INTRODUCTION
Coverage
QUALITY SYSTEM
Personnel Training
DESIGN AND DEVELOPMENT PLANNING
Interface
Structure of Plans
DESIGN INPUT
Input Checklists
DESIGN REVIEW
Combination Devices
Preparation For Reviews
Why Design Reviews
Types Of Design Review Meetings
Design Review Requirements
End Of Initial Design
DESIGN OUTPUT
Documenting Design Output
Acceptance Criteria
Design Output Approval
DESIGN VERIFICATION AND VALIDATION
Design Evaluation versus Specifications
Software Validation
Labeling Verification
DESIGN TRANSFER
DESIGN CHANGES
DESIGN HISTORY FILE
EXHIBITS
Design Input Requirements Procedure
The Safe Medical Devices Act of 1990 added design validation requirements
to the GMP requirements in section 520(f) of The Act. Section
820.30 of the Quality System (QS) regulation lists the design
control requirements that manufacturers should satisfy to be in
compliance. This chapter describes design controls and provides
guidance to assist manufacturers in complying with design control
requirements.
"Design Control Guidance for Medical Device Manufacturers"
is another document that may assist manufacturers in understanding
the intent of the design control requirements. This manual interprets
the language of the QS regulation and explains the underlying
concepts in practical terms. "Do It By Design: An Introduction
to Human Factors in Medical Devices" is a document that contains
background information about human factors as a discipline, describes
and illustrates device problems and discusses human factors principles
and methods as a part of the design control system. Both of these
manuals are possible resources for manufacturers who are either
developing or improving their design control system. These manuals
are also available through DSMA.
Coverage
The design controls section 820.30 of the QS regulation applies
to the design of products, and processes and changes to existing
designs and processes. Changes to existing designs should be made
in accordance with design control requirement even if the original
design was not subject to these requirements. Design controls
are not retroactive to completed portions of ongoing design programs.
Each manufacturer of any class III or class II device, and class
I devices automated with computer software and those listed below
shall establish and maintain procedures to control the design
of the device in order to make certain that specified design requirements
are met. Manufacturers of other Class I devices should develop
and document their devices under their own design control system
because the documentation is needed to help meet the device master
record requirements in 820.181 and marketing submission requirements.
Thus, manufacturers of exempt Class I devices are encouraged to
use 820.30, Design Controls, as guidance.
| Classification Section | Class I Devices Subject to Design Controls Listed in Paragraph 820.30(a)(2) |
| 868.6810 | Catheter, Tracheobronchial Suction |
| 878.4460 | Glove, Surgeon's |
| 880.6760 | Restraint, Protective |
| 892.5650 | System, Applicator, Radionuclide, Manual |
| 892.5740 | Source, Radionuclide Teletherapy |
| All Sect. | Devices automated with computer software |
The design requirements for the device are primarily specified
by the manufacturer; however, FDA has a few design requirements
in the 21 CFR Part 801 labeling regulations and in Parts 1000-1050
which cover radiological and electronic products. A few of the
FDA design requirements are in standards. For example, some parameters
for medical gloves are in standards by the American Society for
Testing and Materials (ASTM). (That is, medical gloves are required
to meet these standards in order to be substantially equivalent
to gloves already in commercial distribution.)
Each manufacturer is required to establish and maintain a quality
system that is appropriate for the specific medical device(s)
designed or manufactured [820.5 and 820.1(a)(3)], and that meets
the requirements of Part 820. Therefore, the details of design
control systems will vary depending on the complexity of the product
or process being designed. However, all non-exempt manufacturers
including very small manufacturers and manufacturers that design
less complex devices or processes are expected to define, document
and implement design control procedures and other quality system
procedures as called for in the regulation. One of these, a sample
design input procedure, is exhibited at the end of this chapter.
Manufacturers may establish one design control procedure to cover
the various design control sections in 820.30; or, they may use
one or more procedures for each topic. Multiple procedures may
be easier to develop, update and implement. Medium to large manufacturers
may have several additional procedures to support their main design
control procedures. Design control procedures may be part of the
quality system records (QSR) noted in section 820.186.
Personnel Training
Personnel training in 820.25 is one of the quality system requirements,
which applies to employees that perform any activity covered by
the QS regulation including all design activities.
Manufacturers are required to establish procedures for identifying
training needs and making certain that all personnel are trained
to adequately perform their assigned responsibilities. Design
personnel shall be made aware of device defects which may occur
from the improper performance of their specific jobs. In particular,
personnel who perform verification and validation activities shall
be made aware of defects and errors that may be encountered as
part of their job functions.
Most technical employees need various degrees of training, as
appropriate, in the medical device regulations, safety, labeling,
human factors, verification, validation, design review techniques,
etc.
DESIGN AND DEVELOPMENT
PLANNING
Developing a new device and introducing it into production are
very complex tasks. For many new devices and associated manufacturing
processes that use software, these tasks are further complicated
because of the importance of software, and the possibility of
subtle software errors. Without thorough planning, program control,
and design reviews, these tasks are virtually impossible to accomplish
without errors or leaving important aspects undone. The planning
exercise and execution of the plans are complex because of the
many areas and activities that should be covered. Some of the
key activities are:
To support thorough planning, the QS regulation requires each
manufacturer to establish and maintain plans that describe or
reference the design and development activities and define responsibility
for implementation.
