELOOPCLINICAL CHEMISTRYQJI
12/23/2025K253281UPDOCUPDOC INC.CLINICAL CHEMISTRYNDC
1/9/2026K252906ECHELON ENDOPATH STAPLE LINE REINFORCEMENT (ECH60R)ETHICON ENDO-SURGERY LLCGENERAL AND PLASTIC SURGERYOXC
1/11/2026K252842SAFEOP 3: NEURAL INFORMATIX SYSTEMALPHATEC SPINE, INC.NEUROLOGYGWF
1/14/2026K253585SMARTGUARD TECHNOLOGY; PREDICTIVE LOW GLUCOSE TECHNOLOGYMEDTRONIC MINIMED INCCLINICAL CHEMISTRYQJI
1/16/2026K251934QXR-DETECTQURE.AI TECHNOLOGIESRADIOLOGYMYN
1/23/2026P160007/S058GUARDIAN CONNECT SYSTEMMEDTRONIC MINIMEDCLINICAL CHEMISTRYMDS
1/28/2026K251306SEG PRO V3 (RT-300)EVER FORTUNE.AI, CO., LTD.RADIOLOGYQKB
2/3/2026K253701SMARTGUARD TECHNOLOGY; PREDICTIVE LOW GLUCOSE TECHNOLOGYMEDTRONIC MINIMEDCLINICAL CHEMISTRYQJI
2/4/2026K252301HENAN YADU LEVEL 4 SURGICAL GOWNS (REINFORCED)YADU MEDICAL (HENAN) CO., LTD.GENERAL HOSPITALFYA
2/12/2026P980043/S086HANCOCK II PORCINE BIOPROSTHESISMEDTRONIC, INC.CARDIOVASCULARDYE
2/12/2026P990064/S096MEDTRONIC MOSAIC PORCINE BIOPROSTHETIC HEART VALVEMEDTRONIC, INC.CARDIOVASCULARDYE
2/12/2026P130021/S179MEDTRONIC COREVALVE SYSTEMMEDTRONIC, INC.CARDIOVASCULARNPT
2/12/2026P970031/S078MEDTRONIC FREESTYLE AORTIC ROOT BIOPROSTHESISMEDTRONIC, INC.CARDIOVASCULARLWR
2/12/2026P170006/S040AVALUS BIOPROSTHESISMEDTRONIC INC.CARDIOVASCULARLWR
2/12/2026P140017/S033MELODY TRANSCATHETER PULMONARY VALVE (TPV), ENSEMBLE TRANSCATHETER VALVE DELIVERY SYSTEM (DS)MEDTRONIC INC.CARDIOVASCULARNPV
2/12/2026P200046/S045HARMONY TPV SYSTEMMEDTRONIC, INC.CARDIOVASCULARNPV
2/20/2026K252972CARTO 3 EP NAVIGATION SYSTEM V8.4BIOSENSE WEBSTER, INC.CARDIOVASCULARDQK
3/2/2026K253593CLARIUS EJECTION FRACTION AICLARIUS MOBILE HEALTH CORP.RADIOLOGYQIH
3/3/2026P250008CLAIRE OCT SYSTEMPERIMETER MEDICAL IMAGING AI INC.GENERAL AND PLASTIC SURGERYSHH
3/12/2026K253743MINIMED FLEX PUMPMEDTRONIC MINIMED INCCLINICAL CHEMISTRYQFG
3/17/2026DEN250014TYTO INSIGHTS FOR EARDRUM BULGING DETECTIONTYTO CARE LTD.EAR, NOSE, AND THROATSHL
3/23/2026P250022/S001VOLT PFA SYSTEMABBOTT MEDICALCARDIOVASCULARQZI
3/26/2026K260281VITEK 2 AST-STREPTOCOCCUS CEFUROXIME (<=0.125 - >=8 µG/ML)BIOMERIEUXMICROBIOLOGYLON
3/27/2026K253270CONTOUR PROTðGðAI+MIM SOFTWARE INC.RADIOLOGYQKB
3/28/2026K252360ECG-AI PULMONARY HYPERTENSION (PH) 12-LEAD ALGORITHM (1020)ANUMANA, INC.CARDIOVASCULARSAT
3/30/2026K260812FRANCISELLA TULARENSIS REAL-TIME PCR ASSAYCENTERS FOR DISEASE CONTROL AND PREVENTIONMICROBIOLOGYSGA
4/7/2026K253801ECG-AI CARDIAC AMYLOIDOSIS (CA) 12-LEAD ALGORITHM (1040)ANUMANA, INCCARDIOVASCULARSHP
4/10/2026K253689SYNGO DYNAMICS (VA41F)SIEMENS HEALTHCARE GMBHRADIOLOGYQIH
