Ophthalmic Devices Panel Advisory Committee Meeting Brief Summary for March 5, 2004 |
VITREORETINAL AND EXTRAOCULAR DEVICES BRANCH (VEDB)
Presented by:
Karen Warburton, M.S., Microbiologist
One of the device types that VEDB reviews is the ophthalmic sponge, which is
used during LASIK surgery. We have recently become aware of MDR (Medical Device
Reports) that identified an association between ophthalmic sponges and diffuse
lamellar keratitis. Testing of a sample of ophthalmic sponges from a lot associated
with a cluster of diffuse lamellar keratitis (DLK) cases showed significantly
higher levels of bacterial endotoxin than a different lot. Additional MDRs have
reported an association between microkeratomes and DLK.
Endotoxin has been shown to cause DLK in a rabbit model and there have been reports in the literature implicating endotoxin from sterilizer water reservoirs as a cause of DLK outbreaks. Additionally, a variety of other etiological factors have been suggested. However, endotoxin-contaminated ophthalmic sponges have not previously been identified as a possible cause. Endotoxin contaminated water used during device manufacture is a potential source. Historically, FDA has not required that ophthalmic sponges or other devices used in LASIK surgery be pyrogen (i.e. endotoxin) free and they are typically not labeled as such (although many may in fact be endotoxin free.)
Our Branch is working with other Center offices to make the ophthalmic community aware of this potential cause of DLK through letters to professional organizations and letters to the editor in journals which we anticipate will be published in the near future. We hope to encourage reporting of DLK to FDA and to stimulate both user and FDA investigation of DLK outbreaks so that we can better understand the role that ophthalmic devices and endotoxin play and make changes in our product review policies if warranted.
Clear Lens Extraction for the Correction of Presbyopia
Clear Lens Extraction (CLE) is an intraocular surgical procedure in which the non-cataractous crystalline lens is removed and replaced with a multifocal or accommodative lens for refractive correction. The panel discussed issues related to the appropriate clinical trial design for the evaluation of a multifocal intraocular lens (MIOL) or accommodative IOL (AIOL) in a cohort of subjects who have undergone CLE to correct presbyopia.
CLE and implantation of an MIOL or AIOL has been performed as an “off label” practice for several years, mainly in eyes with high refractive errors. There are currently no Standards or Guidance available for CLE with an intraocular lens, nor ar there any Standards or Guidance for devices solely intended for the correction of presbyopia.
After a presentation of the issues by Drs. Malvina Eydelman and Joseph Blustein of FDA, the panel was asked to discuss questions specific to the issue. There was no vote taken nor consensus requested of the panel. The purpose of the deliberations was to assist the Agency in working with industry to design clinical trial protocols that would establish the reasonable safety and effectiveness of CLE with implantation of an MIOL or AIOL.
A majority of the panel recommended
Contact: Sara M. Thornton, Executive Secretary
(301) 594-2053, ext.127; e-mail smt@cdrh.fda.gov
TRANSCRIPTS MAY BE PURCHASED FROM: (Written Request Only)
Miller Reporting Company, Inc.
507 C Street, N.E.
Washington, D.C. 20002
(202) 546-6666
or
Food and Drug Administration
Freedom of Information Staff (FOI)
5600 Fishers Lane, HFI-35
Rockville, MD 20852
(301) 827-6500 (voice) (301) 443-1726 (fax)
EXECUTIVE SUMMARY MAY BE PURCHASED FROM:
(Written Request Only)
Food and Drug Administration
Freedom of Information Staff (FOI)
5600 Fishers Lane, HFI-35
Rockville, MD 20852
(301) 827-6500 (voice) (301) 443-1726 (fax)
Back to Advisory Committee Meetings Database
Updated March 10, 2004
![]()
CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page
Center for Devices and Radiological Health / CDRH