Novel Medical Devices with Predetermined Change Control Plans: Hypothetical Case Examples
Section 515C of the Federal Food, Drug, and Cosmetic Act (FD&C Act) sets forth certain requirements relating to Predetermined Change Control Plans (PCCPs), and the FDA has issued guidance relating to PCCPs. The hypothetical case examples below illustrate, consistent with these resources, how a PCCP might be used for different types of novel medical devices.
The hypothetical case examples and corresponding information do not include or address all the information expected to be provided in a PCCP submitted as part of a device marketing submission, though they may be representative of the level of information provided in a public device summary about a PCCP.
All modifications described in these examples are premised on the proposed modifications being appropriate for inclusion in a PCCP, meaning that, consistent with sections 515C(a)(2) and 515C(b)(2) of the FD&C Act:
- the modifications would generally otherwise require a new marketing submission (see 21 CFR 807.81(a)(3) or 21 CFR 814.39(a)),
- the modifications would maintain the device within its intended use, and
- for devices subject to 510(k) requirements, the modifications would allow the device to remain substantially equivalent to the predicate device.
In addition, while the information included in the hypothetical case examples does not represent a complete PCCP, the examples contain content that represents the three components of a PCCP: the Description of Modifications, the Modification Protocol, and the Impact Assessment. For recommendations on the content of a PCCP, please see Predetermined Change Control Plans.
DISCLAIMER: These hypothetical case examples are provided for illustrative purposes only. These examples help illustrate key concepts for certain novel medical devices and PCCPs consistent with FDA’s statute, regulations and guidance.
Indications for Use
The device is intended to aid clinicians in the early identification of patients at risk of developing sepsis within 24 hours of patient assessment in acute care hospital settings by analyzing routinely collected electronic health record (EHR) data, including vital signs, laboratory values, and nursing documentation, and generating a real-time risk score and alert to prompt clinical evaluation.
Device Description
The device integrates with a hospital system and incorporates a machine learning model to continuously analyze structured patient data streams and generate a sepsis risk score and corresponding dashboard alert to bedside nurses and attending physicians. The device’s “global” model is trained on a multi-site national dataset.
PCCP
A PCCP could be designed to accommodate local adaptations of the authorized device using site-specific historical patient data from each hospital system to account for differences in patient population demographics, documentation practices, laboratory reference ranges, and institutional treatment protocols. Such a PCCP would not require new marketing submissions for each deployment, provided that each adaptation remains within the defined scope of the Description of Modifications and meets the acceptance criteria of the Modification Protocol.
For purposes of this example, the following modifications are specified:
- Modification 1: Adjust the relative weighting of structured nursing documentation inputs within a prespecified, allowable range (specifically, mental status assessments and skin color/temperature flags) to reflect site-specific documentation conventions (“site-specific input adjustment”).
- Modification 2: Re-training of the model’s final output layer(s) using site-specific EHR data to tune the risk score for a local patient population (“local model fine-tuning”).
- Modification 3: Modify the risk score threshold within a prespecified, allowable range at which a sepsis alert is surfaced to clinical users, to account for site-specific clinical workflows and baseline sepsis prevalence (“site-specific alert threshold adjustment”).
The scope of these modifications in the PCCP is limited to optimizing the device for specific sites with a locally-adapted model; the device containing the global model (without modifications) remains available for sites that do not intend to deploy the device with a locally-adapted model (“locally-adapted device”) or, if needed, to return the device to this configuration if issues occur with the locally-adapted device at a particular site. Modifications are planned to be implemented on an ongoing basis; the first implementation is to modify the device with a locally-adapted model (all modifications described above), and the subsequent iterations are for ongoing optimization of performance of the device with the locally-adapted model (one or more modifications; interdependencies defined below).
The scope of these modifications is limited to the device as deployed with the locally-adapted model. The base model architecture, feature set, and risk score calculation are maintained with these modifications. The inputs/output data types are also maintained; only structured, coded nursing documentation fields are to be reweighted (free-text fields are excluded). Once modifications are made to locally adapt the device, it will be trained, tuned, and locked prior to release of the device to the deploying site. Local adaptations are implemented manually, necessitating prespecified minimum local data and review and approval by the manufacturer prior to deployment at each new site.
