SynCardia Systems Inc.
CardioWest Total Artificial Heart (TAH)
DIRECTIONS FOR USE

1992 E. Silverlake Road
Tucson, AZ 85713
520-545-1234
CAUTION: Federal
(USA) law restricts this device to sale by or on order of a physician.
Table
of Contents
Page
1.0 Device Description.…………………………………………...…... 3
1.1 Implantable TAH…………………………...………...…… 3
1.2 External
Console………………………………….….......... 4
2.0 Indications for Use………………………………………....…..… 4
3.0 Contraindications………………………………………………... 4
4.0 Warnings ...........................………………………………………. 5
5.0 Precautions……………………………………………........…..... 6
6.0 Summary
of Clinical Study……………………………………… 7
6.1 Trial
Success……………………………………………….. . 7
6.2 Hemodynamics……………………………………………….. 7
6.3 Adverse
Events……………………………………………….. 8
6.4 TAH
Reliability……………………………………………….. 9
7.0 Implant Procedures.......................................................................... 10
7.1 Preparation.............................................................................. 10
7.2 Removal
of Native Ventricl
es................................................. 11
7.3 Preparing
the Atria................................................................... 13
7.4 Outflow
Connectors.................................................................. 14
7.5 Connect
Artificial Ventricles.................................................... 15
8.0 Operator’s Manual for Console………………………………… 20
8.1 Warnings
for Console Operation…………............................. 20
8.2 Readying
Console for Clinical Use............................................ 21
8.3 TAH
Startup Procedure...................................................……… 21
9.0 Explant Procedures…………………………………...……...…… 23
10.0 System Components............................................................................. 24
Appendix
A: Patient Selection and Management……………………………….. 25
Patient Selection...……………………………....…...…….……………… 25
Anticoagulation………………………………..…......…………….....…... 25
Exit Site…………………………………………………………………… 26
Appendix
B: Outline of Training Program…………………………………….. 27
Appendix
C: Materials Matrix…………………………………………………… 28
1.0 Device Description
The
SynCardia CardioWest Total Artificial Heart (TAH) system is a pulsatile
biventricular device that replaces a patient's native ventricles and valves and
pumps blood to both the pulmonary and systemic circulation. The system consists of the implantable
CardioWest TAH and an external console connected by drivelines (Figure 1).

Figure
1: CardioWest TAH System
1.1
The Implantable CardioWest TAH
The implantable CardioWest TAH consists two
artificial ventricles, each made of a semi-rigid polyurethane housing with four
flexible polyurethane diaphragms separating the blood chamber from the air
chamber. The diaphragms allow the artificial ventricle to fill and then eject
blood when compressed by air from the external console. Mechanical valves, mounted in the inflow
(27mm) and outflow (25mm) ports of each artificial ventricle, control the
direction of blood flow. The maximum
dynamic stroke volume of each ventricle is 70 ml, which allows for generating a
flow rate up to 9.5 liters per minute.
The left artificial ventricle is connected via the
left atrial inflow connector to the left atrium, and via the aortic outflow
cannulae to the aorta. The right artificial ventricle is connected via the
right atrial inflow connector to the right atrium and via the pulmonary artery
outflow cannulae to the pulmonary artery. Each artificial ventricle’s driveline
conduit is tunneled through the chest wall. The right and left artificial
ventricle’s driveline conduits are attached to seven-foot pneumatic drivelines
that connect to the back of the external console.
The external console operates and monitors the
CardioWest TAH. The console includes a monitoring computer that provides
noninvasive diagnostic and monitoring information to the user. Device rate,
dynamic stroke volumes, and calculated cardiac outputs are displayed on a
beat-to-beat basis. Drive pressure and flow waveforms, along with cardiac
output trends are provided. Patient
related alarms (e.g., low cardiac output) are also displayed on the computer
screen.
A separate alarm panel on the console provides
information on critical drive pressure and backup air and battery status. In
addition, an alarm is generated if the computer is not monitoring the patient.
All alarms generate audio and visual feedback to the user.
A backup air supply (two air tanks) and electrical
power (backup power supply and console battery) are automatically activated if
the external compressed air and/or AC power are interrupted. This can occur
during patient transport or in the event of a failure in the hospital’s air or
electrical supply.
The controller is the major component of the external console, and supplies pulses of pneumatic pressure to the right and left drivelines, which connect into the air chambers of the respective implanted artificial ventricles. These pulses cause the diaphragms to distend and thereby eject blood from the right artificial ventricle into the pulmonary circulation (typically 50-70mmHg) and from the left artificial ventricle into the systemic circulation (typically 180-200mmHg).
2.0 Indications
for Use
The SynCardia Systems, Inc., CardioWest Total
Artificial Heart (hereinafter called the CardioWest TAH) System is indicated
for use as a bridge to transplantation in cardiac transplant candidates at risk
of imminent death from non-reversible biventricular failure. The CardioWest
TAH System is intended for use inside the hospital.
