PMA Final Decisions Rendered for July 2005

related information

Some of the documents on this section of the CDRH Website are available only in PDF format and may not be accessible to those with certain disabilities. If you cannot access the documents you are interested in, please use the accessibility link for assistance. Below are Premarket Approvals (PMA), Product Development Protocols (PDP), Supplement and Notice Decisions. This list is generated on a monthly basis.

A PDF document that contains the "Approval letter and Summary of Safety and Effectiveness" is being added to this listing for each PMA. The PMA number will appear as a link if this document is available. Information about the PDF Reader is available.

PMA Original Approvals                  

P030004
7/21/05

Onyx® Liquid Embolic System (LES)

Micro Therapeutics, Inc.
Irvine, CA
92618

Approval for the Onyx Liquid Embolic System (LES). The device is indicated for presurgical embolization of brain arteriovenous malformations.


PMA Supplemental Approvals

N12159/S020
7/22/05
180-Day

SURGICEL™/ SURGICEL™ NU-KNIT/ SURGICEL™ Fibrillar Absorbable Hemostat

Johnson & Johnson Wound Management Worldwide
A Div of ETHICON, Inc.
Somerville, NJ
08876

Approval for a sterilization site located at Steris Isomedix Services, Whippany, New Jersey.

P830079/S016
7/22/05
180-Day

INSTAT™ Collagen Hemostat

Johnson & Johnson Wound Management Worldwide
A Div of ETHICON, Inc.
Somerville, NJ
08876

Approval for a sterilization site located at Steris Isomedix Services, Whippany, New Jersey.

P870038/S005
7/20/05
Real-Time

Starr-Edwards Silastic Ball Prosthesis, Models 1260 & 6120

Edwards Lifesciences
Irvine, CA
92614

Approval for an alternate supplier of polyethylene tetrafluoroethylene (PTFE) thread, namely W.L. Gore, and for a change from multifilament PTFE thread to monofilament PTFE thread used in the final sewing of the cloth to complete the sewing ring of the Starr-Edwards Silastic Ball Prosthesis, Models 1260 & 6120.

P880047/S009
7/22/05
180-Day

INTERCEED™ (TC7) Absorbable Barrier

GYNECARE, A Div. Of ETHICON, Inc.
Somerville, NJ
08876

Approval for a sterilization site located at Steris Isomedix Services, Whippany, New Jersey.

P890003/S082
7/26/05
180-Day

Medtronic 2020A CardioSight Reader and the Model 2491 Device Data Management Application (DDMA) Software

Medtronic, Inc.
Minneapolis, MN
55432

Approval for the Medtronic 2020A CardioSight Reader and the Model 2491 Device Data Management Application (DDMA) Software for use in the transfer of patient and device data.

P890003/S087
7/18/05
Real-Time

Medtronic Model 2490G CareLink Monitors and Model 2491 DDMA Software for InSync Sentry Model 7299, InSync Maximo Model 7304, and the V-V timing feature for both the InSync Sentry Model 7297 and the InSync Maximo Model 7303

Medtronic, Inc.
Minneapolis, MN
55432

Approval for the CareLink Monitor Model 2490G for use with additional pulse generator models, specifically the InSync Sentry Model 7299, InSync Maximo Model 7304, and the V-V timing feature for both the InSync Sentry Model 7297 and the InSync Maximo Model 7303. The CareLink system is indicated for use by a patient to interrogate their implanted device, and transfer the data to a private, secure server by use of a standard (analog) telephone connection.

P910066/S021
7/19/05
Real-Time

SpinaLogic 1000 Bone Growth Stimulator

dj Orthopedics, LLC
Vista, CA
92083

Approval for the described coil modifications including geometric changes, turn schedule changes, and manufacturing changes. The device, as modified, will be marketed under the trade name SpinaLogic 1000 Bone Growth Stimulator and is indicated as an adjunct electrical treatment to primary lumbar spinal fusion surgery for one or two levels.

P950020/S010
7/18/05
180-Day

Flextome Cutting Balloon™ Monorail (MR)/Over-the-Wire (OTW)

Boston Scientific IVT
San Diego, CA
92123

Approval for design, manufacturing, and labeling changes in the Cutting Balloon, both Over-the-Wire (OTW) and Monorail (MR) configurations. The device, as modified, will be marketed under the trade name Flextome Cutting Balloon™ and is indicated for dilatation of stenoses in coronary arteries for the purpose of improving myocardial perfusion in those circumstances where a high pressure balloon resistant lesion is encountered.

