SUMMARY MINUTES
OF THE
108th MEETING OF
THE
OPHTHALMIC DEVICES
PANEL
OPEN SESSION
March 5, 2004
Hilton Washington
DC North/Gaithersburg
Gaithersburg, MD
Ophthalmic Devices Panel Meeting
March
5, 2004
Attendees
Chair
Jayne S. Weiss, M.D
Kresge Eye Institute
Wayne State University
Executive Secretary
Sara M. Thornton
Food and Drug Administration
Voting Members
Arthur Bradley, Ph.D.
Indiana University
Michael R. Grimmett, M.D.
Bascom Palmer Eye Institute
University of Miami School of Medicine
Allen C. Ho, M.D.
Wills Eye Hospital
Thomas Jefferson School of Medicine
William D. Mathers, M.D.
Casey Eye Institute
Oregon Health Sciences University
Timothy T. McMahon, O.D., F.A.A.O.
University of Illinois at Chicago School of Medicine
Consultants
Neil M. Bressler,
M.D.
Wilmer Eye Institute
The Johns Hopkins University School of Medicine
Jeremiah Brown, Jr., M.D.
Walter Reed Army Institute of Research Laboratory
Alexander J.
Brucker, M.D.
Scheie Eye Institute
University of Pennsylvania School of Medicine
Frederick L. Ferris III, M.D.
National Eye Institute
National Institutes of Health
Leo J. Maguire, M.D.
Mayo Medical School
Janine A. Smith, M.D.
National Eye
Institute
Walter J. Stark,
M.D.
Wilmer Eye Institute
The Johns Hopkins University School of Medicine
Industry Representative
Ronald E. McCarley
Ophtec USA, Inc.
Joseph Blustein,
M.D., M.P.H.
Don Calogero,
M.S.
Malvina B.
Eydelman, M.D.
Gene N.
Hilmantel, O.D., M.S.
A. Ralph
Rosenthal, M.D.
Karen Warburton,
M.S.
Panel Chair Jayne S. Weiss, M.D., called the meeting to order at 9:04 a.m. Panel Executive Secretary Sara M. Thornton noted that the meeting’s purpose is to discuss general issues relating to the use of intraocular lenses (IOLs) for the correction of presbyopia after clear lens extraction (CLE). She welcomed the participants and introduced the two new consultant members of the panel, Neil M. Bressler, M.D., and Jeremiah Brown, Jr., M.D. She asked panel members to introduce themselves and noted that the consumer representative, Glenda V. Such, M.Ed., would not be in attendance.
Ms. Thornton read the conflict of interest statement. Full waivers had been granted to Alexander J. Brucker, M.D. for his interest in firms in matters that could be affected by the panel’s recommendations. The Agency took into consideration other matters concerning Dr. Brucker, Michael R. Grimmett, M.D., Dr. Bressler, Allen C. Ho, M.D., Frederick L. Ferris III, M.D., and Dr. Weiss for their interests in firms at issue but in matters not related to the day’s agenda; they could participate fully in the panel’s deliberations.
BRANCH UPDATE
Karen F. Warburton, M.S., presented an update from the Vitreoretinal and Extraocular Devices Branch (VEDB) on recent medical device reports that have identified a connection between ophthalmic sponges and diffuse lamellar keratitis (DLK). Tests from a sample of ophthalmic sponges from one lot associated with a number of DLK cases have shown significantly higher levels of bacterial endotoxins when compared to another lot of sponges. Endotoxin-contaminated ophthalmic sponges have not been previously identified as a possible cause of DLK; however, reports in the literature have suggested that endotoxins from sterilizer water reservoirs may have caused DLK outbreaks.
The FDA does not require that ophthalmic sponges or other devices used in LASIK surgery be free from endotoxins or pyrogens, and they are not are typically labeled as such (although they may in fact be pyrogen- or endotoxin-free). The VEDB is working with other Center offices to apprise the ophthalmic community of these developments, specifically through letters to professional organizations and letters to editors of appropriate journals.
FDA TEAM PRESENTATION
Malvina B. Eydelman, M.D., began the FDA team presentation on CLE for the correction of presbyopia, an intraocular surgical procedure where a noncataractous lens is removed and replaced with a multifocal intraocular lens (MIOL) to allow for both near and distance vision. The sole purpose of this procedure is for refractive correction.
Dr. Eydelman underscored several points about CLE to panel members:
· CLE is not currently approved in the United States for any indication.
· CLE has been performed as an “off label” procedure for several years, mainly in eyes with high refractive errors.
· There are currently no standards or guidances available for CLE with IOL implantation.
