Cybersecurity Resources for Medical Device Manufacturers: Guidance, Resources, and FAQs
Report a medical device cybersecurity issue
The FDA has published premarket and postmarket guidance documents that are intended to provide recommendations for comprehensive management of medical device cybersecurity risks, promote continuous improvement throughout the total product life cycle, and encourage device developers to update marketed and distributed medical devices to reduce risk. This page provides answers to frequently asked questions (FAQs) related to cybersecurity in medical devices, including Section 524B of the Federal Food, Drug, and Cosmetic (FD&C) Act.
On this page:
- Cybersecurity Guidance
- Section 524B FD&C Act, Ensuring Cybersecurity in Medical Devices
- Premarket Considerations
- Postmarket Considerations
- Additional Resources
Cybersecurity Guidance
| Final Guidance: Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions | This guidance provides FDA’s recommendations to industry regarding cybersecurity device design, labeling, and the documentation that FDA recommends be included in premarket submissions for devices with cybersecurity risk. Additionally, this guidance has been updated to address FDA’s recommendations regarding section 524B of the FD&C Act for cyber devices (see Section VII). This document supersedes the guidance, of the same title, issued on September 27, 2023. (6/27/2025) |
| Final Guidance: Postmarket Management of Cybersecurity in Medical Devices | Provides recommendations to industry for structured and comprehensive management of postmarket cybersecurity vulnerabilities for marketed and distributed medical devices throughout the product lifecycle. (12/27/2016) |
For more guidances, search “cybersecurity” in FDA guidance documents.
Section 524B of the FD&C Act
Congress granted the FDA explicit cybersecurity regulatory authority in 2022 by adding Section 524B to the FD&C Act. This authority took effect on March 29, 2023. Under Section 524B, medical devices that meet the definition of a “cyber device” must include certain information within their premarket submissions to demonstrate that “there is a reasonable assurance that the device and related systems are cybersecure.” This information includes:
- A plan to address postmarket cybersecurity vulnerabilities and exploits, including coordinated vulnerability disclosure and related procedures
- Processes and procedures to provide reasonable assurance of device cybersecurity, including postmarket updates and patches to devices and related systems
- Software Bill of Materials (SBOM)
Section 524B, Ensuring Cybersecurity of Devices, also authorizes the FDA to issue regulations with additional cybersecurity requirements that the FDA determines are necessary to demonstrate reasonable assurance of medical device cybersecurity.
Q1: Who is required to comply with section 524B of the FD&C Act? What types of premarket submissions does this apply to?
A: Under section 524B(a) of the FD&C Act, a person who submits a premarket application or submission— including 510(k), premarket approval application (PMA), Product Development Protocol (PDP), De Novo request, or Humanitarian Device Exemption (HDE)— for a device that meets the definition of a cyber device, as defined under section 524B(c), is required to submit information to ensure that the cyber device meets the cybersecurity requirements under section 524B(b). This includes Special and Abbreviated 510(k) submissions as well as PMA and HDE supplements.
Q2: What is a cyber device?
A: Section 524B(c) of the FD&C Act defines "cyber device" as a device that (1) includes software validated, installed, or authorized by the sponsor as a device or in a device, (2) has the ability to connect to the internet, and (3) contains any such technological characteristics validated, installed, or authorized by the sponsor that could be vulnerable to the cybersecurity threats. If manufacturers are unsure whether their device is a cyber device, they may contact the FDA. Further, the guidance Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions discusses section 524B of the FD&C Act in Section VII “Cyber Devices.”
Q3: Does this law only apply to future medical devices, rather than retroactively?
A: The cybersecurity requirements do not apply to an application or submission submitted to the FDA before March 29, 2023. If a cyber device was previously authorized, and the manufacturer is making a change to the device that requires premarket review by the agency, the law would apply for the new premarket submission.
Q4: What requirements apply to manufacturers of cyber devices under section 524B of the FD&C Act?