The plans should be consistent with the remainder of the design
controls. For example, the design controls section of the quality
system requires a design history file (DHF) [820.30(j)] that contains
or references the records necessary to demonstrate that the design
was developed in accordance with the:
1. approved design plan, and
2. regulatory requirements.
Thus, the design control plans should agree with, and require
meeting, the quality system design control requirements. One of
the first elements in each design plan should be how you plan
to meet each of the design control requirements for the specific
design you plan to develop; that is, the design plans should support
all of the required design control activities. Such plans may
reference the quality system procedures for design controls in
order to reduce the amount of writing and to assure agreement.
Interface
Design And Development Planning section 820.30(b) states:
"The plans shall identify and describe the interfaces with
different groups or activities that provide, or result in, input
to the design and development process..."
If a specific design requires support by contractors such as developing
molds, performing a special verification test, clinical trials,
etc., then such activities should be included or referenced in
the plan and proactively implemented in order to meet the interface
and general quality system requirements. Of course, the interface
and general requirements also apply to needed interaction with
manufacturing, marketing, quality assurance, servicing or other
internal functions.
Proactive interface is a important aspect of concurrent engineering.
Concurrent engineering is the process of concurrently, to the
maximum feasible extent, developing the product and the manufacturing
processes. This valuable technique for reducing problems, cost
reduction and time saving cannot work without proactive interface
between all involved parties throughout all stages of the development
and initial production program.
Structure of Plans
Each design control plan should be broad and complete rather than
detailed and complete. The plan should include all major activities
and assignments such as responsibility for developing and verifying
the power supplies rather than detailing responsibility for selecting
the power cords, fuseholders and transformers. Broad plans are:
Over the years, several manufacturers have failed to follow this
advice and opted for writing detailed design control procedures.
They reported being unable to finish writing the over-detailed
procedures and were unable to implement them.
Regardless of the effort in developing plans, they usually need
updating as the development activities dictate. Thus, the QS regulation
requires in 820.30(a) that the plans shall be reviewed, updated,
and approved as the design and development evolves. The details
of updating are left to the manufacturer; however, the design
review meetings are a good time and place to consider, discuss
and review changes that may need to be made in the design development
plan.
Design input means the physical and performance requirements of
a device that are used as a basis for device design [820.3(f)].
Section 820.30(c) Design Input, requires that each manufacturer
shall establish and maintain procedures to make certain that the
design requirements relating to a device are appropriate and address
the intended use of the device, including the needs of the user
and patient. Also, a design requirement in 820.130 requires that
each manufacturer shall make certain that device packaging and
shipping containers are designed and constructed to protect the
device from alteration or damage during the customary conditions
of processing, storage, handling, and distribution. The intent
of 820.130 is to add the broad conditions that are considered
for a package design. Packaging design activities should be done
according to design controls. Likewise, the design of the content
and physical parameters of labeling are covered by design controls.
Manufacturers that are exempt from design controls shall labeling
and packaging specifications in the DMR (820.181) and are encouraged
to use the QS design controls as guidance.
The input procedures shall address incomplete, ambiguous, or conflicting
requirements. The design input requirements shall be documented
and shall be reviewed and approved by a designated individual(s).
The approval, including the date and signature of the individual(s)
approving the requirements, shall be documented.
Under a design control system, manufacturers should identify device
requirements during the design input phase or beginning of the
design activity. Design input includes determining customer needs,
expectations and requirements plus determining regulatory, standards,
and other appropriate requirements. These various requirements
are documented by the manufacturer in a set of device requirements.
A set of design input requirements, when converted to engineering
terminology, finalized and accepted as part of the device master
record is called a device or product specification.
The design input phase usually is a continuum because intensive
and formal input requirements activities usually occur near the
beginning of the feasibility phase and continue to the early physical
design activities. After the initial design input phase there
are also intensive and formal activities to reduce the input requirements
to engineering-type input specifications -- usually called a product
or device specification.
At the opposite end of the design program, the last event is initial
production which may be pilot production or the beginning of routine
production. Whether a manufacturer starts with pilot or routine
production depends on the nature of the new device and associated
production. Pilot devices may be distributed after design validation
of initial units is completed if they meet all of the device master
record and other GMP requirements. Some manufacturers, however,
use the pilot models in training programs for technical writers,
production and service personnel, etc. Pilot models are also commonly
used in early marketing displays.
After the concept of the new device design is established, the
following basic design input questions should have been answered:
1. What is the real need for the new device?
2. Where will the new device be used?
3. Who will use the new device?
4. How will the new device be used?
5. With what devices will the new device be used?
6. How long will the new device be used? and
7. Other questions related to the specific device to be developed.
Designing a device and verifying that it meets customer requirements
are expensive and time consuming activities. Therefore, to control
these activities and increase the probability of achieving desired
safety and performance characteristics, device, software, and
process requirements and specifications should be thoroughly reviewed
and approved before physical design and development begins. As
the design evolves, the hardware, software, packaging, labeling,
etc., shall be verified [820.30(f)] and reviewed [820.30(e)] versus
their latest specifications to verify that design input requirements
have been met.