3.0
Contraindications
The CardioWest TAH is contraindicated for us in patients with body
surface areas < 1.7m˛.
4.0
Warnings
1)
Setup
and operation of this device should only be undertaken by personnel trained in
accordance with the SynCardia training program. A thorough understanding of the technical principles, clinical
applications, and risks associated with the device is necessary. Prior to use,
refer to this IFU and to the CardioWest TAH Operator’s Manual for important
operating instructions.
2)
Sterile
components of the CardioWest TAH are intended for single use only. Do not use
if package is damaged. Do not re-sterilize or reuse.
3)
Safe
use of this system has not been established in pregnant patients.
4)
Do
not subject patients implanted with the CardioWest TAH to magnetic resonance
imaging (MRI) scans.
5)
Safety
and effectiveness in populations other than those of idiopathic and ischemic
cardiomyopathies has not been established.
6)
Do
not use this device if the implantable artificial ventricles cannot fit in the
chest area vacated by the natural ventricles. Inferior vena cava and left
pulmonary venous compression are possible consequences.
7)
Do
not allow any catheter to get near the inflow valves of the CardioWest TAH. If
a catheter gets into an inflow valve, the valve could become stuck, limiting
flow. Confirm by x-ray after catheter insertion. A percutaneously inserted
central catheter may migrate into the inflow valve when the patient raises
their arm.
8)
There
is a potential for air embolism. De-air the artificial ventricles to minimize
the possibility of air inadvertently entering the device.
9)
Do
not allow the external drivelines to become kinked. If there is any low cardiac
output alarm, inspect the external drivelines for kinking.
10)
A
reduction in the maximum stroke volume on the external console’s monitoring
computer to below 50 milliliters may indicate a failure of one of the
diaphragms in an artificial ventricle of the CardioWest TAH.
5.0 Precautions
1)
Measures
should be taken to prevent infection or sepsis. Use strict aseptic techniques
during implantation.
2)
The
outflow grafts must be pre-clotted before use.
3)
When
closing the chest, a reduction in device output may indicate inflow
obstruction. Reposition the artificial ventricles by anchoring to a rib or
moving into the left plural space.
4)
Do
not use an antifibinolitic agent like Aprotinine or Amicar with an active
clotting agent like FEIBA.
5)
Use
only water-soluble antiseptic cleaners around the exit site. Ointments may
delay tissue in-growth into the driveline conduits.
6)
Each
external console contains a primary and a backup controller. An additional
external console should also be available for use.
7)
A
sudden reduction in CardioWest TAH flow may be due to a kink in the pneumatic
drivelines, or some inflow obstruction to the CardioWest TAH, such as
tamponade. Defibrillation or CPR will
not be effective.
6.0 Summary of Clinical Study
The
multi-center clinical (5) study focused on use of the CardioWest TAH as a
bridge to cardiac transplantation in transplant eligible patients at risk of
imminent death from non-reversible biventricular failure. Ninety-five patients (ages 16-67) were
implanted with the CardioWest TAH and 35 patients were controls. Of the 95
patients implanted, 81 (70 males, 11 females) met all inclusion/exclusion
criteria and were designated the core implant group. All patients were
in NYHA Class IV at time of enrollment. The control group did not
receive the TAH but met study inclusion/exclusion criteria. Both groups were on maximal medical therapy
and at imminent risk of death before a donor heart could be obtained.
6.1
Trial Success
Treatment success was defined as patients who, at 30
days post transplant, were 1) alive; 2) NYHA Class I or II, 3) not bedridden;
4) not ventilator dependent; and 5) not requiring dialysis.
Trial success was achieved in 56 (69%) of the 81
core patients and in 13 (37%) of the 35 control patients. The difference was
highly significant (p=0.0019). Statistical significance was also
found with respect to survival to transplant (p=0.0008) and survival to 30 days
post transplant (p=0.0018). Of the core patients, 64 of the 81 (79%) reached
transplant after an average of 79 days (range 1-414). 16 of the 35 (46%) controls reached transplant after an average
of 9 days (range 1-44). 58 (72%) core patients and 14 (40%) controls survived
to 30 day post transplant.
There are not data to document success of the device
in patients with severe pulmonary edema.
6.2
Hemodynamics
The hemodynamic performance of the CardioWest TAH
was assessed through a comparison of pre- and post-implant values of cardiac
index, systolic arterial blood pressure, and central venous pressure.
Hemodynamic indices were effectively restored to near normal values. Average
cardiac index increased from 1.9 to 3.0 L/min/m˛, average systolic blood
pressure increased from 93mmHg to 120mmHg, and average CVP decreased from
20mmHg to 14mmHg.
