SARS-CoV-2 Viral Mutations: Impact on COVID-19 Tests

The SARS-CoV-2 virus has mutated over time, resulting in genetic variation in the population of circulating viral strains over the course of the COVID-19 pandemic. Molecular, antigen, and serology tests are affected by viral mutations differently due to the inherent design differences of each test.

This page provides information regarding the impact of viral mutations on COVID-19 tests, recommendations for clinical laboratory staff and health care providers, and information about certain tests for which the FDA has identified potential impacts on performance due to SARS-CoV-2 genetic mutations. The FDA will list tests on this page as the FDA's analyses identify tests for which performance may be impacted for known SARS-CoV-2 variants.

The FDA will update this page as significant new information becomes available.

On this page:


Genetic Variations: Background and Considerations

A mutation (also referred to as viral mutation or genetic mutation) of the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) virus is a change in the genetic sequence of the SARS-CoV-2 virus when compared with a reference sequence such as Wuhan-Hu1 (the first genetic sequence identified) or USA-WA1/2020 (the first identified in the United States). A new variant (virus variant or genetic variant) of SARS-CoV-2 may have one or more mutations that differentiate it from the reference sequences or predominant virus variants already circulating in the population. Variants of SARS-CoV-2 can have different characteristics. For example, some may spread more easily or show signs of resistance to existing treatment options and some may have no impact when compared with previous and currently circulating virus.

The presence of mutations in the SARS-CoV-2 virus in a patient sample can potentially impact test performance. The impact of mutations on a test's performance is influenced by several factors, including the sequence of the variant, the design of the test, and the prevalence of the variant in the population.

The FDA has collaborated with stakeholders to better understand the public health impact of new SARS-CoV-2 variants and their impact on test performance, has been routinely monitoring publicly available databases, and has coordinated efforts to evaluate the impact of new virus variants on tests that have received Emergency Use Authorization (EUA).

In February 2021, the FDA issued the Policy for Evaluating Impact of Viral Mutations on COVID-19 Tests to provide a policy and recommendations on evaluating the potential impact of emerging and future viral mutations of SARS-CoV-2 on COVID-19 tests for the duration of the COVID-19 public health emergency, including considerations for test designs to minimize the impact of viral mutations and recommendations for ongoing monitoring.

On September 23, 2021, the FDA revised the EUAs of certain authorized molecular, antigen, and serological tests to establish additional Conditions of Authorization in response to the continued emergence of new variants of SARS-CoV-2. These conditions have also been included in other EUAs, such that all EUAs previously issued for COVID-19 molecular, antigen, and serology tests include these conditions, and they continue to be included in new EUAs. These additional conditions require test developers to update their authorized labeling and evaluate the impact of SARS-CoV-2 viral mutations on their test's performance, among other things.

The FDA will update this page as significant new information becomes available, including the information provided by EUA holders in their authorized labeling.


General Information for Clinical Laboratory Staff and Health Care Providers

Clinical laboratory staff and health care providers should be aware that false negative results may occur with any molecular test for the detection of SARS-CoV-2, particularly if a mutation occurs in the part of the virus' genome assessed by that test.

Changes in the viral genome can result in changes to viral proteins and, therefore, can also impact the performance of an antigen or serology test.

The FDA recommends clinical laboratory staff and health care providers who use SARS-CoV-2 tests:

  • Be aware that genetic variants of SARS-CoV-2 arise regularly, and false negative test results can occur.
  • Be aware that molecular tests that use multiple genetic targets to determine a final result are less likely to be impacted by increased prevalence of genetic variants.
  • Consider negative results in combination with clinical observations, patient history, and epidemiological information.
  • Be aware of currently circulating SARS-CoV-2 variants since test performance may be impacted by certain variants.
  • Consider repeat testing with a different EUA authorized or FDA cleared molecular diagnostic test (with different genetic targets) if COVID-19 is still suspected after receiving a negative test result.

In addition to these general recommendations, the FDA is providing recommendations for the use of specific tests impacted by genetic variation in the sections below.


Omicron Variant: Molecular Tests That May Be Impacted

The FDA's analysis to date has identified certain EUA-authorized molecular tests whose performance may be impacted by mutations in the SARS-CoV-2 omicron variant.