Dr. Eydelman presented information on MIOLs, noting that only one MIOL is approved for use in the United States but several others are under investigation. Only two IOLs are approved for improving near visual acuity in patients with presbyopia. Because an estimated 1.5 billion people worldwide live with presbyopia, devices that are approved to correct presbyopia will have a major impact on the public’s health.
Designing a clinical trial to establish the efficacy and safety of devices to correct presbyopia offers numerous challenges, Dr. Eydelman said. The trial should be as easy and straightforward as practically possible for all parties involved, while also recognizing the impact on public health. To assist in the process of designing the clinical trail, Dr. Eydelman asked the panel to consider the following issues: control population; inclusion and exclusion criteria; acceptable adverse event rates; sample size; study duration; variables investigated; efficacy endpoints; and quality of life assessment.
Joseph N. Blustein, M.D., M.P.H., continued the presentation with a literature review on CLE for presbyopia, CLE with monofocal IOLs, CLE for hyperopia, CLE for high myopia, and retinal detachment risks. With very few articles specifically addressing CLE for presbyopia, the literature review was expanded to include CLE for other refractive corrections. Two articles reported on CLE for presbyopia, (Dick, et al. J Refract Surg 2002; Packer, et al. J Cataract Refract Surg 2002)evaluated the Array Multifocal IOL. Both studies only reported 6 month follow-up results.
The expanded literature review included articles on CLE with monofocal IOLs for ametropia, hyperopia, and myopia. Most of these studies were for short duration and small numbers. There were a considerable number of short term complications including one study reporting a greater than 50% posterior capsule opacification requiring YAG capsulotomy. Dr. Blustein noted that while several studies looking at CLE for high myopia reported efficacy, these studies were limited because of their short follow-up times and the exclusion of patients lost to follow-up. However, one study of CLE for high myopia (Colin, et al. Ophthalmology 1999) did report a follow-up period of 7 years. This study had shown a retinal detachment rate of 2 % at 4 years and 8.1 % at 7 years. This study also showed an increase in posterior capsular opacification from 37% at 4 years to 61% after 7 years. Despite the efficacy of CLE for high myopia, the potential complications still outweight the benefits (O’Brien, et al.).
A literature review looking at long-tem retinal detachemtn (RD) rates after cataract extraction (CE), provided the following additional information. About 40% of all RDs occur post-CE. The increased risk of RD after CE persists over time, even at 10 years after CE. Younger age at the time of CE, torn posterior capsule at the time of CE, later posterior capsule opacification requiring YAG capsulotomy and male gender are associated with an increase risk of RD.
Dr. Eydelman presented on the recommended measure to be perormed for multifocal IOLs under study, including endothelial cell loss, contrast sensitivity and functional performance. She finished the presentation by going over the eight questions the FDA is asking panel members to consider, given the information from the literature review. (See list of questions below.)
OPEN PUBLIC HEARING SESSION
Dr. Weiss opened the next session by noting that the FDA encourages the open public hearing speakers to advise the panel of any financial relationship they may have with any sponsor, its products and, if known, its direct competitors.
Adrian Glasser, Ph.D., associate professor at the College of Optometry at the University of Houston, reported that he appeared at the meeting as a consultant to industry as well as a scientist. He presented information on accommodation restoration concepts related to pseudophakic accommodation measurement and accommodative IOLs after CLE. Dr. Glasser’s presentation also included a list of questions about the FDA’s management of clinical trials for accommodative IOLs, subjective testing of accommodation, measurement of accommodation, and comparison of performance with a standard IOL.
Panel member Arthur Bradley, Ph.D., asked Dr. Glasser whether pharmacologically induced accommodation could substitute for voluntary accommodation. Dr. Glasser said that he would not suggest it as a substitute or the sole means to identify whether an accommodative IOL would produce an accommodative change, but as one more tool to assess the accommodative ability of an IOL.
Stephen Lane, M.D., clinical professor of ophthalmology at the University of Minnesota, noted that he appeared at the panel meeting as a consultant. He stated that the Agency could proceed along two pathways: allowing market forces to work through the practice of medicine; or proceeding with formal clinical trails. Ultimately, he asserted, a formal refractive lens exchange clinical study is the most advisable, featuring controls and addressing the outcomes important to patients.
Dr. Lane agreed with many of the trial design considerations mentioned in Dr. Eydelman’s presentation, for example, including measurement parameters; risk factor assessments (endothelial cell loss and RD).. He asked the panel to consider a number of proposals, including minimizing the study size and duration by employing exclusion criteria derived from RD surveys; applying study subjects’ own preoperative data to provide the best method of control; using minimum preoperative endothelial cell counts as an exclusion criteria based on the FDA phakic IOL requirement; and employing appropriate quality of life assessments through surveys. He noted that unless reasonable and practical considerations are used to design the trial, this increasingly popular procedure will be performed outside the scope of the best interests of everyone.
