Blood
Human Chorionic Gonadotropin (hCG) Assays: What Laboratorians
Should Know about False-Positive Results
(posted November 10, 2005)
(Back to Safety Tips for Laboratorians)
1. What Are the Scope and Purpose of This Communication?
The Food and Drug Administration’s (FDA) Office of In Vitro
Diagnostic Device Evaluation and Safety (OIVD) has developed this
communication in response to reports sent to FDA by manufacturers
and users of hCG assays of adverse events associated with falsely
elevated human chorionic gonadotropin (hCG) results. This phenomenon
has also been described in the scientific literature. hCG false-positive
results have led to errors in patient diagnosis and treatment with
severe clinical consequences including unnecessary interventions
such as chemotherapy and major surgical procedures. In this document,
an hCG false-positive result is defined as the detection of hCG
by immunoassay in the absence of actual hCG in blood.
The purpose of this communication is: a) to remind laboratorians and
clinicians of analytical interfering factors and clinical conditions
that may affect the performance of the hCG immunoassay testing, b) to
provide techniques to identify any suspected interference, and c) to
recommend methods to reduce or eliminate the interfering factors.
2. What Is hCG and What Does hCG
Immunoassay Measure?
hCG is a glycoprotein consisting of two
dissimilar subunits (α and β)
with eight carbohydrate side chains. This combination results in
significant heterogeneity in the hCG structure.1 The
hCG molecule is produced by trophoblast cells of the placenta;
by trophoblast cells in gestational trophoblastic diseases such
as hydatidiform moles, choriocarcinoma,
and placental site trophoblastic tumors; and by trophoblast cells
in testicular germ cell malignancies.1,2
The gonadotropesi in the human
pituitary also synthesize and secrete hCG (known as pituitary hCG),
but this molecular form is slightly different from placental hCG.1 Many nontrophoblastic tumors also express hCG.3
hCG immunoassays measure the presence (qualitative) or amount
(quantitative) of hCG molecule- regular and variant molecular forms-
in blood or urine. All hCG immunoassays are based on the sandwich
principal; at least one antibody binds and immobilizes hCG and
a second antibody raised to a distant epitope and labeled with
an enzyme, dye, or chemilluminescence agent, marks the presence
of hCG or quantifies it. 2,4 Most
hCG assays are designed to primarily detect the regular placental
form of the hCG molecule since regular hCG is considered to be
the key marker for pregnancy.2
i
Gonadotropes are cells in the anterior pituitary
which produce the gonadotropin luteinizing hormone and follicle stimulating
hormone . 3. What Clinical Conditions Are Associated with Detectable
Levels of hCG?
In non-pregnant women, hCG levels are normally undetectable. During
early pregnancy,
the placenta produces hCG and its level in the blood doubles every
two to four days.5 Low levels
of hCG are present following an early, spontaneous abortion. In
gestational trophoblastic diseases, serum hCG is persistently elevated.