Tests with Reduced Ability to Detect the SARS-CoV-2 Omicron Variant

Tide Laboratories DTPM COVID-19 RT-PCR Test

  • Test Name (Link to EUA): DTPM COVID-19 RT-PCR test
  • Manufacturer: Tide Laboratories, LLC
  • The FDA's Analysis: This test is expected to fail to detect the SARS-CoV-2 omicron variant (B.1.1.529) due to a nine-nucleotide deletion in the N-gene, spanning positions 28370-28362. The single genetic target of this test covers the portion of the N-gene where the deletions occur.
  • Potential Impact: Since this is a single target test, the test is expected to fail to detect the SARS-CoV-2 omicron variant, resulting in false negative results in patients with the omicron variant. This nine-nucleotide deletion appears to be specific to the omicron variant; therefore, based on initial bioinformatics analysis, the performance of the test is not expected to be impacted for other known SARS-CoV-2 variants.
  • Notes: The FDA continues to gather additional information and work with the manufacturer to address this issue.
     
    Recommendations for Clinical Laboratory Staff and Health Care Providers Using This Test 
     
    Be aware that this test is being performed in the following laboratories, though it is not believed to be used for high volume testing:
     
    Laboratory Name, City, State
    • Centennial State Lab - Golden Lab, Golden, CO
    • Centennial State Lab, Colorado Springs, CO
    • Community Health & Wellness, Atlanta, GA
    • Covenant, Bay St. Louis, MS
    • Crescent City Surgical Center, Metaire, LA
    • Dr. Douglas Mehaffie MD, LLC, Laplace, LA
    • Elite Diagnostics, Crown Point, IN
    • Flexsus, LLC, Columbus, OH
    • Harvest Labs, Inc, Crowley, LA
    • Health First Urgent Care, Richland, WA
    • Lab Works, LLC, Homewood, AL
    • Lab24, Redding, CA
    • Laboratory Specialists of Michigan, Dearborn, MI
    • Lakeside North Family Medicine, LLC, Guntersville, AL
    • Matthews Vu Medical Group, Colorado Springs, CA
    • Med Lake Laboratory LLC, Milledgeville, GA
    • Medlab, Inc, Buffalo, NY
    • Nibbana LLC, Tigard, OR
    • Nona Scientific, Tampa, FL
    • PCR-DX-AL, Tuscaloosa, AL
    • PCR-DX-CO, Fort Collins, CO
    • PCR-DX-GA, Gainesville, GA
    • Pinnacle Genetics Labs Inc, Miami, FL
    • PremierTox 2.0 Inc, Russell Springs, KY
    • QualiTox Lab, McKees Rocks, PA
    • Southern Immediate Care, Anniston, AL
    • St. Jude Laboratories, Columbia, MD
    • Stone Clinical Labs, New Orleans, LA
    • Stone Clinical Labs of FL, Freeport, FL
    • TF Medical LLC, Tuscaloosa, AL
    • Tide Laboratories, LLC, Fort Payne, AL
    • True Labs LLC, Coral Springs, FL
    • United Clinical Laboratory, West Palm Beach, FL

Tests with S-Gene Drop Out - SARS-CoV-2 Detection Should Not Be Impacted

  • The FDA's Analysis: For each of the tests listed in the below table, one genetic target is expected to have significantly reduced sensitivity due to a mutation in the SARS-CoV-2 omicron variant (B.1.1.529). A specific deletion in the spike (S) gene (Δ69-70) results in an S-gene drop out, also referred to as an S-gene target failure (SGTF), where the genetic target that covers the portion of the S-gene where the deletion occurs fails to detect the virus.
  • Potential Impact: Since these tests are designed to detect multiple genetic targets, the overall test sensitivity should not be impacted. The pattern of SARS-CoV-2 detection with an S-gene drop out may provide a signal that the omicron variant may be present. However, the presence of the S-gene drop out pattern does not mean an individual with SARS-CoV-2 has the omicron variant.
  • Notes: These tests are identified based on initial bioinformatics analyses and may not yet have been evaluated against the omicron variant in the laboratory.