Input Checklists
Device requirements should identify all of the desired performance,
physical, safety and compatibility characteristics of the proposed
device and, ultimately, the finished device. Design input also
includes requirements for labeling, packaging, manufacturing,
installation, maintenance and servicing. The final device specifications
should cover ALL of the device characteristics. The device specifications
may incorporate other specifications by reference such as reference
to the manufacturer's list of specifications for a type of device,
to specific paragraphs in standards, or to all of a standard,
etc. with respect to a referenced specification. It should be
very clear exactly what is going to be met. A failure to properly
address characteristics or factors such as immunity from transients
in the power source, thermal stress, electromagnetic compatibility
(EMC), packaging protection, shipping stability, proper maintenance,
etc., can have disastrous consequences.
It is possible to diligently develop device requirements and still
forget one or more elements in the final specification. Hopefully,
no key factors will be left out. To reduce the probability of
a requirement or characteristic being left out, a specification
checklist(s) may be used during the design input phase. A checklist
should be developed that is broad based but also germane to the
product line of the manufacturer. If used, a checklist should
be part of a standard operating procedure such as a Design Input
Specification Procedure.
The input requirements should cover any standards that the manufacturer
plans for the device to meet. In the United States, information
about essentially all national and international standards may
be obtained from the American National Standards Association (ANSI),
11 West 42nd Street, New York, New York, 10036, phone 212-642-4900.
ANSI is a private organization, which monitors most of the standards
activity in the United States and foreign activity in which U.S.
citizens "officially" participate. Thus, ANSI can supply
addresses and other information about all well established standards
writing groups. Also, ANSI has for sale many different types of
standards including quality system standards. For example, the
International Electrotech Commission has a draft design review
standard, "Guide on Formal Design Review" (plus a supplement),
which should be helpful to product assurance/design control personnel.
The QS regulation requires that the input procedures shall address
incomplete, ambiguous, or conflicting requirements. Thus, every
reasonable effort should made to collect all of the requirements
from which the designers can generate detailed design specifications
that are clear, correct and complete.
At the end of the major aspects of the design input stage, the
design input requirements shall be documented and shall be reviewed
and approved by a designated individual(s). The approval, including
the date and signature of the individual(s) approving the requirements,
shall be documented.
A documented device specification or set of specifications derived
from the input requirements should exist at the beginning of the
physical design project. The device and other related specifications
should be kept current as the design of the device,
packaging, labeling and manufacturing processes evolve during
the development program. As the physical design evolves, the specifications
usually become more specific and more detailed.
The device specification will undergo changes and reviews as the
device design evolves. However, one goal of market research and
initial design reviews is to establish complete device requirements
and specifications that will minimize subsequent changes.
Old versions of the input requirements and later the input specifications
are put in the design history file (DHF) or indexed in the computer
as part of the DHF to help show that the design plan was followed.
Design review [820.30(e)] is one of the key design control elements
in a quality system. The objectives of design review are stated
in the definition of design review in 820.3(h) as follows:
Design review means a documented, comprehensive, systematic examination
of a design to evaluate the adequacy of the design requirements,
to evaluate the capability of the design to meet these requirements,
and to identify problems.
To meet the systematic design review requirement, device design
and design reviews should progress through defined and planned
phases starting with the design input phase and continuing through
validation of initial production units or lots. Subsequent activities
are usually design changes.
To meet the design review comprehensive requirement, assessments
should include a formal review of the main device and subsystems,
including accessories, components, software, labeling, and packaging;
production and resource needs; and installation and service, if
needed. The scope includes performance, physical safety, compatibility
with other devices, overall device system requirements, human
factors, and environmental compatibility.
Even though users or medical practitioners will be aware of direct
medical requirements, they may not be fully aware of physical
safety, compatibility, system, human factors, and environmental
requirements. Thus, the reviews of the design input and the design
should extend beyond merely satisfying user-stated requirements
in order to assure that safety and effectiveness goals are met.
As the development program progresses, the reviews should cover
producibility and production documentation such as assembly drawings,
manufacturing instructions, test specifications, test procedures,
etc.
The extent and frequency of design reviews depends on the complexity
and significance of the device being evaluated.
When the design program is a redesign of an existing device, a
special effort should be made to assure that data obtained from
previous failures, complaints, and service records are made available
and reviewed by those responsible for design, design input and
design review.
Combination Devices
Marketing submissions to FDA for drug delivery, drug coated, etc.,
devices are required to have appropriate data that supports combination
claims. The verification of combination devices requires interaction
between device, drug or other manufacturers. Records of this interaction,
such as design review meeting minutes, are required in order to
meet the interface requirements of 820.30(b), Design and Development
Planning. The labeling and particularly the cross-labeling of
combination devices should be carefully analyzed during verification
and validation activities, and design review meetings.
Preparation For Reviews
The designated moderator or other designated employee should announce
the formal review meetings with appropriate lead time and include
an agenda.
Persons who are making presentations should prepare and distribute
information to help clarify review issues and help expedite the
review. However, the intent of the quality system is not that
presentations be so formal and elaborate that designers are spending
excessive time on presentations rather than on designing a safe
and effective device.
Persons who plan to attend a review meeting should come prepared
to discuss the key issues on the agenda and issues related to
the current design phase. Design review meetings are a great educational
forum. However, design review meetings should not be used as a
primary tool to educate or bring new employees or unprepared employees
up-to-speed. To do so detracts from the intent of the meeting
and detracts from the intent of the GMP requirements. Obviously,
design review is also an excellent educational tool. However,
new, or new-to-the-project employees should be primarily oriented
by other means that do not detract from the primary function of
design review meetings.
Why Design Reviews
Design reviews are conducted for design definition, selection
and adequacy; communication; and resolution of problems and issues.
