![]()
CLiRpathä Excimer
Laser Catheters
Cool
Laser Revascularization for Peripheral Artery Therapy
Extreme (OTW) and
Vitesse (Rx) Catheter Models
DRAFT Instructions For Use
1..... Description....................................................................... 1
2..... Indications for Use....................................................... 2
3..... Contraindications........................................................ 2
4..... Warnings............................................................................ 2
5..... Precautions..................................................................... 2
6..... Potential Adverse Events........................................ 3
7..... Clinical Studies.............................................................. 3
8..... Individualization of Treatment.............................. 3
9..... Operator's Manual...................................................... 4
10.. How
supplied................................................................... 4
10.1... Sterilization........................................................ 4
10.2... Inspection Prior to Use............................... 4
10.3... Procedure Set Up........................................... 4
10.4... Compatibility........................................................ 4
11.. Directions
for use........................................................ 4
12.. company
information.................................................. 6
.....
Spectranetics CLiRpathä excimer laser catheters are
constructed of multiple fiber optics arranged around a guidewire lumen and are
intended for use in the peripheral
vasculature for recanalization of obstructed arteries.
For
Extreme, over the wire (OTW) catheters, a side arm adapter located at the proximal end of the usable
length facilitates the use of the laser catheter over the appropriate sized
guidewire (0.014", 0.016", 0.018", 0.025”, and 0.035”).
For
Vitesse, rapid exchange (Rx) catheters,
the guidewire lumen begins at the distal tip and is concentric with the fiber
array, and exits the laser catheter 9 cm away from the distal tip which has
direct patient contact.
For
Vitesse-E, eccentric catheters,
the laser catheter
consists of eccentrically aligned optical fibers and a stainless steel torque
device encased within a polyester shaft. There
are two major portions of the laser catheter shaft, the proximal portion which
terminates at the laser connector, and the distal portion which terminates at
the tip having direct patient contact. The torque
device extends from the torque handle, located at the y-adapter, through the
entire 140 cm of the distal portion of the catheter, and terminates in the
distal tip. There is a mechanism within the torque handle which limits the
turns to five full rotations in each direction. The torque handle also has an
indicator displaying its range of motion. The laser catheter is packaged with
the indicator in the center of its range (see Figure 1). The torque response is
6:1, six turns of the torque handle result in one 360° turn of the distal tip.
A radiopaque marker band with radiolucent window is located on the distal tip
of the laser catheter to aid localization within the coronary vasculature in
conjunction with fluoroscopy (see Figure 2).
Figure 1

Figure 2
The multifiber laser catheters transmit
ultraviolet energy from the Spectranetics CVX-300® to the
obstruction in the artery. The ultraviolet energy is delivered to the tip of
the laser catheter to photo-ablate fibrous, calcific, and atheromatous lesions,
thus recanalizing diseased vessels (photo ablation is the process by which
energy photons cause molecular bond disruption at the cellular level without
thermal damage to surrounding tissue). The Spectranetics laser catheters have a
proprietary lubricious coating to ease their trackability through arteries.
|
Device OD Outer Diameter (mm) |
Model Number |
Max. Guidewire Lumen (in.) |
Max. Tip Outside Diameter (in.) |
Min. Tip Inside Diameter (in.) |
Max. Shaft Diameter (in.) |
Min Tail Tube Length (cm) |
Min Working Length (cm) |
|
Extreme (OTW) Catheter Specifications |
|||||||
|
0.9
mm |
110-001 |
0.014 |
0.038 |
0.0155 |
0.047 |
183 |
130 |
|
0.9
mm |
110-002 |
0.014 |
0.038 |
0.0155 |
0.047 |
183 |
130 |
|
1.4
mm |
114-001 |
0.014 |
0.056 |
0.017 |
0.056 |
183 |
131 |
|
2.0
mm |
120-001 |
0.018 |
0.077 |
0.021 |
0.076 |
183 |
131 |
|
2.2
mm |
222-005 |
0.035 |
0.088 |
0.037 |
0.089 |
168 |
120 |
|
2.5
mm |
225-004 |
0.035 |
0.1 |
0.037 |
0.098 |
168 |
100 |
|
Extreme
II (OTW) Catheter Specifications |
|||||||
|
2.0
mm |
220-006 |
0.035 |
0.0775 |
0.026 |
0.083 |
168 |
131 |
|
2.3
mm |
223-001 |
0.035 |
0.092 |
0.039 |
0.094 |
168 |
120 |
|
2.5
mm |
225-010 |
0.035 |
0.099 |
0.039 |
0.101 |
168 |
100 |
|
Vitesse
(Rx) Catheter Specifications |
|||||||
|
0.9
mm |
110-003 |
0.014 |
0.038 |
0.0155 |
0.049 |
183 |
131 |
|
1.4
mm |
114-009 |
0.014 |
0.057 |
0.0175 |
0.062 |
183 |
131 |
|
1.7
mm |
117-016 |
0.014 |
0.0685 |
0.0175 |
0.072 |
183 |
131 |
|
2.0
mm E |
120-008 |
0.018 |
0.0785 |
0.0205 |
0.084 |
183 |
129 |
|
2.0
mm |
120-009 |
0.014 |
0.08 |
0.0175 |
0.084 |
183 |
131 |
Glossary
of Special Terms
Retrograde
Fashion = In the
direction opposite to blood flow.