P950027/S006
7/19/05
Real-Time

Hyalgan®

Fidia Farmaceutical, S.p.A.
Via Ponte Della Fabbrica, Italy

Approval for modifying the specifications for the active ingredient (sodium hyaluronate) by eliminating the abnormal toxicity test and replacing the rabbit pyrogen test with the bacterial endotoxins test (Limulus Amoebocyte Lysate (LAL) test) and modifying the specifications for the finished product to replace the rabbit pyrogen test with the LAL test.

P960040/S066
7/29/05
Real-Time

VENTAK PRIZM 2 DR AICD Model 1861

Guidant Corporation CRM
St. Paul, MN
55112

Approval to replace the existing Polyimide tubing that provides insulation in the pulse generator header assembly with Polyetheretherke-tone (PEEK) tubing.

P960040/S067
7/27/05
Real-Time

VITALITY HE Models T180 and T182

Guidant Corporation CRM
St. Paul, MN
55112

Approval for 1) An alternate magnetic reed switch manufactured by Hermetic and an associated change to the equipment used in the rework of the component; and 2) Programmer Software Application Model 2845 v2.8.

P960058/S045
7/25/05
180-Day

HiResolution Bionic Ear System

Advanced Bionics Corporation
Sylmar, CA
91342

Approval for an updated Auria PowerPak assembly configuration.

P960058/S046
7/14/05
Real-Time

HiResolution Bionic Ear System

Advanced Bionics Corporation
Sylmar, CA
91342

Approval for a new Auria earhook, the iConnect, to provide an ear-level, cable-free interface for third-party miniature FM receivers.

P970003/S050
7/15/05
Panel

VNS Therapy System

Cyberonics, Inc.
Houston, TX
77058

Approval for the VNS Therapy System. The device is indicated for the adjunctive long-term treatment of chronic or recurrent depression for patients 18 years of age or older who are experiencing a major depressive episode and have not had an adequate response to four or more adequate antidepressant treatments.

P970015/S029
7/22/05
Real-Time

LT-Cage® Lumbar Tapered Fusion Device

Sofamor Danek USA
Memphis, TN
38132

Approval for the addition of the 14mm x 26mm cage titanium LT-CAGE® Lumbar Tapered Fusion Device. The device remains indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level. LT-CAGE® devices are to be used with autogenous bone graft and implanted via a laparoscopic or an open anterior approach. Patients receiving the LT-CAGE® Lumbar Tapered Fusion Device should have had at least six months of nonoperative treatment prior to treatment with the LT-CAGE® device.

P970015/S031
7/27/05
180-Day

INTER FIX™ Threaded Fusion Device

Medtronic Sofamor Danek USA
Memphis, TN
38132

Approval for the post approval study. The device remains indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level. INTER FIX™ implants and INTER FIX™ RP implants are to be used with autogenous bone graft and implanted via an open anterior approach.

P980016/S048
7/26/05
180-Day

Medtronic 2020A CardioSight Reader and the Model 2491 Device Data Management Application (DDMA) Software

Medtronic 2020A CardioSight Reader and the Model 2491 Device Data Management Application (DDMA) Software

Approval for the Medtronic 2020A CardioSight Reader and the Model 2491 Device Data Management Application (DDMA) Software for use in the transfer of patient and device data.

P980022/S011
7/18/05
Panel

Guardian RT

Medtronic MiniMed
Northridge, CA
91325

Approval for the Guardian RT. The device is indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin in adults (ages 18 and older) with diabetes mellitus for the purpose of improving diabetes management. It alerts if a glucose level falls below or rises above preset values. Values are not intended to be used directly for making therapy adjustments, but rather to provide an indication of when a finger stick may be required. All therapy adjustments would be based on measurements obtained using a home glucose monitor and not on Guardian values. Guardian RT provides real-time glucose values that allow users to track patterns in glucose concentrations and to possibly identify episodes of low and high blood glucose episodes. It also stores the data so that it can be analyzed to track patterns. Glucose data can be further downloaded to PC software for analysis of historical glucose values.

P980023/S019
7/18/05
180-Day

Kentrox Family of Implantable Cardioverter Defibrillator Leads, RV, SL, RV-S, SLS

BIOTRONIK, Inc.
Lake Oswego, OR
97035

Approval for modifications to the Kentrox family of Implantable Cardioverter Defibrillator (ICD) leads. The devices, as modified, will be marketed under the trade names Kentrox RV, Kentrox RV Steroid, Kentrox SL, Kentrox SL Steroid, Kentrox RV-S, Kentrox RV-S Steroid, and Kentrox SLS Steroid and are indicated for use in conjunction with a BIOTRONIK ICD.