However, one to five years preceding malignant gestational
trophoblastic disease, persistently low levels of hCG (e.g., <50
IU/L) may be present.6
Pituitary hCG circulates in low concentrations in men and premenopausal
women, and the concentrations rise in perimenopausal women and
older women.1 A variable proportion of many nontrophoblastic tumors,
such as transitional cell carcinoma of the bladder and urinary
tract, renal cancer, prostate cancer, cancers of the gastrointestinal
system, neuroendocrine tumors, lung cancer, breast cancer, gynecological
cancers, and hematological cancers, also express hCG at various
levels.3
4. What Is the Intended Use of hCG Immunoassays?
hCG immunoassays are intended to measure hCG, a placental hormone,
in blood or urine for the early detection of pregnancy. Other uses
include aiding in the diagnosis, prognosis, management, and treatment
of certain tumors. The level of regulatory oversight of these assays
varies with the intended use of the device. When they are used
for the early detection of pregnancy, they are regulated as Class
II devices. When they are intended for uses other than early detection
of pregnancy, they are regulated as Class III devices.7
FDA has cleared a large number of hCG assays (professional and
home use) for the early detection of pregnancy. The diagnosis and
monitoring of early pregnancy is the only application for hCG assays
that is cleared by FDA although the performance of these assays
has not been established in association with ectopic pregnancy
or molar pregnancy. Even though not approved by FDA, the measurement
of hCG for the diagnosis and monitoring of ectopic pregnancy, molar
pregnancy, gestational trophoblastic diseases1,3,4,8,
and testicular cancer9 is
considered the standard of care in the United States.
5. What Factors Can Cause False-Positive
hCG Assay Results?
Clinically significant hCG false-positive
results are rare but do occur.10 False-positive
results may occur because of analytical interfering factors and/or
device malfunctions. Therefore, it is important that laboratorians,
working in conjunction with physicians who order hCG tests, are
aware of the underlying causes of falsely elevated results so that
they can assist physicians to properly utilize hCG results in the
patient management.
Since device malfunctions have broad potential sources and root
causes, it is outside the scope of this document to address them.
However, the analytical interfering factors are extensively discussed
below.
6. What Are the Analytical Interfering Factors That May
Lead to Falsely Elevated hCG Results?
Some of the analytical interfering factors that may lead to falsely
elevated hCG results include but are not limited to:
- heterophile antibodies, human anti-animal antibodies, rheumatoid
factor, and autoantibodies11
- fibrin clots in serum as a result of incompletely clotted specimens
- interference from other endogenous components in the blood
such as bilirubin, hemoglobin, and lipids12
7. What Are the Endogenous Interfering Antibodies (Heterophile
and Human Anti-Animal Antibodies)?
Heterophile antibodies are produced against poorly defined antigens,
frequently foreign proteins. The general term "heterophile
antibodies" is sometimes used in the literature interchangeably
to refer to heterophile antibodies, human anti-animal antibodies,
rheumatoid factor, and autoantibodies.
Human anti-animal antibodies are circulating human antibodies
reactive to animal proteins.11 Circulating antibodies with specificities
for a wide range of animal immunoglobulins have been reported such
as mouse, rat, rabbit, and others. 11 The most common human anti-animal
antibody interferent is human anti-mouse antibodies ( HAMA ), which
causes both positive and negative interferences in mouse monoclonal
antibody-based assays.11
Circulating heterophile antibodies and anti-animal antibodies
have the potential to interfere with two-site (sandwich) or competitive
immunoassays, such as hCG assays, by cross-linking the capture
and label antibodies in the absence of specific analyte.11,13 The
estimated prevalence of interfering antibodies in the general population
is up to 40% of normal serum samples. 13,14,15
Most modern immunoassays contain nonspecific blocker immunoglobulins
(which originate from the same species as the analyte-specific
antibodies) in order to limit the effect of the interfering antibodies.16 However,
in some instances the blocking proteins can not sufficiently neutralize
the interfering antibodies. Thus, analytical errors may occur.
An individual may acquire these antibodies from a variety of sources
including the use of mouse monoclonal antibodies in diagnostic
imaging and cancer therapy; exposure to microbial antigens; exposure
of veterinarians, farm workers, and food preparers to foreign proteins;
exposure of workers to animals in research laboratories and veterinary
facilities; the presence of domestic animals in the home; autoimmune
diseases which can give rise to autoantibodies such as rheumatoid
factor; blood transfusion; vaccination; and maternal transfer across
placenta to the fetus. 11,13,17
hCG immunoassays may suffer from the interference of heterophile
antibodies/human anti-animal antibodies/autoantibodies which in
turn may lead to false-positive results. 3,18
8. How Can Laboratorians Identify Potentially False-Positive
Results in a Specimen Tested for hCG?
Laboratorians should suspect the occurrence of false-positive
results if at least one of the following events occurs: 1,10,18
- The blood test result is not reproducible on the same or different
assay system.
