The Reveal Rapid HIV-1 Antibody Test must be used in accordance with this package insert to ensure accurate results.
2. The FDA has approved the Reveal Rapid HIV-1 Antibody Test for serum or plasma specimens only. Use of other types of specimens may not yield accurate results.
3. Test results are to be read and interpreted immediately following the final washing step with Universal Buffer. A delay in reading test results may yield inaccurate results.
4. Specimens that do not pass through the membrane within thirty (30) seconds after centrifugation (see Testing Procedure, step 2) are unsuitable for testing with the Reveal Rapid HIV-1 Antibody Test.
5. Lipemic samples or specimens contaminated with bacteria may not pass through the membrane within thirty (30) seconds, and therefore may be unsuitable for testing with the Reveal Rapid HIV-1 Antibody Test.
6. Limited studies were conducted to determine the potential effect of interfering substances and unrelated medical conditions on the performance of the Reveal Rapid HIV-1 Antibody Test.
7. The specificity of the Reveal Rapid HIV-1 Antibody Test for serum specimens in low-risk populations has not been evaluated.
8. Limited studies were conducted to determine the performance of the Reveal Rapid HIV-1 Antibody Test on fresh serum and plasma specimens.
9. A Reactive test result using the Reveal Rapid HIV-1 Antibody Test suggests the presence of anti-HIV-1 antibodies in the specimen. The Reveal Rapid HIV-1 Antibody Test is intended to be used as an aid in the diagnosis of infection with HIV-1. AIDS and AIDS-related conditions are clinical syndromes and their diagnosis can only be established clinically. Results of the MedMira Reveal Rapid HIV-1 Antibody Test should not be used in isolation, but in conjunction with the clinical status, history, and risk factors of the individual being tested.
10. The intensity of the red dot (Reactive test result) does not necessarily correlate with the antibody titre of the specimen.
11. A Non-Reactive test result with the Reveal Rapid HIV-1 Antibody Test indicates the absence of detectable antibodies to HIV in the specimen. However, a Non-Reactive test result does not exclude the possibility of exposure to, or infection with HIV. Following a recent exposure to HIV, it may take several months for the antibody response to reach detectable levels, during which time testing for antibodies to HIV will not be indicative of true infection status. A comprehensive risk history and clinical judgement should be considered before concluding that an individual is not infected with HIV.
VII. Alternative Practices and Procedures
The detection of antibodies to HIV-1 in humans is primarily performed using laboratory-based assays for serum, plasma, oral fluid or urine. The majority of these tests use principles similar to that of the Reveal Rapid HIV-1 Antibody Test; utilizing peptides, recombinant antigens, isolated proteins or viral lysate immobilized onto a solid phase support to capture antibodies in a patient sample. In many cases, the detection of the captured antibodies is accomplished using an instrument to reveal a coloured or chemiluminiescent endpoint from enzymatic reactions. All reactive screening test results require supplemental, more specific testing.
The Reveal Rapid HIV-1 Antibody Test differs from traditional laboratory-based testing in that it requires no extra instrumentation, and a protein A-colloidal gold conjugate colorimetric detection agent is used for qualitative visual interpretation of test results.
VIII. Marketing History
There is no marketing history for the Reveal Rapid HIV-1 Antibody Test. This product has not been marketed domestically or internationally.
IX. Potential Adverse Effects of the Device on Health
There have been no adverse effects of the Reveal Rapid HIV-1 Antibody Test indicated during studies performed to date.
X. Summary of Preclinical Studies
Sensitivity of the Reveal Rapid HIV-1 Antibody Test in the Detection of HIV-1 From Various Geographic Regions
The sensitivity of the Reveal Rapid HIV-1 Antibody Test for the detection of antibodies to HIV-1 Group M subtypes (A,B,C,D,E,F,G) from various geographic regions was assessed by testing 1026 confirmed HIV-1 antibody positive serum and plasma specimens obtained from various parts of the world. Of these 1026 specimens, 1024 were Reactive using the Reveal Rapid HIV-1 Antibody Test. Two confirmed HIV-1 antibody positive specimens from Canada were Non-Reactive using the Reveal Rapid HIV-1 Antibody Test.