The average perfusion pressure (mean aortic pressure
minus CVP) increased from 49mmHg to 63mmHg, which was associated with recovery
of renal and hepatic function.
6.3
Adverse Events
Adverse events collected for all 81 core patients while on the CardioWest TAH device are presented in descending order below. The adverse events represent 17.6 device years of experience for an overall event rate of 1.9 events per month while on the device awaiting transplant. By comparison, 89 events occurred in 31 of the control patients (88.9%) during a total of 299 days awaiting transplant, for an overall event rate of 3.6 events per month while waiting for transplant.
Table 1
Incidence of Adverse Events in Core Patients During Device Implantation,
in Decreasing Order of Frequency
(Represents 17.6 years or 6411 days on the device)
|
Adverse Event |
Number of Events |
Number (%) of Patients |
|
Any Adverse Event |
400 |
76 (93.8%) |
|
Infection |
125 |
58 (71.6%) |
|
Bleeding |
55 |
34 (42.0%) |
|
Respiratory Dysfunction |
44 |
24 (29.6%) |
|
Hepatic Dysfunction |
30 |
29 (35.8%) |
|
Neurological Event |
26 |
20 (24.7%) |
|
Renal Dysfunction |
23 |
21 (25.9%) |
|
Reoperation |
18 |
17 (21.0%) |
|
Device Malfunction |
18 |
15 (18.5%) |
|
Peripheral Thromboembolism |
14 |
9 (11.1%) |
|
Reduced Blood Pressure |
14 |
12 (14.8%) |
|
Cardiac Index |
11 |
7 (8.6%) |
|
Technical/Procedural |
11 |
3 (3.7%) |
|
Fit Complication |
5 |
5 (6.2%) |
|
Hemolysis |
3 |
3 (3.7%) |
|
Miscellaneous |
3 |
3 (3.7%) |
6.4
CardioWest TAH Reliability:
Reliability testing was conducted to determine with reasonable assurance how long a device would perform as intended, without failure.
Three separate sets of in vitro reliability testing were conducted. In one test, four TAH units were run for a period of 180 days. During this time there were no failures or abnormalities observed.
In a second in vitro reliability trial, four TAH units were tested in a “run to failure” study design and are ongoing. After 35 months of testing, there were no failures or abnormalities observed.
A third test was initiated using three TAH units which had expired their 3 year sterilization expiration date. This provided information about the effects of long-term storage on the fatigue resistance properties of the TAH. After 24 months of testing, there were no failures or abnormalities observed.
In conclusion, a total of eleven units have been run for various lengths of time over the last six years with no device-related failures. The cumulative number of days used for calculation was 6715 and there have been no failures or signs of appreciable wear observed. When the 11 units are used to calculate reliability with a 90% confidence, the reliability at 30, 60 and 365 days is as reported in the table below.
Table 2
Reliability Test Results with 90% Confidence
|
# days
run |
MTBF* |
Reliability
in number of days run |
||
|
30 |
60 |
365 |
||
|
6715 |
2916 |
0.99 |
0.98 |
0.88 |
7.0 Implant
Procedures
This section contains the Implant Procedures. Patients receiving the CardioWest TAH are prepared for the implant per standard hospital procedures for any cardiac surgery. An arterial line, a central line, and standard artificial ventilation are required prior to the start of surgery. Transesophogeal echocardiography is recommended.
7.1 Preparation
·
Pass
the CardioWestTAH sterile components into the sterile field.
·
After
a standard median sternotomy is performed and before starting heparin, 1)
prepare the arterial outflow connectors, 2) trim atrial inflow connectors to
appropriate size, and 3) tunnel the artificial ventricle conduits through the
skin.
·
Preclot
the two arterial outflow connectors three times with the patient's blood before
giving the heparin. After exposure to
the blood (approx. 30 cc for each connector each time) stretch connector, let
dry for about 5 minutes and preclot again.
The connectors are coated on the outside with biologic glue
(cryoprecipitate with calcium and topical thrombin). Stretch again and let dry.
This is done before cannulation so there is plenty of time to obtain
sufficient preclotting of the outflow connectors. If the patient has been heparinized before deciding to implant
the CardioWest TAH, the arterial outflow connectors should be preclotted with a combination of
heparinized blood, protamine, and thrombin.
·
Trim
the two inflow connectors. Cut edges of
the atrial quick connects for the atrial anastomoses to a radius extending out
from the connector for
5-7 mm. Cut in a completely circular
fashion. Then stretch and invert
them.
·
Pass the drivelines
conduits through their subcutaneous pathways before heparinization of the
patient. Position the left-sided
ventricle conduit in the epigastrium at the level of the midclavicular line and
approximately 2 inches below the costal margin. Make a semicircular skin flap incision on the left midclavicular
line approximately 5 to 10 cm below the costal margin.
·
Place a long clamp
through the subcutaneous tissue, rectus fascia, rectus muscle, and into the
chest as a chest tube would be placed.