- The blood test result is not linear after serial dilutions
and rerun of the sample.
- The blood test produces a positive hCG result, but a parallel
urine test produces a negative result. It is important to note
that this criterion can be applied only to blood samples with
hCG ≥ 50 IU/L. The reason is that the sensitivity of urine
tests is approximately 25 IU/L; therefore, if blood hCG values
are less than 50 IU/L, urine levels may be too low to be detected.
- Test result turns negative following treatment of the sample
with a heterophile antibody blocking agent.
- The patient’s clinical presentation does not match the
hCG positive result.
9. Can Results of Different hCG Assay Systems Be Compared?
For assay results to be easily comparable,
they must measure the same molecular form of the analyte and be
calibrated using the same reference material. There are standards
issued by the World Health Organization (WHO) to calibrate immunoassays
for hCG and its subunits3 All of the more recent hCG assays cleared
by FDA are calibrated using one of the WHO standards.
In addition to regular hCG, there are
at least five major variants of the hCG molecule present in serum
and urine samples, plus a β-core
fragment which is only present in urine.1,2 Although all of
these forms are detectable in serum and/or urine samples during
pregnancy, variable detection or lack of detection of cleaved molecules,
free subunits, and fragments is a major cause of inter-assay variation
in hCG results.2,3
Different hCG assays use antibodies to different sites on the
hCG molecule.18 Therefore, different tests may measure very different
combinations of hCG-related molecules. This may not be important
for the detection and monitoring of pregnancy since most hCG assays
are designed to detect primarily placental hCG, which is the predominant
hCG molecule form during pregnancy.18 However, assay differences
in terms of antibody binding sites may create variability in test
results when the tests are used for detection and monitoring of
other conditions such as hydatidiform moles or gestational trophoblastic
diseases where other forms of hCG molecule are predominantly present.18
In addition, the size of hCG molecules circulating in the blood
of individuals may vary as a result of differences in the protein
and carbohydrate structure. 10 Also, some individuals may produce
aberrant forms of hCG that cross-react with the hCG assays. 10
Others may partly break down circulating hCG into non-biologically-active
forms that react differently with the various assay systems.10
In these conditions, other substances than native, biologically
active hCG may be recognized by the assay system, and this can
sometimes account for differences in measurements reported by different
assays.10
10. If an hCG Test Result Does Not Match the Patient’s
Clinical Presentation, What Should Laboratorians Consider
Doing?
If an hCG test result does not match the
patient’s clinical
presentation (which will depend on the assay’s intended use),
it is possible that the test result is falsely elevated due to
analytical interfering factors and/or device malfunction. The laboratory
should investigate the presence of any interfering factors, the
possibility of device malfunction, and verify the test result using
the following guidelines.
The American College of Obstetricians and Gynecologists (ACOG),
the Committee on Gynecologic Practice, published the following
opinion in 200210in order for laboratorians to rule out the presence
of heterophile antibodies and other interfering substances in hCG
immunoassays.
- If the serum value of hCG is ≥50 IU/L, a quantitative or
qualitative hCG urine test can be performed to rule out the presence
of heterophile antibodies. Because heterophile antibodies are
not present in urine, if the urine test is negative and the serum
test is persistently positive, it confirms the serum immunoassay
interference.
- Wide variations between repeat runs of the same assay could
result from the presence of interfering factors. The assay can
be rerun with serial dilutions of the serum. Heterophile antibodies
are directed to reagents in the immunoassay and not to hCG molecules
so their interaction with the hCG curve will not be linear. Lack
of linearity confirms assay interference.
- Assay malfunction due to various inherent assay factors can
result in false-positive results. Repeating the test using a
different assay system will confirm a false-positive result if
the repeat result is negative.
- The serum can be pre-absorbed to remove heterophile antibodies
before performing the hCG assay. The interference is confirmed
if the result becomes negative after pre-absorbing the serum.