Panel members had a number of questions for Dr. Lane about RD rates in younger populations after cataract surgery, and the concern that there is little information on this population. Dr. Lane suggested that younger subjects and subjects with higher myopia be eliminated from any study. But, he added that there is an entire presbyopic patient population that would ultimately want this procedure for strictly presbyopic reasons.
Ms. Thornton read into the record a letter to the Agency from Randall J. Olson, M.D., chair, Department of Ophthalmology and Visual Sciences and director of the John A. Moran Eye Center at the University of Utah Health Sciences Center. Dr. Olson’s letter commented on the use of IOLs to correct presbyopia after CLE. He believes that, even given the issue surrounding RD in high myopes, a study of clear lensectomy “does not seem warranted, in that our cataract database is already so large and so inclusive.” As well, a study of patients with truly clear lens would be difficult, he wrote, because so few of these patients have truly clear lenses.
PANEL DELIBERATIONS
Dr. Weiss began the discussion of
the eight questions on which the FDA needs panel feedback and advice.
1. A) Do you recommend a control population for studies of clear lens extraction (CLE) in the correction of presbyopia, or do you believe the study subject’s own preoperative data is sufficient for comparison?
B) If you
recommend a control population, which one of the following controls do you
believe to be appropriate?
·
Historical control of subjects that have undergone
CLE for correction of presbyopia
·
Historical control of subjects that have undergone
CLE for correction of composite of all refractive indications
·
Historical control of postcataract extraction subjects
·
Historical control of subjects with no previous
ocular surgery
·
Active control of subjects with no previous ocular
surgery
·
Other
Most panel members agreed that some type of
control population is necessary for a study of CLE for presbyopia.
A majority of panel members recommended that
the study include historical controls to measure safety and active controls to
measure efficacy. A number of panel members felt that the study would still
have to stratify patients by such risk factors as refractive error, age,
endothelial count, and axial length. Some panel members expressed concern that,
while ideal, active randomized controls may be too burdensome for sponsors.
However, Dr. Bressler argued that randomization could ultimately be easier for
patients and sponsors with a large enough sample size followed for three years.
2. Should the clinical study inclusion/exclusion
criteria limit subject enrollment based on the criteria listed below? If yes,
please discuss the appropriate ranges of each limiting criteria for inclusion
in the study.
A) Refractive error/axial length
(hyperopia, emmetropia, myopia)
B) Patient’s age
C) Degree of accommodative loss (based
on what measurement?)
D) Preoperative endothelial cell count
E) Any other factors (e.g., best-corrected
visual acuity [BCVA])
Panel members recommended a refractive range
from +6 to +8 to -10 to -14 diopters. Dr. Bressler recommended and others
supported using the refractive range of what 95% of the refractive errors are
in the US population. Eyes with
pathologic conditions associated with high myopia or extreme hyperopia
(nanophthalmos) should be excluded. Most panel members were in favor of
including emmetropes. Since this is for
presbyopia, panel members supported a minimum age of between 40 and 45.
Discussion as to what degree of accommodative
loss and/or near vision with or without correction was felt to be necessary,
but should be determined by the product.
There was no consensus on which measurement to use.
Members agreed that the study should have a
minimum age-dependent endothelial cell count as an inclusion criterion. Additional inclusion/exclusion criteria
proposals included corneal astigmatism, axial lengths, corneal curvature, and a
minimum level of preoperative visual acuity for distance.
3. A) What should be the
primary safety endpoint for the study?
·
Retinal detachment rates
·
Endothelial cell loss
·
Any others
B) What
should be the acceptable adverse event rate associated with this safety
endpoint?
A majority of panel members suggested that
the primary safety endpoint be the incidence of RD with 0.3% as an acceptable
rate with a pre-PMA sample size of 321 followed for 3 years after CLE. The panel came to a consensus at 1,500
endothelial cell density at age 75, for the Agency to use to calculate
inclusion criteria.
Panel member Jeremiah Brown, Jr., M.D., and
others suggested that the sponsor stratify the population by nonmyopes and
myopes. However, Gene N. Hilmantel, O.D., M.S., FDA participant, cautioned that
the smaller the acceptable rate of RD for the trial, the larger the required
sample size. A number of panel members noted that the CLE procedure is so safe
now that it would take an enormous number of subjects to produce calculable
risk.
4. A) In order to
adequately determine the rates of the adverse events/complications of concern,
what do you feel is the appropriate sample size and the appropriate follow-up
period for a CLE study prior to the submission of the PMA?
B) Do you
believe a postmarket study is indicated? If so, what is the appropriate type of
study, sample size and length of follow-up for such a study?