P980035/S044
7/26/05
Real-Time

Kappa 700/650/600, Kappa 900/800, and EnPulse E1/E2 Implantable Pulse Generators

Medtronic, Inc.
Minneapolis, MN
55432

Approval for design modifications to the IPG reed switch.

P990066/S020
7/1/05
180-Day

Senographe DS & 2000D Full Field Digital Mammography System

GE Healthcare
Waukesha, WI
53188

Approval for software changes to the Senographe DS and (2000D) FFDM Acquition WorkStation (AWS) and Review WorkStation to allow users to select the Premium View (PV) algorithm as primary image processing.

P990071/S007
7/12/05
180-Day

Stockert 70 RF Generator for Cardiac Ablation

Biosense Webster, Inc.
Diamond Bar, CA
91765

Approval for replacing Software Version 1.034B with version 1.035. The device, as modified, will be marketed under the trade name Stockert 70 RF Generator for Cardiac Ablation and is indicated for use in conjunction with a compatible catheter for cardiac ablation procedures.

P990075/S016
7/27/05
Real-Time

Saline-Filled Mammary Prosthesis and Spectrum® Post-Operatively Adjustable Saline-Filled Mammary Prosthesis

Mentor Corporation
Santa Barbara, CA
93111

Approval for 1) new Style 2000 Moderate Plus Profile, which is considered a modification to the existing product ine, with sizes 175cc-800cc; 2) additional sizes of 400cc, 450cc, 500cc, 600cc, 650cc and 800cc for Style 1600; and 3) additional sizes of 350cc, 400cc, 700cc, 770cc, and 800cc for Style 3000. There are no changes to the indications for use.

P000025/S019
7/25/05
Real-Time

COMBI 40+ System with C40+ or PULSARCI100 Cochlear Implants

MED-El Corporation
Durham, NC
27713

Approval for the COMBI 40+ System with C40+ or PULSARCI100 Cochlear Implants to update non-essential labeling and a manual for troubleshooting FM systems which may be used ith the TEMPO+.

P000040/S005
7/12/05
Special

HTA™ System

Boston Scientific Corporation
Marlborough, MA
01752

Approval for manufacturing process and quality control changes to the adapters used to connect a hysteroscope to the HTA™ System Procedure Sheath.

P010012/S057
7/27/05
Real-Time

CONTAK RENEWAL Model H135; CONTAK RENEWAL 3 Models H170 and H175; CONTAK RENEWAL 3 HE Models H177 and H179 and Programmer Software Application Model 2845 v.2.8

Guidant Corporation CRM
St. Paul, MN
55112

Approval for 1) An alternate magnetic reed switch manufactured by Hermetic and an associated change to the equipment used in the rework of the component; and 2) Programmer Software Application Model 2845 v2.8.

P010013/S014
7/22/05
Real-Time

NovaSure® Impedance Controlled Endometrial Ablation System

Cytyc Surgical Products
Palo Alto, CA
94303

Approval for a modification to the device performance protocol used to establish the shelf life of the NovaSure Impedance Controlled Endometrial Ablation System. The protocol used to assess package integrity, i.e., PPQ-012, previously approved is unchanged.)

P010031/S020
7/26/05
180-Day

Medtronic 2020A CardioSight Reader and the Model 2491 Device Data Management Application (DDMA) Software

Medtronic, Inc.
Minneapolis, MN
55432

Approval for the Medtronic 2020A CardioSight Reader and the Model 2491 Device Data Management Application (DDMA) Software for use in the transfer of patient and device data.

P010031/S024
7/18/05
Real-Time

Medtronic Model 2490G CareLink Monitors and Model 2491 DDMA Software for InSync Sentry Model 7299, InSync Maximo Model 7304, and the V-V timing feature for both the InSync Sentry Model 7297 and the InSync Maximo Model 7303

Medtronic, Inc.
Minneapolis, MN
55432

Approval for the CareLink Monitor Model 2490G for use with additional pulse generator models, specifically the InSync Sentry Model 7299, InSync Maximo Model 7304, and the V-V timing feature for both the InSync Sentry Model 7297 and the InSync Maximo Model 7303. The CareLink system is indicated for use by a patient to interrogate their implanted device, and transfer the data to a private, secure server by use of a standard (analog) telephone connection.