- Because of the presence of hCG molecules with different sizes,
aberrant forms of hCG, and non-biologically active forms of hCG
circulating in blood of different individuals, the amount of
true hCG may be measured differently or incorrectly by different
immunoassays. Repeating the hCG test in a different assay system
can detect this problem.
- Patients with evidence of hCG assay interference should be
notified of the risk for recurrent false-positive results. These
patients should be instructed to inform their future health care
providers of this issue, and the information should be recorded
in their medical records.
In addition to the ACOG Committee’s opinion, the following
methods11 may also be applied to reduce or remove the effect of
interfering antibodies in hCG assays after ruling out the possibility
of technical errors and analyzer malfunction:
- Use commercially available heterophile blocking reagents.
- Remove endogenous immunoglobulins by adding endogenous immunoglobulin-free
serum samples to the specimen. For example, use normal mouse
(animal) sera as blocking reagent.
11. If an hCG Test Result Does
Not Match the Patient’s
Clinical Presentation, What Should Physicians Consider Doing?
In cases when an hCG test result does
not match the patient’s
clinical presentation, it is important that the physician gather
all the available information and reassess the patient. The physician
may:
- consider the possibility that some other clinical condition
may be causing an elevated hCG level
- communicate with the laboratory staff about the test result
and ask the laboratory to rule out technical errors, analytical
interfering factors, and device malfunction.
- consider repeating the blood draw and retesting
- consider testing with urine
- review the clinical presentation and consider additional diagnostic
testing and bear in mind that the hCG test result is only one
piece of the diagnostic puzzle.
12. Is It Important That Laboratorians Follow Manufacturer’s
Recommended hCG Assay Instructions?
Yes. It is critical that laboratorians
follow manufacturer’s
assay instructions for use found in the device package insert.
It is important that patient specimens be collected and processed
according to the manufacturers’ recommendations because improper
collection, handling, and preparation of specimens can impact the
accuracy of results. The following general recommendations are
especially useful in order to avoid false-positive results due
to interfering factors:
- Store unused collection tubes and blood specimens according
to the manufacturers’ recommendations.
- Follow manufacturers’ instructions for using collection
tubes with anticoagulants. Some may contain insufficient anticoagulant
and lead to elevated or decreased results.
- Mix the content of tubes properly at the time of blood collection
to prevent incomplete clot formation (serum) and platelet clumping
or clotting (plasma).
- Process specimens according to the tube manufacturer’s
recommendations. Different types of tubes may have different
requirements.
- Use a refrigerated, horizontal centrifuge head for best results.
Use the centrifuge settings recommended by the tube manufacturer.
- Inspect samples for clots, fibrin, particulate matter, and
other debris prior to processing them on an analyzer. Cellular
debris from grossly hemolyzed samples may elevate test results.
- Follow the manufacturers’ recommended calibration and/or
maintenance schedules since analyzer malfunction is one of the
common assay interfering factors that leads to inaccurate results.
13. Reporting to FDA
If there are any questions or concerns
regarding the performance of hCG test method, contact the assay
manufacturer. You should report all occurrences of unusual test
performance to the manufacturer, and you are encouraged to also
report them to FDA. To obtain more information about medical device
reporting you can refer to the FDA’s web site at http://www.fda.gov/cdrh/mdr/index.html.
A. All Reports Should Be Sent to:
Food and Drug Administration
Center for Devices and Radiological
Health
Medical Device Reporting
P.O. Box 3002
Rockville, MD 20847-3002
B. For any Questions or Concerns Regarding the Content
of This Communication, You May Contact:
Ms. Veronica Calvin
Food and Drug Administration
Center for Devices and Radiological
Health
Office of In Vitro Diagnostic Device Evaluation and Safety
2098 Gaither Road, HFZ-440
Rockville, MD 20850
Phone: (240) 276-0443
Fax: (240) 276-0652
Email: veronica.calvin@fda.hhs.gov
References
1Seki K, Matsui H, Sekiya
S. Advances in the Clinical Laboratory Detection of Gestational
Trophoblastic Disease. Clinica Chimica Acta 2004; 349(1-2):1-13.