Antegrade
Fashion = In the
direction of blood flow.
Baseline
Angiography = Angiographic
record of blood vessels.
Contralateral
Approach = Arterial
access by a crossover approach.
The CLiRpathä Excimer
Laser Catheters and the Spectranetics CVX-300® Excimer Laser System
are approved for the following indication:
• For
facilitating limb salvage in patients with critical limb ischemia (associated
with Rutherford Categories 4, 5 and 6) who have angiographically evident
culprit stenoses and/or occlusions in the SFA, popliteal and/or infrapopliteal
arteries, who are poor surgical candidates and who are acceptable candidates
for revascularization.
• No known contraindications.
Federal (USA)
law restricts this device to sale by or on the order of a physician with
appropriate training.
Spectranetics Excimer Laser Catheters require CVX-300®
software version 3.7 or higher.
When the laser catheter is in the body, it should be
manipulated only while it is under fluoroscopic observation with radiographic
equipment that provides high quality images.
The
use of the CVX-300® Excimer Laser System is restricted to physicians
who are trained in atherectomy, Percutaneous Transluminal Coronary Angioplasty
(PTCA) and who meet the training requirements listed below. These requirements
include, but are not limited to:
1. Training of laser safety and physics.
2. Review of patient films of lesions that
meet the indications for use.
3. A review of cases demonstrating the
CLiRpath catheters in lesions that meet the indications for use.
4. A review of laser operation followed by a
demonstration of the CVX-300® Excimer Laser System.
5. Hands on training with the CVX-300®
Excimer Laser System and appropriate model.
6.
A fully trained Spectranetics representative will be present
to assist for a minimum of the first three cases.
7. Following the formal training session,
Spectranetics will make available additional training if so requested by the
physician, support personnel, the institution or Spectranetics.
This catheter has been sterilized using
Ethylene Oxide and is supplied STERILE. The device is designated and intended
for SINGLE USE ONLY and must not be
resterilized and/or reused.
Store in a cool, dry place. Protect
from direct sunlight and high temperatures (greater
than 60°C or 140°F).
The sterility of the product is
guaranteed only if the package is unopened and undamaged. Prior to use,
visually inspect the sterile package to ensure that the seals have not been
broken. Do not use the catheter if the integrity of the package has been
compromised. Do not use catheter
product if its “Use Before Date,” found on package labeling, has been passed.
Before use, examine carefully all of
the equipment to be used in the procedure for defects. Do not use any equipment
if it is damaged.
After use, dispose of all equipment in
accordance with applicable specific requirements relating to hospital waste,
and potentially biohazardous materials.
Read the Operator’s Manual (7030-0035
or 7030-0068) thoroughly before operating the Excimer Laser System. Pay
particular attention to the Warnings and Responsibility section of the manual
which explains Notes, Cautions, and Warnings to be followed to ensure safe
operation of the CVX-300®.
During the procedure, appropriate
anticoagulant and vasodilator therapy should be provided to the patient per the
institution’s PTA protocol.
Use of the Spectranetics CVX-300®
Excimer Laser System may contribute to the following complications:
|
Þ
Dissection of
the arterial wall |
Þ Embolization |
|
Þ
Acute reclosure |
Þ Spasm |
|
Þ
Aneurysm
formation |
Þ Thrombus |
|
Þ Nerve Injury |
Þ Hematoma |
|
Þ
Perforation |
Þ Arrhythmia |
|
Þ
AV fistula
formation |
Þ Death |
No long term adverse effects of peripheral excimer laser atherectomy are
known at this time.
