
(You are encouraged to copy and distribute this information)
Issued: May 10, 2005
This is to remind radiology personnel and physicians that serious injury or death can occur when patients with implanted neurological stimulators undergo MRI procedures, and to recommend preventive actions.
Background
The FDA has received several reports of serious injury, including coma and permanent neurological impairment, in patients with implanted neurological stimulators who underwent magnetic resonance imaging (MRI) procedures. The mechanism for these adverse events is likely to involve heating of the electrodes at the end of the leadwires, resulting in injury to the surrounding tissue. Although these reports involved deep brain stimulators and vagus nerve stimulators, similar injuries could be caused by any type of implanted neurological stimulator, such as spinal cord stimulators, peripheral nerve stimulators, and neuromuscular stimulators.
Recommendations
If you are a physician who implants or monitors patients with implanted neurological stimulators:
If you are a radiologist or health care professional who uses MRI equipment:
Reporting Adverse Events
FDA requires hospitals and other user facilities to report deaths and serious injuries associated with the use of medical devices. If you suspect that a reportable adverse event has occurred involving a patient with an implanted device who has undergone an MRI procedure, you should follow the reporting procedure established by your facility.
We also encourage you to report adverse events related to MRI and medical devices that do not meet the requirements for mandatory reporting. You can report these directly to the device manufacturer. You can also report these events to MedWatch, the FDA’s voluntary reporting program. You may submit reports to MedWatch by phone at 1-800-FDA-1088; by FAX at 1-800-FDA-0178; by mail to MedWatch, Food and Drug Administration, 5600 Fishers Lane, Rockville, MD 20857-9787; or online at http://www.fda.gov/medwatch/report.htm.
Getting More Information
If you have questions about this notification, please contact Nancy Pressly, Office of Surveillance and Biometrics (HFZ-510), 1350 Piccard Drive, Rockville, Maryland, 20850, Fax at 240-276-3356, or by e-mail at phann@cdrh.fda.gov. You may also leave a voice mail message at 240-276-3357 and we will return your call as soon as possible.
FDA medical device Public Health Notifications are available on the Internet at http://www.fda.gov/cdrh/safety.html. You can also be notified through email on the day the safety notification is released by subscribing to our list server. To subscribe, visit: http://service.govdelivery.com/service/subscribe.html?code=USFDACDRH_10.
Sincerely yours,
|
Updated May 10, 2005
![]()
CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page
Center for Devices and Radiological Health / CDRH