2 Cole LA, Khanlian SA, Sutton
JM, Davies S, Stephens ND. Hyperglycosylated hCG (Invasive Trophoblast
Antigen, ITA) a Key Antigen for Early Pregnancy Detection. Clinical
Biochemistry 2003; 36:647-655.
3 Stenman UH, Alfthan H, Hotakainen
K. Human Chorionic Gonadotropin in Cancer. Clinical Biochemistry 2004;
37:549-561.
4 Berkowitz RS, Goldstein
DP. Chorionic Tumors. The New England Journal
of Medicine 1996; 335(23):1740-1748.
5 Henry JB. Clinical Diagnosis
and Management by Laboratory Methods. Nineteenth Edition.
Saunders, 1996.
6 Khanlian SA, Smith HO, Cole
LA. Persitent Low Levels of Human Chorionic Gonadotropin: A Premalignant
Gestational Trophoblastic Disease. American Journal of Obstetrics
and Gynecology 2003; 188(5):1254-1259.
7 Code of Federal Regulation
21 Parts 800 to 1299 revised as of April 1, 2005 Food and Drug
Administration 862.1155.
8 Azzazy HME, Romero LF, Hall
L, Ruttle D, Christenson RH. Two-Center Clinical Evaluation of
a New Automated Fluorometric Immunoassay for the Quantitative Analysis
of Total β-Human Chorionic Gonadotropin. Clinical Biochemistry 2003;
36:523-528.
9 Trojan A, Joller-Jemelka
H, Stahel RA, Jacky E, Hersberger M. False-Positive Human Serum
Chorionic Gonadotropin in a Patient with a History of Germ Cell
Cancer. Oncology 2004; 66:336-338.
10 Committee on Gynecologic
Practice Number 278, November 2002. ACOG Committee Opinion Avoiding
Inappropriate Clinical Decisions Based on False-Positive Human
Chorionic Gonadotropin Test Results. International Journal
of Gynecology & Obstetrics 2003; 80:231-233.
11 Kricka LJ. Human Anti-Animal
Antibody Interferences in Immunological Assays. Clinical Chemistry 1999;
45(7):942-956.
12Tietz NW. Clinical
Guide to Laboratory Tests. Second Edition. Saunders, 1990.
13Yeo KTJ, Storm CA, Li
Y, Jayne JE, Brough T, Quinn-Hall KS, Fitzmaurice TF. Performance
of the Enhanced Abbott AxSYM Cardiac Troponin I Reagent in Patients
with Heterophilic Antibodies. Clinica Chimica Acta 2000;
292:13-23.
14 Gibson TN, Hanchard B.
False Positive Troponin I in a Case of Metastatic Small Cell Bronchogenic
Carcinoma Complicated by Pulmonary Thromboembolism. West Indian
Medical Journal 2001; 50(2):171-172.
15 Fleming SM, O’Byrne
L, Finn J, Grimes H, Daly KM. False-Positive Cardiac Troponin I
in a Routine Clinical Population. The American Journal of Cardiology 2002;
89:1212-1215.
16Fitzmaurice TF, Brown
C, Rifai N, Wu AHB, Yeo KTJ. False Increase of Cardiac Troponin
I with Heterophilic Antibodies. Clinical Chemistry 1998;
44(10):2212-2214.
17 Roongsritong C, Warraich
I, Bradley C. Common Causes of Troponin Elevations in the Absence
of Acute Myocardial Infarction – Incidence and Clinical Significance. Chest 2004;
125(5):1877-1884.
18Cole LA, Sutton JM. HCG
Tests in the Management of Gestational Trophoblastic Diseases. Clinical
Obstetrics and Gynecology 2003; 46(3):523-540.
|