4/15/2026K260282VITEK 2 AST-STREPTOCOCCUS INDUCIBLE CLINDAMYCIN RESISTANCEBIOMERIEUX, INC.MICROBIOLOGYLON
4/15/2026K253573ASTAR BC G- KITQ-LINEA ABMICROBIOLOGYSAN
4/17/2026K260696MTS GEPOTIDACIN 0.016-256 µG/MLLIOFILCHEM S. R. L.MICROBIOLOGYJWY
4/17/2026K252945NOVASIGHT HYBRID SYSTEMCONAVI MEDICAL INC.CARDIOVASCULAROBJ
4/20/2026K260212ENSITE X EP SYSTEMABBOTT MEDICALCARDIOVASCULARDQK
4/21/2026K260447ETEST GEPOTIDACIN (GEP) (0.016-256 µG/ML)BIOMðRIEUX SAMICROBIOLOGYJWY
4/21/2026K260216CD HORIZON MODULEX FENESTRATED SCREW SET; CD HORIZON MODULEX SPINAL SYSTEMMEDTRONIC SOFAMOR DANEK USA, INCORTHOPEDICNKB
4/30/2026K253976ILET ACE PUMPBETA BIONICS, INCCLINICAL CHEMISTRYQFG
5/4/2026K254115ARTERAAI BREASTARTERA INC.PATHOLOGYSHW
5/6/2026K253152BP1000 ULTRA-COMPACT FINGERTIP BLOOD PRESSURE MONITOR (BP1000)BIOZEN, LLCCARDIOVASCULARDXN
5/11/2026K260842HARDYDISK AST GEPOTIDACIN 10µG (GEP10)HARDY DIAGNOSTICSMICROBIOLOGYJTN
5/14/2026K254013SUBTLEHD-PET (1.X)SUBTLE MEDICAL, INC.RADIOLOGYLLZ
5/19/2026P250027GUARDANT360 LIQUID CDXGUARDANT HEALTH, INC.PATHOLOGYPQP
5/21/2026K260582ROSA® SHOULDER SYSTEMORTHOSOFT INC. (D/B/A) ZIMMER CASRADIOLOGYOLO
5/21/2026K254131BACDEEPHEALTH, INC.RADIOLOGYQIH
5/21/2026K260234MSK GOSMART ALFA TEKNOLOJI SAN. VE TIC. A.S.RADIOLOGYQIH
5/22/2026DEN250057PCWP ANALYSIS SOFTWARECARDIOSENSE, INC.CARDIOVASCULARSIF
5/27/2026K254102GLUCOMMANDERGLYTEC, LLCCLINICAL CHEMISTRYNDC
6/11/2026P960030/S084PASSIVE PLUS DX ENDOCARDIAL STEROID ELUTING, PASSIVE-FIXATION PACING LEADSABBOTT MEDICALCARDIOVASCULARNVN
6/11/2026P960013/S133TENDRIL DX MODEL 1388T & 1388K ENDOCARDIAL PACING LEADSABBOTT MEDICALCARDIOVASCULARNVN
6/12/2026K253088ONESTIM-DUO CARDIAC STIMULATOR (MP5003-4CO)MICROPACE PTY LTDCARDIOVASCULARDQK
6/12/2026K260406BRAINOMIX 360 HYPERDENSITYBRAINOMIX LIMITEDRADIOLOGYQIH
6/18/2026K260636REMANUFACTURED LIGASURE XP MARYLAND JAW SEALER/DIVIDER WITHOUT NANO-COATINGSTRYKER SUSTAINABILITY SOLUTIONSGENERAL AND PLASTIC SURGERYGEI
6/24/2026K261260HD CAMERA HEAD OLYMPUS CH-S200-08-LBOLYMPUS MEDICAL SYSTEMS CORPORATIONGASTROENTEROLOGY-UROLOGYFET
6/25/2026K253786SYNGO.CT CORONARY COCKPITSIEMENS MEDICAL SOLUTIONS USA, INC.RADIOLOGYQIH
6/26/2026K253316PHAST INSTRUMENT AND PHAST BLOOD CULTURE GRAM-NEGATIVE PANELPHAST CORP.MICROBIOLOGYSAN
6/30/2026P030031/S143BIOSENSE WEBSTER NAVISTAR/CELSIUS THERMO COOL DIAGNOSTIC/ABLATION DEFLECTABLE TIP CATHETERSBIOSENSE WEBSTER INCCARDIOVASCULAROAD
6/30/2026P240006/S016VARIPULSE PLATFORM (VARIPULSE CATHETER; TRUPULSE GENERATOR; STERILE INTERFACE CABLE; NGEN PUMP)BIOSENSE WEBSTER INCCARDIOVASCULARQZI