Reactivity with Seroconversion Panels
Seven seroconversion panels were tested in comparison to a licensed anti-HIV-1,2 EIA. Each panel consisted of a series of sequential specimens obtained from a single individual undergoing seroconversion. Five of the 7 panels were obtained from a commercial source, while the remaining 2 were from clinical settings. The 7 seroconversion panels consisted of 36 specimens. In this study, the Reveal Rapid HIV-1 Antibody Test detected seroconversion similarly to the FDA-licensed HIV-1,2 EIA (Table 1).
Table 1. Performance of the Reveal Rapid HIV-1 Antibody Test with Seroconversion Panels
| Specimen |
Information |
|
|
| Panel |
Relative Day of Bleed |
Reveal test |
Licensed anti-HIV-1,2, EIA |
| AF |
1 |
NR |
NR |
| 3 |
NR |
NR |
| 8 |
NR |
NR |
| 10 |
NR |
NR |
| 16 |
R |
NR |
| 29 |
R |
RR |
| 34 |
R |
RR |
| 29 |
R |
RR |
| 43 |
R |
RR |
| D |
1 |
NR |
NR |
| 22 |
NR |
NR |
| 50 |
R |
NR |
| 93 |
R |
RR |
| 100 |
R |
RR |
| H |
1 |
NR |
NR |
| 8 |
NR |
NR |
| 13 |
NR |
NR |
| 20 |
NR |
NR |
| 27 |
NR |
NR |
| 29 |
R |
RR |
| M |
1 |
NR |
NR |
| 23 |
R |
RR |
| E |
1 |
NR |
NR |
| 8 |
NR |
NR |
| 22 |
NR |
NR |
| 36 |
NR |
NR |
| 43 |
NR |
NR |
| 50 |
NR |
NR |
| 64 |
NR |
NR |
| 85 |
NR |
NR |
| 92 |
NR |
RR |
| 127 |
R |
RR |
| TORONTO PANEL 1 |
1 |
NR |
RR |
| 22 |
R |
RR |
| TORONTO PANEL 6 |
1 |
R |
NR |
| 72 |
R |
RR |
NR= Non-Reactive; R=Reactive; RR=Repeatedly Reactive
Reactivity with Low Titre HIV-1 Antibody Performance Panel
A low titre HIV-1 antibody panel consisting of 15 specimens, obtained from a commercial source, was tested in comparison with licensed anti-HIV EIA tests. The results of the study are shown in Table 2. The Reveal Rapid HIV-1 Antibody Test was capable of detecting antibodies to HIV-1 similarly to the licensed anti-HIV EIA tests.
Table 2: Comparison of the Reveal Rapid HIV-1 Antibody Test and Licensed Anti-HIV EIA Tests Using a Low Titre HIV-1 Antibody Performance Panel
Panel
Member |
Reveal Test |
Licensed Anti-HIV Tests |
| EIA #1 |
EIA #2 |
EIA #3 |
EIA #4 |
EIA #5 |
| 1 |
NR |
NR |
RR |
RR |
NR |
NR |
| 2 |
R |
NR |
RR |
RR |
RR |
NR |
| 3 |
R |
NR |
RR |
NR |
NR |
NR |
| 4 |
R |
RR |
RR |
RR |
RR |
NR |
| 5 |
NR |
NR |
NR |
NR |
NR |
NR |
| 6 |
R |
RR |
RR |
RR |
RR |
NR |
| 7 |
NR |
NR |
RR |
RR |
NR |
NR |
| 8 |
NR |
NR |
RR |
NR |
RR |
NR |
| 9 |
NR |
NR |
RR |
NR |
NR |
NR |
| 10 |
R |
RR |
RR |
RR |
RR |
RR |
| 11 |
R |
RR |
RR |
RR |
RR |
RR |
| 12 |
NR |
NR |
RR |
NR |
NR |
NR |
| 13 |
NR |
NR |
RR |
RR |
NR |
NR |
| 14 |
R |
RR |
RR |
RR |
RR |
RR |
| 15 |
R |
RR |
RR |
RR |
RR |
RR |
NR=Non-Reactive, R=Reactive, RR=Repeatedly Reactive
Interfering Substances and Unrelated Medical Conditions
The effects of anticoagulants, potentially interfering medical conditions, seromarkers and abnormal blood chemistry on the outcome of Reveal Rapid HIV-1 Antibody Test were assessed using a total of 276 HIV-1 antibody positive serum and plasma specimens. Of these, 155 specimens were from confirmed cases of HIV-1 infection and 121 specimens were spiked with HIV-1 antibody positive specimens. These 276 HIV-1 antibody positive specimens represented the following conditions: EDTA (n=73), heparin (n=89), sodium citrate (n=20), hemolyzed (n=2), icteric (n=2), lipemic (n=1), HBsAg (n=3), anti-HBc (n=3), anti-HBs (n=4), anti-HCV (n=3), EBV (n=5), CMV (n=6), measles (n=6), HSV (n=5), rubella (n=4), varicella (n=3), RA (n=5), C reactive protein (n=5), infectious mononucleosis (n=4), antistreptolysin O (n=2), and H. pylori (n=5). The effect of multiple elevated analytes on the sensitivity of the Reveal Rapid HIV-1 Antibody Test was assessed using pairs of specimens (with elevated analyte levels and normal analyte levels) spiked with an HIV-1 positive specimen. Specimens contained elevated levels of the following (n=26): AST (range of elevated levels tested, 96 to 1764 U/L; reference range, 0 to 37 U/L), ALT (136 to 2296 U/L; reference, 0 to 37 U/L), ALP (297 to 550 U/L; reference, 36 to 125 U/L), total bilirubin (34 to 65 ?mol/L; reference, 0 to 20 ?mol/L), LD (350 to 600 U/L; reference, 90 to 180 U/L), uric acid (600 to 940 ?mol/L; reference, 150-450 ?mol/L). Multiple elevated analytes were: AST, ALT, ALP (1); ALT, total bilirubin (4); AST, ALT (2); AST, ALT, total bilirubin (2); AST, ALT, ALP, Uric Acid (1); ALT, LD, AST, ALP (2); ALT, ALP (1).
The results indicated that none of the above conditions interfered with the Reveal Rapid HIV-1 Antibody Test.
The effects of anticoagulants, potentially interfering medical conditions and seromarkers on the outcome of Reveal Rapid HIV-1 Antibody Test were assessed using a total of 271 serum and plasma specimens from known HIV antibody negative individuals. These 271 specimens contained the following: EDTA (n=79), heparin (n=20), sodium citrate (n=26), HBsAg (n=4), anti-HBc (n=4), anti-HBs (n=3), anti-HCV (n=10), HTLV-1 (n=10), CMV (n=10), measles (n=10), HSV (n=10), rubella (n=10), varicella (n= 10), RA (n=10), C reactive protein (n=10), infectious mononucleosis (n=10), antistreptolysin O (n=15), mycoplasma (n=10), and syphilis reagin antibodies (n=10).
The results indicated that none of the above conditions interfered with the Reveal Rapid HIV-1 Antibody Test.
REPRODUCIBILITY
The reproducibility of the Reveal Rapid HIV-1 Antibody Test was studied at three sites using three lots of the device on three different days by three operators per site. Coded panels of 15 samples were used in triplicate for this study. Each panel consisted of strongly and weakly Reactive HIV-1 antibody serum/plasma samples, as well as HIV-1 antibody negative specimens. In addition, three lots of MedMira Positive and Negative Test Controls were included in the panel. A total of 810 tests were performed (270 per site) with a total of 54 tests performed per panel member. The overall reproducibility of the Reveal Rapid HIV-1 Antibody Test was found to be 810/810 = 100%.
XI. Summary of Clinical Studies
SENSITIVITY
Serum Specimens
Sensitivity studies were performed using repository HIV-1 antibody positive serum specimens from individuals known to be infected with HIV-1 and freshly collected serum specimens from routine clinical settings in high-risk populations at four clinical sites. Serum specimens were obtained from 483 individuals known to be infected with HIV-1, as well as 2914 serum specimens from previously unscreened individuals from high-risk populations. All 483 serum specimens from known HIV-1 antibody positive individuals that were Repeatedly Reactive using an FDA-licensed HIV-1,2 EIA and were HIV-1 antibody Western blot positive gave Reactive results with the Reveal Rapid HIV-1 Antibody Test. Of the 2914 previously unscreened serum specimens from individuals from a high-risk population, 124 were Repeatedly Reactive using an FDA-licensed HIV-1,2 EIA, 123 were confirmed positive by Western blot (1 indeterminate result), and 122 were Reactive using the Reveal Rapid HIV-1 Antibody Test. The results of these studies are shown in Table 3A.