Use a similar approach to place the driveline conduit for the prosthetic
right ventricle, approximately 4 to 5cm medial to the left ventricle conduit so
that no necrosis between the two exit sites will result.
·
Enlarge pathway by
opening the clamp and inserting a 1-inch Penrose drain through the
pathway. Place the end of the conduit
in the Penrose drain and advance approximately 8-10 cm. Pull Penrose drains through the pathway that
delivers the driveline conduit.
Position the artificial ventricles lateral to the wound and cover with a towel while the rest
of the procedure takes place. This
provides ample opportunity for small bleeders in the driveline pathway to clot.
7.2 Removal
of the Native Ventricles
·
Cannulation
of the aorta and both superior and inferior vena cava is done in a standard
fashion. Umbilical tape chokers are
used on the cavae. Dissection around
the aorta and pulmonary artery is limited to the proximal portion of the aorta
in anticipation of transplantation, thus leaving some untouched areas that will
not be very fibrotic. Cardiopulmonary
bypass is instituted and the heart is fibrillated. Total bypass is instituted by pulling on the choker tapes.
·
The
heart is fibrillated and excision of the heart begun. The excision is different
from that used for transplantation. It seeks to preserve the annulus of both
the tricuspid and mitral valves. Thus, an incision is made on the ventricular
side of the AV groove of the right ventricle (Figure 2).

Figure
2: First Incision of Ventricle Excision
·
Incision
can be done with a knife and extended with a knife or scissors. It is extended
anteriorly across the right ventricular outflow tract and just proximal to the
pulmonary valve. Posteriorly, it is extended to the interventricular septum and
across the septum, staying on the left side of the arterioventricular (AV)
groove and preserving the entirety of the mitral annulus. The anterior and
posterior lines of incision are dissected apart from each other out to the
level of the pulmonary bifurcation.
·
Trim
the excess muscle on the right and left sides down to near the AV valves. All
chordae are trimmed away, and a 2 mm edge of valve tissue along with the
annulus is left intact. The atrial cuff
generally extends 1 cm beyond the AV valves and consists of residual
ventricular muscle and fat in the AV groove.
The portion of the cuff in the left ventricular outflow tract consists
of the residual anterior leaflet of the mitral valve and some aortic
tissue. Most of the aortic tissue is
trimmed away; however, some is left intact because it is felt to present strong
tissue for the sewing of the inflow connector.
The great vessels are then separated from the remaining ventricular
myocardium above the valvular level.
The great vessels are separated from each other (Figure 3).

Figure
3: Ventricles Removed
Over-sew the coronary sinus entrance into the right
atrium. This prevents backflow of blood
through the coronary sinus and out to the cut vessels on the AV groove.
·
Three
15 by 20 centimeter sheets of membrane are used to create a neo-pericardium to
prevent adhesions. On the right side a sheet is anchored with non-absorbable
suture to the pericardial reflection at the level of the superior vena cava,
pulmonary veins and inferior vena cava. On the left side, a second sheet is
sutured to the pericardial reflection just anterior to the left pulmonary
veins. On the diaphragmatic side, a third sheet is sutured so as to cover the
entire diaphragmatic pericardial surface. The 3 sheets are then folded upon
themselves to keep them out of the operative field while the CardioWest TAH is
implanted.
7.3 Preparing the
Atria
The outer walls of the entire right and left atrial
cuff complex are encircled with Teflon felt buttresses. These are placed in such a way that they can
be used for strengthening the anastomosis to the inflow connector and also to
tamponade and control all possible bleeding from the AV groove portion of the
connector. These are cut to
approximately 10-12 mm in width and are generally 10 cm in length. It most often takes at least two of these to
extend around the entire atrial cuff.
They are placed on the outer edge of the cuff and sewn in place with a
running 3-0 polypropylene (Figure 4). A
long needle is used to accomplish this (MH needle) and, after completing this,
the left and right atrial cuffs are surrounded by Teflon felt buttresses.

Figure 4:
Atrial Sutures
·
The
atrial inflow connector is sewn first. It is inverted and placed inside the
left atrial cuff on the lateral wall.
3-0 polypropylene is used with an MH needle with a running stitch,
taking care to tailor the atrial cuff and the inflow connector into a single
hemostatic suture line. The suture line
includes both free walls of the atrium, buttressed with Teflon felt in the
atrial septum, which has no buttressing material. A similar procedure is done with the right inflow connector. The connector is inverted, placed in the
atrium, the suture line is run, and after completing both suture lines, the
inflow connectors are returned to their normal position (Figure 5).