A: Section 524B(a) of the FD&C Act provides that the sponsor of a premarket submission for a cyber device include information to demonstrate that the cyber device meets the cybersecurity requirements in section 524B(b) of the FD&C Act. The requirements in section 524B(b) of the FD&C Act are:
- Submit a plan to monitor, identify, and address, as appropriate, in a reasonable time, postmarket cybersecurity vulnerabilities and exploits, including coordinated vulnerability disclosure and related procedures;
- Design, develop, and maintain processes and procedures to provide a reasonable assurance that the device and related systems are cybersecure, and make available postmarket updates and patches to the device and related systems; and
- Provide a software bill of materials, including commercial, open-source, and off-the-shelf software components.
The FDA may also issue regulations with other requirements to demonstrate reasonable assurance that the device and related systems are cybersecure. See FAQs 6-9 (below) for additional details on ways manufacturers might demonstrate that their devices are cybersecure.
Q5: When do manufacturers of cyber devices have to submit the information described in section 524B?
A: Manufacturers of cyber devices are required to submit this information starting March 29, 2023, in premarket submissions including 510(k), premarket approval application (PMA), Product Development Protocol (PDP), De Novo, or Humanitarian Device Exemption (HDE). This includes Abbreviated and Special 510(k) submissions and PMA/HDE supplements.
The Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions final guidance does not supersede the previously issued guidance Cybersecurity in Medical Devices: Refuse to Accept Policy for Cyber Devices and Related Systems, however, the policy in the latter guidance expired on October 1, 2023. Beginning October 1, 2023, the FDA expects that sponsors of cyber devices will have had sufficient time to prepare premarket submissions that contain information required by section 524B of the FD&C Act.
Additionally, as part of the FDA’s efforts to modernize the 510(k) Program and implement MDUFA V, starting October 1, 2023, all 510(k) submissions, unless exempted, must be submitted as electronic submissions using eSTAR, as noted in the Electronic Submission Template for Medical Device 510(k) Submissions final guidance. For eSTAR submissions, an eSTAR will be put on a Technical Screening hold if it does not contain accurate responses and relevant attachments in the Cybersecurity section of eSTAR.
Q6: Section 524B(b)(1) of the FD&C Act requires manufacturers of cyber devices to submit plans to manage vulnerabilities and exploits as part of their premarket submissions. What resources are available to manufacturers?
A: The guidance Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions and the guidance Postmarket Management of Cybersecurity in Medical Devices describe recommendations for managing cybersecurity after the device has been introduced into the market.
Q7: Section 524B(b)(2) of the FD&C Act requires, among other aspects, that manufacturers of cyber devices design, develop, and maintain processes and procedures to provide a reasonable assurance that the device and related systems are cybersecure. What resources are available to manufacturers?
A: The guidance Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions provides recommendations on cybersecurity considerations for devices and provides recommendations for documentation in device premarket submissions that may help manufacturers meet their obligations with the 524B(b)(2) requirements.
Q8: Section 524B(b)(2) of the FD&C Act also requires manufacturers of cyber devices to make available postmarket updates and patches to the device and related systems to address vulnerabilities. What resources are available to manufacturers?
A: The guidance Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions discusses throughout plans for patches and updates across the total product life cycle (TPLC). The guidance "Postmarket Management of Cybersecurity in Medical Devices" discusses cybersecurity routine updates and patches and describes patching in the context of remediating cybersecurity vulnerabilities.
Q9: Section 524B(b)(3) of the FD&C Act requires that manufacturers of cyber devices provide a software bill of materials (SBOM) for the commercial, open-source, and off-the-shelf software components contained within the device. What resources are available to manufacturers?
A: The guidance Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions discusses SBOMs in Section V.A.4(b). Additional information about SBOMs can be found in the October 2021 National Telecommunications and Information Administration (NTIA) Multistakeholder Process on Software Component Transparency document Framing Software Component Transparency: Establishing a Common Software Bill of Materials (SBOM).
Q10: Do I need to submit information to demonstrate that I comply with the requirements in section 524B for all device modifications?
A: In keeping with least burdensome principles, the information we recommend that manufacturers of cyber devices provide will generally differ based on the type of change and whether such change impacts the cybersecurity of the device. The guidance Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions discusses the different changes that are likely or unlikely to impact cybersecurity and what documentation is required and recommended in Section VII.D “Modifications.”
Premarket Considerations
The FDA published Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions to provide the FDA’s recommendations to industry regarding cybersecurity device design, labeling, and the documentation that the FDA recommends be included in premarket submissions for devices with cybersecurity risk.
