Magellan Diagnostics Inc. Expands Recall for LeadCare Testing Systems Due to Inaccurate Test Results
The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death.
Recalled Product
- Product Name: LeadCare and LeadCare II Blood Lead Testing Systems
- Serial Numbers: All serial numbers
- Kit Lot Numbers: All Kit Lot Numbers
- Manufacturing Dates: September 2013 to Present
- Distribution Dates: January 1, 2014 to May 22, 2017
- Devices Recalled in the U.S.: 7,058,516 systems nation wide
Device Use
The LeadCare and the LeadCare II Testing Systems detect the amount of lead in a blood sample obtained from finger or heel prick (capillary) or from a vein (venous).
Reason for Recall
Magellan Diagnostics is expanding LeadCare Plus and Ultra Testing Systems recall to include two additional testing systems; LeadCare and LeadCare II Blood Lead Testing Systems.
Magellan Diagnostics is recalling the LeadCare and the LeadCare II Testing Systems because they may underestimate the blood lead levels (BLL) and give inaccurate results when processing venous blood samples. Falsely lower test results may lead to improper patient management and treatment for lead exposure or poisoning. The use of affected product may cause serious adverse health consequences.
This recall accompanies FDA's safety communication from May 17, 2017. Magellan's LeadCare and LeadCare II Testing Systems are two of four blood lead testing systems affected by the recommendations in FDA's safety communication.
The FDA is unable to identify the root cause for the inaccurate results, based on data provided by Magellan. We are conducting studies with the Center for Disease Control and Prevention (CDC) to identify the cause and better characterize the extent of the problem.
Who May be Affected
- Laboratories that use Magellan's LeadCare Testing Systems as part of diagnostic applications.
- Laboratory personnel who interpret the results of Magellan's LeadCare Testing Systems.
- Health care professionals who perform lead tests using Magellan's LeadCare Testing Systems.
- Patients being tested for elevated BLLs with Magellan's LeadCare Testing Systems, in particular children in high-risk environments with BLL in the 5 to14 micrograms per deciliter (μg/dl) range.
What to Do
On May 17 2017, Magellan Diagnostics Inc. sent a "Customer Safety Communication" letter to all affected customers. The letter provided the following information:
- Do NOT use venous blood samples with any LeadCare Blood Lead Testing Systems.
- All LeadCare Blood Lead Testing Systems can be used with capillary blood samples, for example:
- Capillary tubes shipped in LeadCare II test kits
- RAM Scientific SAFE-T-FILL capillary collection tubes
- Complete the "Fax Form Record: Notification on LeadCare Systems" and return it by fax 978-600-1480 or 888-789-8040. Or, fill out the online version of the form, available at www.leadcare2.com/recall
- Share this safety information with appropriate personnel and retain a copy of the letter.
The FDA recommends laboratories and health care professionals take the following actions:
- Discontinue using Magellan's LeadCare System Testing Systems with venous blood samples. At this time, all LeadCare systems can be used with capillary blood samples.
- Laboratories and health care professionals should follow recommendations in the FDA Safety Communication issued on May 17, 2017 rather than previous communications from Magellan Diagnostics on their LeadCare Test Systems, including Magellan's Field Safety Correction Notification dated April 28, 2017.
- Report any adverse events to the FDA and to Magellan Diagnostics.
- If laboratories or health care professionals are concerned about using the LeadCare Test Systems, the alternative options are mass spectrometry or atomic absorption methods. These are not point-of-care tests, and may be available only from larger-capacity laboratories such as reference labs.
The FDA is also encouraging parents and at-risk adults to follow CDC's recommendations listed in their Health Advisory issued on May, 17 2017 regarding any necessary re-testing based on this recall.
Contact Information
Customers with questions are instructed to contact Magellan Diagnostics' Product Support by telephone at 800-275-0102 with any questions related to this recall.
Date Recall Initiated
May 17, 2017
Additional Resources
- LeadCare Plus and Ultra Testing System Recall Notice
- FDA Warns Against Using Magellan Diagnostics LeadCare Testing Systems with Blood Obtained from a Vein: FDA Safety Communication
- CDC's Health Advisory - May 17, 2017
- CDC - Lead
- CDC's Lead Exposure in pregnancy and lactating women
- CDC's Adult Blood Lead Epidemiology & Surveillance (ABLES)
- American Congress of Obstetricians and Gynecologists Opinion on Lead Screening During Pregnancy and Lactation
- U.S. Occupational Safety and Health Administration-Medical Surveillance Guidelines
- American Academy of Pediatrics: Lead Screening
How do I report a problem?
Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program either online, by regular mail or by FAX to 1-800-FDA-0178.