Guidance for Third Parties and FDA Staff
Third Party Review of Premarket Notifications
Appendix 4: Special 510(k): Device Modification
Review Memorandum
To: THE FILE RE: DOCUMENT NUMBER
K________
This 510(k) submission contains information/data on modifications made to the
SUBMITTER’S own Class II, Class III or Class I devices requiring 510(k).
The following items are present and acceptable (delete/add items as necessary):
- The name and 510(k) number of the SUBMITTER’S previously cleared
device. (For a preamendments device, a statement to this effect has been provided.)
- Submitter’s statement that the INDICATION/INTENDED USE
of the modified device as described in its labeling HAS NOT CHANGED
along with the proposed labeling which includes instructions for use, package
labeling, and, if available, advertisements or promotional materials (labeling
changes are permitted as long as they do not affect the intended use).
- A description of the device MODIFICATION(S), including
clearly labeled diagrams, engineering drawings, photographs, user’s
and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL
SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for ____________________________________________________________
- Comparison Information (similarities and differences)
to applicant’s legally marketed predicate device including, labeling,
intended use, physical characteristics, and _____________________________________________________________________________
_____________________________________________________________________________
- A Design Control Activities Summary which includes:
- Identification of Risk Analysis method(s) used to assess the impact
of the modification on the device and its components, and the results
of the analysis
- Based on the Risk Analysis, an identification of the verification and/or
validation activities required, including methods or tests used and acceptance
criteria to be applied
- A declaration of conformity with design controls. The declaration of
conformity should include:
- A statement signed by the individual responsible, that, as required
by the risk analysis, all verification and validation activities were
performed by the designated individual(s) and the results demonstrated
that the predetermined acceptance criteria were met, and
- A statement signed by the individual responsible, that the manufacturing
facility is in conformance with design control procedure requirements
as specified in 21 CFR 820.30 and the records are available for review.
- A Truthful and Accurate Statement, a 510(k) Summary
or Statement and the Indications for Use Enclosure (and Class
III Summary for Class III devices).
The labeling for this modified subject device has been reviewed to verify that
the indication/intended use for the device is unaffected by the modification.
In addition, the submitter’s description of the particular modification(s)
and the comparative information between the modified and unmodified devices
demonstrate that the fundamental scientific technology has not changed. The
submitter has provided the design control information as specified in The New
510(k) Paradigm and on this basis, I recommend the device be determined substantially
equivalent to the previously cleared (or their preamendment) device.
_______________________________
(Reviewer’s Signature) |
___________________
(Date) |
Comments
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________
___________________________________________________________________