A majority of panel members agreed that the
premarket study should include 321 patients followed from preoperative status
to 3 years, and that a postmarket study should be conducted for 5 years.
The panel discussed a postmarket study over 5
years with approximately 2,000 eyes or whatever sample size is needed to
detect.a .05 percent RD rate. Some members believed that this sample size may
be too onerous for industry, but others argued that because the market size
will be so large a large postmarket study is required.
5. Acceptable adverse
event rates for the posterior chamber (PC) IOLs at one year following cataract
extraction are listed in the FDA Grid for PC IOLs.
A) Are these rates
applicable for correction of presbyopia in noncataractous eyes via CLE at one
year postoperative?
B) Should the
acceptable adverse event rates be adjusted for the study duration recommended?
If yes, how should these rates be adjusted?
C) Do
additional adverse events need to be collected? If so, what should be their
acceptable rates be?
A majority of panel members agreed that the
study should include a cumulative adverse event rate, including that for loss
of best-corrected visual acuity, ruptured capsules, vitreous loss, and any
others relevant events on the list of refractive surgical guidance.
Dr. Weiss suggested that FDA grids should be
created that are similar to those for refractive surgical patients with a cumulative
total of acceptable complication rates.
6. FDA believes that all
multifocal IOLs’ safety profile will have to be established in a cataractous
population, prior to initiation of a clinical trial in a non-cataractous
population. MIOLs’ performance in a cataractous population will, therefore, be
known for all tests and substudies outlines in ANSI draft standard for MIOLs.
Which
substudies do you recommend for inclusion in the clear lens extraction protocol
for evaluation of performance in this non-cataractous population?
A) Functional
performance
B) Contrast
sensitivity
C) Defocus
curves
D) Fundus
visualization
E)
Endothelial cell evaluation
F)
Others
A majority of panel members agreed that the
premarket substudies for CLE should include driving simulation , near vision functional
performance, i.e., reading speed, vision
surveys on functionality in low contrast conditions, contrast sensitivity, and
endothelial cell evaluation. New IOLs
should be evaluated for fundus visualization and fundus photos.
7. The only current
performance efficacy endpoint for aphakic posterior chamber IOLs (FDA Grid) is
postoperative BCVA of 20/40 or better in 92.5 percent of the subjects.
A) Is this
applicable to non-cataractous eyes undergoing CLE for the correction of
presbyopia?
B) Are the
predictability (75 percent of eyes with MRSE +1.00 D, 50 percent with
MRSE + 0.5 D) and UCVA (85 percent with 20/40 or better) outcomes
outlined in the FDA’s Draft Guidance for Refractive Implants applicable?
C) Do we need
to establish a performance efficacy endpoint for UCVA at near in this
population of subjects undergoing surgery for correction of presbyopia? If yes,
what do you recommend?
D) What
additional performance efficacy endpoints, if any, need to be set?
A majority of panel members agreed that
criteria should be set for combined near and distance uncorrected visual
acuity.
Most panel members agreed that Question 7A
was not applicable, but they were interested in doing this for safety (distance
BCVA), not efficacy.
Panel members suggested the following
efficacy criteria; distance UCVA of
90 - 95 percent in the range of 20/25 to 20/30. Panel members also suggested the following
efficacy endpoints: low contrast reading tests, and better driving function
tests.
8. How do you recommend we evaluate patient’s quality of life
issues?
A majority of panel members suggested that an
appropriate questionnaire be administered to evaluate all patients to include
questions related to night vision, night driving, and reading. The questions
need to be directed toward refractive surgery as opposed to diseased eye, and
more toward the younger patient, as well.
SECOND OPEN PUBLIC HEARING
SESSION
Ronald E. McCarley, president and CEO of Ophtec USA, Inc., and panel industry representative, noted that most of the CLE performed have been accomplished with monofocal IOLs. He asked why the panel and the Agency would expect this pattern to change, considering the prospective problems in multifocal IOLs. Dr. Eydelman responded that the panel’s discussion covered only multifocal and accommodating IOLs.
ADJOURNMENT
Dr. Eydelman and Ms. Thornton thanked the panel members and participants. Dr. Weiss adjourned the meeting at 3:52 p.m.
I certify that I attended this
meeting of the Ophthalmic Devices Advisory Panel Meeting on March 5, 2004, and
that these minutes accurately reflect what transpired.
______________________________
Sara M. Thornton
Executive Secretary
I approve the minutes of this meeting
as recorded in this summary.
_________________________________
Jayne S. Weiss, M.D
Chair
Summary prepared by
Susan C. Sanderson
5932 Lee Highway
Arlington, Virginia 22207
(703) 237-3696
astrobuddy@mindpring.com