When performing a physics survey or equipment evaluation on a unit with multiple target/filter combinations, what tests or measurements must be performed for each combination?

For a unit with multiple target/filter combinations, the following tests must be performed for each clinically used target/filter combination:
-
Focal spot condition (for different target materials [tracks] only)
-
X-ray field/light field/image receptor/compression paddle alignment (for different target materials [tracks] only)
-
Beam quality and half-value layer
-
System artifacts