Figure 5: Inflow connector inverted for suturing
(left), finished normal position (right)
·
Check
for hemostasis with the plastic leak tester made to fit within the inflow
connector. A syringe (60-100 cc) is
used to inject into a three-way stopcock connected with the tester to test the
left atrial suture line. The surgeon places his hand posterior to the left
atrium and compresses the right and left pulmonary veins, while the assistant
injects saline mixed with a small amount of blood into the left atrium. Observe for leaks. A dental tool is used to
break the seal between the tester and connector. If there are any leaks, sutures are placed at this time. On the right side, fluid is simply injected
into the right atrium under pressure, since the inferior and superior vena cava
are already obstructed by the caval tapes.
Again, closure of leaks with a 3-0 MH polypropylene suture is done at
this time.
7.4 Outflow
Connectors
· Great vessel connections are
made. The pulmonary artery anastomosis
is made first. The lengths of the
outflow connectors are determined by placing the artificial ventricles in
position within the pericardial cavity. Place the outflow connector between the
aortic or pulmonic valve and its respective great vessel and measure the
distance. Cut outflow
connectors to the appropriate lengths, usually 3 to 5 cm.
·
The
pulmonary artery anastomosis is made with a running 4-0 polypropylene suture in
an end-to-end fashion, beginning with lateral wall and running the back wall of
the anastomosis from the inside (Figure 6).

Figure
6: Outflow Connector Suturing
·
A
similar anastomosis is made with the aortic suture line. Then, the outflow connector leak tester is
used, which is inserted into the aortic outflow connector. Saline is injected under pressure, observed
for leaks, and then any leaks are closed with a 4-0 polypropylene suture. The pulmonary artery needs to be
cross-clamped in order to test the integrity of the pulmonary artery to
connector anastomosis. The pulmonary
artery and aortic tester is the same, but smaller, than the one utilized for
the atrial inflow connector.
7.5 Connect Artificial Ventricles
·
Once
adequate hemostasis of all suture lines is established, placement of the
artificial ventricles begins. First,
the left artificial ventricle is connected (Figure 7).

Figure 7: Connect Ventricles
·
Grasp
the left inflow connector with two large Mayo clamps placed side by side, with
a good hold of the connector. The
opposite side of the plastic fitting for the connector of the artificial left
ventricle is placed within the connector, and the operator pulls with the Mayo
clamps and pushes the artificial left ventricle into the inflow connector. The position in which the heart sits for
the duration of the support of the patient is determined by the orientation of
the artificial left ventricle as it is placed into the left atrial inflow
connector. Therefore, a careful
assessment of exactly where the aortic outflow connector will connect to the
artificial left ventricle, and the anticipated position of the artificial left
ventricle should be made before the connection of the atrial inflow connector
is completed. It is then an easy matter
to snap on the aortic outflow connector, taking care not to twist the connector
or aorta. While this is being done, the
artificial left ventricle should be filled with saline through the aortic valve
as well as the outflow connector. Once
the connection is made, the patient is placed in a steep Trendelenburg position and large vent
sites are placed in the highest point of the aortic outflow connector and the
aorta for removal of air.
· The artificial right ventricle is then connected. The atrial connection is made first, again taking care with the orientation of the artificial right ventricle so that the direction of flow from the outlet valve is appropriate for the anatomy of the patient. After the atrial connection is made, the pulmonary outflow connection is made, again, taking care not to twist. Before connecting the pulmonary outflow connector graft, the chokers on the superior and inferior vena cava should be removed. This allows a flow of blood into the right atrium and the right artificial ventricle, and flushes air out as the connection to the pulmonary artery is made (Figure 8).

Figure 8: CardioWest TAH Final Position
·
With
the patient in extremely steep Trendelenburg position and lungs being slowly
ventilated, begin pumping at a very slow rate (40 BPM, 40%SYS, 180mmHg-LDP,
60mmHg-RDP, 0mmHg-VAC). Agitation of
the artificial ventricles, as well as atria, is done at this time. If available, monitor for air bubbles in the
atria and aorta with transesophageal echo to help decide when the device has
been completely de-aired. As air is
slowly removed from the device, increase pumping rate and pressure. Generally, this process takes about 10 minutes
and should be done with patience and attention to remove air before the CardioWest
TAH takes over from the heart-lung machine.
Decrease flow on the heart-lung machine temporarily to help move air
through the lungs and into the device.
Once satisfied that all air is out of the device, close vent sites and begin
full pumping as the heart-lung machine is weaned off.
The patient should be kept in steep Trendelenburg for an additional
15-20 minutes.
·
As
the table is flattened out, try to position the artificial ventricles within
the mediastinum. The pleura on both sides should not be opened and the
pericardium should be left intact for closure.
In smaller patients, there may be a need to force the right ventricle
under the left edge of the sternum.
Care should be taken to examine the left pulmonary veins and the inferior
vena cava for evidence of compression.
This is facilitated with trans-esophageal echo.
