• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Radiation-Emitting Products

  • Print
  • Share
  • E-mail
-

I have been following the ACR manual and have used an optical density range of 0.7-0.8 when doing my screen-film contact test. The film manufacturer recommends that I use a higher optical density. What should I use under the final regulations?

Policy Guidance Help System logo

The final regulations do not reference a specific manual giving facilities flexibility to use procedures that best enable them to meet the requirements. In this specific case, the facility may follow the manufacturer’s recommendation, their medical physicist’s recommendation, or an appropriate manual in performing the screen-film contact test or any other QC test.

Back to Quality Assurance Program

-
-