Medical Devices

CareFusion Recalls AVEA Ventilator Due to an Electrical Issue Which May Cause an Unexpected Shutdown

The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death.

Recalled Product:

Product DescriptionProduct Numbers
AVEA Standard with Compressor ventilator-refurbishedR17312-xx (xx = 0-14)
AVEA Comprehensive ventilator17310-xx (xx = 0 -14)
AVEA Standard ventilator17311-xx (xx = 0 -14)
AVEA Standard with Compressor ventilator17312-xx (x=0 -14)
AVEA Comprehensive ventilator-refurbishedR17310-xx (xx = 0-14)
AVEA Standard ventilator-refurbishedR17311-xx (xx = 0-14)
AVEA Standard with Compressor ventilator-refurbishedR17312-xx (xx= 0-14)
TCA Board16542A
GDE- 1st Generation16222-001-99
GDE16650A
GDE-refurbishedR16650A
AVEA GDE/UIM upgrade kit12283-PMN
Image of recalled CareFusion AVEA ventilator

Image of recalled CareFusion AVEA ventilator

  • Manufacturing Dates: November 13, 2015 to January 4, 2016
  • Distribution Dates: December 16, 2015 to February 15, 2016
  • Devices Recalled in the U.S. 501 units distributed nationwide

Device Use

The AVEA ventilator is intended for continuous breathing support for infants, children and adults. The ventilator is only used in hospitals and other health care facilities.

Reason for Recall

CareFusion is recalling the AVEA Ventilator because of a faulty fuse on the ventilators’ alarm board, which may cause the ventilator to unexpectedly shut down. If the ventilator shuts down, a patient may not receive necessary oxygen. The use of affected product may cause serious adverse health consequences, including death.

Who May be Affected

  • Health care providers using AVEA ventilators
  • All patient groups who may be using these ventilator systems for breathing support

What to Do

On May 17, 2016, CareFusion sent a "Field Safety Notice"disclaimer icon letter to affected health care facilities. The letter instructed health care professionals and consumers to:

  • Complete and return the acknowledgement and receipt form
  • After the response card is returned, CareFusion will schedule onsite remediation for affected devices
  • Report adverse events or quality problems experienced with use of the product to the FDA through:

Contact Information

Health care professionals and consumers with questions are instructed to contact CareFusion at 888-526-6018 or supportcenter@carefusion.com with any questions related to this recall.

Date Recall Initiated:

May 17, 2016

Additional Resources:

CareFusion Press Releasedisclaimer icon

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program either online, by regular mail or by FAX to 1-800-FDA-0178.

Page Last Updated: 10/18/2017
Note: If you need help accessing information in different file formats, see Instructions for Downloading Viewers and Players.
Language Assistance Available: Español | 繁體中文 | Tiếng Việt | 한국어 | Tagalog | Русский | العربية | Kreyòl Ayisyen | Français | Polski | Português | Italiano | Deutsch | 日本語 | فارسی | English