· Check for hemostasis. After protamine has been administered and hemostasis obtained, a trial closure of the sternum is done using towel clips. If the fit of the device is judged adequate by hemodynamic stability and by transesophageal echo examination of the caval and pulmonary venous flows, reopen the chest and bring together the edges of the Gortex sheets to form a tent or neo-pericardium. Take care to make a loose fit, without impingement upon the cavae and tension on the device. Prior to closure of the cephalic part of the neo-pericardium, pass a rectangular piece of Gortex membrane around the proximal ascending aorta and anchor with non-absorbable suture. This is to provide a surgical plane at explant between the aorta and pulmonary artery to facilitate encircling and cross clamping the aorta.
· One chest tube is placed in the neo-pericardium and a second in the native pericardial space. Irrigate with antibiotic solution before closure. Close the sternum and remaining incision in a routine fashion. Check device output, central venous pressure, and device filling when the chest is closed, because chest closure may alter the anatomy, causing pressure on the left-sided pulmonary veins, inferior vena cava, and occasionally the right-sided pulmonary veins. If decreased flow is noted, the chest must be reopened and changes made in the position of the device. One change has been to mobilize the diaphragmatic attachment of the pericardium, allowing the device to sit more leftward in the chest. This requires opening the left pleura, allowing the CardioWest TAH to slightly migrate into the left pleural space. If decreased flow is still observed, the right artificial ventricle may need to be anchored to a rib using umbilical tape (Figure 9).

Figure
9: Solution to a Fit Problem
8.0 Operator’s Manual for
Console
The operator’s manual, Part # 960001, contains detailed information on the setup, operation and troubleshooting of the CardioWest TAH system. A brief description of the contents is given as a reference.
§
Introduction to the External Console:
Describes the system overview, indications for use, and warnings.
§
Features and Operations: Describes the operation of the controller, power
supply, air supply, vacuum pump, UPS, alarms and computer.
§
Unpacking and Initial Setup: Covers unpacking instructions and initial setup.
§
Performance Verification: Describes the console test procedure and the preparation
for standing by for an implant.
§
Clinical Use: Describes console operation, readying system for clinical use, CardioWest
TAH startup, patient transport, transfer to backup controller, and console
replacement.
§
Specifications: Describes the CardioWest TAH physical and performance specifications.
§
Routine Maintenance and System Checkout: Describes console checkout, batteries,
cleaning, and checkout procedure.
§
Field Service Guide: Describes air tank replacement, scheduled servicing,
air tank connector O-ring replacement, controller pilot pressure calibration,
controller replacement, fuse replacement, inactive storage, and crating
instructions.
8.1 Warnings for Console Operation
·
DO NOT operate or adjust system
without proper training.
·
DO NOT operate console on an air
supply of substandard or unknown quality, either from tanks or in-house
compressors.
·
DO NOT use a controller outside of
its planned maintenance cycle.
·
DO NOT intentionally operate a
system having only one functional controller for any longer than is necessary
to switch systems.
·
DO NOT defeat the alarm system by
turning it off, tampering with the alarm mute button, by muffling the audible
alarms, or by any other means.
·
DO NOT expose the system to any
unusual environment, i.e. electric or magnetic fields, dampness or temperature
extremes.
·
DO NOT leave key in primary
controller key switch during an implant.
The key may be kept on Velcro near top right side of the backup air
supply compartment.
·
DO have a backup system in a
state of ready standby.
·
DO set backup controller
parameters to the same values as the primary controller.
·
DO have a controller switch
key attached to the console.
·
DO keep system casters locked
except for transport.
·
DO have a number of spare
charged air tanks on hand.
8.2 Readying Console for
Clinical Use
·
Two
consoles should be in ready standby mode. Ensure that backup batteries are
fully charged.
·
Verify
that each system has been connected to AC power with the SYSTEM POWER switch in
the ON (1) position
·
Confirm
that each controller AC POWER/BATT CHRG lamp is on.
·
Before
moving system to Operating Room, moisten a clean cloth with an antibacterial
agent and wipe down all exterior surfaces of the console.
·
Do not spray any cleaning agent directly on system.
8.3 CardioWest TAH Startup Procedures
·
Turn
SYSTEM POWER switch OFF (0); disconnect system mains power cord and move system
to patient site.
·
Position
the rear side of the console within driveline length of the patient’s chest.
·
Lock
front casters (wheels).
·
Connect
system power cord and turn SYSTEM POWER switch ON (1).
·
Verify
console AC POWER and CHARGE LEDs are green, indicating system has AC power.
·
Verify
that both controller AC POWER/BATT CHRG indicator lights are on.
·
Connect
main air supply and verify pressure is 50–110 psi (340-575 kPa) at system power
interface panel gauge.
·
Open
primary and reserve air tank valves and verify they are fully charged.