7.1
LACI
The LACI (Laser Angioplasty
in Critical limb Ischemia) Registry Group enrolled 155 limbs
exhibiting critical limb ischemia (CLI) of 145 patients who were poor surgical
bypass candidates, at 14 sites.
Outcomes were compared to the 789-patient Control Group of a randomized
trial of prostaglandin E1 versus standard therapies in CLI patients
(Ann Intern Med 1999; 130:412-421). In
the Registry Group, treatment was laser atherectomy plus balloon angioplasty
and optional stenting in the SFA, popliteal and infrapopliteal arteries. In the Control Group, various therapies were
used including bypass surgery, endarterectomy, analgesics and others.
Analysis
The two groups were similar in age,
history of coronary artery disease, proportion of high surgical risk and
proportion of Rutherford Category 4 patients.
The Registry Group had greater history of stroke, hypertension,
diabetes, hypercholesterolemia, and obesity; the Control Group had more men and
more smokers. Procedural details for
the Registry Group are shown in Table 7.1.1, with complications shown in Table
7.1.2.
Table
7.1.1 Procedure Information, Registry
Group
|
Locations
of vascular lesions (n=423 ) |
|
|
SFA |
174
(41%) |
|
popliteal |
64
(15%) |
|
infrapopliteal |
174
(41%) |
|
Angiographic
results |
|
|
Mean lesion stenosis, baseline |
92%
± 12% |
|
Stenosis post-laser |
55%
± 24% |
|
Final residual stenosis |
18%
± 26% |
|
Lesions
per limb (n=155 limbs) |
2.7
± 1.4 |
|
Laser
treatment delivered |
153
(99%) |
|
Laser
pulses delivered per limb |
5371
± 5871 |
|
Balloon
catheter used |
149
(96%) |
|
Stent
implanted |
70
(45%) |
|
Procedure
success |
132
(85%) |
|
Straight
line flow to foot established |
138
(89%) |
Procedure success: £50%
final residual stenosis
Table
7.1.2 Complications, Registry Group, n=155 limbs
|
Procedural
Complications |
|
|
Spasm |
5
(3%) |
|
Major dissection |
6
(4%) |
|
Thrombus |
5
(3%) |
|
Distal embolization |
5
(3%) |
|
Perforation |
4
(3%) |
|
Other |
7
(5%) |
|
In-Hospital
Complications |
|
|
Reocclusion |
2
(1%) |
|
Pseudoaneurysm |
2
(1%) |
|
Renal failure |
2
(1%) |
|
Bleeding |
8
(5%) |
|
Infection |
2
(1%) |
|
Other |
4
(3%) |
A comparison of Serious Adverse Events
(SAEs) (Table 7.1.3) shows similar rates of death, major amputation, and
nonfatal cardiovascular events during the 6-month enrollment period. More frequent reintervention was observed in
the Registry Group, 24/ 145 (17%) versus 34/ 789 (4%) for the Control Group
(p<.001). The Control Group experienced
significantly more surgical interventions (bypass or endarterectomy) than the
Registry Group (p<.001). Primary
outcomes at 6 months are shown in Table 7.1.4.
Amongst limbs with 6-month data, limb salvage was 118/127 (93%) in the
Registry Group and 501/577 (87%) in the Control Group, p = .08.
Table
7.1.3 Adjudicated SAEs, Registry and Control Groups
|
|
Registry Group n=145 |
Control Group n=789 |
p |
|
Death |
15 (10%) |
113 (14%) |
ns |
|
MI
or Stroke* |
1 (1%) |
10 (1%) |
ns |
|
Reintervention |
24 (17%) |
34 (4%) |
<.001 |
|
ALI |
1 (1%) |
DNA |
- |
|
Major
amputation |
9 (6%) |
76 (10%) |
ns |
|
Bypass |
3 (2%) |
- |
- |
|
Endarterectomy |
1 (1%) |
- |
- |
|
Hematoma
with Surgery |
1 (1%) |
6 (0.8%) |
ns |
|
total |
55 (38%) |
239 (30%) |
ns |
MI = myocardial infarction. ALI = acute limb ischemia.
DNA = data not available. ns =
not significant.