The guidance discusses submission documentation recommendations referenced in the following sections, among others:
- Cybersecurity management plan (Section VI.B)
- Cybersecurity risk management report (Section V.A)
- Threat modeling (Section V.A.1)
- Cybersecurity risk assessment (Section V.A.2)
- Security assessment of unresolved software anomalies (Section V.A.5)
- Cybersecurity measures and metrics, if available (Section V.A.6)
- Cybersecurity controls (Section V.B.1, Appendix 1)
- Security architecture views (Section V.B.2, Appendix 2)
- Cybersecurity testing (Section V.C)
- Cybersecurity labeling (Section VI.A)
- Software Bill of Materials including level of support, end of support and vulnerability assessment of software components (Section V.A.4)
- Cyber Device updates reflecting 524B (Section VII)
Q1: Which federal agency is responsible for the regulation of cybersecurity in medical devices?
A: The FDA is responsible for the regulation of cybersecurity in medical devices. However, the FDA works closely with several other federal government agencies to increase security of the U.S. critical cyber infrastructure, including the U.S. Department of Homeland Security (DHS) and Cybersecurity & Infrastructure Security Agency (CISA).
Q2: Is cybersecurity for medical devices optional?
A: Cybersecurity protection by design is not optional and should be included as part of the device Section 524B of the FD&C Act requires manufacturers of cyber devices to meet specific requirements to demonstrate that a device is safe, and effective. Manufacturers are responsible for ensuring the safety and effectiveness of devices and cybersecurity is an integral part of medical device safety and effectiveness. Manufacturers may wish to refer to the FDA recommendations in Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions.
Q3: If my device does not fall within the scope of a cyber device, do I need to follow the premarket guidance?
A: The premarket guidance recommendations are applicable to all medical devices that have or are software.
Q4: If a device isn’t designed to connect to the internet, is it considered a cyber device?
A: Devices that can, whether intentionally or unintentionally, through any means connect to the internet are one of the elements to be considered a cyber device. The FDA considers devices that include any of the following features to have the ability to connect to the internet. This list is illustrative, not exhaustive: wi-fi or cellular; network, server, or cloud service provider connections, Bluetooth or Bluetooth low energy (BLE); other radio frequency (RF) communications; inductive communications; and hardware connectors capable of connecting to the internet, such as USB, ethernet connector (RJ45), or serial port. The FDA considers a “cyber device” to include devices that are or contain software, including software that is firmware or programmable logic, and have the ability to connect to the internet.
Q5: Is a Software Bill of Materials (SBOM) optional for medical device submissions for approval?
A: Section 524B(b)(3) requires manufacturers of cyber devices to provide a Software Bill of Materials (SBOM), including commercial, open-source, and off-the-shelf components. Additionally, the FDA recommends that premarket submissions include SBOM documentation, as mentioned in the Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions Section 4(a).
Q6: Can FDA-regulated medical devices participate in the U.S. Cyber Trustmark program?
A: The Center for Devices and Radiological Health is concerned that including medical devices in the IoT Labeling Program may cause consumer confusion and potentially create conflict where product manufacturers attempt to both qualify for the Cyber Trust Mark and comply with existing statutory and regulatory cybersecurity requirements under other federal laws, such as the Federal Food, Drug, and Cosmetic Act (FD&C Act).
Q7: Does the FDA test medical devices for cybersecurity?
A: The FDA does not conduct testing for medical devices. Cybersecurity testing, including security requirements testing, vulnerability assessment, and penetration testing, is the responsibility of the medical device manufacturer. The FDA reviews cybersecurity testing performed by the manufacturer as part of premarket submissions.
Q8: Are companies that manufacture off-the-shelf (OTS) software, which may be used in medical devices, responsible for validating their software’s secure use in medical devices?
A: The medical device manufacturer chooses to use OTS software, thus bearing responsibility for the security as well as the safe and effective performance of the medical device. Manufacturers may have cybersecurity agreements with the OTS software suppliers established during procurement of the software as part of their purchasing controls under the Quality Management System Regulations, 21 CFR Part 820 .
Q9: Does VEX satisfy FDA recommendations for vulnerability analysis?
