Our facility has multiple mammographic units and uses several different screen-film combinations. What QC tests are we required to perform?

The required QC tests are summarized in the following table.
Table: QC Tests* Required for Facilities Using Multiple Units & Screen-Film Combinations
Test
|
Units Tested
|
S-F Combinations Tested With Each Unit |
|
Focal Spot Condition System Resolution
|
All units** |
For all S-F combinations clinically used with the unit in question. (900.12(e)(5)(iii)
|
| Phantom Image |
All units used to image the standard*** breast
|
All S-F combinations clinically used for the standard breast (900.12(e)(2)) |
| Phantom Image |
All units that are used only for non-standard breasts, and/or magnification work
|
One S-F combination, using clinical techniques that would be used for phantom images submitted to the accreditation body |
| Dose | All units used to image the standard breast |
All S-F combinations clinically used for the standard breast with their corresponding techniques (900.12(e)(5)(vi)
|
| Dose |
All units that are used only for non-standard breasts, and or magnification work
|
One S-F combination, using clinical techniques that would be used for the standard breast |
| Darkroom Fog |
Any unit (one only)
|
All film types clinically used (900.12(e)(4)(i)) |
|
Uniformity of Screen Speed
|
Any unit (one only) |
All clinically used screens (cassettes), one film type (appropriate for cassette used). Also see guidance on grouping of cassettes (Question 1 under Screen Speed Uniformity Annual Quality Control Test). (900.12(e)(5)(viii)).
|
| Screen-Film Contact | Any unit (one only) |
All clinically used screens (cassettes), one film type (900.12(e)(4)(ii))
|
| AEC Performance - Reproducibility | All units |
One S-F combination (typically used with the unit) (900.12(e)(5)(v))
|
| AEC Performance – kVp & Thickness Tracking |
All units used clinically in the 2-6 cm thickness range
|
One S-F combination (typically used with the unit) (900.12(e)(5)(i)) |
| Collimation and System Artifacts | All units |
One S-F combination (typically used with the unit) (900.12(e)(5)(ix))
|
| kVp Accuracy & Reproducibility, HVL, Decompression, and Radiation Output | All units | None |
* This table does not cover other required QC tests that are independent of the x-ray unit or screen-film combination (e.g., the daily processor QC, the fixer retention, and repeat analysis), or additional tests that are required for units with multiple targets and filters.
** In this table, "All units" refers to those that are used clinically.
*** The standard breast referenced in this table is defined in the regulations as "a 4.2 centimeter (cm) thick compressed breast consisting of 50 percent glandular and 50 percent adipose tissue."