A method for implementing the modifications could be designed as follows:
- Data Sufficiency Review: Verify that the prospective site has sufficient historical data meeting the minimum sample size and data quality requirements specified in the PCCP prior to initiating any modification.
- Model Training and Locking: Execute site-specific input adjustment, local model fine-tuning using the prespecified training pipeline, and/or site-specific alert threshold adjustment. The locally-adapted model would be locked (i.e., fixed weights, no further changes) prior to validation testing.
- Performance Validation: Conduct site-specific validation testing on a sufficient, independent test dataset that was sequestered prior to model training and aligns with data re-use requirements, confirming that acceptance criteria are met, as measured against the performance of the authorized device (i.e., the device containing the global model) and an appropriate ground truth comparator, that are prespecified in the PCCP.
- Documentation and Labeling Update: Updated site-specific labeling and performance documentation, including, for example, a summary of performance of the device with the locally-adapted model relative to the performance of the authorized device as well as the locally-adapted device and model version history.
- Deployment: Deploy the locally-adapted device at the site via a controlled software update process consistent with the manufacturer's quality management system (QMS).
- Post-Deployment Monitoring: Initiate site-level post-market surveillance, including ongoing tracking of alert positive predictive value, sensitivity, and clinical response metrics.
Data management PCCP requirements for modifications could include: data is sourced exclusively from the deploying hospital system’s EHR; data is de-identified in accordance with relevant regulations and standards prior to use in local model fine-tuning; and data provenance is documented and retained per the manufacturer’s QMS.
Additional verification and validation activities, such as software verification, could be applicable depending on the details of the modifications.
Whether modifications can be implemented individually or in combination depends on the desired modifications to be made by the deploying hospital system. Modification 1 impacts both Modification 2 and Modification 3, in that it may change how inputs are weighted, which can change the calculation of the risk score and therefore, the generation of the alert. Modification 2 impacts Modification 3, in that the fine-tuning of the model may impact the risk score output, which impacts the generation of the alert. Modification 2 does not impact Modification 1. Modification 3 does not impact Modification 1 or 2.
Collectively, in consideration of these interdependencies, when modifications are made to locally adapt the device, all modifications would be implemented in the following sequence: Modification 1, 2, and 3, as outlined and required by the PCCP. In ongoing optimization of the locally-adapted device, when implementing Modification 1, it would be followed by implementation of Modification 2 and then implementation of Modification 3. In addition, when implementing Modification 2, it would be followed by implementation of Modification 3 to ensure the alert generation threshold is within the prespecified, allowable range. Modification 3 could be implemented individually.
When all modifications are implemented to locally adapt the device and/or when one or more modifications are implemented for ongoing optimization of locally-adapted device performance, the cumulative impact of all modifications is improved performance for the deploying hospital system’s patient population.
Whether modifications are implemented individually or in combination with each other, the methods described above, as well as other elements of the Modification Protocol not described in this example, provide verification and validation of the entire device and continue to ensure the safety and effectiveness of the device. Further, the benefits and risks, including risks of harm, have been assessed to demonstrate that the combination of the proposed modifications is unlikely to introduce additional, unmitigated risks.
Indications for Use
The device is intended to assist cardiologists in detecting clinically significant cardiac arrhythmias, including atrial fibrillation, ventricular tachycardia, and third-degree heart block, in patients undergoing continuous inpatient cardiac telemetry monitoring.
Device Description
The device is integrated into a hospital-based cardiac telemetry monitoring platform. The device receives continuous electrocardiogram (ECG) waveform data from bedside monitors and applies a deep learning model to classify rhythm strips and generate real-time arrhythmia alerts.
PCCP
A PCCP could be designed to accommodate automated, periodic re-training of the deep learning arrhythmia detection model across all deployed sites as new annotated ECG data is accrued, without requiring a new marketing submission for each re-training cycle, provided that all model updates remain within the defined scope of the Description of Modifications and meet the automated validation acceptance criteria of the Modification Protocol.
For purposes of this example, the following modifications are specified:
- Modification 1: Re-training of the full deep learning model on an expanded dataset incorporating newly annotated rhythm strip data from across the deploying institutions to improve performance (“periodic model re-training”).
- Modification 2: Modify the classification threshold applied to model output probabilities for one or more arrhythmia classes to maintain calibrated sensitivity and specificity across the current data distribution (“operating point adjustment”).
