WARNING LETTER
Shanghai Hua Yue Packaging Products Co., Ltd.
- Recipient:
- Shanghai Hua Yue Packaging Products Co., Ltd.
United States
- Issuing Office:
- Center for Devices and Radiological Health
United States
| Public Health Service Food and Drug Administration |
10903 New Hampshire Avenue White Oak Building 66 Silver Spring, MD 20993 |
WARNING LETTER
DEC 24, 2014
VIA UNITED PARCEL SERVICE
Wang Wen Hua
President
Shanghai Hua Yue Packaging Products Co., Ltd.
Room F, 27th Floor, No.58
Xin Jin Qiao Rd.
201206 Shanghai
China
Dear Mr. Hua:
During an inspection of your firm located in Xinsi, Zhelin Town, Feng Xian District, Shanghai, China, on August 25, 2014,through August 28, 2014, an investigator from the United States Food and Drug Administration (FDA) determined that your firm manufactures surgical equipment covers, and surgical sponge counters. Under section 201(h) of the Federal Food, Drug, and Cosmetic Act (the Act), 21 U.S.C. § 321(h), these products are devices because they are intended for use in the diagnosis of disease or other conditions or in the cure, mitigation, treatment, or prevention of disease, or to affect the structure or function of the body.
This inspection revealed that these devices are adulterated within the meaning of section 501(h) of the Act, 21 U.S.C. § 351(h), in that the methods used in, or the facilities or controls used for, their manufacture, packing, storage, or installation are not in conformity with the current good manufacturing practice requirements of the Quality System regulation found at Title 21, Code of Federal Regulations (CFR), Part 820. These violations include, but are not limited to, the following:
1. Failure to ensure ,where the results of a process cannot be fully verified by subsequent inspection and test, that the process shall be validated with a high degree of assurance and approved according to established procedure, as required by 21 CFR 820.75(a). For example, your firm failed to validate its (b)(4) process, which is used to manufacture surgical sponge counters. Specifically, your firm failed to establish validation protocols and execute a validation of the (b)(4) process for (b)(4).
2. Failure to develop, conduct, control, and monitor production processes to ensure that a device conforms to its specifications, as required by 21 CFR 820.70(a). For example, when temperatures exceeded established limits in your firm’s (b)(4), your firm failed to identify and address these failures. Specifically, the temperature parameter for (b)(4) of the (b)(4) process (b)(4). Your firm’s (b)(4) working data record, dated March 14, 2014, shows temperatures beyond specifications in four out of eight days reported.
Our inspection also revealed that your devices are misbranded under section 502(t)(2) of the Act, 21 U.S.C. § 352(t)(2), in that your firm failed or refused to furnish material or information respecting the device that is required by or under section 519 of the Act, 21 U.S.C. § 360i, and 21 CFR Part 803 - Medical Device Reporting. Significant violations include, but are not limited to, the following:
3. Failure to develop, maintain, and implement written Medical Device Reporting (MDR) procedures, as required by 21 CFR 803.17. For example, your firm has not created and implemented MDR procedures.
The eMDR Final Rule requiring manufacturers and importers to submit electronic Medical Device Reports (eMDRs) to FDA was published on February 13, 2014. The requirements of this final rule will take effect on August 14, 2015. If your firm is not currently submitting reports electronically, we encourage you to visit the following web link for additional information about the electronic reporting requirements:
If your firm wishes to discuss MDR reportability criteria or to schedule further communications, it may contact the Reportability Review Team by email at ReportabilityReviewTeam@fda.hhs.gov.
U.S. federal agencies may be advised of the issuance of Warning Letters about devices so that they may take this information into account when considering the award of contracts. Additionally, premarket approval applications for Class III devices to which the Quality System regulation deviations are reasonably related will not be approved until the violations have been corrected.
Please notify this office in writing, within fifteen business days from the date you receive this letter, of the specific steps your firm has taken to correct the noted violations, including an explanation of how your firm plans to prevent these violations, or similar violations, from occurring again. Include documentation of the corrections and/or corrective action (which must address systemic problems) that your firm has taken. If your firm’s planned corrections and/or corrective actions will occur over time, please include a timetable for implementation of those activities. If corrections and/or corrective actions cannot be completed within fifteen business days, state the reason for the delay and the time within which these activities will be completed. Please provide a translation of documentation not in English to facilitate our review. We will notify you regarding the adequacy of your firm’s response and the need to re-inspect your firm’s facility to verify that the appropriate corrections and/or corrective actions have been made.
Your firm’s response should be sent to: Food and Drug Administration, Center for Devices and Radiological Health, Office of Compliance, Field Inspections Support Branch, White Oak Building 66, Rm 2622, 10903 New Hampshire Ave., Silver Spring, MD 20993. Refer to CMS case #445161 when replying. If you have any questions about the contents of this letter, please contact: Daniel Walter, Chief, Foreign Enforcement Branch, at +1 (301) 796-5587 (telephone) or +1 (301) 847-8139 (fax).
Finally, you should know that this letter is not intended to be an all-inclusive list of the violations at your firm’s facility. It is your firm’s responsibility to ensure compliance with applicable laws and regulations administered by FDA. The specific violations noted in this letter and in the Inspectional Observations, FDA 483, issued at the close of the inspection may be symptomatic of serious problems in your firm’s manufacturing and quality management systems. Your firm should investigate and determine the causes of the violations, and take prompt actions to correct the violations and bring the products into compliance.
Sincerely yours,
/S/
Steven D. Silverman
Director
Office of Compliance
Center for Devices and
Radiological Health