P010055/S001
7/7/05
180-Day

ProstaLund® CoreTherm™ System Microwave Thermotherapy for BPH

ProstaLund Operations AB
Lund, Sweden
SE-226 60

Approval for technological modifications to the CoreTherm™ System to comply with the new edition of IEC 6061-1-2 “Medical electrical equipment – Part1: general requirements for safety; electromagnetic compatibility requirements and tests” (2nd edition, 2001).

P020014/S009
7/12/05
180-Day

Essure™ System

Conceptus, Inc.
San Carlos, CA
94070

Approval for modification of device labeling to include 4- and 5-year effectiveness results and other labeling changes.

P030006/S006
7/22/05
Real-Time

Celsion Prolieve™ Thermodilatation System

Celsion Corporation
Columbia, MD
21064

Approval for use of a new computer motherboard to replace an obsolete motherboard currently used in the Prolieve™ Thermodilatation System.

P030010/S001
7/7/05
180-Day

Siemens Mammomat NovationDR Full Field Digital Mammography System

Siemens Medical Solutions, Inc.
Malvern, PA
19355

Approval to change the acquisition workstation (AWS) from the currently approved
Unix-based SUN workstation with Solaris 8 operating system to a Fujitsu/Siemens M420PC with Syngo operating system and the proposed changes allowing the software drivers for the detector interface (DRUL) to be Windows XP based.

P030025/S017
7/8/05
135-Day

TAXUS™ Express2 Paclitaxel-Eluting Coronary Stent System

Boston Scientific Corporation
Maple Grove, MN
55311

Approval to change the in-process coating weight acceptance criteria. The device, as modified, will be marketed under the trade name TAXUS™ Express2 Paclitaxel-Eluting Coronary Stent System and is indicated for improving luminal diameter for the treatment of de novo lesions <28 mm in length in native coronary arteries =2.5 to =3.75 mm in diameter.

P030032/S003
7/1/05
180-Day
Hylaform, Hylaform Plus, Nd Captique Genzyme Corporation
Cambridge, MA
02142
Approval of the post-approval study for the device.
P040020/S001
7/29/05
Real-Time
Acrysof ReSTOR Natural Intraocular Lens (IOL) Models SN60D3 and MN60D3 Alcon Research, Ltd.
Fort Worth, TX
76134
Approval for Acrysof ReSTOR Natural Intraocular Lens (IOL) Models SN60D3 and MN60D3 (Level A Modification: addition of chromophore). The device, as modified, will be marketed under the trade name Acrysof ReSTOR Natural IOL and is indicated for primary implantation for the visual correction of aphakia secondary to the removal of a cataractous lens in adult patients with or without presbyopia, who desire near, intermediate and distance vision with increased spectacle independence. The lens is intended to be placed in the capsular bag.

 

30-Day Notices (135 Day Supplement was not required)

P840064/S028
7/29/05

Provisc® Ophthalmic Viscosurgical Device, Provisc® packaged in Duovisc® Viscoelastic System and DisCoVisc™ Ophthalmic Viscosurgical Device.

Alcon Research, Ltd.
Fort Worth, TX
76134

Change in the bacterial master seed storage media used by the vendor.

P860004/S072
7/19/05

Medtronic® SynchroMed® II Implantable Programmable Infusion Pump

Medtronic, Inc.
Minneapolis, MN
55432

Change in quality control tests used on components in the SynchroMed II pump.

P860004/S073
7/19/05

Medtronic® SynchroMed® II Implantable Programmable Infusion Pump

Medtronic, Inc.
Minneapolis, MN
55432

Change in the welding process on the SynchroMed II inner shield.

P880086/S104
7/8/05

Pulse Generator

St. Jude Medical
Sylmar, CA
91342

Removal of an automatic washing machine step in the current washing process of pacemakers.

P890047/S017
7/29/05

Provisc® Ophthalmic Viscosurgical Device, Provisc® packaged in Duovisc® Viscoelastic System and DisCoVisc™ Ophthalmic Viscosurgical Device.

Alcon Research, Ltd.
Fort Worth, TX
76134

Change in the bacterial master seed storage media used by the vendor.

P910073/S051
7/8/05

Reliance Leads

Guidant Corporation
St. Paul, MN
55112

Use of an alternative silicone adhesive in the manufacture of the Reliance family of leads.

P910077/S052
7/8/05

VENTAK MINI

Guidant Corporation
St. Paul, MN
55112

Changes in procedures for cytotoxicity and pyrogenicity testing for inspections of incoming components.