Table 3A. Detection of Anti-HIV-1 Antibodies in Serum Specimens Obtained From Known HIV-1 Antibody Positive Individuals and Routine Clinical Settings in HIV High-Risk Populations
| Test Group |
Total Samples |
Reveal Reactive |
Licensed EIA Repeatedly Reactive |
Western Blot Positive |
| Known HIV-1 Antibody Positive |
483 |
483 |
483 |
483 |
| High-Risk Population |
2914 |
122 |
124* |
123 |
| Total |
3397 |
605 |
607 |
606 |
* One specimen was Western blot indeterminate.
The overall sensitivity of the Reveal Rapid HIV-1 Antibody Test for the detection of anti-HIV-1 antibodies in serum specimens in these studies was calculated to be 605/606 = 99.8% (95% confidence interval; 99.2-100%), combining the number of Reveal Rapid HIV-1 Antibody Test Reactive results obtained from the study of known HIV-1 antibody positive serum specimens with the number of Reveal Rapid HIV-1 Antibody Test Reactive results obtained from the studies of high-risk populations.
Plasma Specimens
Sensitivity studies were performed using repository HIV-1 antibody positive plasma specimens from individuals known to be infected with HIV-1 and repository plasma specimens from clinically diagnosed AIDS patients. Plasma specimens were obtained from 397 individuals known to be infected with HIV-1, as well as 107 specimens from clinically diagnosed AIDS patients from one clinical site. All 397 plasma specimens were Repeatedly Reactive using an FDA-licensed HIV-1,2 ---, 395 were ---------------- positive (2 ------------------ indeterminate) and 395 gave Reactive results with the Reveal Rapid HIV-1 Antibody Test. Of the 107 specimens from clinically diagnosed AIDS patients, 107 were Repeatedly Reactive using an FDA-licensed HIV-1,2 ------, 104 were positive by ------------------- (3 ----------------- indeterminate) and 103 were Reactive using the Reveal Rapid HIV-1 Antibody Test. The results of these studies are shown in Table 3B.
Table 3B. Detection of Anti-HIV-1 Antibodies in Plasma Specimens Obtained From Known HIV-1 Antibody Positive Individuals and Clinically Diagnosed AIDS Patients In Routine Clinical Settings
| Test Group |
Total Samples |
Reveal Reactive |
Licensed ----- Repeatedly Reactive |
----------- Positive |
| Known HIV-1 Antibody Positive |
397 |
395 |
397* |
395 |
| Clinically Diagnosed AIDS Patients |
107 |
103 |
107** |
104 |
| Total |
504 |
498 |
504 |
499 |
* Two specimens were ----------------- indeterminate, ** Three specimens were -------------------- indeterminate.
The overall sensitivity of the Reveal Rapid HIV-1 Antibody Test for the detection of anti-HIV-1 antibodies in plasma specimens in these studies was calculated to be 498/499 = 99.8% (95% confidence interval; 99.0-100%), combining the number of Reveal Rapid HIV-1 Antibody Test Reactive results obtained from the study of known HIV-1 antibody positive plasma specimens with the number of Reveal Rapid HIV-1 Antibody Test Reactive results obtained from clinically diagnosed AIDS patients.
SPECIFICITY
Serum Specimens
Specificity studies were performed using repository serum specimens obtained from 850 previously screened HIV-1 antibody negative individuals from a high-risk population and 2914 freshly collected serum specimens from previously unscreened individuals from high-risk populations. Specimens were collected from three clinical sites. Of the 850 repository HIV-1 antibody negative serum specimens, 845 gave Non-Reactive results with the Reveal Rapid HIV-1 Antibody Test. Of the 2914 freshly collected serum specimens, 2763 gave Non-Reactive results using the Reveal Rapid HIV-1 Antibody Test. Five of the specimens were found to be Western blot indeterminate. The results of these studies are shown in Table 4A.