For example, the design review of the design input requirements
and subsequent design input specifications for the device, labeling,
packaging and accessories is performed to help select the best
and/or needed characteristics and requirements, usually from among
many available and sometimes conflicting inputs.
The design review of the initial requirements allows input from
all parties. Various people may participate and "buy in"
or "become part of the program." As the design input
and review activities progress, any conflicts are resolved and
the preliminary specifications for the device, accessories, labeling,
and packaging are established. Herein, the device, accessories,
labeling and packaging is called the device system. Because of
the establishment of these input requirements and subsequent specifications,
plus interface and communication during the reviews, all personnel
are directed toward the goal of developing the "exact"
same device system.
As the development progresses and the design and production processes
evolve, design reviews reduce errors, help avoid problems, help
find existing problems, help propose solutions, increase producibility
and reduce production transfer problems. The relentless inquiry
during design reviews will expose needed design input requirements
and/or design corrections that otherwise may have been overlooked.
Throughout the design program and particularly toward the end
of the development cycle, design reviews help assure that the
final design of the device system meets the current design requirements
and specifications.
Types Of Design Review Meetings
Design review meetings may be grouped into two levels such as:
Some of the review meetings need to be total or major program
review meetings because this is the only type of review meeting
that will satisfy all of the GMP review requirements, particularly
the interface requirement for interaction between or among different
organizational groups. However, sub-program, team and contractor
review meetings are design review meetings, are subject to quality
system design controls, and should be conducted in a manner that
meets the GMP requirements. Sub-program or team meetings are encouraged
as these can be very effective and efficient in reviewing and
resolving sub-program issues.
The records of total program and team meetings are part of the
device design history file. The team review records or a summary
of team records and the current design documentation are to be
available, as appropriate, at total program review meetings.
Design review meetings are called under two scenarios:
The preplanned design review meetings and ad hoc meetings are
part of the planning and interaction that are required in 820.30(b),
Design and Development Planning. That is, the manufacturer
should expect, plan for, and encourage appropriate ad hoc meetings
as well as the major design review meetings. Reasonable notes
and copies of significant engineering documents discussed during
total device system, ad hoc, contractor, and other review meetings
are part of the device design history file.
Design Review Requirements
The objectives of design review are stated in the definition noted
above. How these objectives are to be achieved are presented in
the design review requirements. The main design review requirements
are in 820.30(e) of the QS regulation as follows:
Each manufacturer shall establish and maintain procedures to ensure
that formal documented reviews of the design results are planned
and conducted at appropriate stages of the device's design development.
The procedures shall ensure that participants at each design review
include representatives of all functions concerned with the design
stage being reviewed and an individual(s) who does not have direct
responsibility for the design stage being reviewed, as well as
any specialists needed. The results of a design review, including
identification of the design, the date, and the individual(s)
performing the review, shall be documented in the design history
file.
There are four requirements related to design reviews:
1. The meetings should be formal. That is, key attendees are designated
and the meetings are conducted at least once per stage/phase,
are planned, are announced or are periodic, have an appropriate
agenda, notes are recorded, etc., according to the manufacturer
procedure for design reviews.
The design review procedure should be broad and complete in that
it contains information about all of the requirements. However,
the procedure should not be so detailed that it cannot be followed.
Over the years, several manufacturers have failed to follow this
advice, tried to write detailed design QA procedures, and have
reported that they were unable to finish writing the over-detailed
procedures and were unable to implement them.
2. To meet the definition of design review in 820.3(h), the review
should include persons who are intimately knowledgeable about
the technical characteristics of the design such as performance,
safety, compatibility, etc. In many manufacturers this can only
be done by those persons responsible for the design. However,
reviews are to be objective, unbiased examinations by appropriately
trained personnel which should include an individual(s) not responsible
for the design. The moderator of the review meeting may be one
of the persons not responsible for the design.
To meet interface and other review requirements, the review meetings
should, as appropriate, include representatives of R&D, Engineering,
Technical Support Services, Production Engineering, Manufacturing,
Quality Assurance, Marketing, Installation and Servicing, Purchasing
and contractors. Design review should, as applicable and at the
appropriate phase, include those responsible for coordinating
or managing preclinical and clinical studies.
3. Pre- and post-review meeting significant responsibilities and
assignments should be documented [820.30(b)]. These assignments
are not unusual -- they are simply ordinary work required to develop
a new product or modify an existing product. The progress and/or
results of such assignments would typically be reported at the
next review meeting. Documentation is not required for detailed
day-to-day development activities that are part of the designers
routine job.
4. The design review meeting results are made a part of the device
design history file. The results should include minutes and should
include notes, or annotated draft drawings and annotated draft
procedures that played a significant role during the design review.
Such documents help show that plans were followed, verification/validation
was reviewed, and, to some extent, how the design evolved.
The QS regulation does not require that every document mentioned,
referenced or used during a design review be placed in the design
history file.
The device design review meeting minutes should include information
such as:
Manufacturers may use a form to capture some of this information
for minutes such the device, date, moderator, attendees, major
phase, problems, assignments, etc. The device design review minutes
are a key and required part of the design history file. The minutes
also help consolidate development information and the current
minutes are also a brief record of some of the immediate development
tasks to be done.
End Of Initial Design
The design control requirements, particularly design validation, give clear insight into when the initial
design effort is completed. The end of the total design
effort has not been reached until it is known that the initial
production devices, when transferred to production and produced
per the device master record, meet all of the current design specifications.