P940031/S035
7/21/05

VIGOR, DISCOVERY MERIDIAN

Guidant Corporation
St. Paul, MN
55112

Change to Leak Test Software Operator Login for Brady Pulse Generators.

P940031/S036
7/8/05

DISCOVERY, MERIDIAN

Guidant Corporation
St. Paul, MN
55112

Changes in procedures for cytotoxicity and pyrogenicity testing for inspections of incoming components.

P940031/S037
7/27/05

Meridian, Discovery

Guidant Corporation
St. Paul, MN
55112

Changes to the software used in the automated setscrew preload verification process.

P940031/S038
7/11/05

DISCOVERY/ MERIDIAN

Guidant Corporation
St. Paul, MN
55112

Change the distance between the accelerometer’s beam and lid.

P960040/S056
7/7/05

Prizm®, Prizm® 2, Prizm® HE, Vitality®, Vitality® AVT, Vitality® DS, Vitality® 2

Guidant Corporation
St. Paul, MN
55112

Removal of a manual audit of an incoming inspection process for IC wafers

P960040/S058
7/21/05

VITALITY ICD Models T125, T127, T135 and VITALITY AVT ICD Model A155

Guidant Corporation
St. Paul, MN
55112

Removal of the Time-Limited ZOOM Programmer Test from the manufacturing process.

P960040/S059
7/21/05

Prizm ICDs Models 1850, 1851, 1852, 1853, 1857, 1858 and Vitality ICDs Models 1870, 1871, 1872

Guidant Corporation
St. Paul, MN
55112

Change in the set screw preload verification process.

P960040/S061
7/21/05

Prizm, Prizm 2, Vitality, Vitality 2, Vitality AVT

Guidant Corporation
St. Paul, MN
55112

Removal of a second cycle of electrical testing for capacitors.

P960040/S062
7/22/05

CONTAK CD family of Cardiac Resynchronization Therapy Defibrillators

Guidant Corporation
St. Paul, MN
55112

Alternate supplier of the Trim Resistor Arrays used in the hybrid builds of the PULSAR MAX,
VENTAK and CONTAK CD families of devices.

P960040/S063
7/8/05

PRIZM,
VENTAK AV, VENTAK VR, VITALITY

Guidant Corporation
St. Paul, MN
55112

Changes in procedures for cytotoxicity and pyrogenicity testing for inspections of incoming components.