Table 4A. Performance of the Reveal Rapid HIV-1 Antibody Test on Serum Specimens Presumed to be Negative for Antibodies to HIV-1
| Test Group |
Total Samples |
Reveal Non-Reactive |
Licensed ----- Non-Reactive |
True Negative* |
| Known HIV-1 Antibody Negative |
850 |
845 |
850 |
850 |
| High-Risk Population |
2914 |
2763 |
2789 |
2789 |
| Total |
3764 |
3608 |
3639 |
3639 |
* True Negative status was based on negative or indeterminate test results using a licensed -----------------------.
The overall specificity of the Reveal Rapid HIV-1 Antibody Test for serum specimens in these studies was calculated to be 3608/3639 = 99.1% (95% Confidence Interval; 98.8-99.4%), combining the number of Reveal Rapid HIV-1 Antibody Test Non-Reactive results obtained from the study of previously screened HIV-1 antibody negative serum specimens with the number of Reveal Rapid HIV-1 Antibody Test Non-Reactive results obtained from the studies of high-risk populations.
Plasma Specimens
Specificity studies were performed using plasma specimens that had been frozen once, obtained from 1000 previously screened HIV-1antibody negative individuals, and 2011 freshly collected plasma specimens from previously unscreened individuals from low risk-populations. All plasma specimens were collected from the same clinical site. Of the 1000 pre-screened HIV-1 antibody negative plasma specimens, 992 gave Non-Reactive results with the Reveal Rapid HIV-1 Antibody Test. Of the 2011 freshly collected HIV-1 antibody negative plasma specimens, 1978 gave Non-Reactive results using the Reveal Rapid HIV-1 Antibody Test. Four of the specimens were found to be ---------------------------------- indeterminate. The results of these studies are shown in Table 4B.
Table 4B. Performance of the Reveal Rapid HIV-1 Antibody Test on Plasma Specimens Presumed to be Negative for Antibodies to HIV-1
| Test Group |
Total Samples |
Reveal Non-Reactive |
Licensed ----- Non-Reactive |
True Negative* |
| Known HIV-1 Antibody Negative |
1000 |
992 |
1000 |
1000 |
| Low-Risk Population |
2011 |
1978 |
2011 |
2011 |
| Total |
3011 |
2970 |
3011 |
3011 |
* True Negative status was based on negative or indeterminate test results using a licensed -----------------------.
The overall specificity of the Reveal Rapid HIV-1 Antibody Test for plasma specimens in these studies was calculated to be 2970/3011 = 98.6% (95% Confidence Interval; 98.4-98.8%), combining the number of Reveal Rapid HIV-1 Antibody Test Non-Reactive results obtained from the study of previously screened HIV-1 antibody negative plasma specimens with the number of Reveal Non-Reactive results obtained from the studies of low-risk populations.
XII. Conclusions Drawn from the Studies
Risk/Benefit Analysis
Conventional laboratory tests for antibodies to HIV-1 generally utilize complex, multi-step tests requiring use by highly trained individuals with a turnaround time of several days to a few weeks. Additionally, the complexity and cost of complex testing and required equipment may prohibit the universal utilization of conventional testing in medical settings with limited resources and personnel. Less complex HIV tests, such as MedMira's Reveal Rapid HIV-1 Antibody Test, will improve the delivery of medical care and HIV prevention services with substantial time and cost savings. Realizing the utility of rapid tests, the World Health Organization (WHO) recommends the use of alternative testing strategies using rapid and simpler HIV tests. The United States Centers for Disease Control and Prevention (CDC) accompanied these recommendations with studies concluding that large numbers of patients tested for HIV using conventional methods did not return to the medical facility to obtain test results. From a public health perspective, this high non-return rate has great implications for the health and welfare of that individual and his/her contacts. The MedMira Reveal Rapid HIV-1 Antibody Test is a rapid, flow-through diagnostic immunoassay developed to utilize the performance characteristics of conventional diagnostic immunoassays while simplifying the testing procedure to eliminate the requirement for expensive equipment and highly trained personnel, thereby decreasing turnaround time.
Safety
No adverse reactions were observed during any of the studies conducted. All operators conducted testing in accordance with the training provided.