Heart Pump Controller Recall: Abiomed Removes Automated Impella Controllers (AIC)

This is an update to an Early Alert. The affected products and recommendations for what to do with the devices below have not changed. This recall involves removing certain devices from where they are used or sold. The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it. 

Affected Product

Abiomed Automated Impella Controller unit

The FDA is aware that Abiomed has issued a letter to affected customers recommending certain Automated Impella Controllers (AIC) be removed from where they are used or sold. While awaiting service from Abiomed, hospital inventory can continue to be used to ensure continuity of care.

Affected devices:

Device NameProduct NumberUDI-DISerial Numbers
Impella Controller, Packaged, US0042-0000-US00813502010022Limited to Specific Consoles
Impella Optical Controller, Packaged, US0042-0010-US00813502010985
Optical, AIC, Impella Controller, Packaged US0042-0040-US00813502011401

List of Impacted Serial Numbers

The serial number of the AIC is located on the label on the right side of the device as shown below.

Image depicts where to identify the Product Number and Serial Number on the Impella Controller

What to Do

While awaiting service from Abiomed, hospital inventory can continue to be used to ensure continuity of care. Various hardware, firmware, and/or software updates are available but have not been implemented consistently across all potentially impacted units. Users should review the recommendations below while waiting to coordinate the scheduling of removal, updates, and return of impacted units with Abiomed.

On September 18, 2026, Abiomed sent all affected customers a letter recommending the following actions:

  • Hospital inventory can continue to be used while awaiting service from Abiomed to ensure continuity of care. 
  • The Abiomed service team will contact you to coordinate the removal, update, and return of the device(s).
  • Forward to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products).
  • If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice.

Reason for Recall

Abiomed stated that a retrospective review of servicing records identified certain Impella Controllers that require hardware, firmware, and/or software updates to address quality issues with the potential for safety implications. These quality issues include a potential failure of the AIC to recognize the purge cassette during setup or change; a sticking, unreliable button that could impact the operator’s ability to select and confirm menu entries and possible loss of control of the console; an instability that could result in start-up failure; and other software defects. Without the hardware, firmware, and/or software updates, patients could face various problems such as delay of therapy, inadequate or loss of hemodynamic result, and other serious issues.

As of September 8, 2026, Abiomed has reported no serious injuries or deaths associated with this issue.

Device Use

The Automated Impella Controller is the primary user control interface for the Impella Catheters. It controls the Impella Catheter and monitors the catheter for alarms. Impella therapy aims to reduce the work of the heart and provide support for the circulatory system so the heart has time to recover.

Contact Information

Customers in the U.S. with adverse reactions, quality problems, questions about this issue, or if the Serial Number is missing or illegible, should contact the Abiomed Field Service Team, 1-800-422-8666, option 3, or email ra-abm-fieldaction@its.jnj.com for assistance.

Additional FDA Resources   

Unique Device Identifier (UDI)

The unique device identifier (UDI) helps identify individual medical devices sold in the United States from distribution to use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified more quickly, and as a result, problems potentially resolved more quickly.

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program.