Modifications are planned to be implemented on an ongoing basis, when sufficient data has been collected to trigger re-training. Changes may impact layer structure, parameter weights, or input/output format, while maintaining input/output specifications and overall model architecture. Performance of the device with the re-trained model must remain within pre-specified allowable sensitivity/specificity bounds for each arrhythmia class as defined in the labeling. The planned modifications are to be implemented automatically and globally by software at the conclusion of each re-training cycle without human input, action, review or decision-making, following successful completion of automated validation checks.
A method for implementing the modifications could be designed as follows:
- Re-Training Trigger: A re-training cycle is initiated automatically when a prespecified number of newly annotated rhythm strip segments have been accrued in the training pool.
- Data Preparation: De-identified rhythm strip data meeting the annotation and data quality criteria specified in the PCCP are assembled into the re-training dataset. A sufficient, independent test dataset is sequestered prior to model training and data re-use requirements are met.
- Model Re-Training: The deep learning model is re-trained on the assembled dataset using the prespecified training pipeline.
- Operating Point Re-Evaluation: Following re-training, the classification threshold for each arrhythmia class is re-evaluated against the updated model’s output probability distribution on the test dataset, and updated as needed to maintain calibrated sensitivity and specificity within the prespecified allowable bounds. The re-evaluated operating point is applied to the updated model prior to validation and deployment.
- Automated Validation: The performance of the device with the updated model is evaluated against the independent, sequestered test dataset using the automated validation pipeline. Performance metrics are computed for each arrhythmia class as well as for each predefined subgroup and compared against acceptance criteria (i.e., performance of authorized device) that is prespecified in the PCCP.
- Automated Deployment Decision: If all acceptance criteria are met, the device with the updated model is automatically promoted for deployment across the installed base. If any acceptance criterion is not met, deployment is blocked, an alert is generated to the manufacturer, and the device remains unmodified.
- Deployment: The device with the updated model is deployed to all sites via a controlled software update process consistent with the manufacturer’s QMS. Deployment is staggered across sites in accordance with the manufacturer's rollout protocol to allow for early detection of unexpected deployment issues.
- Documentation and Labeling Update: The device's user interface is updated to display the current device and model version and the date of the most recent model update, so that clinical users are aware that a model update has occurred. Instructions for Use are updated to reflect the current device and model version and relevant performance information, including operating point, and the updated Instructions for Use are made accessible within the platform.
- Audit Trail: A complete record of each re-training cycle is automatically generated and retained, including training dataset composition and version, validation results, deployment timestamp, and model version identifier.
- Post-Deployment Monitoring: Automated post-market monitoring tracks per-site and aggregate arrhythmia detection performance on an ongoing basis. Predefined drift thresholds trigger a review by the manufacturer if performance deviates beyond acceptable limits.
Data management PCCP requirements for re-training cycles include: all rhythm strip data is to be de-identified prior to inclusion in the training or test datasets; annotation is to be performed per the prespecified annotation protocol; data contributions from individual sites are logged and version-controlled; training and test datasets are sequestered; and data provenance is documented and retained per the manufacturer's QMS. Data within training and test datasets can be re-used as specified in the PCCP.
Additional verification and validation activities, such as software verification, could be applicable depending on the details of the modifications.
Modifications are intended to be implemented in sequence each time, because Modification 1 impacts Modification 2, in that the model re-training can impact the model’s output probabilities, warranting at a minimum, re-evaluation of the classification thresholds to determine if an update is needed to maintain calibrated sensitivity and specificity within the prespecified allowable bounds. The methods outlined above address Modification 1’s impact on Modification 2 (i.e., Step 4 above), and along with other elements of the Modification Protocol not described in this example, provide verification and validation of the entire device and continue to ensure the safety and effectiveness of the device. Further, the benefits and risks, including risks of harm, have been assessed to demonstrate that the combination of the proposed modifications is unlikely to introduce additional, unmitigated risks.
The automatic implementation of these modifications presents risks in that there is no human-in-the-loop when modifications are implemented, validated, and deployed. As described above, methods are in place to control deployment of the device with an updated model when any acceptance criteria are not met, while post-deployment monitoring can highlight drifts in performance if re-training is not maintaining or improving performance as intended.