Recall -- Firm Press Release

FDA posts press releases and other notices of recalls and market withdrawals from the firms involved as a service to consumers, the media, and other interested parties. FDA does not endorse either the product or the company.


Insulet Corporation Issues Voluntary Recall of OmniPod® Insulin Management System


FOR IMMEDIATE RELEASE – August 27, 2015 – Billerica, MA – On July 8, 2015, Insulet Corporation (Insulet or the Company) initiated a lot-specific voluntary recall of 40,846 boxes (10 Pods per box) of the OmniPod (Pod) Insulin Management System. This field corrective action is due to the possibility that some of the Pods from these lots may have a higher rate of failure than Insulet's current manufacturing standards. This recall does not affect the OmniPod Personal Diabetes Manager (PDM).

There are two ways in which these Pods can fail at a rate that is higher than Insulet's current standard. The cannula may either completely retract or fail to fully deploy, which may result in the patient not receiving the expected insulin dose. Or the Pod may trigger an audible alarm indicating it will no longer deliver insulin and will need to be replaced. Both situations can result in the interruption of insulin delivery that can cause hyperglycemia, which, if left untreated, can result in diabetic ketoacidosis (DKA).

The affected Pod lots have resulted in 90 Medical Device Reports of which 13 required medical intervention. No serious injuries or deaths have been reported in patients using OmniPod devices from the affected lots.

Please check to determine if you have Pods from any of the lots listed below. The lot number is located on the Pod tray lid label and is also laser etched on the side of each individual Pod. The lot number is also located on the box of OmniPods. Consumers who have Pods from the affected lots should stop using them and return the pods for replacement.

The following OmniPod lots have been voluntarily recalled:

Distribution Catalog Number Description Lot Number
United States POD-ZXP420 OmniPod® Insulin Management System L40806
International 14810 OmniPod® Insulin Management System L40771

OmniPods from the affected lots were distributed to customers from December 2013 to March 2015.

Insulet has notified its distributors and customers by email, FedEx, and phone calls and is arranging for return and replacement of all recalled product. Consumers with questions may contact the Company via telephone at 1-855-407-3729 at any time.

Insulet has notified all applicable regulatory agencies of this voluntary action, including U.S. Food and Drug Administration and the Competent Authorities in Austria, Germany, Italy, Netherlands, Norway, Sweden, Switzerland and the UK.

Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.