·
Set
primary and backup controllers to values listed below:
Left
Drive Pressure = 0 mmHg; Right Drive Pressure = 0 mmHg
Vacuum = 0 mmHg
Heart
Rate = 40 bpm
Systolic
Duration = 33%
Be
sure vacuum remains off until the patient’s mediastinum is closed.
·
Turn
computer on. Wait for WCOMDU to load.
Select Patient Monitoring Mode. Enter patient identification
requested. Inhibit WCOMDU alarms during
startup.
·
Be
sure that LDP, RDP and VACUUM are zero. Turn primary CONTROLLER key switch On
and press controller ALARM RESET button.
·
Turn
ALARM SYSTEM key switch to ON. Mute
console hardware alarms until LDP > 90 mmHg and RDP > 20 mmHg.
·
Verify
controller is operating normally; connect left ventricle driveline to
controller upon order by surgeon. After
left ventricle is connected and de-aired, await instructions from surgeon to
start. To start left ventricle, raise LDP to about 100 mmHg. You should see a
slight overload of the cardiopulmonary bypass waveform and WCOMDU may show a
small output.
·
Connect
right ventricle driveline to controller. After right ventricle is connected and
de-aired, await surgeon’s instructions to start the right ventricle. To start right ventricle, raise RDP to 40
mmHg.
·
As
perfusionist begins to slow venous return, CardioWest TAH filling should
increase. As filling increases, adjust
drive pressures, heart rate and systolic duration to prevent full fill and to
provide full ejection. Normal range is
LDP= 170-210 mmHg, RDP= 60-100 mmHg, % Systole=50-60. Be vigilant during the weaning process, you may need to make
rapid adjustments. Observe WCOMDU waveforms for signs of flow obstruction and
other cardiac output information.
·
After
chest is closed, vacuum may be started (normal range is approximately 10
mmHg). Do not exceed 30 mmHg vacuum.
·
Remove
key from the primary controller key switch before moving patient.
·
Pneumatic
drive ejection pressures should be set to achieve full ejection. Pressure
tracings on the monitoring computer can be viewed to assure the right drive
pressure is set to overcome the pulmonary systolic pressure, and the left drive
pressure is set to overcome the aortic systolic pressure.
·
The
CardioWest TAH rate should be set to achieve a stroke volume between 50 and 65
milliliters on the monitoring computer. CardioWest TAH beat rates should be
between 100 and 130 beats per minute.
9.0 Explantation Procedures
Explantation of the device should be handled like
any other redo cardiac procedure. Great
care should be taken in the separation of the sternum from the device, the
great vessel connector, and the drivelines.
Explantation may be easier if the device is covered with a Gortex
membrane.
Cardiopulmonary bypass is
initiated with dual caval cannulation with tourniquets, the aorta is
cross-clamped, and the CardioWest TAH is turned off. The artificial ventricles are separated from the atrial inflow
cannula. The great vessels outflow
connectors are amputated at the level of the connector/great vessel
anastomosis. The artificial ventricles
are transected at the base to the driveline conduit connection, and the CardioWest
TAH is removed from the operating field. The driveline conduits are pulled
through the skin. The remaining atria
inflow connectors are still in the remaining portion of ventricular muscle where
they were initially sutured. They are
removed by transecting the AV groove throughout. The remaining atria and great vessels can now be trimmed to
accept the donor heart.
10.0 System Components
The CardioWest TAH system is
comprised of the following:
·
Implant
Kit - Part # 500101 (Sterile)
Contains left artificial ventricle, right artificial
ventricle, 2 inflow connectors, 2 outflow connectors, and an ancillary pack
with drivelines, inflow pressure test plug, outflow pressure test plug, locking
ties, and 2 de-airing needles (all sterile). All sterile components are
packaged in double aseptic transfer packages.
·
Surgical
Spares Kit - Part # 500177 (Sterile)
Contains inflow connector, outflow connector,
drivelines, inflow pressure test plug, outflow pressure test plug, and locking
ties.
·
Circulatory
Support System (External Console) – Part # 500207 (Non-sterile)
·
Air
Tank - Part # 390004 (Non-sterile)
There should be two complete
implant kits, one surgical spares kit, two circulatory support systems and
eight air tanks.
Appendix A
Patient Selection and
Management
Management and coordination of successful CardioWest
TAH support requires a multidisciplinary team that has experience with
circulatory support systems. Teams can include surgeons, cardiologists, heart
transplant coordinators, perfusionists, engineers, nurses, cardiac
rehabilitation therapists and coagulation specialists. The following reports the experience of the largest enrolling
clinical site, University Medical Center, Tucson, Arizona:
Patient Selection
Successful
bridge to transplant with the CardioWest TAH involved selecting patients who were
transplant eligible and who additionally were assessed in two main areas: 1)
evaluation of fit of the CardioWest TAH in the patient’s chest, and 2) evaluation
of the potential for reversal of any end organ dysfunction.
