• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Radiation-Emitting Products

  • Print
  • Share
  • E-mail
-

Our facility has five FFDM units (not all same model or manufacturer), two laser printers (not same model or manufacturer) and four review workstations (not same model or manufacturer). We run clinical images through all the accessory components. Do we have to perform periodic QC testing (e.g., phantom image test) on all combinations of units with the printers and review workstations?

Policy Guidance Help System logo

Our facility has five FFDM units (not all of the same model or from the same manufacturer), two laser printers (not the same model or from the same manufacturer) and four review workstations (not all of the same model or from the same manufacturer). We run clinical images through all the accessory components interchangeably. Do we have to perform periodic QC testing (e.g., weekly phantom image test) on all possible combinations of units with the printers and review workstations?

 

If the facility’s medical physicist provides the facility with documentation that the FFDM units and accessory components are matched (see below), then the facility may perform its periodic QC tests using any review workstation or laser printer, as long as each accessory component is tested at its recommended frequency. For example, if the QC manual specifies performing the phantom image test on a weekly basis, the facility may produce a phantom image using each of its x-ray units and evaluate them on any review workstation or laser printer at least once each week of use. One mechanism by which the medical physicist can consider the FFDM units and the accessory components to be matched is for the medical physicist to initially perform and obtain passing results for all applicable QC tests through all possible combinations. (In this scenario each accessory component must be operating within its own pre-established action limits set forth in the applicable QC manual.) 21 CFR 900.12(e)(6). This method will reduce the number of QC tests that must be performed. Note:  At least one example of each required QC test must be evaluated using each accessory component at the required periodic frequency. 21 CFR 900.12(e)(6).

Back to Other Modalities Quality Control Tests

-
-