This fact can only be determined by performing design validation
on one or more samples of the finished production units as required
by 820.30(g). Initial production and subsequent validation are
well defined stages; and, therefore, design review(s) shall be
performed as required by 820.30(e), Design Review.
Thus the design validation of initial production should be followed
by a "final" design review to meet the design review
requirement. If the validation of the final design and subsequent
design review(s) reveal design problems, then design changes
are required to correct these problems. Design changes require
another design verification and, where appropriate, validation
and review of all parts or the affected parts of the device system.
Design output per 820.3(g) means the results of a design effort at each design phase and at the end of the total design effort. The finished design output is the basis for the device master record. The total finished design output consists of the device, its packaging and labeling, and the device master record.
Device master record (DMR) means a compilation of records containing
the procedures and specifications for a finished device.
The design output at each phase are documents and physical design
elements that are either complete or are used to move the design
effort into the next phase. For example, the first design output
will usually be the design requirements document. From the requirements
and their engineering knowledge, the designers will derive the
preliminary design specifications. Then the physical design begins.
For example, the designers may begin the selection of known routine
components that are part of the design and begin documenting their
purchasing and acceptance requirements documented to meet 820.50
Purchasing Controls, (b) Purchasing Data which requires that each
manufacturer shall establish and maintain data that clearly describe
or reference the specified requirements, including quality requirements,
for purchased or otherwise received product and services.
Other components will be selected as the design evolves. The design
output for some special or new components, or components in unusual
applications, will include verification protocols, purchasing
and acceptance requirements.
Many of the design output documents are documents that directly
form part of the DMR. The remaining DMR documents are created
by quality assurance, production engineering, process engineering,
technical writing, installation and servicing, etc., using design
output data and information. For example, the finished device
final-test methods and some installation and/or servicing test
methods and data forms may be derived from the design verification
protocol(s). When all of these design and documentation activities
are completed, the DMR is complete. When the DMR is complete and
initial production units, including packaging, meets all specifications,
the total finished design output exists.
To generate the design output per the QS regulation in 820.30(d),
three activities are required. Each of these is listed and discussed
below.
1. Each manufacturer shall establish and maintain procedures for
defining and documenting design output in terms that allow an
adequate evaluation of conformance to design input requirements.
2. Design output procedures shall contain or make reference to
acceptance criteria and ensure that those design outputs that
are essential for the proper functioning of the device are identified.
3. Design output shall be documented, reviewed, and approved before
release. The approval, including the date and signature of the
individual(s) approving the output, shall be documented.
Documenting Design Output (1)
Documenting design output in terms that allow an adequate evaluation
of conformance to design input requirements is a significant requirement
and design activity. A common technique for achieving this conformance
is listed below.
Each of these documents has a different drawing number but the
line/paragraph numbers are the same. The first of these documents
may be used as the beginning format for the next one. Therefore,
it is almost impossible to leave out an element. Thereafter, when
the verification is performed and documented, conformance or lack
of conformance from input to output is known.
Acceptance Criteria (2)
The verification documents and data contain more information than
is typically needed for production evaluation and acceptance of
components, in-process items and finished devices. Therefore,
it is easy to copy and modify verification documents to meet the
quality system requirement that: design output procedures shall
contain or make reference to acceptance criteria and ensure that
those design outputs that are essential for the proper functioning
of the device are identified. In fact, this technique of deriving
test procedures from the verification protocols also yields the
test method(s) and data form(s) needed to meet the DMR requirements
for QA procedures and acceptance criteria in 820.181(c).
Design Output Approval (3)
The third and final output requirement is that: design output
shall be documented, reviewed, and approved before release. The
approval, including the date and signature of the individual(s)
approving the output, shall be documented. This means that:
Design output also includes the physical design which, of course,
is not intended to be signed, and dated. The approval for the
physical design is the validation that is done on initial production
units.
DESIGN VERIFICATION AND VALIDATION
Each manufacturer shall establish and maintain procedures for
verifying the device design. Design verification [820.30(f)] shall
confirm that the design output meets the design input requirements.
The results of the design verification, including identification
of the design, method(s), the date, and the individual(s) performing
the verification, shall be documented in the DHF.
Validation [820.30(g)] means confirmation by examination and provision
of objective evidence that the particular requirements for a specific
intended use can be consistently fulfilled.
Process validation means establishing by objective evidence that
a process consistently produces a result or product meeting its
predetermined specifications.
Design validation means establishing by objective evidence that
device specifications conform with user needs and intended use(s).
Verification means confirmation by examination and provision of
objective evidence that specified requirements have been fulfilled.
Each manufacturer shall establish and maintain procedures for
validating the device design. Design validation shall be performed
under defined operating conditions on initial production units,
lots, or batches, or their equivalents. Design validation shall
ensure that devices conform to defined user needs and intended
uses and shall include testing of production units under actual
or simulated use conditions. Design validation shall include software
validation and risk analysis, where appropriate. The results of
the design validation, including identification of the design,
method(s), the date, and the individual(s) performing the validation,
shall be documented in the DHF.
Design verification is always done versus specifications. Therefore,
to control the specifications and increase the probability of
achieving desired safety and performance characteristics, device,
software, labeling, packaging and any other specifications should
be complete and thoroughly reviewed before development commences.
As the hardware and software designs evolve, they should be evaluated
versus their current specifications.
