Skip NavigationFDA Logo links to FDA home pageCenter for Devices and Radiological Health, U.S. Food and Drug AdministrationHHS Logo links to Department of Health and Human Services website
FDA Home Page | CDRH Home Page | Search | A-Z Index U.S. Food and Drug AdministrationCenter for Devices and Radiological Health Questions?
horizonal rule

NON-INVASIVE BLOOD PRESSURE (NIBP) MONITOR GUIDANCE

Related Information

This document is intended to provide guidance in the preparation of a regulatory submission.
It does not bind the FDA or the regulated industry in any manner.


Office of Device Evaluation
Division of Cardiovascular, Respiratory, and Neurological Devices
Circulatory Support and Prosthetic Devices Group


Document issued on: March 10, 1997


While this guidance document represents a final document, comments and suggestions may be submitted at any time for Agency consideration by writing to:

Charles Ho (HFZ-450)
Food and Drug Administration
9200 Corporate Boulevard
Rockville, Maryland 20850.

For questions regarding the use or interpretation of this guidance, contact Charles Ho, at 240-276-4080 or by electronic mail at charles.ho@fda.hhs.gov.



U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES
Public Health Service
Food and Drug Administration
Center for Devices and Radiological Health









Scope

This guidance is intended to aid in the preparation or review of premarket notification (510(k)) applications for some of the devices regulated under:

CFR Section: 21 CFR §870.1130, Noninvasive Blood Pressure Measurement System
Class: II
Panel: Circulatory System Devices Panel (74)
Product Code: DXN

This guidance applies to non-invasive blood pressure (NIBP) monitors covered by the ANSI/AAMI SP10-1992 standard for electronic or automated sphygmomanometers (SP10 standard). Included in the SP10 standard are automated NIBP monitors which measure blood pressure at the arm, wrist, or finger of the patient using a standard oscillometric measurement method.

This guidance does not apply to NIBP monitors excluded by the SP10 standard, and those which use a non-oscillometric (or non-standard oscillometric) measurement method. Despite this limitation, the information contained in this guidance may be helpful to any NIBP monitor application.

This guidance is complementary to the requirements of 21 CFR § 807.87. Other information not identified in this guidance may be required in a 510(k) application. This guidance is subordinate to all other applicable statutes, regulations, and policies.

Recommended Information and Testing

1. Device Description

The description should include sufficient information to define the design, capabilities, and function of the device, and the scope of the 510(k) submission. Minimal information includes:

If the 510(k) application is for a modification to an existing device, the manufacturer should provide the specifications for the original device and a detailed and complete description of the similarities and differences between the two versions of the device. Lastly, the manufacturer should provide a table comparing the predicate device to the new device for the items listed above.

2. In-Vitro and Clinical Performance Testing

Substantial equivalence can be demonstrated by showing either 1) sufficient comparison testing with a legally-marketed predicate device, 2) conformance to the SP10 standard, or 3) conformance to any foreign or domestic standard which meets or exceeds the requirements of the SP10 standard.

Comparison Testing

It is strongly recommended that substantial equivalence be demonstrated by showing conformance to the SP10 standard. However, if the manufacturer chooses to provide comparative testing, the provided data should meet the General Requirements (listed below) and account for the following:

SP10 Standard Testing

To show conformance to the SP10 standard, the manufacturer should list each of the requirements of the standard and describe how the device conforms to each requirement. For every requirement which necessitates clinical or in-vitro testing, the test protocol, test data and results, and analysis should be provided and clearly identified. The necessary detail for each element of a test report is described below in General Requirements. For devices with unique features or intended uses, additional testing beyond the SP10 standard may be necessary.

If the SP10 standard is chosen by the manufacturer, conformance to the entire standard is necessary. Conformance to portions of the standard is insufficient to permit the standard's use or to allow a labeling claim to that effect. Therefore, if only part of the standard is met, the manufacturer should refer to the Comparison Testing section of this guidance.

Foreign Standards

If the manufacturer chooses to conform to a standard other than the SP10, it is recommended that they list each requirement of the SP10 standard, compare the foreign standard to the SP10 requirements, and clearly identify where the foreign standard does not meet the requirements of the SP10 standard (if at all). Justification for any differences should be based on valid scientific or statistical analyses and supported by testing if necessary.

Additional Testing

In addition to the testing described above, it is necessary that the manufacturer evaluated the following:

General Requirements

In general, any in-vitro test report should include the following:

In general, any clinical data necessitates consideration or inclusion of the following:

3. In-Vitro Safety Testing

Environmental Testing

The manufacturer should evaluate the ability of the device to function after exposure to the environmental hazards expected when used by an abusive user. Tests for some of these hazards may be found in SP10, IEC 601-1, and IEC 529. These hazards minimally include:

Software

To demonstrate the quality of the software used in or with the device, the following is necessary:

Electrical Safety

Any appropriate standard for electrical safety may be used. If the SP10 standard is used, the manufacturer should conform to the standard's requirements or justify a modification to the standard.

Electromagnetic Compatibility

Electromagnetic compatibility (EMC) testing is necessary to demonstrate that the device (1) will not adversely interfere with the performance of other electronic devices (emissions), and (2) will perform as expected in the presence of other electronic devices or other sources of electromagnetic interference (EMI) in the intended environment of use (immunity). To demonstrate EMC for the device, the following information is necessary:

Any omitted tests or deviations from the requirements of the chosen standard require justification. In addition, the manufacturer should provide a list of all known or suspected EMI incidents associated with the device, the results of any related investigations, a description of any corrective action taken, and any device labeling that references EMC or EMI.

Biocompatibility

Identification of all patient and operator contacting materials and conformance to the International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," is necessary. For materials that are widely used in the same or similar applications, supportive information demonstrating the material's use in other medical devices or products may be acceptable in establishing biocompatibility. However, any references should be to the same vendor and material, and account for any changes to the material due to subsequent processes or manufacturing (e.g., sterilization, forming, melting).

Sterilization

Usually, there are no sterile components to NIBP monitors. If sterile components are identified, refer to the 510(k) Sterility Review Guidance #K90-1 (February 1990) or the most recent sterilization policy or guidance.

Packaging

The manufacturer should describe all packing for the device. This description should include a description of the design, materials, and the sealing method.

For package integrity, any appropriate standard may be used. If the SP10 standard is used, the manufacturer should conform to the standard's requirements or justify any modification.

Shelf Life

Usually a shelf life is not necessary for NIBP monitors. However, if the device contains any sterile or degradable components, shelf life data may be necessary.

4. Labeling

Conformance to the labeling regulations and policies is necessary. Appropriate labeling guidances are available through the Division of Small Manufacturers Assistance (DSMA) at its toll-free number (800) 638­2041 or at its internet address: http://www.fda.gov/cdrh/dsmamain.html.

If the SP10 standard is used, the labeling requirements of the standard should be included or justification provided for any modifications.

5. Regulatory Requirements

Either a Summary of Safety and Effectiveness or a 510(k) Statement is necessary as described in 21 CFR § 807.92 and 21 CFR § 807.93, respectively.

A "Truthful and Accuracy Statement" is necessary according to 21 CFR § 807.87 (j).

An "Indications for Use Statement" is necessary according to Office of Device Evaluation policy. A format for this statement can be provided to the manufacturer by DSMA.

Updated April 15, 1997

horizonal rule

CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page

Center for Devices and Radiological Health / CDRH