Table
7.1.4 Primary Outcomes, Registry and Control Groups
|
|
Registry Group n (%) |
Control Group n (%) |
p |
|
Patients
enrolled |
145 (100%) |
673 (100%) |
. |
|
Primary
Endpoint (see Note) |
110 (76%) |
494 (73%) |
ns |
|
Death,
any cause |
15 (10%) |
96 (12%) |
ns |
Note: Primary
endpoint: limbs without major
amputation, death, lost-to-follow-up, or withdrawal. ns = not significant.
The risks and benefits described above should be carefully
considered for each patient before use of
the CliRpath catheters.
Additionally, recanalization of native arteries may be attempted in patients presenting with
reoccluded synthetic bypass grafts.
Patient selection and clinical techniques should be
conducted according to instructions provided in Section 2., “Indications for
Use,” and Section 9, “Operator’s Manual.”
ENERGY PARAMETERS
The devices
described in this document can be operated within the following energy ranges
on the CVX-300®:
|
Device O.D. |
Model No. |
Fluence |
Repetition Rate |
Laser On/ Off Time |
|
Extreme (OTW) Catheters |
|
|
||
|
0.9
mm |
110-001 |
30-60 |
25-40 |
5
sec on/10 sec off |
|
0.9
mm |
110-002 |
30-80 |
25-80 |
10
sec on/5 sec off |
|
1.4
mm |
114-001 |
30-60 |
25-40 |
5
sec on/10 sec off |
|
2.0
mm |
120-001 |
30-60 |
25-40 |
5
sec on/10 sec off |
|
2.2
mm |
222-005 |
30-60 |
25-40 |
5
sec on/10 sec off |
|
2.5
mm |
225-004 |
30-50 |
25-40 |
5
sec on/10 sec off |
|
Extreme II (OTW) Catheters |
|
|
||
|
2.0
mm |
220-006 |
30-60 |
25-40 |
10
sec on/5 sec off |
|
2.3
mm |
223-001 |
30-60 |
25-40 |
10
sec on/5 sec off |
|
2.5
mm |
225-010 |
30-50 |
25-40 |
10
sec on/5 sec off |
Vitesse (Rx) Catheters
|
|
|
|
|
|
0.9
mm |
110-003 |
30-60 |
25-40 |
5
sec on/10 sec off |
|
1.4
mm |
114-009 |
30-60 |
25-40 |
5
sec on/10 sec off |
|
1.7
mm |
117-016 |
30-60 |
25-40 |
5
sec on/10 sec off |
|
2.0
mm E |
120-008 |
30-60 |
25-40 |
5
sec on/10 sec off |
|
2.0
mm |
120-009 |
30-60 |
25-40 |
5
sec on/10 sec off |
Recommended
calibration settings: 45 Fluence, 25 Hz.
For
single use only. Do not re-sterilize
and/or reuse.
The Spectranetics laser catheters are supplied sterile.
Sterility is guaranteed only if the package is unopened and undamaged.
Before use, visually inspect the sterile package to ensure
that seals have not been broken. All equipment to be used for the procedure,
including the catheter, should be examined carefully for defects. Examine the
laser catheter for bends, kinks or other damage. Do not use if it is damaged.
Some or all of the following additional
materials, which are not included in the laser catheter package, may be required
for the procedure (these are single use items only—do not resterilize or
reuse):
§ Introducer sheaths and/or femoral guiding catheter(s) in the appropriate size and configuration to select the peripheral artery and facilitate largest laser catheter to be used.
§ Tuohy-Borst “y” adapter or Hemostatic valve(s).
§ Sterile normal saline
§ Standard contrast media
§ 0.014", 0.016", 0.018”, 0.025”, or 0.035" guidewires
Do not use
in combination with any other laser system.
Guidewire
Compatibility
See
Catheter Specification Table in Section 1.
Using
sterile technique, open the sterile package. Remove the packaging wedges from
the tray and gently lift the laser catheter from the tray while supporting the
black laser connector, also known as the proximal end, proximal coupler, or
proximal connector. Please note that the proximal end of the laser catheter
connects only to the CVX-300®, and is not meant to have any patient
contact.
Connect
the proximal end of the laser catheter to the CVX-300® and position
the laser catheter in the laser system extension pole. Calibrate the laser
catheter following the instructions provided in the CVX-300®
Operator's Manual (7030-0035 or 7030-0068).
1. Use standard femoral puncture technique to insert a 5 Fr. to 9 Fr. (depending on the largest interventional device to be used during treatment) introducer sheath into the common femoral artery in anterograde or retrograde fashion for contralateral approaches. Heparinize intravenously using the PTA protocol for heparinization.