Verification and validation should be done with test equipment
calibrated and controlled according to quality system requirements.
Otherwise, there is limited confidence in the data.
Verification and validation should also be done according to a
written protocol(s). The protocol(s) should include defined conditions
for the testing. The protocol(s) should be approved before being
used. Test protocol(s) are not perfect for a design, particularly
a new design. Therefore, the designers and other verification
personnel carefully annotate any ongoing changes to a protocol.
Likewise, the verification personnel should record technical comments
about any deviations or other events that occurred during the
testing. The slightest problem should not be ignored. During design
reviews, the comments, notes and deviations may be as important
as test data from the formal protocol(s).
Design Evaluation versus Specifications
The original design of devices and any subsequent changes should
be verified by appropriate and formal laboratory, animal, and
in vitro testing. Risk analysis should be conducted to identify
possible hazards associated with the design. Failure Mode Effects
Analysis and Fault Tree Analysis are examples of risk analysis
techniques.
Appropriate laboratory and animal testing followed by analysis
of the results should be carefully performed before clinical testing
or commercial distribution of the devices. The manufacturer should
be assured that the design is safe and effective to the extent
that can be determined by various scientific tests and analysis
before clinical testing on humans or use by humans. For example,
the electrical, thermal, mechanical, chemical, radiation, etc.,
safety of devices usually can be determined by laboratory tests.
Clinical testing is not needed for many substantially equivalent
devices (See 21 CFR Part 807 Subpart E Premarket Notification
Procedure). Where it is needed, such as for complex substantially
equivalent devices or new devices, clinical testing on humans
should meet the applicable requirements in the Investigational
Device Exemption (IDE) regulations (21 CFR Parts 812 and 813).
The general IDE regulation (21 CFR Part 812) exempts a manufacturer
during the "premarketing phase" from the following provisions
of the FD&C Act:
Don't be misled by this list of exemptions -- being exempted from
these provisions does not mean that a manufacturer may develop
a new device under uncontrolled conditions and then test it on
humans. Devices being clinically tested are not exempt from section
501(c) of the FD&C Act, which states that a device is adulterated
if it does not meet a manufacturer's quality claims. Devices being
manufactured for use in clinical studies under an IDE are exempt
ONLY from the production section of the QS regulation. They are
not exempt from design controls listed in 820.30. In addition,
the IDE regulation has labeling requirements in 812.5 and quality
assurance requirements in 812.20(b)(3) that shall be met. Further,
manufacturers should remember that human subjects are also protected
through the courts via product liability laws and actions. In
summation, protection of manufacturer interests, human test subjects,
practitioners, and patients requires that all medical devices
be developed, evaluated, and manufactured under a total quality
system.
Laboratory testing to force a failure takes considerable time
and the "culprit" may not fail during the testing. Another
evaluation technique is Failure Mode and Effects Analysis (FMEA)
in which failures are assumed to occur. FMEA is useful for evaluating
reliability, safety, and general quality where, for example, the
evaluator assumes that:
The probability of each failure actually occurring and, if it
does, the resulting effect are analyzed. Then, where needed and
feasible, hazards and faulty performance are designed out of the
device or reduced; or compensated or prevented/reduced by interlocks,
warning signs, explicit instructions, alarms, etc. Risks, of course,
cannot always be removed from medical devices, but they should
be known and controlled to the extent feasible with existing technology.
Failure Mode and Effects Analysis (FMEA) is a very powerful and
costeffective technique. Note that it takes very little
time to assume that a component or subsystem is going to fail
versus the time required to test to failure. The idea is not to
promote one method above the other because a reasonable amount
of both actual testing and failure mode and effects analysis should
be done before a device is clinically tested and/or placed into
production.
Besides using FMEA there are also other human factor and validation process techniques that can be used in developing an overall risk analysis. These techniques include: timelines, workload analysis, failure analysis, alternative calculations, testing including animal testing, auditing the design output, design reviews, demonstrations, and comparing a new design to a proven design etc. The users should be considered components when developing a fault tree and failure mode effects analysis.
All evaluation results should be reviewed by product development
personnel who compare the tests and FMEA results with specifications,
including safety and performance standards, to make certain that
the desired level of intrinsic quality has been designed into
the device. Also, the appropriate design of manufacturing processes,
including validation where appropriate, is needed to assure that
production can achieve the level of quality designed into the
device.
Software Validation
Software is evaluated and reviewed versus the software specifications
during the ongoing development of the device design. When a "final"
prototype(s) is available, the software and hardware are validated
to make certain manufacturer specifications for the device and
process are met. Some aspects of hardware evaluation were discussed
above. Aspects specific to software are covered below.
Before testing the software in actual use, the detailed code should
be visually reviewed versus flow charts and specifications. All
cases, especially decision points and error/limit handling, should
be reviewed and the results documented.
In all cases, algorithms should be checked for accuracy. Recalls
have occurred because algorithms were incorrectly copied from
a source and, in other cases, because the source algorithm was
incorrect. During the development phase, complex algorithms may
need to be checked by using a test subroutine program written
in a highorder language, if the operational program is written
in a lowlevel language.
The validation program is planned and executed such that all relevant
elements of the software and hardware are exercised and evaluated.
The testing of software usually involves the use of an emulator
and should include testing of the software in the finished device.