P960040/S064
7/27/05
Prizm, Prizm HE, Prizm 2, Vitality, Vitality AVT, Vitality DS, Vitality EL Guidant Corporation
St. Paul, MN
55112
Changes to the software used in the automated setscrew preload verification process.
D970003/S033
7/7/05
Insignia I Entra, Nexus I Entra, Insignia I Ultra Guidant Corporation
St. Paul, MN
55112
Removal of a manual audit of an incoming inspection process for IC wafers
D970003/S034
7/8/05
INSIGNIA 1 Implantable Pacemakers, NEXUS 1 Implantable Pacemakers Guidant Corporation
St. Paul, MN
55112
Change in test software at incoming inspection for integrated circuits used with the INSIGNIA and NEXUS pacemaker family of devices.
D970003/S035
7/21/05
INSIGNIA, PULSAR/ PULSAR MAX, VIRTUS, INTELLIS, NXTDS Guidant Corporation
St. Paul, MN
55112
Change to Leak Test Software Operator Login for Brady Pulse Generators.
D970003/S036
7/22/05
PULSAR MAX family of Implantable pacemakers Guidant Corporation
St. Paul, MN
55112
Alternate supplier of the Trim Resistor Arrays used in the hybrid builds of the PULSAR MAX,
VENTAK and CONTAK CD families of devices.
D970003/S037
7/8/05
INSIGNIA 1 Implantable Pacemakers Guidant Corporation
St. Paul, MN
55112
Change the visual inspection criteria for jumper chip components to view weld melt from the side.
D970003/S038
7/27/05
Pulsar, Pulsar Max, Pulsar Max II, Discovery II, Insignia I Entra, Insignia I Ultra, Insignia I Plus, Nexus I Entra, Intellis II Guidant Corporation
St. Paul, MN
55112
Changes to the software used in the automated setscrew preload verification process.
D970003/S039
7/8/05
DISCOVERY, INSIGNIA, INTELIS II, NEXUS PULSAR Guidant Corporation
St. Paul, MN
55112
Changes in procedures for cytotoxicity and pyrogenicity testing for inspections of incoming components.
D970003/S040
7/11/05
DISCOVERY/ PULSAR MAX/INSIGNIA/ INTELIS Guidant Corporation
St. Paul, MN
55112
Change the distance between the accelerometer’s beam and lid.
D970012/S026
7/15/05
Model 700 Inflatable Penile Prosthesis American Medical Systems
Minnetonka, MN
55343
Move a quality control step (Receiving Inspection Screening Test) to after a cross-linker (catalyst) is premixed into the silicone dispersion by the supplier and received by American Medical Systems.
P980018/S004
7/13/05
HercepTest™ DakoCytomation Denmark A/S
Glostrup, Denmark
DK-2600
Change in the supplier of the reagent for immunization of rabbits and subsequent changes in manufacturing process and quality control procedures.
P000021/S008
7/11/05
Dimension tPSA Flex Reagent Cartridge Dade Behring, Inc.
Newark, DE
19714
Change to prevent loss of reagent due to microbial contamination.
P000033/S002
7/7/05
IntraCoil® Self-Expanding Peripheral Stent Ev3, Inc.
Plymouth, MN
55442
Change in the manufacturing process to the label printing and label verification process.
P000037/S006
7/11/05
On-X® Prosthetic Heart Valve Medical Carbon Research Institute, LLC Change in the cleaning methods used to remove manufacturing materials.
P010012/S048
7/7/05
Contak Renewal® Guidant Corporation
St. Paul, MN
55112
Removal of a manual audit of an incoming inspection process for IC wafers
P010012/S049
7/21/05
Contak CD CRT-D Models H115, H119 and Contak Renewal CRT-D Models H120, H125 Guidant Corporation
St. Paul, MN
55112
Change in the set screw preload verification process.
P010012/S050
7/21/05
Contak Renewal, Contak Renewal 3 RF Guidant Corporation
St. Paul, MN
55112
Removal of a second cycle of electrical testing for capacitors.
P010012/S051
7/22/05
VENTAK family of Implantable Cardioverter Defibrillators Guidant Corporation
St. Paul, MN
55112
Alternate supplier of the Trim Resistor Arrays used in the hybrid builds of the PULSAR MAX,
VENTAK and CONTAK CD families of devices.
P010012/S052
7/8/05
CONTAK CD, CONTAK RENEWAL, EASYTRAK Guidant Corporation
St. Paul, MN
55112
Changes in procedures for cytotoxicity and pyrogenicity testing for inspections of incoming components.
P010012/S053
7/27/05
Contak CD 2, Contak Renewal 3 DS, Contak Renewal 3 HE Guidant Corporation
St. Paul, MN
55112
Changes to the software used in the automated setscrew preload verification process.
P020027/S004
7/13/05
Dimension® FPSA Flex® Reagent Cartridge Dade Behring, Inc.
Newark, DE
19714
Change to prevent loss of reagent due to microbial contamination.
P030005/S009
7/21/05
Vitality AVT ICDs Models A135, A155 Guidant Corporation
St. Paul, MN
55112
Change in the set screw preload verification process.
P030005/S010
7/8/05
CONTAK RENEWAL TR Guidant Corporation
St. Paul, MN
55112
Changes in procedures for cytotoxicity and pyrogenicity testing for inspections of incoming components.
P030026/S007
7/18/05
VITROS Immunodiagnostic Products Anti-HBc IgM Reagent Pack and Calibrators Ortho-Clinical Diagnostics, Inc.
Rochester, NY
14626
Change in the acceptance criteria for an initial manufacturing step.
P040024/S001
7/12/05
Restylane Injectable Gel Medicis
Scottsdale, AZ
85258
Changes to processing methods used in the manufacturing of Restylane Injectable Gel.

Summary of PMA Originals & Supplements Approved
Originals: 1
Supplements: 37

Summary of PMA Originals Under Review
Total Under Review: 85
Total Active: 36
Total On Hold: 49
Number Greater Than 180 Days: 3

Summary of PMA Supplements Under Review
Total Under Review: 272
Total Active: 140
Total On Hold: 132
Number Greater Than 180 Days: 5

Summary of All PMA Submissions Received
Originals: 5
Supplements: 72

Summary of PMA Supplement PMA Approval/Denial Decision Times
Number of Approvals: 37
Number of Denials: 0
Average Days Fr Receipt to Decision (Total Time): 116.2
FDA Time: 102.3 Days MFR Time: 13.9 Days

Updated October 12, 2005