Document issued on: July 18, 2006
This document supersedes Humanitarian Device Exemptions (HDE) Regulation: Questions and Answers, issued July 12, 2001
For questions regarding this document contact Elisa Harvey at 301-594-1190 x106 or elisa.harvey@fda.hhs.gov
|
|
U.S. Department of Health and Human Services Program Operations Staff |
Written comments and suggestions may be submitted at any time for Agency consideration to the Division of Dockets Management, Food and Drug Administration, 5630 Fishers Lane, Room 1061, (HFA-305), Rockville, MD, 20852. When submitting comments, please refer to the exact title of this guidance document. Comments may not be acted upon by the Agency until the document is next revised or updated.
Additional copies are available from the Internet at: http://www.fda.gov/cdrh/ode/guidance/1381.pdf. You may also send an e-mail request to dsmica@fda.hhs.gov to receive an electronic copy of the guidance or send a fax request to 240-276-3151 to receive a hard copy. Please use the document number (1381) to identify the guidance you are requesting.
| This guidance represents the Food and Drug Administration's (FDA's) current thinking on this topic. It does not create or confer any rights for or on any person and does not operate to bind FDA or the public. You can use an alternative approach if the approach satisfies the requirements of the applicable statutes and regulations. If you want to discuss an alternative approach, contact the FDA staff responsible for implementing this guidance. If you cannot identify the appropriate FDA staff, call the appropriate number listed on the title page of this guidance. |
This guidance answers commonly asked questions about Humanitarian Use Devices (HUDs) and applications for Humanitarian Device Exemption (HDE) authorized by section 510(m)(2) of the Federal Food, Drug, and Cosmetic Act (the act). This update of the version issued in 2001 reflects the comments FDA has received. It also clarifies postmarket aspects related to the regulation of HDEs.
FDA's guidance documents, including this guidance, do not establish legally enforceable responsibilities. Instead, guidances describe the Agency's current thinking on a topic and should be viewed only as recommendations, unless specific regulatory or statutory requirements are cited. The use of the word should in Agency guidances means that something is suggested or recommended, but not required.
We believe we should consider the least burdensome approach in all areas of medical device regulation. This guidance reflects our careful review of the relevant scientific and legal requirements and what we believe is the least burdensome way for you to comply with those requirements. However, if you believe that an alternative approach would be less burdensome, please contact us so we can consider your point of view. You may send your written comments to the contact person listed in the preface to this guidance or to the CDRH Ombudsman. Comprehensive information on CDRH's Ombudsman, including ways to contact him, can be found on the Internet at http://www.fda.gov/cdrh/ombudsman/.
The questions below are numbered consecutively, 1-30 to be consistent with the 2001 version of this document, but we’ve added subheadings for ease of use.
A Humanitarian Device Exemption (HDE) is an application that is similar to a premarket approval (PMA) application, but exempt from the effectiveness requirements of sections 514 and 515 of the act. FDA approval of your HDE authorizes you to market your Humanitarian Use Device (HUD).
As defined in 21 CFR 814.3(n), a HUD is a “medical device intended to benefit patients in the treatment or diagnosis of a disease or condition that affects or is manifested in fewer than 4,000 individuals in the United States per year.”1
FDA makes its determination when you a request a HUD designation. You should submit your request for a HUD designation before submitting an application for a HDE. You should include FDA’s HUD designation letter in your application.
Submit 2 copies of your request for a HUD designation to:
Office of Orphan Products Development (OOPD), HF-35
Food and Drug Administration
5600 Fishers Lane
Rockville, MD 20857
In accordance with 21 CFR 814.102(a), your request must include:
FDA’s Office of Orphan Products Development is available at (301) 827-3666, if you have questions about the HUD designation.
See 21 CFR 814.104(a) for more information on each of the above items.
FDA recognizes that, in some cases, the number of contacts with the device may exceed one per patient. We believe a device that involves multiple patient contacts may still qualify for HUD designation as long as the total number of patients is less than 4,000 per year in the US.
FDA will consider a HDE application when no comparable device is available to treat or diagnose the condition, or if a comparable device is available under an approved HDE application, or being studied under an approved Investigational Device Exemption (IDE)) (21 CFR 814.104(b)(2)). FDA cannot grant a HDE for a HUD device, however, once we allow a comparable device with the same intended use to be marketed through the premarket approval (PMA) process or the premarket notification (510(k)) process.