2. Perform baseline angiography by injecting contrast medium through the introducer sheath or guiding catheter. Obtain images in multiple projections, delineating anatomical variations and morphology of the lesion(s) to be treated.
3. Introduce a 0.014", 0.016", 0.018”, 0.025”, or 0.035" guidewire to the peripheral vasculature via the introducer sheath or guiding catheter. Cross the target lesion with the guidewire.

Note:
As an alternate method of recanalization, laser ablation can be used in a step-by-step manner where the guidewire and then a laser catheter are sequentially advanced and activated (mm by mm) until the occlusion or stenosis is crossed.
4. Size the laser catheter appropriately:
|
Catheter
Size |
Vessel
Diameter and Location |
|||
|
Reference Vessel Normal |
Total Occlusion or Bend ³30° |
|||
|
Distal |
Proximal |
Distal |
Proximal |
|
|
0.9 mm 110-001, 110-003 |
³1.5 mm |
³1.5 mm |
Cannot be determined |
³1.5 mm |
|
0.9 mm 110-002 |
³2.0 mm |
³2.0 mm |
Cannot be determined |
³2.0 mm |
|
1.4 mm |
³2.0 mm |
³2.0 mm |
Cannot be determined |
³2.2 mm |
|
1.7 mm |
³2.3 mm |
³2.3 mm |
Cannot be determined |
³2.5 mm |
|
2.0 mm |
³2.6 mm |
³2.6 mm |
Cannot be determined |
³3.0 mm |
|
2.2 mm |
³2.9 mm |
³2.9 mm |
Cannot be determined |
³3.2 mm |
|
2.3 mm |
³2.9 mm |
³2.9 mm |
Cannot be determined |
³3.2 mm |
|
2.5 mm |
³3.1 mm |
³3.1 mm |
Cannot be determined |
³3.5 mm |
5. Inject 5-10cc of heparinized saline solution through the laser catheter to flush the guidewire lumen. Attach a rotating hemostatic valve to the guidewire port into the guidewire lumen (See Figure 3). Introduce the distal tip of the Spectranetics laser catheter over the selected guidewire. Under fluoroscopic control, guide the laser catheter to the lesion. The laser catheter’s radiopaque band marker indicates its position relative to the lesion.


Figure 3 (not to scale)
6. Inject contrast medium solution through the introducer sheath or guiding catheter to verify the positioning of the laser catheter under fluoroscopy.
7. Following confirmation of the laser catheter’s position in contact with the target lesion and using normal saline or Lactated Ringer’s solution:
a. Flush all residual contrast media from the introducer sheath or guide catheter and in-line connectors,
b. Flush all residual contrast media from the lasing site and vascular structures adjacent to the lasing site, prior to activating the CVX-300® laser system.
c. Please refer to the Saline Infusion Protocol section of this Instructions for Use and perform saline flush and infusion per the instructions.
8. Depress the footswitch, activating the CVX-300®, and slowly, less than 1 mm per second, advance the laser catheter allowing the laser energy to remove the desired material. Release the footswitch to deactivate the CVX-300®.

Note
Advancing the laser catheter through moderately calcified lesions may require more pulses of laser energy than fibrous atherosclerotic tissue.
9. Pull back the laser catheter and inject contrast medium through the guiding catheter and examine the lesion via fluoroscopy.
10. Repeat steps 6 through 9 as needed to complete treatment.

Note
If the laser catheter is removed from the vessel for any reason, thoroughly clean the laser catheter outer surface, inner lumen, and tip with heparinized saline to prevent blood from sticking. Blood remaining on the laser catheter may diminish the efficiency of the laser catheter.
11. There is no need to remove the laser catheter from the patient in order to increase or decrease either the fluence or pulse repetition rate; as the laser catheter was previously calibrated. Refer to the CVX-300® Excimer Laser System Operator’s Manual, 7030-0035 or 7030-0068.

Caution
All patients should be monitored for blood pressure and heart rate during the procedure.
12. Following laser atherectomy, perform follow up angiography and balloon angioplasty if needed.
13. Recommended pharmacology follow up to be prescribed by the physician.
EXCIMER
LASER SALINE INFUSION PROTOCOL
Note
This technique requires two operators. It is recommended that the primary physician operator advance the laser catheter and operate the laser system foot pedal. A scrub assistant should manage the saline infusion control syringe and (if appropriate) depress the fluoroscopy pedal.