Once
the CardioWest TAH is implanted, and there were no fit issues, flow was
maximized through the CardioWest TAH. The controller average settings during
the trial were: left drive pressure of 180-220 mmHg, right drive pressure of
50-70mmHg, device rate of 110-130 BPM, percent systole of 50-55%, and diastolic
vacuum of 8-12 mmHg. With these settings an average device output was 6.5-7.5
LPM, with a CVP of 8-12 mmHg.
The
CardioWest TAH is specified for patients with body surface areas of at least
1.7 m˛. At a cardiac index of 2.5
l/min/m˛, the calculated flow would be 4.25 liters/min. This is the flow used to simulate
hypotensive conditions tested during product reliability testing. The TAH console is pre-set with an alarm to
indicate flows <3.5 l/min.
With
normalized hemodynamics, device outputs remained relatively constant, changing
as the CVP fluctuated. This “Starling like response”, (where an increase in CVP
fills the CardioWest TAH with more volume, which is ejected on the next beat,
increasing device output), required no controller adjustments. Constant device
output and high flow under normal CVP provided good washing of the artificial
ventricles.
Anticoagulation
therapy
The
level of anticoagulation will vary depending on the patient’s coagulation
status. In general, the patient required systemic anticoagulation, similar to
that used for patients with mechanical valves. The following guidelines were
used by the largest enrolling clinical site:
Pre-operative
baseline
Results
of PT, PTT, bleeding time, TEG, platelet count, platelet aggregation studies
and fibrinogen were obtained.
Intra-operative
period
Heparinized
for CBP per usual routine. Protamine
was used for reversal per usual routine.
Post-operative
period (immediate)
Dipyridamole
was started at 100 mg -250 mg PO or NG every 6 hours. The dose was adjusted to balance platelet aggregation factors:
collagen factor kept positive; ADP, epinephrine, arachnadonic acid factors were
kept negative. If all factors were
positive, the dose of dipyridamole was maximized. If only one factor other than collagen is positive, dipyridamole
or ASA were increased until only collagen was positive. Platelet aggregation studies were checked
twice per week.
ASA was started when
platelet aggregation showed any factor other than collagen was positive,
usually within 24 hours post-operative.
The ASA dose was started between 81-650 mg PO per day. If all platelet aggregation factors were
positive and the dipyridamole was already started, 325 mg of ASA was used per
day to start. If only one factor, other
than collagen, was positive, ASA was started at 81 mg per day. The dipyridamole was adjusted according to
the results of Platelet Factor 4 and Beta Thromboglobulin. If these tests were elevated, the platelets
are very active and the dipyridamole needed to be increased. ASA is adjusted with the platelet
aggregation and bleeding time studies.
The bleeding time was kept between 10 - 20 min if possible.
If
the collagen was negative, too much dipyridamole or ASA was being given; daily
dosages of one or both were decreased to prevent bleeding.
Pentoxifylline
400 mg was started PO every 8 hours in the early post-operative period (2-3
days). Pentoxifylline could be
increased if fibrinogen increased above normal.
Post-operative (chest tubes
pulled)
IV Heparin was started at
25,000 units in 250 cc of D5W at 500-1000 units per hour, when chest tubes were
discontinued. IV Heparin was continued to maintain PTT at 50-55 sec for 2
weeks, then converted to Coumadin to keep INR 2.5-3.5 or PT 18-22 sec then IV
Heparin was stopped.
Exit Site Management
Take care to keep driveline
exit sites clean and dry. Infections should be treated according to hospital
protocol.
Appendix B
Outline of Training Program
Operation of this device
should only be undertaken by personnel trained in accordance with the SynCardia
Training Program. The training will include the following topics:
1)
Indications
and Contraindications
2)
System
Overview
3)
Implant
Procedures
4)
Operation
of the console
5)
Explant
Procedures
6)
Patient
Management
7)
Summary
of Clinical Studies
8)
Animal
Procedure – a minimum of one implant needs to be performed.
9)
Cadaver
Study – a minimum of one study to practice surgical technique and to become
familiar with fitting the device.
10)
Surgical
Proctors – surgeons may view an implantation of the device at a center of
excellence or have oversight during the initial case.
Appendix C
Materials Matrix
The
CardioWest TAH ventricle components are manufactured from the raw materials as
defined in the matrix below. The
artificial ventricles have met the test requirements of ISO 10993, Biological Evaluation of Medical Devices.
CardioWest TAH Patient Contacting Materials Matrix
|
Component |
Material |
|
Ventricle and diaphragm |
Segmented polyurethane |
|
|
Nylon |
|
Inflow connector |
Segmented polyurethane |
|
|
Polyester fabric |
|
Outflow connector |
Segmented polyurethane |
|
|
Polyethylene material |
|
Valves |
Titanium and pyrolitic carbon (Medtronic Hall Heart Valves) |
|
Drivelines |
Polyvinyl chloride tubing |