A “comparable device” need not be identical to the device submitted under your HDE application. In determining whether a comparable device exists, FDA will consider:
In accordance with 21 CFR 814.104(b)(5), if you charge more than $250 for the device, FDA requires you obtain a report by an independent certified public accountant (CPA), or an attestation by a responsible individual of your organization, verifying the amount does not exceed the cost of research, development, fabrication, and distribution. If the amount charged is $250 or less, this requirement is waived.
Yes, but FDA will primarily focus on those manufacturing practices the agency deems most relevant to the safety of the device.
Yes. If you believe you cannot comply with or should not be held to the QSR requirements, you may request an exemption. In evaluating your request, FDA will give overriding consideration to the risks posed by the device, the potential risks that a manufacturing defect might pose, and the public health need for the device.
Under the Section 520(m)(2), FDA will grant or deny a HDE application within a total of 75 days from the date of receipt. This includes a 30 day filing period during which we determine whether the HDE application is sufficiently complete to permit substantive review. If FDA notifies you that your application is incomplete and requests additional information, the 75 day time frame will reset upon our receipt of the additional information.
The review timeframe for HDE amendments, supplements, and reports is 75 days, the same as for HDE original applications, except for a change submitted in a 30-day notice (21 CFR 814.39(f)).
HDE amendments, supplements, and reports are generally subject to the same regulations as those for PMAs.2
No, user fees for HDEs are waived under MDUFMA.
No. You are not required to submit the names and addresses of the reviewing IRBs to FDA. As required in 21 CFR 814.126(b)(2), however, you must maintain records of:
The healthcare provider should be responsible for obtaining IRB approval before he or she uses a HUD to treat or diagnose patients.
Yes. In accordance with Section 520(m)(4) of the act and section 21 CFR 814.124(a), IRB approval is required before a HUD is used in a facility, except in emergencies where the physician determines that approval cannot be obtained in time to prevent serious harm or death to the patient. As is true for IDEs, the IRB may be a local IRB or it may be an independent or national IRB. In addition, a local IRB may defer in writing to another similarly constituted IRB that has agreed to assume responsibility for review of the use of the HUD.
IRBs should be responsible for initial as well as continuing review of the HUD. For initial review of a HUD, IRBs are required to perform a full board review. For continuing review, however, IRBs may use the expedited review procedures (21 CFR 56.110) unless the IRB determines that full board review should be performed. The agency believes that the expedited review procedures are appropriate for continuing review since the initial review would have been performed by the full board and use of a HUD within its approved labeling does not constitute research.
No. The IRB is not required to review and approve individual uses of a HUD, although it may do so. The IRB may use its discretion to determine how to approve use of the HUD. The IRB may approve use of the HUD, for instance, without any further restrictions, under a protocol, or on a case-by-case basis. In reviewing the use of a HUD, IRBs should be cognizant that the FDA recommends that the use of the device not exceed the scope of the indication approved in the HDE.
Neither the act nor the regulations require informed consent for use of a HUD. Because a HDE provides for marketing approval, use of the HUD does not constitute research or an investigation, which would normally require informed consent.
Although informed consent is not required, there is nothing in the statute or regulation that preempts a state or institution from requiring prospective informed consent.
Most HDE holders, however, develop patient labeling that incorporates information to assist a patient in making an informed decision about the use of the HUD. That is, the patient labeling generally contains a discussion of the potential risks and benefits of the HUD and any procedures associated with its use. Patient labeling also should state that the device is a humanitarian use device and effectiveness for the labeled indication has not been demonstrated.
You are required under 21 CFR Part 803 Medical Device Reporting to submit a report to FDA whenever a device with an approved HDE may have caused or contributed to a death or serious injury, or has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
In accordance with 814.126(b)(1), you must provide FDA with updated information on a periodic basis demonstrating that the HUD designation is still valid, based on the most current and authoritative information available. If, as a result of information in your report, FDA believes that the HUD designation may no longer apply, we may contact you for additional information. In addition, you must provide FDA with information on the number of devices shipped or sold since initial marketing approval, the clinical experience with the device and a summary of any changes made to the device. See 21 CFR 814.126(b)(1) for more information on each of the above items.
No. In accordance with 21 CFR 814.106, if you are seeking a new indication for use of an approved HUD, you must first obtain a HUD designation for the new indication for use and then submit a new original HDE application. In your new application, any information or data submitted in your HDE for the original indication may be incorporated by reference.
