Safety

Central Venous Catheters and Pressure Monitoring Sets and Trays by Cook Medical: Recall - Catheter Tip Fracture and/or Separation

UPDATED 03/30/2016. Recall Classified as Class I, Class I Recall Notice added.

[Posted 02/11/2016]

AUDIENCE: Risk Manager

ISSUE: During an internal inspection, a catheter exhibited the potential for catheter tip fracture and/or separation. Further investigation revealed that the technique used by the product assembler while tipping the catheter likely contributed to this nonconformance. No reports of illness or injury have been associated with this issue to date.

Potential adverse events that may occur as a result of catheter tip fracture and/or separation include loss of device function, the need for medical intervention to retrieve a separated segment, and complications resulting from a separated tip occluding blood flow to end organs. Examples of such complications include stroke, kidney injury, or damage to the intestines or limbs.

The Single Lumen Central Venous Catheters and Pressure Monitoring Sets and Trays in this recall were distributed globally between April 24, 2015, and October 23, 2015. Product can be identified by the part number and lot number that are provided on the outer package product label. See the Press Release for a listing of affected lot numbers.

BACKGROUND: These products are intended for use in venous or arterial pressure monitoring, blood sampling, and administration of drugs and fluids.

RECOMMENDATION: Cook Medical notified its customers and distributors by recall notification letters in January. All customers and distributors should quarantine and discontinue use of all affected units and return the affected product to the company as soon as possible for credit.

Healthcare professionals and patients are encouraged to report adverse events or side effects related to the use of these products to the FDA's MedWatch Safety Information and Adverse Event Reporting Program:

  • Complete and submit the report Online: www.fda.gov/MedWatch/report
  • Download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178

[03/30/2016 - Recall Notice - FDA]
[02/10/2016 - Press Release - Cook Medical]
 

Page Last Updated: 03/30/2016
Note: If you need help accessing information in different file formats, see Instructions for Downloading Viewers and Players.