A. Before
the laser procedure, warm a 500cc bag of 0.9% normal saline (NaCl) to 37°C. It
is not necessary to add heparin or potassium to the saline solution. Connect
the bag of warmed saline to a sterile intravenous line and terminate the line
at a port on a triple manifold.
B. If
applicable, cannulate the ostium of the artery with an appropriate “large
lumen” guide catheter in the usual fashion. It is recommended that the guide
catheter not have side holes.
C. Under
fluoroscopic guidance, advance the laser catheter into contact with the lesion.
If necessary, inject contrast to help position the tip of the laser catheter.
If contrast appears to have become entrapped between the laser catheter tip and the lesion, the laser catheter may be retracted slightly (1-2mm) to allow antegrade
flow and contrast removal while flushing the system with saline. However, before lasing, ensure that the
laser catheter tip is in contact with the lesion.
D.
Expel any residual contrast from
the control syringe back into the contrast bottle. Clear the triple manifold of
contrast by drawing up saline through the manifold into the control syringe.
E. Remove
the original control syringe from the manifold and replace it with a fresh 20cc
luer-lock control syringe. This new 20cc control syringe should be primed with
saline prior to connection to reduce the chance for introducing air bubbles.
(Merit Medical and other vendors manufacture 20cc control syringes.)
F. Flush
all traces of blood and contrast from the manifold, connector tubing,
y-connector, and introducer sheath or guide catheter, with at least 20-30cc of
saline (several syringes of saline). When this initial flushing is completed,
refill the 20cc control syringe with saline.
G. Under
fluoroscopy, confirm that the tip of the laser catheter is in contact with the
lesion (advance the laser catheter if necessary), but do not inject contrast.
H. When
the primary operator indicates that he/she is ready to activate the laser
system, the scrub assistant should turn the manifold stopcock off to pressure
and inject 10cc of saline as rapidly as possible (within 1-2 seconds). This
bolus injection is to displace and/or dilute blood down to the level of the
capillaries and limit back-bleeding of blood into the laser ablation field.
I. After
the injection of the initial 10cc bolus and without stopping the motion of
injection, the scrub assistant should next slow down the rate of injection to
2-3cc/second. This portion of the saline infusion is to displace and/or dilute
the antegrade blood flow entering the laser ablation field. At the instant the scrub assistant slows
down the injection rate, the primary operator should activate the laser system by
depressing the foot pedal and begin a lasing sequence.
J. Terminate
the saline injection at the end of the lasing train. Turn the manifold stopcock
back to pressure and refill the control syringe with 20cc of saline in
preparation for the next lasing sequence.
K. Each
subsequent laser train should be preceded by a bolus of saline and performed
with continuous saline infusion as described in steps H-J.
L. If
contrast is used to assess treatment results during the course of a laser
treatment, repeat steps D - G prior to
reactivation of the laser system (before activating the laser as described in
steps H – J).

Note
Depending on which approach is used, antegrade or contralateral, saline can be administered through the sheath (antegrade approach) or laser catheter inner lumen (contralateral approach). When the contralateral approach is used, smaller diameter guidewires are suggested to allow adequate saline infusion at the treatment site.
The company's standard one-year product
warranty and remedy are exclusive and expressly in lieu of all other warranties
expressed or implied either in fact or by operation of law, statutory or
otherwise, including warranties of merchant ability and fitness for use or for
any particular purpose and of all other liabilities or obligations on the part
of the company relating in any way to the CVX-300® Excimer Laser
System, whether arising from personal injury, property damage or otherwise. The
company neither assumes nor authorizes any other person to assume for it any
other liability in connection with the sale, installation, service or use of
the CVX-300® Excimer Laser System. Notwithstanding the generality of
the foregoing, (a) the company shall have no liability whatsoever for special,
consequential, incidental or punitive damages of any kind arising out of the
sale, installation, service or use of the CVX-300® Excimer Laser
System, and (b) the company's liability shall in no event exceed the original
purchase price of the CVX-300® Excimer Laser System.
![]()
Spectranetics Corporation 96
Talamine Court Colorado
Springs, Colorado 80907-5186 USA Telephone 719-633-8333 Spectranetics
International BV Plesmanstraat
6 3833
LA Leusden The
Netherlands Telephone 31-33-434-7050