7/1/2026K253527TRI-STAPLE 2.0 BLACK CIRCULAR RELOAD 21MM EXTRA THICK (SIGCIR21XT); TRI-STAPLE 2.0 BLACK CIRCULAR RELOAD 25MM EXTRA THICK (SIGCIR25XT)COVIDIEN LLCGENERAL AND PLASTIC SURGERYGAG
7/1/2026K261021EASYWHIP FAMILYWINTER INNOVATIONS, INC.GENERAL AND PLASTIC SURGERYGAT
7/2/2026K253884KOIOS DS (3.8)KOIOS MEDICAL INC.RADIOLOGYPOK
7/7/2026K260973INFEROPERATE SUITEBEIJING INFERVISION HEALTHCARE MEDICAL TECHNOLOGY CO., LTD.RADIOLOGYQIH
7/8/2026K260160VITEK REVEAL GN AST ASSAY AND VITEK REVEAL AST SYSTEMBIOMERIEUX, INC.MICROBIOLOGYSAN
7/9/2026K261962SMARTGUARD TECHNOLOGY; PREDICTIVE LOW GLUCOSE TECHNOLOGYMEDTRONIC MINIMEDCLINICAL CHEMISTRYQJI
7/10/2026K261949NYES TUFF CUFF L3 DISPOSABLE GOWNSNEW YORK EMBROIDERY STUDIOGENERAL HOSPITALFYA
7/13/2026K261085RAPID VESSEL OCCLUSION (RAPID VO)ISCHEMAVIEWRADIOLOGYQAS
7/13/2026K254044HEARING ASSISTMETA PLATFORM TECHNOLOGIES, LLC.EAR, NOSE, AND THROATSCR
7/15/2026K253606NANOZOOMER S20MD SLIDE SCANNER SYSTEM; NANOZOOMER S360MD SLIDE SCANNER SYSTEMHAMAMATSU PHOTONICS K.K.PATHOLOGYPSY
7/20/2026K261011SLEEP APNEA FEATURE (V2.0)SAMSUNG ELECTRONICS CO., LTD.ANESTHESIOLOGYQZW
7/22/2026K260964NANOZOOMER S540MD SLIDE SCANNER SYSTEMHAMAMATSU PHOTONICS K.K.PATHOLOGYPSY
7/22/2026K261113OP.N SPINE NAVIGATION AND ROBOTIC GUIDANCEECENTIAL ROBOTICSORTHOPEDICOLO
7/23/2026K261853GEPOTIDACIN GEP 10 µGLIOFILCHEM INC.MICROBIOLOGYJTN
7/24/2026K261781THERMO SCIENTIFIC OXOID CEFEPIME/ZIDEBACTAM DISC (30/30 µG) FPZ60THERMO FISHER SCIENTIFIC (OXOID LTD.)MICROBIOLOGYJTN
7/29/2026K261210CLEERLY PLAQUECLEERLY, INC.RADIOLOGYQIH
7/30/2026K260980CMANGIO® V2.0CUREMETRIX INC.RADIOLOGYQIH
7/31/2026K261187HEARING AID FEATURESAMSUNG ELECTRONICS CO., LTD.EAR, NOSE, AND THROATSCR
8/6/2026K261059OVERJET MULTI-IMAGE CARIES AND CHARTING ASSISTOVERJET INCRADIOLOGYMYN
8/7/2026K253554APERIO GT 450 DX; APERIO GT 180 DXLEICA BIOSYSTEMS IMAGING, INC.PATHOLOGYPSY
8/7/2026P060019/S059IBI THERAPY COOL PATH ABLATION CATHETER & IBI-1500T9 RFIRVINE BIOMEDICAL, INC.CARDIOVASCULAROAD
8/7/2026P040042/S059THERAPY DUAL 8 CARDIAC ABLATION SYSTEM,THERAM 8MM THERMISTER ABLATION CATHETER SAFIRE TX ABLATION CATHETERIRVINE BIOMEDICAL,INC.(IBI)CARDIOVASCULAROAD
8/7/2026P110016/S095THERAPY COOL PATH DUO/ SAFIRE BLU DUO ABLATION CATHETER AND IBI 1500T9-CP V1.6 CARDIAC ABLATION GENERATORABBOTT MEDICALCARDIOVASCULAROAD
8/7/2026P130026/S102TACTICATH QUARTZ SETST JUDE MEDICALCARDIOVASCULAROAE
