U.S. flag An official website of the United States government
  1. Home
  2. Safety
  3. Recalls, Market Withdrawals, & Safety Alerts
  4. AmEx Pharmacy Issues Voluntary Nationwide Recall for one Lot of Bevacizumab 1.25mg/0.05mL 31 G Syringe Due to Reported Defective Delivery System
  1. Recalls, Market Withdrawals, & Safety Alerts

COMPANY ANNOUNCEMENT

AmEx Pharmacy Issues Voluntary Nationwide Recall for one Lot of Bevacizumab 1.25mg/0.05mL 31 G Syringe Due to Reported Defective Delivery System

When a company announces a recall, market withdrawal, or safety alert, the FDA posts the company's announcement as a public service. FDA does not endorse either the product or the company.

Read Announcement

Summary

Company Announcement Date:
FDA Publish Date:
Product Type:
Drugs
Generic Drugs
Reason for Announcement:
Recall Reason Description
Product Defect
Company Name:
AmEx Pharmacy
Brand Name:
Brand Name(s)
AmEx Pharmacy
Product Description:
Product Description
Bevacizumab 1.25mg/0.05mL 31G Injectable

Company Announcement

AmEx Pharmacy is voluntarily recalling one Lot of Bevacizumab 1.25mg/0.05mL 31G Injectable to the consumer/user level. The Monoject Syringe of this product may become difficult to express, and when additional force is applied, while the needle is in the eye, may cause injury to the patient.

The additional force needed to express the drug product could potentially result in damage to the eye while the needle is in the eye. To date, AmEx Pharmacy has received three reports associated with the Lot being recalled as either being difficult to express, two of which, resulted in an Adverse Drug Event.

The product is used for Wet Age-related Macular Degeneration and Diabetic Retinopathy. It is individually wrapped and labeled in a Tyvek pouch which is then placed in a labeled amber bag to protect from light. The affected Lot of Bevacizumab 1.25mg/0.05m 3 lG Injectable is 190212AB, BUD 5/13/2019. The product can be identified by referencing the Lot number 190212AB, which prominently appears on all labeling. This specific Lot was distributed nationwide to ophthalmologist clinics in the following states: PA, IL, TX, WI, KS, TN, IN, & AZ.

AmEx Pharmacy is notifying its consignees by telephone and overnight mail and is arranging for return/replacement of all recalled product. Administering physicians that have product which is being recalled should stop use, remove from inventory and return to AmEx Pharmacy.

Consumers with questions regarding this recall can contact AmEx Pharmacy at (800) 644 - 9431 or by email at pharmacist@amexpharmacy.com during normal business hours Monday through Friday 9:00a.m. - 6:00p.m, EST. Consumers should contact their physician or healthcare provider if they have experienced any problems that may be related to being administered this drug product.

Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.


Company Contact Information

Consumers:
800-644-9431