Absorb GT1 Bioresorbable Vascular Scaffold (BVS) by Abbott Vascular: Letter to Health Care Providers - FDA Investigating Increased Rate of Major Adverse Cardiac Events

UPDATED October 31, 2017. The FDA issued an update to the March 18, 2017 letter to health care providers to inform the health care community that interim study results through three years from the pivotal clinical trial (ABSORB III) continue to show an increased rate of major adverse cardiac events and BVS scaffold thrombosis in patients receiving the Absorb GT1 Bioresorbable Vascular Scaffold (BVS), when compared to patients treated with the approved metallic XIENCE drug-eluting stent.

The FDA was made aware that the manufacturer has stopped global sales of the Absorb GT1 Bioresorbable Vascular Scaffold as of September 14, 2017. The FDA’s recommendations for health care providers outlined in the previous letter remain unchanged. Although health care providers with available inventory may continue to implant the BVS, they should carefully consider its safety and effectiveness and only use it if they believe it is in the best interest of their patients.  See the updated letter to health care providers for more information. 

[Posted 03/18/2017]

AUDIENCE: Cardiology, Surgery, Risk Manager

ISSUE: The FDA is informing health care providers treating patients with Absorb GT1 Bioresorbable Vascular Scaffold (BVS) that there is an increased rate of major adverse cardiac events observed in patients receiving the BVS, when compared to patients treated with the approved metallic XIENCE drug-eluting stent.

The FDA’s initial review of two-year data from the BVS pivotal clinical study (the ABSORB III trial) shows an 11 percent rate of major adverse cardiac events (e.g., cardiac death, heart attack, or the need for an additional procedure to re-open the treated heart vessel) in patients treated with the BVS at two years, compared with 7.9 percent in patients treated with the already-approved Abbott Vascular’s metallic XIENCE drug-eluting stent (p = 0.03). This study also shows a 1.9 percent rate of developing blood clots (thrombosis) within the BVS versus 0.8 percent within the XIENCE stent at 2 years. These observed higher adverse cardiac event rates in BVS patients were more likely when the device was placed in small heart vessels.

The FDA is working with Abbott Vascular, Inc. to conduct additional analyses to better understand the cause(s) of the higher cardiac event and device thrombosis rates in patients treated with BVS compared to the XIENCE stent. The FDA will continue to monitor the performance of the BVS in ongoing clinical studies and in reports submitted to FDA through MedWatch. We will update this communication when additional information or analyses become available.

BACKGROUND: The BVS is used to open heart blood vessels (coronary arteries) blocked by scar tissue (plaque) in order to increase blood flow to the heart muscle. The BVS is implanted during an angioplasty procedure. It gradually dissolves and is fully absorbed by the body over time.

RECOMMENDATION: The FDA recommends that health care providers:

  • Follow the instructions for target heart vessel selection (e.g., avoiding BVS use in small heart vessels) and optimal device implantation that are included in the BVS physician labeling.
  • Advise patients experiencing any new cardiac symptoms such as irregular heartbeats, chest pain, or shortness of breath to seek clinical care. For more information about risks associated with the BVS, refer to the BVS physician labeling.
  • Advise BVS patients to follow the recommendations for DAPT prescribed by their health care providers.
  • Report any adverse events related to the BVS that come to your attention. If you suspect a problem with the BVS, we encourage you to file a voluntary report through MedWatch, the FDA Safety Information and Adverse Event Reporting Program. Health care personnel employed by facilities that are subject to the FDA's user facility reporting requirements should follow the reporting procedures established by their facilities.
    • Complete and submit the report Online: www.fda.gov/MedWatch/report
    • Download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178

[10/31/2017 - UPDATED Letter to Health Care Providers - FDA]