8/7/2026P220013/S033TACTIFLEX ABLATION CATHETER, SENSOR ENABLED, TACTISYS QUARTZ EQUIPMENT, TACTISYS QUARTZ, TACTIFLEX RADIOFREQUENCY CABLE, AMPERE RADIOFREQUENCY GENERATOR, COOL POINT PUMPABBOTT MEDICALCARDIOVASCULAROAE
8/7/2026P960016/S094LIVEWIRE(R) CARDIAC ABLATION SYSTEMST. JUDE MEDICALCARDIOVASCULARLPB
8/7/2026P040014/S051IBI THERAPY CARDIAC ABLATION SYSTEM ERS/ 1500T RF GENERATORIRVINE BIOMEDICAL, INC.CARDIOVASCULARLPB
8/20/2026K262113HARDYDISK AST TEBIPENEM 10µG (TBP10)HARDY DIAGNOSTICSMICROBIOLOGYJTN
8/21/2026K253699TEMPUS ECG-PHTEMPUS AI, INC.CARDIOVASCULARSAT
8/25/2026K261595CEEVRA REVEAL 4CEEVRA, INC.RADIOLOGYQIH
8/26/2026K260811HEARTVUE.PROTONHEARTVUE.AI, INC.RADIOLOGYQIH
8/26/2026K261153LITHOVUE ELITE DIGITAL FLEXIBLE URETEROSCOPE SYSTEMBOSTON SCIENTIFIC CORPORATIONGASTROENTEROLOGY-UROLOGYFGB
8/26/2026K260407PRINCIPLE & WILL ANSI/AAMI PB 70 (STERILE/NON-STERILE) LEVEL 3 SURGICAL GOWNS (REINFORCED) AND PRINCIPLE & WILL ANSI/AMMI PB70 (STERILE/NON-STERILE) LEVEL 4 SURGICAL GOWNS/ISOLATION GOWNSPRINCIPLE & WILL CO., LTD.GENERAL HOSPITALFYA
9/2/2026P210011/S002XT CDXTEMPUS LABS, INC.PATHOLOGYPQP
9/2/2026K261313VIVID EXPLORER/VIVID ADVANCED/VIVID FOCUSGE MEDICAL SYSTEMS ULTRASOUND AND PRIMARY CARE DIAGNOSTICSRADIOLOGYIYN
9/3/2026DEN250044STEMI AI ECG MODELPOWERFUL MEDICAL INC.CARDIOVASCULARSHS
9/8/2026K260062MODI SYSTEMDREAMED DIABETES , LTD.CLINICAL CHEMISTRYQRX
9/9/2026K252486CHESTVIEW US (1.1)GLEAMERRADIOLOGYMYN
9/11/2026K262763ILET DOSING DECISION SOFTWAREBETA BIONICS, INCCLINICAL CHEMISTRYQJI
9/14/2026K261517QCT LUNG (V1.0)QURE.AI TECHNOLOGIES PVT. LTD.RADIOLOGYQDQ
9/16/2026K261623ECHOPAC SOFTWARE ONLY / ECHOPAC PLUG-IN (V211)GE MEDICAL SYSTEMS ULTRASOUND AND PRIMARY CARE DIAGNOSTICSRADIOLOGYQIH
9/16/2026K260906A2Z-ABDO-TRIAGEA2Z RADIOLOGY AI INC.RADIOLOGYQAS
9/16/2026K261896IOS CAMAPS FXCAMDIABCLINICAL CHEMISTRYQJI
9/22/2026P220013/S034TACTIFLEX ABLATION CATHETER, SENSOR ENABLED, TACTISYS QUARTZ EQUIPMENT, TACTISYS QUARTZ, TACTIFLEX RADIOFREQUENCY CABLE, AMPERE RADIOFREQUENCY GENERATOR, COOL POINT PUMPABBOTT MEDICALCARDIOVASCULAROAE
9/23/2026K260597VIRTUE QAPRONOVA SOLUTIONSRADIOLOGYLHN
9/25/2026K254297TEMPUS ECG-MRTEMPUS AI, INC.CARDIOVASCULARSBQ
9/28/2026K260619VITEK 2 AST-STREPTOCOCCUS CEFEPIME (<=0.06 – >=4 µg/mL)BIOMERIEUX, INC.MICROBIOLOGYLON
9/30/2026K263035CDC HUMAN INFLUENZA VIRUS REAL-TIME RT-PCR DIAGNOSTIC PANEL (INCLUDES INFLUENZA A/B TYPING KIT, INFLUENZA A SUBTYPING KIT, INFLUENZA A(H5) SUBTYPING KIT, AND INFLUENZA B LINEAGE GENOTYPING KIT)CENTERS FOR DISEASE CONTROL AND PREVENTIONMICROBIOLOGYOZE

Examples of Devices with PCCPs


Additional Resources