If FDA makes the determination that more than 4,000 individuals in the US are affected or manifest a certain disease or condition per year, the agency may consider whether the HDE should be withdrawn. FDA intends to consider factors such as the number of patients with the disease or condition, the feasibility of conducting a pivotal clinical trial (to demonstrate reasonable assurance of safety and effectiveness), and the public health need for the device.
If FDA subsequently approves a PMA or clears a 510(k) for the HUD or another comparable device with the same indication, we may rescind the HDE approval. Once a comparable device becomes legally marketed through PMA approval or 510(k) clearance to treat or diagnosis the disease or condition, there may no longer be a need for the HUD and the HUD can no longer meet the requirements of Section 520(m)(2)(B) of the act.
A HDE holder may collect safety and effectiveness data to support a PMA for the HDE-approved indication without an IDE. FDA considers the study exempt from the requirement for an approved IDE as long as the HUD is used in accordance with the approved indication for use and labeling. IRB approval (21 CFR Part 56) and informed consent (21 CFR Part 50) are still needed, however, as required for FDA-regulated clinical studies. Clinical investigation, however, of a HUD beyond its approved indication (e.g., for a broader or different indication) requires an approved IDE. In addition to the requirement of having an FDA-approved IDE, sponsors of these trials must also comply with the regulations governing IRBs (21 CFR Part 56) and informed consent (21 CFR Part 50).
Yes. In an emergency situation, a HUD may be used off-label, but FDA recommends you follow certain patient protection measures before use. Because IRB review and approval is required before a HUD is used within its approved labeling, FDA believes similar procedures should apply if you use a HUD outside of its approved labeling. That is, in an emergency situation, a HUD may be used off-label to save the life or protect the physical well-being of a patient; however, in this situation, FDA recommends that the physician and HDE holder follow the same emergency use procedures that govern the use of unapproved devices.
According to this policy, before the device is used, if possible, the physician should obtain the IRB chairperson’s concurrence, informed consent from the patient or his/her legal representative, an independent assessment by an uninvolved (i.e., not the referring) physician, and institutional clearance. In addition, the physician should obtain authorization from the HDE holder before the emergency use of the HUD. After the emergency use occurs, the physician should submit a follow-up report on the patient’s condition and information regarding the patient protection measures to the HDE holder, who would then submit this information as a HDE report to the FDA.
If a HUD is used in an emergency situation, the physician should devise a schedule for monitoring the patient, taking into account the specific needs of the patient and the limited information available about the risks and benefits of the device. See Guidance on IDE Policies and Procedures for further discussion of the post-treatment procedures for emergency use cases, including the submission of a follow-up report to FDA.3
Yes, a HUD may be used for compassionate use. As in the case of emergency use, FDA recommends that the physician ensure that patient protection measures discussed above are addressed before the device is used. In addition, we recommend you first obtain FDA approval for compassionate use.
FDA believes that a physician who wishes to use a HDE-approved device for compassionate use should provide the HDE holder with:
We also recommend the HDE holder submit the above information in a HDE report for FDA approval before the use occurs to help ensure adequate patient protection. FDA will review the information in the most expeditious manner possible and issue a letter to the HDE holder.
If the physician undertakes a compassionate use, he or she should devise a schedule for monitoring the patient, taking into consideration the specific needs of the patient and the limited information available regarding the risks and benefits of the device for this unapproved use. See Guidance on IDE Policies and Procedures for further discussion of the post-approval procedures for compassionate use cases, including the submission of a follow-up report to FDA.3
1 See 61 FR 33233, June 26, 1996.
2 See 21 CFR 814.106 HDE Amendments and resubmitted HDEs, 21 CFR 814.108 Supplemental applications, and 21 CFR 814.126 Postapproval requirements and reports for information about the respective exceptions to 21 CFR 814.37 PMA amendments and resubmitted PMAs, 21 CFR 814.39 PMA supplements, and 21 CFR 814.82 Postapproval requirements.
3 See Guidance on IDE Policies and Procedures: Emergency Use of Unapproved Medical Devices, Chapter III Expanded Access to Unapproved Devices, www.fda.gov/cdrh/ode/idepolcy.html. For emergency use of a HUD, we believe the HDE holder assumes the responsibilities of the IDE sponsor described in the guidance
Updated July 25, 2006
![]()