    Recommendations for Clinical Laboratory Staff and Health Care Providers Using These Tests
    • Be aware that the S-gene drop out pattern of detection (reduced sensitivity with the S-gene target) when using these tests is consistent with certain mutations, including those in the omicron variant.
    • S-gene drop out has been observed in other variants and is not specific to the omicron variant.  S-gene drop out has also been observed without the presence of a mutation in the S-gene, due to the sensitivity of the S-gene target.  Therefore, these tests cannot be used as a diagnostic tool or specific identifier of the omicron variant. 
    • The presence of the S-gene drop out detection pattern can signal that sequencing should be considered to characterize the variant in that specimen.
    • S-gene drop out is typically not observed in the delta variant. Since delta is the primary variant circulating in the United States, specimens with an S-gene drop out detection pattern may potentially be omicron variants and should be prioritized for sequencing confirmation.
    • If local or state clinical laboratories have access to quick turnaround whole genome sequencing services, these labs should consider further characterizing the specimen with genetic sequencing when this pattern is identified. If such services are not readily available, local or state clinical laboratories should consider reaching out to the Centers for Disease Control and Prevention at EOCevent177@cdc.gov for additional information. 
Test Developer Test Name (Link to Most Recent Letter of Authorization)
Acupath Laboratories, Inc. Acupath COVID-19 Real-Time (RT-PCR) Assay
Aeon Global Health Aeon Global Health SARS-CoV-2 Assay
AIT Laboratories SARS-CoV-2 Assay
Alpha Genomix Laboratories Alpha Genomix TaqPath SARS-CoV-2 Combo Assay
Altru Diagnostics, Inc. Altru Dx SARS-CoV-2 Assay
Bandar Enterprises, LLC dba Cuur Diagnostics Cuur Diagnostics SARS-CoV-2 Molecular Assay
Biocerna LLC SARS-CoV-2 Assay
Boston Heart Diagnostics Boston Heart COVID-19 RT-PCR Test
Cormeum Laboratory Services Cormeum SARS-CoV-2 Assay
DxTerity Diagnostics, Inc. DxTerity SARS-CoV-2 RT-PCR Test
Express Gene LLC, DBA: Express Gene Molecular Diagnostics Laboratory Express Gene 2019-nCoV RT-PCR Diagnostic Panel
Gravity Diagnostics, LLC Gravity Diagnostics SARS-CoV-2 RT-PCR Assay
Gravity Diagnostics, LLC Gravity Diagnostics SARS-CoV-2 RT-PCR for use with DTC kits
HealthQuest Esoterics HealthQuest Esoterics TaqPath SARS-CoV-2 Assay
Helix OpCo LLC Helix COVID-19 Test
Thermo Fisher Scientific Inc. TaqPath COVID-19, FluA, FluB Combo Kit *
Thermo Fisher Scientific Inc. TaqPath COVID-19 Combo Kit  
Northeastern University College of Science, Life Sciences Testing Center Life Sciences Testing Center COVID-19 Test
P23 Labs, LLC. P23 Labs TaqPath SARS-CoV-2 Assay
Patients Choice Laboratories, LLC PCL SARS-CoV-2 Real-Time RT-PCR Assay
Infinity BiologiX LLC Infinity BiologiX TaqPath SARS-CoV-2 Assay
Southern California Permanente Medical Group Kaiser Permanente High Throughput SARS-CoV-2 Assay
Tempus Labs, Inc. iC SARS-CoV-2 Test
Cleveland Clinic Robert J. Tomsich Pathology and Laboratory Medicine Institute SelfCheck COVID-19 TaqPath Multiplex PCR
Thermo Fisher Scientific Inc. Amplitude Solution with the TaqPath COVID-19 High-Throughput Combo Kit
Verily Life Sciences Verily COVID-19 RT-PCR Test
Applied DNA Sciences, Inc. Linea COVID-19 Assay Kit

* The TaqPath COVID-19, FluA, FluB Combo Kit is expected to have an S-gene drop out due to the presence of a mutation in the S-gene, without a significant impact to overall test sensitivity. However, due to the design of the test, there will not be a distinct S-gene detection pattern to signal the potential presence of the omicron variant.


Other Variants: Molecular Tests That May Be Impacted

The FDA's analysis to date has identified the following EUA-authorized molecular tests whose performance may be impacted by other SARS-CoV-2 viral mutations:

Accula SARS-CoV-2 Test

  • Test Name (Link to EUA): Accula SARS-CoV-2 Test
  • Manufacturer: Mesa Biotech Inc.
  • The FDA's Analysis: Performance may be impacted when a SARS-CoV-2 patient sample having certain viral mutations is tested. The FDA previously alerted health care providers about a potential performance impact due to a genetic mutation at positions 28881-28883 (GGG to AAC). In addition, there is potential impact on performance of the test due to a genetic mutation at positions 28877-28878 (AG to TC) in patient samples.
  • Potential Impact: While the impact does not appear to be significant, the FDA is providing this information out of an abundance of caution.
  • Notes: The FDA's analysis included information provided by the manufacturer.

    Recommendations for Clinical Laboratory Staff and Health Care Providers Using This Test

    Be aware that the current instructions for use for the Accula SARS-CoV-2 Test include the exact variant location, in silico analysis of the primer binding, and observed performance impact due to genetic mutation at positions 28881-28883 (GGG to AAC) when compared with the perfect match target, but do not yet include specific information about impact from a genetic mutation at positions 28877-28878 (AG to TC).