The testing includes normal operation of the complete device;
and this phase of the validation program may be completed first
to make certain that the device meets the fundamental performance,
safety and labeling specifications. Concurrently or afterward,
the combined system of hardware and software should be challengedwith abnormal inputs and conditions. As appropriate, these inputs
and conditions include such items as:
The results of the software and combined device system validation are included in the design reviews.
Labeling Verification
During verification, the complete device is exercised such that
all labeling, displays, and outputs are generated, reviewed, and
the results documented. During the verification, all displayed
prompts and instructions are checked versus the manufacturer's
and FDA's labeling requirements and versus the operator manual.
Printed labeling and screen displays should be checked to see
if they are directed to the user and not to the system designers,
which is a common fault found in labeling. Displayed text should
be short and to the point. Because displays are brief, keywords
should be carefully selected to match system characteristics,
yet transfer the maximum information to the user. The text of
references to controls or other parts of the system should match
the labeling on the device. Data, identifications, or other key
information displayed should be current, complete, unambiguous,
and accurate.
During verification, all prompts and instructions should be followed
exactly by the device test or other operators and such action
should result in correct operation of the device. Prompts and
instructions should appropriately match the instructions in the
operator's manual. The evaluation should include verification
that any screen or other displays meet the requirements of, and
have been approved per, the manufacturer's policy/procedure for
design of labeling.
Patient and procedure data on printouts should be correct; therefore,
printouts should undergo a verification similar to that performed
for the screen or other displays. In addition, the printouts should
be evaluated with respect to their "cold" information
transfer characteristics. Will the printouts be quickly and clearly
understood a few weeks later when the reader is not reading the
displays, operating the device, or looking at the patient? All
printouts should also meet the manufacturer's design control policy/procedure
requirements for labeling. Likewise, patient data or other key
information transmitted to a remote location should be correct;
therefore, it should be checked for accuracy, completeness, and
identification. Annotated copies of verified labeling, printouts,
etc. and associated notes and any checklists should be placed
in the design history file.
The overall device specifications usually have requirements that
cover user/operator error prevention and control. Along with operator
training, such errors are controlled by:
Also, for some devices, it may be important to control the order
in which data can be entered by the operator. In emergency situations
or because of distractions, it may be important to present the
operator with a brief history or status report of recent actions.
During the verification, the listed items should be evaluated
versus the specifications, and checked for completeness and appropriateness.
A checklist or matrix may be used to aid in the review of labeling.
The design controls require that each manufacturer shall establish and maintain procedures to ensure
that the device design is correctly translated into production
specifications.
It is common practice for sections of a design to be transferred
before the entire design is completed. The QS regulation does
not prevent such split or multiple transfers. Transfer is to be
performed only for completed elements of the design -- multiple
transfers may not be used to bypass any design, labeling or other
GMP requirements.
A significant part of the transfer requirement is met when the
design output is being created. That is, some of the design output
documents are part of the DMR and are used directly for production.
The remaining DMR documents are based on design output information.
A procedure is needed to cover the generation of the remaining
device master record documents based on information in the design
output documents.
Design transfer should assure that the section of the design being
transferred:
Design transfer may include training of production, installation
and service employees and such training should be covered by or
referenced by the transfer procedure.
Changes to a design element are controlled per 820.30(i) Design
Changes which states that: each manufacturer shall establish
and maintain procedures for the identification, documentation,
validation or where appropriate verification, review, and approval
of design changes before their implementation.
The original design activities and subsequent change control activities
for the design are both done under the full set of the
quality system design controls. A manufacturer may not use a design
change control procedure to bypass part of the design controls.
Thus, it is difficult to describe change control before design
transfer because both activities are done under design controls.
Most of the details of the change control system are left to the
manufacturer to develop, document and implement. As the design
activity progresses toward the final stage, it is expected that
the degree of change control will increase.
Those elements of the design that have been verified and accepted
obviously should be under change control. A design that has been
submitted to FDA for marketing clearance should be under change
control. A design undergoing clinical trials should be under change
control or the clinical data may not be accepted by FDA. A design
that is released for production should be under design
and general change control.
After design activities are begun and the physical design evolves
into an accepted entity, subsequent changes to the device specification(s)
are proposed, evaluated, reviewed, approved, and documented per
all of 820.30. The revised specification(s) becomes the current
design goal in accordance with the manufacturer procedures for:
design control, design change control, and document control.
Design history file (DHF) means a compilation of records which
describes the design history of a finished device [820.3(e)].
The DHF covers the design activities used to develop the device,
accessories, major components, labeling, packaging and production
processes.
The design controls in 820.30(j) require that each manufacturer
shall establish and maintain a DHF for each type of device. Each
type of device means a device or family of devices that are manufactured
according to one DMR. That is, if the variations in the family
of devices are simple enough that they can be handled by minor
variations on the drawings, then only one DMR exists. It is common
practice to identify device variations on drawings by dash numbers.
For this case, only one DHF could exist because only one set of
related design documentation exists. Documents are never created
just to go into the DHF.
The QS regulation also requires that the DHF shall contain or
reference the records necessary to demonstrate that the design
was developed in accordance with the approved design plan and
the requirements of this part. As noted, this requirement cannot
be met unless the manufacturer develops and maintains plans that
meet the design control requirements. The plans and subsequent
updates should be part of the DHF. In addition, the QS regulation
specifically requires that:
the results of a design review, including identification of the
design, the date, and the individual(s) performing the review,
shall be documented in the DHF.
design verification shall confirm that the design output meets
the design input requirements. The results of the design verification,
including identification of the design, method(s), the date, and
the individual(s) performing the verification, shall be documented
in the DHF.