A: No. The vulnerability assessment in the guidance is intended to assess the risk a vulnerability poses to the medical device system; both on its own and in the larger context of how it could be impacted by other risks or vulnerabilities present in the system (such as vulnerability chaining). The output of this assessment may inform the appropriate VEX information but VEX does not provide sufficient information on its own to replace the vulnerability assessment recommended in the guidance.
Postmarket Considerations
The FDA is responsible for providing recommendations to manufacturers to help safeguard patient safety from cybersecurity risks after authorization. The Postmarket Management of Cybersecurity in Medical Devices guidance provides recommendations, emphasizing that manufacturers and other stakeholders should monitor, identify, and address cybersecurity risks as part of their postmarket management of a medical device.
This guidance provides recommendations for remediating and reporting cybersecurity vulnerabilities for controlled and uncontrolled risks of patient harm. Manufacturers are recommended to have a structured and systematic risk management approach that includes:
- methods to identify, characterize and assess a cybersecurity vulnerability
- methods to analyze, detect and assess threat sources
- methods to remediate uncontrolled risk of patient harm posed by a device cybersecurity vulnerability to an acceptable level
In addition to the postmarket guidance, manufacturers are required to make available updates and patches to address certain types of vulnerabilities defined in Section 524B for any cyber devices submitted for approval after March 29, 2023.
Q1: Should medical device manufacturers update medical devices for cybersecurity?
A: Yes, medical device manufacturers are required to create and maintain procedures to ensure that their device and system remain cybersecure.
Q2: Can health care delivery organizations update and patch medical devices for cybersecurity?
A: The FDA recognizes that health care delivery organizations may need to patch or update devices and/or supporting infrastructure to reduce cybersecurity risks. Recognizing that changes may require a risk assessment, the FDA recommends working closely with medical device manufacturers to communicate changes that are requested or necessary to support cybersecurity within their environment.
Q3: Does a cybersecurity patch constitute a change to the device design which must be approved by the FDA?
A: Changes that are made solely to strengthen cybersecurity and do not have any other impact on the device likely will not require a premarket submission. Manufacturers should perform a risk-based assessment to determine whether a change could significantly affect safety or effectiveness of the device and whether a submission is needed.
Q4: Do low severity vulnerabilities in my product need to be addressed?
A: A device should be designed to eliminate or mitigate known vulnerabilities. For marketed devices, if comprehensive design mitigations are not possible, compensating controls should be considered. Manufacturers are expected to assess and address uncontrolled vulnerabilities, regardless of their severity.
Q5: Are manufacturers responsible for supporting the cybersecurity within medical devices they have deemed end-of-life or end-of-support?
A: Manufacturers must maintain a process to intake and evaluate applicable complaints as well as report pursuant to FDA’s requirements for Medical Device Reports (MDRs), 21 CFR Part 803. Manufacturers must maintain a process to evaluate risks, including assessing for cybersecurity vulnerabilities, and take appropriate actions for products affected by a cybersecurity issue that may present a risk to health.
Q6: Is the FDA responsible for the validation of changes made to address cybersecurity vulnerabilities?
A: The medical device manufacturer is responsible for the validation of all software design changes, including changes to address cybersecurity vulnerabilities.
Additional Resources
- Data Normalization Challenges and Mitigations in Software Bill of Materials (SBOM) Processing White Paper The FDA contracted with MITRE to develop this publication. The publication discusses normalization (data standardization and formatting) challenges and mitigations for generating and digesting (obtaining, importing, and analyzing) SBOMs, and advantages of using a standard nomenclature and formats to ensure that data from various sources is consistent.
- Playbook for Threat Modeling Medical Devices | MITRE: The FDA contracted with MITRE to develop this playbook. This playbook is an educational resource that discusses best practices for understanding basic threat modeling concepts and processes, and how to apply them to medical devices.
Related Content
- Collaborations on Medical Device Cybersecurity
- Reports, White Papers, and Other Resources
- Safety Communications and Alerts Related to Cybersecurity
Contact the FDA’s Medical Device Cybersecurity Division
If you would like to partner with the FDA’s Medical Device Cybersecurity Division, please email us at CyberMed@fda.hhs.gov.