Linea COVID-19 Assay Kit

  • Test Name (Link to EUA): Linea COVID-19 Assay Kit
  • Manufacturer: Applied DNA Sciences, Inc.
  • The FDA's Analysis: One of the two targets of the test has significantly reduced sensitivity due to certain mutations, including one of the mutations in the B.1.1.7 variant (UK VOC-202012/01).
  • Potential Impact: Since this test is designed to detect multiple genetic targets, the overall test sensitivity should not be impacted. The pattern of detection when certain mutations are present may help with early identification of new variants in patients to reduce further spread of infection.
  • Notes: The FDA's analysis included information provided by the manufacturer.

    Recommendations for Clinical Laboratory Staff and Health Care Providers Using This Test

    Be aware that one positive target and one negative target showing the S-gene drop out (reduced sensitivity with the S-gene target) when using the Linea COVID-19 Assay Kit is consistent with certain mutations, including those in the B.1.1.7 variant. If local or state clinical laboratories have access to quick turnaround whole genome sequencing services these labs should consider further characterizing the specimen with genetic sequencing when this pattern is identified. If such services are not readily available, local or state clinical laboratories should consider reaching out to the Centers for Disease Control and Prevention at EOCevent177@cdc.gov for additional information. 

TaqPath COVID-19 Combo Kit

  • Test Name (Link to EUA): TaqPath COVID-19 Combo Kit (may also be labeled as the TaqPath COVID-19 Combo Kit Advanced)
  • Manufacturer: Thermo Fisher Scientific, Inc.
  • The FDA's Analysis: One of three targets of the test has significantly reduced sensitivity due to certain mutations, including one of the mutations in the B.1.1.7 variant (UK VOC-202012/01).
  • Potential Impact: Since this test is designed to detect multiple genetic targets, the overall test sensitivity should not be impacted. The pattern of detection when certain mutations are present may help with early identification of new variants in patients to reduce further spread of infection.
  • Notes: The FDA's analysis included information provided by the manufacturer and multiple reports from clinical laboratories.

    Recommendations for Clinical Laboratory Staff and Health Care Providers Using This Test

    Be aware that two positive targets and one negative target showing the S-gene drop out (reduced sensitivity with the S-gene target) when using the TaqPath COVID-19 Combo Kit is consistent with certain mutations, including those in the B.1.1.7 variant. If local or state clinical laboratories have access to quick turnaround whole genome sequencing services these labs should consider further characterizing the specimen with genetic sequencing when this pattern is identified. If such services are not readily available, local or state clinical laboratories should consider reaching out to the Centers for Disease Control and Prevention at EOCevent177@cdc.gov for additional information. 

Xpert Xpress SARS-CoV-2, Xpert Xpress SARS-CoV-2 DoD, Xpert Omni SARS-CoV-2

  • Test Name (Link to EUA): Xpert Xpress SARS-CoV-2; Xpert Xpress SARS-CoV-2 DoD; Xpert Omni SARS-CoV-2
  • Manufacturer: Cepheid
  • The FDA's Analysis: While it is generally unexpected that a single point mutation will impact test performance for most SARS-CoV-2 molecular tests, the FDA's analysis indicates that the Cepheid tests are impacted by a single point mutation in the target area of the test. There are reports noting that two independent single point mutations reduce the test's sensitivity for detecting the N2 target. The E target is still detected when enough virus is present leading to a "presumptive positive" result in the Xpert Xpress SARS-CoV-2 and Xpert Xpress SARS-CoV-2 DoD tests. Detection of the E target without detecting the N2 target will be reported as "positive" in the Xpert Omni SARS-CoV-2
  • Potential Impact: Since this test is designed to detect multiple genetic targets, and these mutations do not lead to a false negative patient result (instead reporting as "presumptive positive" or "positive" based on detection of the conserved E gene target), the impact on test performance does not appear to be significant. However, the FDA is providing this information out of an abundance of caution. The FDA's analysis suggests that the impact of a single point mutation on the test performance is associated with the unique chemistry of the Cepheid tests.
  • Notes: The FDA continues to gather additional information and work with the manufacturer to address this issue.

    Recommendations for Clinical Laboratory Staff and Health Care Providers Using This Test

    Be aware of the current instructions for use for each test, especially the "Possible Results" and "Results and Interpretation" tables, which describe when a result is positive, presumptive positive, negative, and invalid, and how to interpret each result. Specifically, "SARS-CoV-2 Presumptive Positive" results may indicate the presence of SARS-CoV-2 nucleic acids, and repeat testing may be indicated; refer to the authorized instructions for use. Clinical laboratory staff and health care providers should contact Cepheid if they have any questions or concerns or suspect an issue with their Cepheid test. 