Typical documents that may be in, or referenced in, a DHF are
listed below:
The DHF contains documents such as the design plans and input
requirements, preliminary input specs, validation data and preliminary
versions of key DMR documents. These are needed to show that plans
were created, followed and specifications were met.
The DHF is not required to contain all design documents or to
contain the DMR, however, it will contain historical versions
of key DMR documents that show how the design evolved.
Does the DHF have value for the manufacturer? Yes, when problems
occur during redesign and for new designs, the DHF has the
"institutional" memory of previous design activities.
The DHF also contains valuable verification and validation protocols
that are not in DMR. This information may be very valuable in
helping to solve a problem; pointing to the correct direction
to solve a problem; or, most important, preventing the manufacturer
from repeating an already tried and foundtobeuseless
design.
Design Input Requirements Procedure
A sample Design Input Requirements procedure is presented which
covers basic activities for obtaining data on requirements that
is needed for employees to develop device specifications. This
procedure uses the multiple specification approach; however, a
single combined specification would use a very similar procedure.
This procedure should be modified to meet specific needs before
being adopted by a manufacturer.
C O M P A N Y L O G O
Title: Design Input Requirements Procedure SOP #: ___________ Page:__1_of_2_____
Prepared by:__________________________________ App: ____________ Date:____________
Prep. Date: __________________________________Rev: ____________ Date:____________
ECN History:_____________________________________________________________________
_________________________________________________________________________________
POLICY - Design specifications covering all design requirements
shall be established for all proposed devices before any significant
physical design activities are started.
SCOPE - This policy applies to all devices and accessories
developed by the manufacturer or developed by a contractor for
us. For purchase of completed designs, refer to SOP ####. The
device specification(s) must exist or be generated regardless
of the source of the design.
CONFIDENTIALITY - Device development plans and activities
are always confidential. Market research reports and documents
such as specifications with parameter data shall be marked confidential.
Design control procedures, standard SOPs, blank forms, and required
design review and design verification/validation records may be
shown to, and may be copied by, FDA investigators as required
by the QS regulation. Design parameters are not covered by the
QS regulation. Therefore, confidential specification characteristics
and parameters in the copies of these documents shall be blacked
out unless the document is being collected during an inspection
related to a marketing submission.
RESPONSIBILITY
Marketing and Engineering have the primary responsibility for
determining safety and performance requirements and developing
input specifications; however, all departments are expected to
support the development of input requirements and subsequent specifications.
MARKETING - Marketing shall plan and conduct all customer
contacts to obtain information on customer desires, needs, expected
pricing, opinions about existing devices, etc.
To the maximum extent feasible, market research shall be conducted
in a manner to reduce leaking of manufacturer confidential information
and plans.
Design review meetings shall normally precede and follow all significant
outside market research activities. Initial market research activities
shall be previewed with top management.
Market research results are to be documented and marked confidential.
PRODUCTION - Production has primary responsibility for
assuring producibility and establishing manufacturing requirements.
Some of these requirements may be general during the early design
stages.
ENGINEERING - Engineering is expected to supply design
input information on most requirements. Such inputs may parallel
data obtained by market research.
Engineering has primary responsibility for specifying what technology to use.
Engineering shall analyze input data on requirements and reduce it to preliminary specifications.
Engineering has primary responsibility for addressing incomplete,
ambiguous, or conflicting requirements and shall see that such
issues are appropriately discussed at design reviews.
Page 2 of 2
RA & QA - RA and QA managers or their designees shall
attend all design input or specification review meetings to provide
input on, and to assure that, regulatory, manufacturer, quality,
safety, performance, etc., procedures are followed and that requirements
are met.
SPECIFICATIONS
STRUCTURE - Multiple specifications shall be used except
for very simple devices. A separate specification shall be developed
for accessories, labeling, packaging, etc. An overall device specification
shall be developed and shall include an index that points to supporting
specifications. The specifications, among other factors, shall
address:
1. Performance and Efficacy;
2. Human Factors;
3. Chemical Safety;
4. Electrical Safety;
5. Mechanical Safety;
6. Radiation Safety;
7. Thermal Safety;
8. Biocompatibility;
9. Device Compatibility;
10. System Compatibility;
11. Environmental Compatibility;
12. Packaging (in a separate specification document);
13. Any FDA design requirements in the Part 801 and Part 1000-1050
regulations; and
14. Labeling in a separate document and, as appropriate, in the
device primary specification.
CHECKLISTS - Checklists of requirements germane to our
product line may be used to develop and support specifications.
If used, such checklists become part of this procedure
and part of the design documentation.
DESIGN REVIEW - Each specification shall undergo
design review before it is approved for physical design activities
or is used as a background document to support further market
research. Such reviews shall be documented.
APPROVAL - The Marketing manager and Engineering manager
shall approve all input specifications after these have
been subjected to design review.
DOCUMENTATION - The approved specifications shall be given
document numbers and become part of the device master record for
the new device.
CHANGE CONTROL - The Engineering manager shall decide when
design activities have progressed to the stage that the various
specifications shall be subject to our Design Change Control Procedure.
However, for our organization, design change control can start
NO later than the FIRST of the following events:
- clearance of a 510(k), or
- start of a clinical investigation.
Updated January 1, 1997
![]()